In-vivo dosimetry with Gafchromic films for patients with sarcoma

Agnieszka Walewska1, Paweł Kukołowicz1

Abstract

Background: The aim of the study was to evaluate Gafchromic films as in-vivo detectors for intensity modulated radiation therapy for patients with sarcomas and to assess the quality of irradiation in these patients.

Material and methods: Phantom measurements were used to validate the measurements performed with the Gafchromic. The uncertainty of the measurement method and that in determining the reference dose value obtained from the treatment planning system (TPS) were independently estimated. In-vivo measurements were performed in 21 patients with sarcomas who were irradiated with dynamic techniques using a 5 × 5 Gy. For each patient, measurements were taken at four points using films placed on the skin under a 1 cm bolus. The results of the measurements obtained in the 96 treatment sessions were analysed. The treatment quality was assessed based on the differences between the doses calculated using the TPS and those measured.

Results: The uncertainty of measurements was less than 0.8% (one standard deviation). Owing to differences in the dose gradient, the uncertainty of the reference dose reading from the TPS had an individual value at each measurement point. The uncertainties were less than 3% for more than 95% of the points; 93% of the in vivo measurements showed a difference of less than 7% between the measurements and calculations.

Conclusions: Gafchromic films can be used for in vivo dosimetry using dynamic techniques. This method made it possible to detect errors of 7% with a probability of approximately 95%. The results obtained for 21 patients with sarcoma demonstrated high-quality preparation and delivery of irradiation.

Article available in PDF format

View PDF Download PDF file

References

  1. Essers M, Mijnheer BJ. In vivo dosimetry during external photon beam radiotherapy. Int J Radiat Oncol Biol Phys. 1999; 43(2): 245–259.
  2. Su FC, Shi C, Papanikolaou N. Clinical application of GAFCHROMIC EBT film for in vivo dose measurements of total body irradiation radiotherapy. Appl Radiat Isot. 2008; 66(3): 389–394.
  3. Su FC, Shi C, Papanikolaou N. Clinical application of GAFCHROMIC EBT film for in vivo dose measurements of total body irradiation radiotherapy. Appl Radiat Isot. 2008; 66(3): 389–394.
  4. Avanzo M, Rink A, Dassie A, et al. In vivo dosimetry with radiochromic films in low-voltage intraoperative radiotherapy of the breast. Med Phys. 2012; 39(5): 2359–2368.
  5. Marcos M, Devic S. Poster — Thur Eve — 53: Novel Technique for the Measurement of Ultra-Superficial Doses Using Gafchromic Film. Med Phys. 2014; 41(8_Part 3): 17–18.
  6. Nakano M, Hill RF, Whitaker M, et al. A study of surface dosimetry for breast cancer radiotherapy treatments using Gafchromic EBT2 film. J Appl Clin Med Phys. 2012; 13(3): 3727.
  7. Moylan R, Aland T, Kairn T. Dosimetric accuracy of Gafchromic EBT2 and EBT3 film for in vivo dosimetry. Australas Phys Eng Sci Med. 2013; 36(3): 331–337.
  8. Morales JE, Hill R, Crowe SB, et al. A comparison of surface doses for very small field size x-ray beams: Monte Carlo calculations and radiochromic film measurements. Australas Phys Eng Sci Med. 2014; 37(2): 303–309.
  9. Avanzo M, Drigo A, Ren Kaiser S, et al. Dose to the skin in helical tomotherapy: results of in vivo measurements with radiochromic films. Phys Med. 2013; 29(3): 304–311.
  10. Devic S, Seuntjens J, Abdel-Rahman W, et al. Accurate skin dose measurements using radiochromic film in clinical applications. Med Phys. 2006; 33(4): 1116–1124.
  11. Bufacchi A, Carosi A, Adorante N, et al. In vivo EBT radiochromic film dosimetry of electron beam for Total Skin Electron Therapy (TSET). Phys Med. 2007; 23(2): 67–72.
  12. Mijnheer B, Jomehzadeh A, González P, et al. Error detection during VMAT delivery using EPID-based 3D transit dosimetry. Phys Med. 2018; 54: 137–145.
  13. Urso P, Lorusso R, Marzoli L, et al. Practical application of Octavius -4D: Characteristics and criticalities for IMRT and VMAT verification. J Appl Clin Med Phys. 2018; 19(5): 517–524.
  14. Miften M, Olch A, Mihailidis D, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218. Med Phys. 2018; 45(4): e53–e83.
  15. Vieillevigne L, Molinier J, Brun T, et al. Gamma index comparison of three VMAT QA systems and evaluation of their sensitivity to delivery errors. Phys Med. 2015; 31(7): 720–725.
  16. Yu L, Tang TLS, Cassim N, et al. Analysis of dose comparison techniques for patient-specific quality assurance in radiation therapy. J Appl Clin Med Phys. 2019; 20(11): 189–198.
  17. Huang YC, Yeh CY, Yeh JH, et al. Clinical practice and evaluation of electronic portal imaging device for VMAT quality assurance. Med Dosim. 2013; 38(1): 35–41.
  18. GAFChromic™ EBT3 film specifications. http://www.gafchromic.com/documents/EBT3_Specifications.pdf.
  19. GAFChromic™ EBT-XD film specifications. http://www.gafchromic.com/documents/EBTXD_Specifications_Final.pdf.
  20. Matney J, Parker B, Neck D, et al. Evaluation of a commercial flatbed document scanner and radiographic film scanner for radiochromic EBT film dosimetry. J Appl Clin Med Phys. 2010; 11(2): 198–208.
  21. Lewis D, Chan MF. Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry. Med Phys. 2015; 42(1): 416–429.
  22. Lewis DF, Chan MF. Technical Note: On GAFChromic EBT-XD film and the lateral response artifact. Med Phys. 2016; 43(2): 643–649.
  23. Micke A, Lewis DF, Yu X. Multichannel film dosimetry with nonuniformity correction. Med Phys. 2011; 38(5): 2523–2534.
  24. Lewis D, Micke A, Yu X, et al. An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan. Med Phys. 2012; 39(10): 6339–6350.
  25. Dybek M, Winiecki J, Iwanicki T, et al. Zalecenia Polskiego Towarzystwa Fizyki Medycznej: Kontrola Systemów Planowania Leczenia 3D W Radioterapii Wiązkami Zewnętrznymi Fotonów I Elektronów. Pol J Med Phys Eng. 2014; 20(1): 1–32.
  26. Aland T, Moyland E, Kairn T, et al. Effect of verification imaging on in vivo dosimetry results using Gafchromic EBT3 film. Phys Med. 2016; 32(11): 1461–1465.
  27. Ding GX, Alaei P, Curran B, et al. Image guidance doses delivered during radiotherapy: Quantification, management, and reduction: Report of the AAPM Therapy Physics Committee Task Group 180. Med Phys. 2018; 45(5): e84–e99.
  28. Ding G, Munro P. Characteristics of 2.5 MV beam and imaging dose to patients. Radiother Oncol. 2017; 125(3): 541–547.
  29. Wang L, Cmelak AJ, Ding GX. A simple technique to improve calculated skin dose accuracy in a commercial treatment planning system. J Appl Clin Med Phys. 2018; 19(2): 191–197.
  30. Chow JCL, Grigorov GN, Barnett RB. Study on surface dose generated in prostate intensity-modulated radiation therapy treatment. Med Dosim. 2006; 31(4): 249–258.
  31. Chung H, Jin H, Dempsey JF, et al. Evaluation of surface and build-up region dose for intensity-modulated radiation therapy in head and neck cancer. Med Phys. 2005; 32(8): 2682–2689.
  32. Higgins PD, Han EY, Yuan JL, et al. Evaluation of surface and superficial dose for head and neck treatments using conventional or intensity-modulated techniques. Phys Med Biol. 2007; 52(4): 1135–1146.
  33. Vera-Sánchez J, Ruiz-Morales C, González-López A. Monte Carlo uncertainty analysis of dose estimates in radiochromic film dosimetry with single-channel and multichannel algorithms. Phys Med. 2018; 47: 23–33.
  34. Sorriaux J, Kacperek A, Rossomm JA, et al. Evaluation of Gafchromic EBT3 films characteristics in therapy photon, electron and proton beams, J. Phys Med. 2013; 29(6): 599–606.
  35. Dąbrowski R, Drozdyk I, Kukołowicz P. High accuracy dosimetry with small pieces of Gafchromic films. Rep Pract Oncol Radiother. 2018; 23(2): 114–120.