Evaluation of plan quality improvements in PlanIQ-guided Autoplanning

Bojarajan Perumal, Harikrishna Etti Sundaresan, Vaitheeswaran Ranganathan, Natarajan Ramar, Gipson Joe Anto, Samir Ranjan Meher
DOI: 10.1016/j.rpor.2019.08.003
·
Rep Pract Oncol Radiother 2019;24(6):533-543.

open access

Vol 24, No 6 (2019)
Original research articles
Published online: 2019-11-01
Submitted: 2019-02-26

Abstract

Aim

Philips recently integrated PlanIQ with Autoplan® in Pinnacle3 TPS (V16.2). The objective of the present work is to quantitatively demonstrate how this integration improves the plan quality.

Background

Pinnacle3 Autoplan® is the tool that generates the treatment plans with clinically acceptable plan quality with less manual intervention. In the recent past, a new tool called PlanIQ (Sun Nuclear Corp.) was introduced for a priori estimation of the best possible sparing of an organ at risk (OAR) for a given patient anatomy. Philips has recently integrated PlanIQ tool with Autoplan® for a seamless and efficient planning workflow.

Materials and methods

We have performed this evaluation in Pinnacle3 TPS (V.16.2) for the VMAT treatment technique. All plans were created using Varian True beam machine with the dual arc technique. Basically, we created two sets of VMAT plans using 6MV photons. In the first set of VMAT plans (AP_RTOG), we used OAR goals from either RTOG guidelines to perform optimization using Autoplan®. Subsequently, we exported the same dataset to the PlanIQ system to perform feasibility analysis on the OAR goals. These newly obtained OAR goals from PlanIQ were used to generate the other set of plans (AP_PlanIQ plans). We compared the dosimetric results from these two sets of plans in five cases, such as brain, head & neck, lung, abdomen and prostate.

Results

We compared the dosimetric results for AP_RTOG and AP_PlanIQ plans. We used RTOG guidelines to evaluate the plans and observed that while both sets of plans were meeting the RTOG guidelines in terms of OAR sparing, the AP_PlanIQ plans were significantly better in terms of OAR sparing as compared to AP_RTOG plans without any compromise in the target coverage.

Conclusion

The results indicate that, although Autoplan helps achieve the user-defined goals without much manual intervention, the plan quality (OAR sparing) can be significantly improved without taking many iterative steps when PlanIQ suggested clinical goals are used in the Autoplan-based optimization.

Advances in knowledge

At present, there are no published material available about the efficacy of the integration of PlanIQ with Autoplanning®. In the present work, our objective is to evaluate the improvements in plan quality resulting from this integration.

Abstract

Aim

Philips recently integrated PlanIQ with Autoplan® in Pinnacle3 TPS (V16.2). The objective of the present work is to quantitatively demonstrate how this integration improves the plan quality.

Background

Pinnacle3 Autoplan® is the tool that generates the treatment plans with clinically acceptable plan quality with less manual intervention. In the recent past, a new tool called PlanIQ (Sun Nuclear Corp.) was introduced for a priori estimation of the best possible sparing of an organ at risk (OAR) for a given patient anatomy. Philips has recently integrated PlanIQ tool with Autoplan® for a seamless and efficient planning workflow.

Materials and methods

We have performed this evaluation in Pinnacle3 TPS (V.16.2) for the VMAT treatment technique. All plans were created using Varian True beam machine with the dual arc technique. Basically, we created two sets of VMAT plans using 6MV photons. In the first set of VMAT plans (AP_RTOG), we used OAR goals from either RTOG guidelines to perform optimization using Autoplan®. Subsequently, we exported the same dataset to the PlanIQ system to perform feasibility analysis on the OAR goals. These newly obtained OAR goals from PlanIQ were used to generate the other set of plans (AP_PlanIQ plans). We compared the dosimetric results from these two sets of plans in five cases, such as brain, head & neck, lung, abdomen and prostate.

Results

We compared the dosimetric results for AP_RTOG and AP_PlanIQ plans. We used RTOG guidelines to evaluate the plans and observed that while both sets of plans were meeting the RTOG guidelines in terms of OAR sparing, the AP_PlanIQ plans were significantly better in terms of OAR sparing as compared to AP_RTOG plans without any compromise in the target coverage.

Conclusion

The results indicate that, although Autoplan helps achieve the user-defined goals without much manual intervention, the plan quality (OAR sparing) can be significantly improved without taking many iterative steps when PlanIQ suggested clinical goals are used in the Autoplan-based optimization.

Advances in knowledge

At present, there are no published material available about the efficacy of the integration of PlanIQ with Autoplanning®. In the present work, our objective is to evaluate the improvements in plan quality resulting from this integration.

Get Citation

Keywords

IMRT; Autoplan®; PlanIQ; Priori-Estimation of objectives; Optimization; VMAT

About this article
Title

Evaluation of plan quality improvements in PlanIQ-guided Autoplanning

Journal

Reports of Practical Oncology and Radiotherapy

Issue

Vol 24, No 6 (2019)

Pages

533-543

Published online

2019-11-01

DOI

10.1016/j.rpor.2019.08.003

Bibliographic record

Rep Pract Oncol Radiother 2019;24(6):533-543.

Keywords

IMRT
Autoplan®
PlanIQ
Priori-Estimation of objectives
Optimization
VMAT

Authors

Bojarajan Perumal
Harikrishna Etti Sundaresan
Vaitheeswaran Ranganathan
Natarajan Ramar
Gipson Joe Anto
Samir Ranjan Meher

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: journals@viamedica.pl