Dosimetric impact of uncorrected systematic yaw rotation in VMAT for peripheral lung SABR
Abstract
Aim
This study aimed to evaluate the dosimetric impact of uncorrected yaw rotational error on both target coverage and OAR dose metrics in this patient population.
Background
Rotational set up errors can be difficult to correct in lung VMAT SABR treatments, and may lead to a change in planned dose distributions.
Materials and methods
We retrospectively applied systematic yaw rotational errors in 1° degree increments up to −5° and +5° degrees in 16 VMAT SABR plans. The impact on PTV and OARs (oesophagus, spinal canal, heart, airway, chest wall, brachial plexus, lung) was evaluated using a variety of dose metrics. Changes were assessed in relation to percentage deviation from approved planned dose at 0 degrees.
Results
Target coverage was largely unaffected with the largest mean and maximum percentage difference being 1.4% and 6% respectively to PTV D98% at +5 degrees yaw.
Impact on OARs was varied. Minimal impact was observed in oesophagus, spinal canal, chest wall or lung dose metrics. Larger variations were observed in the heart, airway and brachial plexus. The largest mean and maximum percentage differences being 20.77% and 311% respectively at −5 degrees yaw to airway D0.1cc, however, the clinical impact was negligible as these variations were observed in metrics with minimal initial doses.
Conclusions
No clinically unacceptable changes to dose metrics were observed in this patient cohort but large percentage deviations from approved dose metrics in OARs were noted. OARs with associated PRV structures appear more robust to uncorrected rotational error.
Keywords: Lung SABRVMATSetup errorRotational error