open access

Vol 25, No 2 (2020)
Original research articles
Published online: 2020-03-01
Submitted: 2019-02-25
Get Citation

EGSnrc application for IMRT planning

Sitti Yani, Ilmi Rizkia, - Kamirul, Mohamad Fahdillah Rhani, Mohammad Haekal, Freddy Haryanto
DOI: 10.1016/j.rpor.2020.01.004
·
Rep Pract Oncol Radiother 2020;25(2):217-226.

open access

Vol 25, No 2 (2020)
Original research articles
Published online: 2020-03-01
Submitted: 2019-02-25

Abstract

The aim of this study was to describe a detailed instruction of intensity modulated radiotherapy (IMRT) planning simulation using BEAMnrc-DOSXYZnrc code system (EGSnrc package) and present a new graphical user interface based on MATLAB code (The MathWorks) to combine more than one. 3ddose file which were obtained from the IMRT plan.

This study was performed in four phases: the commissioning of Varian Clinac iX6MV, the simulation of IMRT planning in EGSnrc, the creation of in-house VDOSE GUI, and the analysis of the isodose contour and dose volume histogram (DVH) curve from several beam angles. The plan paramaters in sequence and control point files were extracted from the planning data in Tan Tock Seng Hospital Singapore (multileaf collimator (MLC) leaf positions – bank A and bank B, gantry angles, coordinate of isocenters, and MU indexes).

VDOSE GUI which was created in this study can display the distribution dose curve in each slice and beam angle. Dose distributions from various MLC settings and beam angles yield different dose distributions even though they used the same number of simulated particles. This was due to the differences in the MLC leaf openings in every field. The value of the relative dose error between the two dose ditributions for “body” was 51.23 %. The Monte Carlo (MC) data was normalized with the maximum dose but the analytical anisotropic algorithm (AAA) data was normalized by the dose in the isocenter.

In this study, we have presented a Monte Carlo simulation framework for IMRT dose calculation using DOSXYZnrc source 21. Further studies are needed in conducting IMRT simulations using EGSnrc to minimize the different dose error and dose volume histogram deviation.

Abstract

The aim of this study was to describe a detailed instruction of intensity modulated radiotherapy (IMRT) planning simulation using BEAMnrc-DOSXYZnrc code system (EGSnrc package) and present a new graphical user interface based on MATLAB code (The MathWorks) to combine more than one. 3ddose file which were obtained from the IMRT plan.

This study was performed in four phases: the commissioning of Varian Clinac iX6MV, the simulation of IMRT planning in EGSnrc, the creation of in-house VDOSE GUI, and the analysis of the isodose contour and dose volume histogram (DVH) curve from several beam angles. The plan paramaters in sequence and control point files were extracted from the planning data in Tan Tock Seng Hospital Singapore (multileaf collimator (MLC) leaf positions – bank A and bank B, gantry angles, coordinate of isocenters, and MU indexes).

VDOSE GUI which was created in this study can display the distribution dose curve in each slice and beam angle. Dose distributions from various MLC settings and beam angles yield different dose distributions even though they used the same number of simulated particles. This was due to the differences in the MLC leaf openings in every field. The value of the relative dose error between the two dose ditributions for “body” was 51.23 %. The Monte Carlo (MC) data was normalized with the maximum dose but the analytical anisotropic algorithm (AAA) data was normalized by the dose in the isocenter.

In this study, we have presented a Monte Carlo simulation framework for IMRT dose calculation using DOSXYZnrc source 21. Further studies are needed in conducting IMRT simulations using EGSnrc to minimize the different dose error and dose volume histogram deviation.

Get Citation

Keywords

Monte carlo; EGSnrc; IMRT planning; VDOSE

About this article
Title

EGSnrc application for IMRT planning

Journal

Reports of Practical Oncology and Radiotherapy

Issue

Vol 25, No 2 (2020)

Pages

217-226

Published online

2020-03-01

DOI

10.1016/j.rpor.2020.01.004

Bibliographic record

Rep Pract Oncol Radiother 2020;25(2):217-226.

Keywords

Monte carlo
EGSnrc
IMRT planning
VDOSE

Authors

Sitti Yani
Ilmi Rizkia
- Kamirul
Mohamad Fahdillah Rhani
Mohammad Haekal
Freddy Haryanto

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.