Vol 21, No 1 (2016)
Original research articles
Published online: 2016-01-01

open access

Page views 301
Article views/downloads 405
Get Citation

Connect on Social Media

Connect on Social Media

Analyzing the performance of ArcCHECK diode array detector for VMAT plan

Rajesh Thiyagarajan12, Arunai Nambiraj2, Sujit Nath Sinha3, Girigesh Yadav3, Ashok Kumar3, Vikraman Subramani1, - Kothandaraman3
DOI: 10.1016/j.rpor.2015.10.004
Rep Pract Oncol Radiother 2016;21(1):50-56.

Abstract

Aim

The aim of this study is to evaluate performance of ArcCHECK diode array detector for the volumetric modulated arc therapy (VMAT) patient specific quality assurance (QA). VMAT patient specific QA results were correlated with ion chamber measurement. Dose response of the ArcCHECK detector was studied.

Background

VMAT delivery technique improves the dose distribution. It is complex in nature and requires proper QA before its clinical implementation. ArcCHECK is a novel three dimensional dosimetry system.

Materials and methods

Twelve retrospective VMAT plans were calculated on ArcCHECK phantom. Point dose and dose map were measured simultaneously with ion chamber (IC-15) and ArcCHECK diode array detector, respectively. These measurements were compared with their respective TPS calculated values.

Results

The ion chamber measurements are in good agreement with TPS calculated doses. Mean difference between them is 0.50% with standard deviation of 0.51%. Concordance correlation coefficient (CCC) obtained for ion chamber measurements is 0.9996. These results demonstrate a strong correlation between the absolute dose predicted by our TPS and the measured dose. The CCC between ArcCHECK doses and TPS predictions on the CAX was found to be 0.9978. In gamma analysis of dose map, the mean passing rate was 98.53% for 3% dose difference and 3[[ce:hsp sp="0.25"/]]mm distance to agreement.

Conclusions

The VMAT patient specific QA with an ion chamber and ArcCHECK phantom are consistent with the TPS calculated dose. Statistically good agreement was observed between ArcCHECK measured and TPS calculated. Hence, it can be used for routine VMAT QA.

Article available in PDF format

View PDF Download PDF file



Reports of Practical Oncology and Radiotherapy