open access
Verification of geometry reconstruction and dose calculation modules of the PLATO radiotherapy planing system
open access
Abstract
Purpose
Verification of reconstruction of application geometry and of calculations of dose distribution in the PLATO v. 13.7 system for brachytherapy treatment planning.
Materials and methods
The correctness of geometry reconstruction was verified using a Perspex phantom with lead markers. Dose calculations were verified by comparing values calculated by the system against values calculated „manually” and against published benchmark values.
Results
The average difference between marker co-ordinates in the phantom, as measured mechanically, and as reconstructed by the PLATO system, did not exceed 0.4 mm. The maximum relative difference between values of dose at given points, calculated by the PLATO system and values of dose calculated „manually” at those points, did not exceed 0.3%. The PLATO system was found to correctly reproduce the published benchmark data.
Conclusions
The geometry reconstruction software module unit and the dose distribution algorithm of the PLATO v. 13.7 planning system used at the Centre of Oncology in Kraków, were found to operate correctly.
Abstract
Purpose
Verification of reconstruction of application geometry and of calculations of dose distribution in the PLATO v. 13.7 system for brachytherapy treatment planning.
Materials and methods
The correctness of geometry reconstruction was verified using a Perspex phantom with lead markers. Dose calculations were verified by comparing values calculated by the system against values calculated „manually” and against published benchmark values.
Results
The average difference between marker co-ordinates in the phantom, as measured mechanically, and as reconstructed by the PLATO system, did not exceed 0.4 mm. The maximum relative difference between values of dose at given points, calculated by the PLATO system and values of dose calculated „manually” at those points, did not exceed 0.3%. The PLATO system was found to correctly reproduce the published benchmark data.
Conclusions
The geometry reconstruction software module unit and the dose distribution algorithm of the PLATO v. 13.7 planning system used at the Centre of Oncology in Kraków, were found to operate correctly.
Keywords
Quality Assurance; Brachytherapy; Treatment Planning; PLATO; Selectron; microSelectron


Title
Verification of geometry reconstruction and dose calculation modules of the PLATO radiotherapy planing system
Journal
Reports of Practical Oncology and Radiotherapy
Issue
Pages
141-147
Published online
2001-01-01
DOI
10.1016/S1507-1367(01)70470-1
Bibliographic record
Rep Pract Oncol Radiother 2001;6(3):141-147.
Keywords
Quality Assurance
Brachytherapy
Treatment Planning
PLATO
Selectron
microSelectron
Authors
Eleonora Góra
Jan Lesiak
Remigiusz Barańczyk
Bożena Rozwadowska-Bogusz
Michał Waligórski