Vol 3, No 2 (1998)
Original papers
Published online: 1998-01-01

open access

Page views 131
Article views/downloads 135
Get Citation

Connect on Social Media

Connect on Social Media

Verification of doses in electron beam radiotherapy

W. Łobodziec1, M. Ganowicz1, A. Orlef1, Z. Maniakowski1, A. Konefał1
DOI: 10.1016/S1507-1367(98)70162-2
Rep Pract Oncol Radiother 1998;3(2):25-26.

Abstract

Entrance doses measured in 50 patients treated by electron beams were compared with planned values as the percentage difference between the given dose and its planned value. Electron beams of energy 6, 9, 12, 15, 18 and 22 MeV were generated by the Clinac 2300 C/D linear accelerator (Varian). Semiconductor EDE-5 diodes connected to a DPD-510 (Scanditronix) electrometer were applied for entrance dose measurements. In our investigations, the mean values of doses administered agreed well with the values of planned doses. The observed relatively large discrepancies in a few patients (up to 12%) should be carefully investigated.

Article available in PDF format

View PDF Download PDF file



Reports of Practical Oncology and Radiotherapy