Sensitivity of the IQM and MatriXX detectors in megavolt photon beams

Oluwaseyi Michael Oderinde, Freek du Plessis
DOI: 10.1016/j.rpor.2019.07.007
·
Rep Pract Oncol Radiother 2019;24(5):462-471.

open access

Vol 24, No 5 (2019)
Original research articles
Published online: 2019-09-01
Submitted: 2019-01-18

Abstract

Aim

This study focused on evaluating the sensitivity of integral quality monitoring (IQM®) system and MatriXX detectors. These two detectors are recommended for radiotherapy pre-treatment quality assurance (QA).

Background

IQM is a large wedged-shaped ionisation chamber mounted to the linear accelerator (linac) head in practice. MatriXX consists of an array of ionisation chambers also attached to the linac head.

Materials and methods

In this study, the dosimetric performance and sensitivity of MatriXX and IQM detectors were evaluated using the following characteristics: reproducibility, linearity, error detection capability and three-dimensional conformal radiotherapy (3D-CRT) plans of the head and neck, thorax and pelvic regions.

Results

This study indicates that the signal responses of the large ionisation chamber device (IQM) and the small pixel array of ionisation chambers device (MatriXX) are reproducible, linear and sensitive to MLC positional errors, backup jaw positional errors and dose errors. The local percentage differences for dose errors of 1%, 2%, and 3% were, respectively, within 0.35–8.23%, 0.78–16.21%, and 1.10–24.41% for the IQM device. While for the MatriXX detector, the ranges were between 0.24–3.19, 0.57–6.43 and 0.81–12.95, respectively. Since IQM is essentially a double wedge-shaped large ionisation chamber, its reproducibility and detection capability are competitive to that of MatriXX. In addition, the sensitivity of the two QA systems increases with an increase in escalation percentage, and the signal responses are patient plan specific.

Conclusions

The two detectors response signals have good correlations and they are accurate for pre-treatment QA. Statistically, (P<0.05) there is a significant difference between the IQM and MatriXX response to dose errors.

Abstract

Aim

This study focused on evaluating the sensitivity of integral quality monitoring (IQM®) system and MatriXX detectors. These two detectors are recommended for radiotherapy pre-treatment quality assurance (QA).

Background

IQM is a large wedged-shaped ionisation chamber mounted to the linear accelerator (linac) head in practice. MatriXX consists of an array of ionisation chambers also attached to the linac head.

Materials and methods

In this study, the dosimetric performance and sensitivity of MatriXX and IQM detectors were evaluated using the following characteristics: reproducibility, linearity, error detection capability and three-dimensional conformal radiotherapy (3D-CRT) plans of the head and neck, thorax and pelvic regions.

Results

This study indicates that the signal responses of the large ionisation chamber device (IQM) and the small pixel array of ionisation chambers device (MatriXX) are reproducible, linear and sensitive to MLC positional errors, backup jaw positional errors and dose errors. The local percentage differences for dose errors of 1%, 2%, and 3% were, respectively, within 0.35–8.23%, 0.78–16.21%, and 1.10–24.41% for the IQM device. While for the MatriXX detector, the ranges were between 0.24–3.19, 0.57–6.43 and 0.81–12.95, respectively. Since IQM is essentially a double wedge-shaped large ionisation chamber, its reproducibility and detection capability are competitive to that of MatriXX. In addition, the sensitivity of the two QA systems increases with an increase in escalation percentage, and the signal responses are patient plan specific.

Conclusions

The two detectors response signals have good correlations and they are accurate for pre-treatment QA. Statistically, (P<0.05) there is a significant difference between the IQM and MatriXX response to dose errors.

Get Citation

Keywords

Sensitivity; Quality assurance; IQM detector; MatriXX detector

About this article
Title

Sensitivity of the IQM and MatriXX detectors in megavolt photon beams

Journal

Reports of Practical Oncology and Radiotherapy

Issue

Vol 24, No 5 (2019)

Pages

462-471

Published online

2019-09-01

DOI

10.1016/j.rpor.2019.07.007

Bibliographic record

Rep Pract Oncol Radiother 2019;24(5):462-471.

Keywords

Sensitivity
Quality assurance
IQM detector
MatriXX detector

Authors

Oluwaseyi Michael Oderinde
Freek du Plessis

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: journals@viamedica.pl