Detailed analysis of dose difference in using water as tissue-equivalent material in Cf brachytherapy

Gholamhossein Izadi Vasafi, Mohammad Mehdi Firoozabadi, Mahdi Ghorbani
DOI: 10.1016/j.rpor.2019.10.002
·
Rep Pract Oncol Radiother 2019;24(6):660-666.

open access

Vol 24, No 6 (2019)
Original research articles
Published online: 2019-11-01
Submitted: 2019-01-20

Abstract

Aim

The purpose of this study is to analyse how small variations in the elemental composition of soft tissue lead to differences in dose distributions from a 252Cf brachytherapy source and to determine the error percentage in using water as a tissue-equivalent material.

Background

Water is normally used as a tissue-equivalent phantom material in radiotherapy dosimetry.

Materials and methods

Neutron energy spectra, neutron and gamma-ray dose rate distributions were calculated for a 252Cf AT source located at the center of a spherical phantom filled with various types of tissue compositions: adipose, brain, muscle, International Commission on Radiation Units and Measurements (ICRU) report No. 44 9-component soft tissue and water, using Monte Carlo simulation.

Results

The obtained results showed differences between total dose rates in various tissues relative to water varying between zero and 4.94%. The contributions of neutron and total gamma ray doses to these differences are, on average, 81% and 19%, respectively. It was found that the dose differences between various soft tissues and water depend not only on the soft tissue composition, but also on the beam type emitted from the 252Cf source and the distance from the source.

Conclusion

Assuming water as a tissue-equivalent material, although leads to overestimation of dose rate (except in the case of adipose tissue), is acceptable and suitable for use in 252Cf brachytherapy treatment planning systems based on the recommendation by the ICRU that the uncertainties in dose delivery in radiotherapy should be lower than 5%.

Abstract

Aim

The purpose of this study is to analyse how small variations in the elemental composition of soft tissue lead to differences in dose distributions from a 252Cf brachytherapy source and to determine the error percentage in using water as a tissue-equivalent material.

Background

Water is normally used as a tissue-equivalent phantom material in radiotherapy dosimetry.

Materials and methods

Neutron energy spectra, neutron and gamma-ray dose rate distributions were calculated for a 252Cf AT source located at the center of a spherical phantom filled with various types of tissue compositions: adipose, brain, muscle, International Commission on Radiation Units and Measurements (ICRU) report No. 44 9-component soft tissue and water, using Monte Carlo simulation.

Results

The obtained results showed differences between total dose rates in various tissues relative to water varying between zero and 4.94%. The contributions of neutron and total gamma ray doses to these differences are, on average, 81% and 19%, respectively. It was found that the dose differences between various soft tissues and water depend not only on the soft tissue composition, but also on the beam type emitted from the 252Cf source and the distance from the source.

Conclusion

Assuming water as a tissue-equivalent material, although leads to overestimation of dose rate (except in the case of adipose tissue), is acceptable and suitable for use in 252Cf brachytherapy treatment planning systems based on the recommendation by the ICRU that the uncertainties in dose delivery in radiotherapy should be lower than 5%.

Get Citation

Keywords

Brachytherapy; 252Cf; Tissue composition; Dose distribution

About this article
Title

Detailed analysis of dose difference in using water as tissue-equivalent material in Cf brachytherapy

Journal

Reports of Practical Oncology and Radiotherapy

Issue

Vol 24, No 6 (2019)

Pages

660-666

Published online

2019-11-01

DOI

10.1016/j.rpor.2019.10.002

Bibliographic record

Rep Pract Oncol Radiother 2019;24(6):660-666.

Keywords

Brachytherapy
252Cf
Tissue composition
Dose distribution

Authors

Gholamhossein Izadi Vasafi
Mohammad Mehdi Firoozabadi
Mahdi Ghorbani

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: journals@viamedica.pl