Evaluation of dose distributions in the overlapping regions of the modulated volumetric arcs used in total body irradiation in pediatric patients

Eduardo Carrasco Solis1, Antonella A. Anticona Verástegui1, J. Fernando Márquez Pachas1, Alberto E. Gonzales-Ccoscco23, Cecilia Querebalu Garcia4, Yesenia Miranda Tunque5, Gustavo Sarria Bardales5, Carlos Herrera Castillo5, Edward Artemio Meca Castro6, José Santa Cruz Delgado1, Fernando Carrasco7, Ana Perini83, Lucio Neves83, Arthur S.B.Z. Alves3, Alessa Maschio3, Jhonny Antonio Benavente Castillo1, Galo Patino Camargo1, William S. Santos39

Abstract

Background: The application of total body irradiation (TBI), using intensity-modulated techniques with dynamic arcs, creates overlap at the treatment arc boundaries, which can lead to dosimetric variations. Therefore, this research aimed to evaluate the dose distributions in the overlap regions of the arcs used in total body irradiation in pediatric patients.

Material and methods: This dosimetric evaluation was conducted on a CIRS 715-TY0710 pediatric body simulator, where a treatment plan with three isocenters was designed, generating two 4 cm overlap regions along the line connecting the isocenters. The plan was verified using the gamma criteria of 3%/3mm and 5%/5mm with the Octavius 4D software.

Results/Conclusions: The planning results showed 98.2% dose coverage over 95% of the planning target volume (PTV) with adequate tolerance for risk organs. Regarding the gamma analysis, the acceptance rate was greater than 95% at the treatment isocenters with criteria of 3%/3 mm and 5%/5 mm. However, there is a decrease in the approved points in the overlap zone when evaluated with a 3%/3 mm criterion, while maintaining an approval percentage above 95% when evaluated with a 5%/5 mm criterion.

Article available in PDF format

View PDF Download PDF file

References

  1. Stanley D, McConnell K, Iqbal Z, et al. Dosimetric Evaluation Between the Conventional Volumetrically Modulated Arc Therapy (VMAT) Total Body Irradiation (TBI) and the Novel Simultaneous Integrated Total Marrow Approach (SIMBa) VMAT TBI. Cureus. 2021; 13(6): e15646.
  2. Yaray K, Damulira E. Evaluation of volumetric modulated arc therapy (VMAT) - based total body irradiation (TBI) in pediatric patients. Rep Pract Oncol Radiother. 2021; 26(4): 518–527.
  3. Fog LS, Hansen VN, Kjær-Kristoffersen F, et al. A step and shoot intensity modulated technique for total body irradiation. Tech Innov Patient Support Radiat Oncol. 2019; 10: 1–7.
  4. Welsh JS, Limmer JP, Howard SP, et al. Precautions in the use of intensity-modulated radiation therapy. Technol Cancer Res Treat. 2005; 4(2): 203–210.
  5. Liang X. Investigation of AP/PA Recumbent Technique for Total Body Irradiation (Doctoral dissertation). Duke University, Durham 2020.
  6. Losert C, Shpani R, Kießling R, et al. Novel rotatable tabletop for total-body irradiation using a linac-based VMAT technique. Radiat Oncol. 2019; 14(1): 244.
  7. Pierce G, Balogh A, Frederick R, et al. Extended SSD VMAT treatment for total body irradiation. J Appl Clin Med Phys. 2019; 20(1): 200–211.
  8. Springer A, Hammer J, Winkler E, et al. Total body irradiation with volumetric modulated arc therapy: Dosimetric data and first clinical experience. Radiat Oncol. 2016; 11: 46.
  9. Muñoz-Moral E, Rodríguez-Laguna A, Jiménez-Acosta JA. Implementation of total body irradiation using VMAT. AIP Conference Proceedings. 2021; 2348(1): 050029.
  10. Li J, Doemer A, Harrison AS, et al. Dose Calculation using Monte Carlo Algorithm of Monaco Treatment Planning System. Int J Radiat Oncol Biol Phys. 2008; 72(1): S674.
  11. Olmos JG, Juan CV, Llinares RP, et al. OCTAVIUS 4D: Establishment of initial reference state and routine clinical use of a device for VMAT verifications. Rev Física Méd. 2015; 16(3).
  12. Pal B, Pal A, Bag S, et al. Comparative performance analysis of 2D and 3D gamma metrics for patient specific QA in VMAT using Octavius 4D with 2D-Array 1500. Phys Med. 2021; 91: 18–27.
  13. VeriSoft. https://www.ptwdosimetry.com/en/products/verisoft.
  14. Morrison C, Symons K, Woodings S, et al. Verification of junction dose between VMAT arcs of total body irradiation using a Sun Nuclear ArcCHECK phantom. J Appl Clin Med Phys. 2017; 18(6): 177–182.
  15. Tas B, Durmus IF, Okumus A, et al. Total-body irradiation using linac-based volumetric modulated arc therapy: Its clinical accuracy, feasibility and reliability. Radiother Oncol. 2018; 129(3): 527–533.
  16. Chakraborty S, Cheruliyil S, Bharathan R, et al. Total Body Irradiation using VMAT (RapidArc): A Planning Study of a novel treatment delivery method. Int J Cancer Ther Oncol. 2015; 3(2): 03028.