Provisional stenting technique in the left main bifurcation setting: Computational fluid dynamics and optical coherence tomography pilot study in humans

Marek Milewski1, Evelyn Linyi Koh2, Paweł Gąsior1, Shaoliang Shawn Lian2, Zhu Yinling2, Tomasz Pawłowski1, Nicolas Foin34, Elvin Khedi15, Wojciech Wojakowski1, Hui Ying Ang2
Pubmed: 38767164
Pol Heart J 2024;82(6):651-653.

Abstract

Not available

SHORT COMMUNICATION

Provisional stenting technique in the left main bifurcation setting: Computational fluid dynamics and optical coherence tomography pilot study in humans

Marek Milewski1Evelyn Linyi Koh2Paweł Gąsior1Shaoliang Shawn Lian2Zhu Yinling2Tomasz Pawłowski1Nicolas Foin34Elvin Khedi15Wojciech Wojakowski1Hui Ying Ang2
1Division of Cardiology and Structural Heart Diseases, Medical University of Silesia, Katowice, Poland
2Department of Biomedical Engineering, National University of Singapore, Singapore
3National Heart Research Institute Singapore, National Heart Centre Singapore, Singapore
4Duke-NUS Medical School, Singapore, Singapore
5Erasmus Hospital, Université libre de Bruxelles (ULB), Brussels, Belgium

Correspondence to:

Hui Ying Ang, MD, PhD,

Department of Biomedical Engineering,

National University of Singapore,

Singapore 119077, Singapore,

phone: +65 6704 23 43,

e-mail: anghuiyingg@gmail.com

Copyright by the Author(s), 2024

DOI: 10.33963/v.phj.100572

Received: October 19, 2023

Accepted: May 5, 2024

Early publication date: May 16, 2024

INTRODUCTION

Due to complex morphology of bifurcation lesions, they are associated with a higher number of adverse events following stent implantation, such as stent thrombosis and in-stent restenosis [1]. Because of the differences in diameter between the pro­ximal and distal parts of the vessel and the side branch (SB), appropriate implantation techniques, such as proximal optimization (POT) and frequently SB optimization, are necessary to achieve optimal procedural results. In left main (LM) stenting, the provisional, single-stent strategy is associated with outcomes comparable to the more complicated double-stent strategy [2]. Previous meta-analyses of randomized trials demonstrated similar clinical outcomes regardless of SB optimization with the kissing-balloon inflation (KBI) technique [3]. However, up to date no large randomized trials focusing on the impact of final POT after KBI were performed. We previously conducted benchtop tests that demonstrated significant differen­ces in flow disturbances at SB ostia with and without KBI optimization in the LM setting [4]. Therefore, in this pilot human study, we aimed to evaluate the impact of KBI followed by final POT in LM interventions based on optical coherence tomography (OCT) ima­ges and computational fluid dynamics (CFD) reconstructions.

METHODS

This pilot study was conducted in the Division of Cardiology and Structural Heart Diseases at the Medical University of Silesia in Katowice between February and July 2021. Patients’ cli­nical data are provided in the Supplementary material. The study received approval from the Bioethical Committee and adhered to the Declaration of Helsinki. All patients underwent surgery using a provisional stenting technique according to recommendations of the European Society of Cardiology, with a single stent (Xience Sierra, Abbott, Santa Clara, CA, US) positioned from the LM to the left anterior descending artery. The following steps were performed in each patient: predilatation, stent implantation, POT, KBI, and final POT. All balloon inflations were carried out using non-compliant balloons sized according to the distal segment of the target vessel. OCT images were obtained before stenting and after each stage of the procedure using the ILUMEN OPTIS system (Abbott, US). Subsequently, all OCT images were analyzed by a qualified medical staff at 1 mm intervals. Malapposition was defined as a distance between the strut blooming and lumen contour of more than 81 μm. Floating struts were described as all struts observed in the opening angle of the SB. Furthermore, to understand the effects of overhanging struts at the boundary intersection between the main and side branches, CFD analysis was performed. Two-dimensional models were created from the cross-sectional OCT pullbacks using Solidworks. The models underwent simulations using fluid computational software (Fluent, ANSYS). For analysis, a region of interest was created based on the SB diameter intersecting the main branch (MB), encompassing any overhanging struts, in each analysis model [5]. Parameters of interest analyzed in these regions included maximum shear rate (SR) and area of high SR.

Statistical analysis

IBM SPSS Statistics 29 (version 29; IBM, Armonk, NY, US) was used to perform the statistical analysis for exploratory purposes. Results were shown as a mean value (range and standard deviation). The values represent a normal distribution; therefore, they were analyzed with parametric tests (one-way analysis of variance for dependent samples). The differences were considered meaningful if a P-value was <0.05. If the P-value was in the range between 0.05 and 0.10, the differences were considered a statistical tendency.

RESULTS AND DISCUSSIONS

In this pilot study, we aimed to evaluate the effects of POT, KBI, and final POT on morphometric OCT parameters and flow disturbances in the SB ostium using CFD analysis in the LM setting. Because of their morphological comple­xity, bifurcation lesions, especially in the LM region, are prone to a higher number of adverse events [1]. In our study, OCT analysis demonstrated a positive trend in the LM minimal lumen area, starting from 4.99 mm2 (3.586.11; SD, 1.29) before stent implantation to 8.76 mm2 (5.8812.06; SD, 3.11) after POT, 8.78 mm2 (6.8411.6; SD, 2.51) after KBI, and 10.29 mm2 (8.7612.36; SD, 1.86) after final POT; P = 0.064. However, after performing the correction for multiple comparisons, none of the compared pairs differed significantly from each other. We observed a significant difference in the LM minimal stent area (MSA) after each stage, with 8.88 mm2 (6.8312.52; SD, 3.16) after POT, 9.24 mm2 (7.1113.0; SD, 3.27) after KBI, and 11.42 mm2 (9.5213.85; SD, 2.21) after final POT; P <0.05, with positive tendency between POT and KBI (P = 0.093). This is an important finding since a smaller post-implantation minimal lumen area and minimal stent area (indicating stent under-expansion) were found to be independent predictors of stent failure, including ST and in-stent restenosis [6]. The number of malapposed struts registered in each frame in the region of interest decreased numerically after each stage, with no statistical significance: 4.67 (011; SD, 5.69) after POT, 4 (010; SD, 5.29) after KBI, and 2.33 (05; SD, 2.52) after final POT; P = 0.35. The number of floating struts in the SB region decreased significantly: 39 (1855; SD, 19) after POT, 28.33 (1241; SD, 14.84) after KBI, and 26.67 (940; SD, 15.95) after final POT, P <0.05. We also observed a positive trend in floating struts between POT and final-POT (P = 0.06). Previously published bench-test studies demonstrated a correlation between struts protruding into the SB ostium and increased thrombogenicity [4].

The SR was defined as the local gradient in velocity between two surfaces that contain the fluid. Blood flow disruption and the extension of high shear rate regions may be caused by the presence of struts in the SB ostia.

In our study, CFD analysis showed a significant difference in the number of maximum shear rates at SB ostia after stent implantation: 390.07 s-1 (115.5616.7; SD, 254.01), subsequent POT: 1522.1 s-1 (450.53049.6; SD, 1358.2) and KBI: 2370.77 s-1 (1489.53843.3; SD, 1283.47) with a reduction after final POT: 454.1 s-1 (356.4518.6; SD, 86.05), P <0.05, with a positive tendency between POT and KBI (P = 0.08). In areas with SR >1000 s-1, we did not observe a significant difference with 0 mm2 (00; SD, 0) at baseline, 0.02 mm2 (00.07; SD, 0.04) following implantation with subsequent POT, 0.02 mm2 (0.0040.066; SD, 0.04) after KBI, and 0 mm2 (00; SD, 0) after final POT; P = 0.39. Representative images of OCT and CFD are presented in Figure 1.

Figure 1. Representative images. A. Before stent implantation: left OCT 2D, middle OCT 3D, right CFD. B. After POT: left OCT 2D, middle OCT 3D, right CFD. C. After KBI: left OCT 2D, middle OCT 3D, right CFD. D. After final POT: left OCT 2D, middle OCT 3D, right CFD
Abbreviations: CFD, computational fluid dynamics; KBI, kissing-balloon inflation; OCT, optical coherence tomography; POT, proximal optimization

In the distal LM lesions, POT plays a crucial role in the MB optimization proximally to the SB and is mandatory to improve stent expansion and strut apposition [9]. Also, it allows adjusting the stent shape to the fractal geometry of the vessel and correcting the flow dynamics. However, despite the theoretical reduction of floating struts at the SB ostia following KBI, its clinical impact is still debated. Nevertheless, it is important to emphasize that KBI may be responsible for ellipsoid stent distortion of the proximal MB and its excessive overexpansion, which might be associated with higher risk of MB-related adverse events [10]. Finally, data from a randomized trial comparing provisional stenting with or without KBI did not demonstrate any improvements in clinical outcomes for the KBI strategy [3].

A few limitations are important to highlight. First, the number of patients analyzed in this preliminary study was small. Additionally, our study assessed only one DES platform, thus the results might differ with other devices.

In conclusion, based on these preliminary results, we demonstrated that performing KBI without following stent optimization may disrupt stent geometry unless it is followed by final POT, which is associated with nume­rical improvements of OCT morphometric parameters as well as a reduction of floating struts at SB ostia and the area of high shear rate in CFD analysis. This shows the potential of KBI followed by final POT in further redu­cing the number of adverse events in patients with distal LM lesions. Nevertheless, larger studies evaluating the impact of SB optimization are crucial to further explore these preliminary findings.

Article information

Conflict of interest: None declared.

Funding: None.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.

REFERENCES

  1. Iakovou I, Schmidt T, Bonizzoni E, et al. Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents. JAMA. 2005; 293(17): 21262130, doi: 10.1001/jama.293.17.2126, indexed in Pubmed: 15870416.
  2. Hildick-Smith D, Egred M, Banning A, et al. The European bifurcation club left main coronary stent study: a randomized comparison of stepwise provisional vs. systematic dual stenting strategies (EBC MAIN). Eur Heart J. 2021; 42(37): 38293839, doi: 10.1093/eurheartj/ehab283, indexed in Pubmed: 34002215.
  3. Zhong M, Tang B, Zhao Q, et al. Should kissing balloon inflation after main vessel stenting be routine in the one-stent approach? A systematic review and meta-analysis of randomized trials. PLoS One. 2018; 13(6): e0197580, doi: 10.1371/journal.pone.0197580, indexed in Pubmed: 29949587.
  4. Milewski M, Ng CK, Gąsior P, et al. Polymer coating integrity, thrombogenicity and computational fluid dynamics analysis of provisional stenting technique in the left main bifurcation setting: Insights from an in-vitro model. Polymers (Basel). 2022; 14(9): 1715, doi: 10.3390/polym14091715, indexed in Pubmed: 35566886.
  5. Gasior P, Lu S, Ng CK, et al. Comparison of overexpansion capabilities and thrombogenicity at the side branch ostia after implantation of four different drug eluting stents. Sci Rep. 2020; 10(1): 20791, doi: 10.1038/s41598-020-75836-6, indexed in Pubmed: 33247219.
  6. Doi H, Maehara A, Mintz GS, et al. Impact of in-stent minimal lumen area at 9 months poststent implantation on 3-year target lesion revascularization-free survival: a serial intravascular ultrasound analysis from the TAXUS IV, V, and VI trials. Circ Cardiovasc Interv. 2008; 1(2): 111118, doi: 10.1161/CIRCINTERVENTIONS.108.784660, indexed in Pubmed: 20031665.
  7. Lu S, Ng J, Ang H, et al. Is there light at the end of the thin-strut tunnel?: In vitro insights on strut thickness impact on thrombogenicity in bio­resorbable stents or scaffolds. JACC Cardiovasc Interv. 2018; 11(7): 714716, doi: 10.1016/j.jcin.2018.01.237, indexed in Pubmed: 29622153.
  8. Kumar S, Molony D, Khawaja S, et al. Stent underexpansion is associated with high wall shear stress: a biomechanical analysis of the shear stent study. Int J Cardiovasc Imaging. 2023; 39(7): 13751382, doi: 10.1007/s10554-023-02838-6, indexed in Pubmed: 37119348.
  9. Burzotta F, Lassen JF, Louvard Y, et al. European Bifurcation Club white paper on stenting techniques for patients with bifurcated coronary artery lesions. Catheter Cardiovasc Interv. 2020; 96(5): 10671079, doi: 10.1002/ccd.29071, indexed in Pubmed: 32579300.
  10. Mortier P, Hikichi Y, Foin N, et al. Provisional stenting of coronary bifurcations: Insights into final kissing balloon post-dilation and stent design by computational modeling. JACC Cardiovasc Interv. 2014; 7(3): 325333, doi: 10.1016/j.jcin.2013.09.012, indexed in Pubmed: 24650404.