Performance characteristics and commissioning of MOSFET as an in-vivo dosimeter for high energy photon external beam radiation therapy

A. GOPIRAJ, RAMESH S. BILLIMAGGA, Velayudham RAMASUBRAMANIAN
DOI: 10.1016/S1507-1367(10)60001-6
·
Rep Pract Oncol Radiother 2008;13(3):114-125.

open access

Vol 13, No 3 (2008)
Untitled
Published online: 2008-05-01
Submitted: 2008-04-01

Abstract

Aim

In vivo dosimetry is an essential tool of quality assurance programmes in radiotherapy. In fact, the assessment of the final uncertainty between the prescribed dose and the dose actually delivered to the patient is an effective way of checking the entire dosimetric procedure. Metal oxide semiconductor field effect transistors (MOSFETs) have recently been proposed for use in radiation therapy. The purpose of this work is to study the performance characteristics and to carry out the commissioning of MOSFET as an in-vivo dosimeter for high-energy photon external beam radiation therapy.

Material and Methods

Characterization and commissioning of low sensitivity TN502RD and high sensitivity TN1002RD MOSFETs for entrance and exit dosimetry respectively for application in in-vivo dosimetry in radiotherapy was carried out. The MOSFETs were characterized in terms of reproducibility, short-term constancy, long-term constancy, linearity, angular dependence, energy dependence, source to skin distance (SSD) dependence and field size dependence.

Results

The reproducibility of standard sensitivity MOSFET is about 1.4% (1 SD) and 1.98% (1 SD) for high sensitivity detectors. The linearity of both MOSFETs was excellent (R2 = 0.996). The response of MOSFETs varies linearly for square fields from 3 × 3 cm2 to 30 × 30 cm2. For beam incidence ranging from ±45° the MOSFET response varies within ±3%. Commissioning of both MOSFETs was carried out in terms of entrance dose calibration factor, exit dose calibration factor, SSD correction factor, field size correction factor, wedge correction factor and shielding tray correction factor. The average calibration factor for low and high sensitivity MOSFET detectors is 0.9065 cGy/mV and 0.3412 cGy/mV respectively. The average SSD correction factors are quite small and vary between 0.968 and 1.027 for both types of detectors for the range of clinical SSDs from 80 cm to 120 cm. The field size correction factor varies from 1.00 to 1.02 for both types of detectors. The wedge and the shielding tray correction factors for both the detectors also show quite small variation. MOSFET characteristics are suitable for in vivo dosimetry of entrance and exit dose measurement relevant to 6 MV treatment.

Conclusion

It can be concluded that MOSFET dosimetry's low energy dependence, high sensitivity and immediate readout make it a good replacement for TLD in radiation therapy dosimetry.

Abstract

Aim

In vivo dosimetry is an essential tool of quality assurance programmes in radiotherapy. In fact, the assessment of the final uncertainty between the prescribed dose and the dose actually delivered to the patient is an effective way of checking the entire dosimetric procedure. Metal oxide semiconductor field effect transistors (MOSFETs) have recently been proposed for use in radiation therapy. The purpose of this work is to study the performance characteristics and to carry out the commissioning of MOSFET as an in-vivo dosimeter for high-energy photon external beam radiation therapy.

Material and Methods

Characterization and commissioning of low sensitivity TN502RD and high sensitivity TN1002RD MOSFETs for entrance and exit dosimetry respectively for application in in-vivo dosimetry in radiotherapy was carried out. The MOSFETs were characterized in terms of reproducibility, short-term constancy, long-term constancy, linearity, angular dependence, energy dependence, source to skin distance (SSD) dependence and field size dependence.

Results

The reproducibility of standard sensitivity MOSFET is about 1.4% (1 SD) and 1.98% (1 SD) for high sensitivity detectors. The linearity of both MOSFETs was excellent (R2 = 0.996). The response of MOSFETs varies linearly for square fields from 3 × 3 cm2 to 30 × 30 cm2. For beam incidence ranging from ±45° the MOSFET response varies within ±3%. Commissioning of both MOSFETs was carried out in terms of entrance dose calibration factor, exit dose calibration factor, SSD correction factor, field size correction factor, wedge correction factor and shielding tray correction factor. The average calibration factor for low and high sensitivity MOSFET detectors is 0.9065 cGy/mV and 0.3412 cGy/mV respectively. The average SSD correction factors are quite small and vary between 0.968 and 1.027 for both types of detectors for the range of clinical SSDs from 80 cm to 120 cm. The field size correction factor varies from 1.00 to 1.02 for both types of detectors. The wedge and the shielding tray correction factors for both the detectors also show quite small variation. MOSFET characteristics are suitable for in vivo dosimetry of entrance and exit dose measurement relevant to 6 MV treatment.

Conclusion

It can be concluded that MOSFET dosimetry's low energy dependence, high sensitivity and immediate readout make it a good replacement for TLD in radiation therapy dosimetry.

Get Citation

Keywords

MOSFET; in vivo dosimetry; entrance dose; exit dose

About this article
Title

Performance characteristics and commissioning of MOSFET as an in-vivo dosimeter for high energy photon external beam radiation therapy

Journal

Reports of Practical Oncology and Radiotherapy

Issue

Vol 13, No 3 (2008)

Pages

114-125

Published online

2008-05-01

DOI

10.1016/S1507-1367(10)60001-6

Bibliographic record

Rep Pract Oncol Radiother 2008;13(3):114-125.

Keywords

MOSFET
in vivo dosimetry
entrance dose
exit dose

Authors

A. GOPIRAJ
RAMESH S. BILLIMAGGA
Velayudham RAMASUBRAMANIAN

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: journals@viamedica.pl