Vol 25, No 4 (2020)
Original research articles
Published online: 2020-07-01

open access

Page views 227
Article views/downloads 286
Get Citation

Connect on Social Media

Connect on Social Media

Software simulation of tumour motion dose effects during flattened and unflattened ITV-based VMAT lung SBRT

Marta Adamczyk1, Marta Kruszyna-Mochalska12, Anna Rucińska3, Tomasz Piotrowski12
DOI: 10.1016/j.rpor.2020.06.003
Rep Pract Oncol Radiother 2020;25(4):684-691.

Abstract

Purpose

Restricted studies comparing different dose rate parameters are available while ITV-based VMAT lung SBRT planning leads to perform the analysis of the most suitable parameters of the external beams used. The special emphasis was placed on the impact of dose rate on dose distribution variations in target volumes due to interplay effects.

Methods

Four VMAT plans were calculated for 15 lung tumours using 6 MV photon beam quality (flattening filter FF vs. flattening filter free FFF beams) and maximum dose rate of 600 MU/min, 1000 MU/min and 1400 MU/min. Three kinds of motion simulations were performed finally giving 180 plans with perturbed dose distributions.

Results

6FFF-1400 MUs/min plans were characterized by the shortest beam on time (1.8 ± 0.2 min). Analysing the performed motion simulation results, the mean dose (Dmean) is not a sensitive parameter to related interplay effects. Looking for local maximum and local minimum doses, some discrepancies were found, but their significance was presented for individual patients, not for the whole cohort. The same was observed for other verified dose metrics.

Conclusions

Generally, the evaluation of VMAT robustness between FF and FFF concepts against interplay effect showed a negligible effect of simulated motion influence on tumour coverage among different photon beam quality parameters. Due to the lack of FFF beams, smaller radiotherapy centres are able to perform ITV-based VMAT lung SBRT treatment in a safe way. Radiotherapy department having FFF beams could perform safe, fast and efficient ITV-based VMAT lung SBRT without a concern about significance of interplay effects.

Article available in PDF format

View PDF Download PDF file



Reports of Practical Oncology and Radiotherapy