Vol 6, No 3 (2001)
Original papers
Published online: 2001-01-01

open access

Page views 156
Article views/downloads 146
Get Citation

Connect on Social Media

Connect on Social Media

Verification of geometry reconstruction and dose calculation modules of the PLATO radiotherapy planing system

Eleonora Góra1, Jan Lesiak1, Remigiusz Barańczyk1, Bożena Rozwadowska-Bogusz1, Michał Waligórski12
DOI: 10.1016/S1507-1367(01)70470-1
Rep Pract Oncol Radiother 2001;6(3):141-147.

Abstract

Purpose

Verification of reconstruction of application geometry and of calculations of dose distribution in the PLATO v. 13.7 system for brachytherapy treatment planning.

Materials and methods

The correctness of geometry reconstruction was verified using a Perspex phantom with lead markers. Dose calculations were verified by comparing values calculated by the system against values calculated „manually” and against published benchmark values.

Results

The average difference between marker co-ordinates in the phantom, as measured mechanically, and as reconstructed by the PLATO system, did not exceed 0.4 mm. The maximum relative difference between values of dose at given points, calculated by the PLATO system and values of dose calculated „manually” at those points, did not exceed 0.3%. The PLATO system was found to correctly reproduce the published benchmark data.

Conclusions

The geometry reconstruction software module unit and the dose distribution algorithm of the PLATO v. 13.7 planning system used at the Centre of Oncology in Kraków, were found to operate correctly.

Article available in PDF format

View PDF Download PDF file