On the use of AAA and AcurosXB algorithms for three different stereotactic ablative body radiotherapy (SABR) techniques: Volumetric modulated arc therapy (VMAT), intensity modulated radiation therapy (IMRT) and 3D conformal radiotherapy (3D-CRT)

Abdulrahman Tajaldeen123, Prabhakar Ramachandran12, Salem Alghamdi3, Moshi Geso1
DOI: 10.1016/j.rpor.2019.02.008
Rep Pract Oncol Radiother 2019;24(4):399-408.

Abstract

Aim

The purpose of this study was to investigate the dosimetric characteristics of three stereotactic ablative body radiotherapy (SABR) techniques using the anisotropic analytical algorithm (AAA) and Acuros XB algorithm. The SABR techniques include coplanar volumetric modulated arc therapy (C-VMAT), non-coplanar intensity modulated radiation therapy (NC-IMRT) and non-coplanar three-dimensional conformal radiotherapy (NC-3D CRT).

Background

SABR is a special type of radiotherapy where a high dose of radiation is delivered over a short time. The treatment outcome and accuracy of the dose delivered to cancer patients highly depend on the dose calculation algorithm and treatment technique.

Materials and methods

Twelve lung cancer patients underwent 4D CT scanning, and three different treatment plans were generated: C-VMAT, NC-IMRT, NC-3D CRT. Dose calculation was performed using the AAA and Acuros XB algorithm. The dosimetric indices, such as conformity index (CI), homogeneity index, dose fall-off index, doses received by organs at risk and planning target volume, were used to compare the plans. The accuracy of AAA and Acuros XB (AXB) algorithms for the lung was validated against measured dose on a CIRS thorax phantom.

Results

The CIs for C-VMAT, NC-IMRT and NC-3D CRT were 1.21, 1.28 and 1.38 for the AAA, respectively, and 1.17, 1.26 and 1.36 for the Acuros XB algorithm, respectively. The overall dose computed by AcurosXB algorithm was close to the measured dose when compared to the AAA algorithm. The overall dose computed by the AcurosXB algorithm was close to the measured dose when compared to the AAA algorithm.

Conclusion

This study showed that the treatment planning results obtained using the Acuros XB algorithm was better than those using the AAA algorithm in SABR lung radiotherapy.