Assessment and evaluation of MV image guidance system performance in radiotherapy

Nithya Kanakavelu, E. James Samuel
DOI: 10.1016/j.rpor.2015.01.002
·
Rep Pract Oncol Radiother 2015;20(3):188-197.

open access

Vol 20, No 3 (2015)
Original research articles
Published online: 2015-05-01
Submitted: 2014-05-22

Abstract

Background and aim

The clinical use of imaging system in image guided radiotherapy (IGRT) necessitates performing periodic quality assurance of the system to be confident in applying corrections for patient set-up errors. We aim to develop and implement a quality assurance (QA) programme for megavoltage (MV) based image guidance system and assess its long term performance for a period of 3 years.

Materials and methods

Periodic QA tests were performed for the MV planar and cone beam computed tomography (CBCT) imaging system to assess the system safety, mechanical and geometrical accuracy, image quality and dose. The tests were performed using the equipment supplied by the manufacturer along with the image guidance system and using simple methods developed in-house. The test results were compared with expected or baseline values established during commissioning.

Results

The safety system was found to be functional. The results of mechanical and geometrical tests were in good agreement with the expected results. The system mechanical positioning was stable and reproducible within ±2[[ce:hsp sp="0.25"/]]mm accuracy. The image quality and the imaging dose of the planar and CBCT imaging were found to agree with the baseline values and the manufacturer specifications.

Discussion

Throughout the three-year period, all the QA tests were within the specification. The mechanical and geometrical tests are most crucial as they directly affect the patient positioning accuracy.

Conclusion

We conclude that the MV image guidance system is efficient to perform IGRT and insist to perform periodic QA tests and calibration for the system.

Abstract

Background and aim

The clinical use of imaging system in image guided radiotherapy (IGRT) necessitates performing periodic quality assurance of the system to be confident in applying corrections for patient set-up errors. We aim to develop and implement a quality assurance (QA) programme for megavoltage (MV) based image guidance system and assess its long term performance for a period of 3 years.

Materials and methods

Periodic QA tests were performed for the MV planar and cone beam computed tomography (CBCT) imaging system to assess the system safety, mechanical and geometrical accuracy, image quality and dose. The tests were performed using the equipment supplied by the manufacturer along with the image guidance system and using simple methods developed in-house. The test results were compared with expected or baseline values established during commissioning.

Results

The safety system was found to be functional. The results of mechanical and geometrical tests were in good agreement with the expected results. The system mechanical positioning was stable and reproducible within ±2[[ce:hsp sp="0.25"/]]mm accuracy. The image quality and the imaging dose of the planar and CBCT imaging were found to agree with the baseline values and the manufacturer specifications.

Discussion

Throughout the three-year period, all the QA tests were within the specification. The mechanical and geometrical tests are most crucial as they directly affect the patient positioning accuracy.

Conclusion

We conclude that the MV image guidance system is efficient to perform IGRT and insist to perform periodic QA tests and calibration for the system.

Get Citation

Keywords

Image guidance system; Quality assurance; Megavoltage imaging

About this article
Title

Assessment and evaluation of MV image guidance system performance in radiotherapy

Journal

Reports of Practical Oncology and Radiotherapy

Issue

Vol 20, No 3 (2015)

Pages

188-197

Published online

2015-05-01

DOI

10.1016/j.rpor.2015.01.002

Bibliographic record

Rep Pract Oncol Radiother 2015;20(3):188-197.

Keywords

Image guidance system
Quality assurance
Megavoltage imaging

Authors

Nithya Kanakavelu
E. James Samuel

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: journals@viamedica.pl