Vol 81, No 11 (2023)
Original article
Published online: 2023-10-19

open access

Page views 555
Article views/downloads 448
Get Citation

Connect on Social Media

Connect on Social Media

Discontinuation of cardiac implantable electronic device therapy after transvenous lead extraction

Andrzej Kutarski1, Wojciech Jacheć2, Jarosław Kosior3, Sebastian Stec4, Dorota Nowosielecka5, Anna Polewczyk67
Pubmed: 37937353
Kardiol Pol 2023;81(11):1113-1121.

Abstract

Background: Patients with cardiac implantable electronic devices (CIEDs) may no longer be eligible for continued therapy.
Aims: The study aimed to assess the circumstances under which CIED reimplantation may not be necessary after transvenous lead extraction (TLE).
Methods: A retrospective analysis of 3646 TLE procedures was performed with assessment of indications for device reimplantation.
Results: Reimplantation was not performed immediately after TLE in 169 (4.6%) and, in long-term follow-up, in 146 (4.0%) of patients. No further need for CIED reimplantation was mostly associated with establishment of stable sinus rhythm (2.4%), conversion of sinus node dysfunction to chronic atrial fibrillation (AF; 1.4%), or improvement in left ventricular ejection fraction (LVEF) (0.9%). Independent prognostic factors were in the pacing groups: LVEF (odds ratio [OR], 1.03; 95% confidence interval [CI], 1.01–1.05; P <0.001), AF (OR, 3.8; 95% CI, 2.4–15.7; P <0.001), patients’ age during first CIED implantation (OR, 0.97; 95% C, 0.96–0.98; P <0.001), and New York Heart Association (NYHA) class (OR, 0.616; 95% CI, 0.43–0.86; P <0.01); in the cardioverter-defibrillator group: LVEF (OR, 1.06; 95% CI, 1.04–1.09; P <001). Non-reimplanted patients had more complex procedures and more frequent complications, but survival after TLE was better in this group of patients.
Conclusions: Reassessment of the need for continuation of CIED therapy should be considered in all patients following lead extraction and also before planned device replacement as TLE delay increases implant duration, complexity, and procedural risk. The predictors of non-reimplantation are a younger age during the first CIED implantation, lower NYHA class, presence of AF, and higher LVEF in pacemaker carriers, and, in the defibrillator group, only higher LVEF. A decision not to reimplant does not negatively affect the long-term prognosis.

Original article

Discontinuation of cardiac implantable electronic device therapy after transvenous lead extraction

Andrzej Kutarski1Wojciech Jacheć2Jarosław Kosior3Sebastian Stec4Dorota Nowosielecka5Anna Polewczyk67
1Department of Cardiology, Medical University, Lublin, Poland
22nd Department of Cardiology, Zabrze, Faculty of Medical Science in Zabrze, Medical University of Silesia in Katowice, Katowice, Poland
3Department of Cardiology, Masovian Specialistic Hospital of Radom, Radom, Poland
4Division of Electrophysiology, Cardioneuroablation, Catheter Ablation and Cardiac Pacing, Subcarpathian Centre for Cardiovascular Interventions, Sanok, Poland
5Department of Cardiology, the Pope John Paul II Province Hospital of Zamosc, Zamość, Poland
6Department of Physiology, Pathophysiology and Clinical Immunology, Collegium Medicum of Jan Kochanowski University, Kielce, Poland
7Department of Cardiac Surgery, Świętokrzyskie Cardiology Center, Kielce, Poland

Correspondence to:

Anna Polewczyk MD, PhD,

Department of Physiology, Pathophysiology and Clinical Immunology,

Collegium Medicum of Jan Kochanowski University,

Żeromskiego 5, 25–369 Kielce, Poland,

phone: +48 600 024 074,

e-mail: annapolewczyk@wp.pl

Copyright by the Author(s), 2023

DOI: 10.33963/v.kp.97655

Received: May 30, 2023

Accepted: September 19, 2023

Early publication date: November 30, 2023

Abstract
Background: Patients with cardiac implantable electronic devices (CIEDs) may no longer be eligible for continued therapy.
Aims: The study aimed to assess the circumstances under which CIED reimplantation may not be necessary after transvenous lead extraction (TLE).
Methods: A retrospective analysis of 3646 TLE procedures was performed with assessment of indications for device reimplantation.
Results: Reimplantation was not performed immediately after TLE in 169 (4.6%) and, in long-term follow-up, in 146 (4.0%) of patients. No further need for CIED reimplantation was mostly associated with establishment of stable sinus rhythm (2.4%), conversion of sinus node dysfunction to chronic atrial fibrillation (AF; 1.4%), or improvement in left ventricular ejection fraction (LVEF) (0.9%). Independent prognostic factors were in the pacing groups: LVEF (odds ratio [OR], 1.03; 95% confidence interval [CI], 1.011.05; P <0.001), AF (OR, 3.8; 95% CI, 2.415.7; P <0.001), patients’ age during first CIED implantation (OR, 0.97; 95% C, 0.960.98; P <0.001), and New York Heart Association (NYHA) class (OR, 0.616; 95% CI, 0.430.86; P <0.01); in the cardioverter-defibrillator group: LVEF (OR, 1.06; 95% CI, 1.041.09; P <001). Non-reimplanted patients had more complex procedures and more frequent complications, but survival after TLE was better in this group of patients.
Conclusions: Reassessment of the need for continuation of CIED therapy should be considered in all patients following lead extraction and also before planned device replacement as TLE delay increases implant duration, complexity, and procedural risk. The predictors of non-reimplantation are a younger age during the first CIED implantation, lower NYHA class, presence of AF, and higher LVEF in pacemaker carriers, and, in the defibrillator group, only higher LVEF. A decision not to reimplant does not negatively affect the long-term prognosis.
Key words: implantable electronic device, long-term prognosis, loss of indications for pacing, superfluous CIED, unnecessary cardiac transvenous lead extraction

What’s new?

To the best of our knowledge, this is the first report discussing the problem of loss of indications for cardiac device therapy in such a large study population. We have demonstrated that reassessment of the need to continue therapy with cardiac implantable electronic devices (CIEDs) should be considered in all patients referred for transvenous lead extraction for both infectious and non-infectious indications. The present study documents for the first time that the loss of the indications for CIED therapy has no effect on long-term survival. The study also indicates that a significant delay in the decision to remove the unnecessary system is associated with greater procedure complexity and higher risk of major complications. Considering the increase in lead removal difficulty from decade to decade, it seems better to remove unnecessary CIEDs earlier than put off removal to a future time.

Introduction

Transvenous lead extraction (TLE) is an inherent part of lead management strategy [1–3]. Indications for lead removal are well-known and generally accepted [1–3]. The TLE guidelines recommend or even mandate reassessment of indications for cardiac implantable electronic device (CIED) therapy continuation before a new implantation [1–3]. Also, the practical guidelines on the CIED implantation technique endorsed by the Heart Rhythm Society recommend reviewing the justification for continuing electrotherapy before any generator replacement due to battery deployment [4]. It is a widely known fact that about 15%–35% of CIED patients do not meet the criteria for continuation of cardiac pacing following device removal [5–10]. This applies to patients with infectious [5–7], non-infectious [13–15], and mixed indications [8–12] for lead extraction The guidelines for CIED implantation [4] and lead extraction [1–3] address the issue of CIED therapy discontinuation; however, most patients no longer needing device therapy receive a new CIED, which puts off device removal to a future time. The present study aimed to identify the reasons for resigning from reimplantation and the need for new CIED implantation during long-term follow-up as well as to assess long-term survival in such patients. An additional goal was examination of complexity and outcome of redundant system removal.

Methods

Study population

We performed retrospective analysis of data from 3646 subsequent patients undergoing lead extraction procedures (described in our previous reports [16–20]) in three high-volume centers in the years 20062021. Indications for TLE and lead extraction techniques were described previously [16–20]. Definitions of TLE efficacy and technical and procedural complications are provided in the supplementary materials.

Detailed inclusion criteria

The study population was divided into five groups based on the reimplantation decision.

Group 1. Patients with unnecessary pacemaker, normal sinus rate (NSR). A pacemaker was deemed unnecessary when there was an extremely low percentage (<1%) or absence of any pacing in pacemaker memory, stable normal sinus rhythm, and normal atrioventricular (AV) conduction over a long period (pacemaker memory). In borderline cases, pacemakers were programmed to VVI with a rate of 35 bpm for 6 months (minimum pacing rate test). In case of doubt on the doctor’s or patients’ side, pacing was switched off (impulse amplitude 0.1 V and duration 0.1 ms) for the next 6 months (“switch off test”). This time-consuming, “conventional” approach was used in 79 patients. In the remaining 7 patients, possible influences of the autonomic nervous system on heart rhythm were evaluated and cardiac neuroablation was performed before device removal.

Group 2. Patients with an unnecessary pacemaker, permanent atrial fibrillation (AF), normal heart rate (NHR). A pacemaker was unnecessary in this group when we observed an extremely low percentage (<1%) or absence of ventricular pacing in pacemaker memory, stable AF with normal AV conduction (NHR) over a long period (pacemaker memory). In doubtful (non-infective) cases, the pacemaker was programmed to VVI with a rate of 35 bpm for 6 months. This time-consuming, “conventional” approach (“minimum ventricular pacing output test” and “switch off ventricular pacing test”) was used in 34 patients. In patients with CIED-related infection, the system was removed, and Holter monitoring on 2 or 3 occasions was performed to reassess the need for permanent pacing.

Group 3. Patients with unnecessary implantable cardioverter-defibrillator (ICD)/cardiac resynchronization therapy (CRT), improved left ventricular ejection fraction (LVEF), NHR. Indications for CIED reimplantation were lost in the case of improvement in LVEF > 40%, no ventricular arrhythmia in device memory, no indications for pacing in ICD patients, and no indications for pacing and narrow QRS complexes in CRT-dual (CRT-D) patients (transient arrhythmias or significant impairment of left ventricular contractility were probably caused by an inflammatory process in these cases).

The decision made by patients and their physicians resulted in resignation from ICD or CRT reimplantation.

Group 4. All non-reimplanted patients from groups 1, 2, and 3 were taken together for general comparisons with the controls.

Group 5. A control group of patients in whom leads or entire devices were removed for infectious or non-infectious indications, and a new CIED (A [atrial], V [ventricular], D [dual], CRT-P [cardiac resynchronization therapy-pacing] in 2448 [70.4%], ICD in 78 [22.4%], and CRT-D in 251 [7.2%]) was implanted immediately or after infection treatment.

The patient’s and his/her doctor’s consent to discontinue pacemaker/ICD-CRT-D therapy was mandatory.

Analysis of non-reimplantation by type of CIED and multivariate analysis of prognostic factors for discontinuation of indications for CIED, as well as analysis of the risk of death in the non-reimplantation group and information on reimplantation during long-term follow-up are presented in supplementary materials.

Dataset and statistical methods

Due to non-Gaussian distribution, all continuous variables were presented as medians and interquartile ranges (IQR). The categorical variables were presented as numbers and percentages. The patients were divided into 5 groups. The significance of differences between groups (1, 2, 3, 4 vs. 5) was determined using the nonparametric χ2 test with Yates’ correction or the unpaired Mann-Whitney U test, as appropriate. The Bonferroni correction was applied, assuming a P-value of <0.0125 as statistically significant. The uni- and multivariable logistic regression were used to determine prognostic factors for the loss of indications for CIED reimplantation. In the multivariable model, data that achieved P <0.05 under univariable analysis were included. Analysis was performed in the two subgroups divided depending on the kind of CIED: conventional pacemakers and cardioverter defibrillator groups. The parameters that reached P <0.05 under multivariable regression analysis were presented.

The Cox regression model was used to determine the impact of not meeting any longer the criteria for device implantation on long-term survival following TLE. Clinical and CIED-related data were analyzed. In the multivariable mo­del, the data that achieved P <0.05 under univariable analysis were included. The parameters that reached P <0.05 under multivariable regression analysis were presented.

To present the relationship between no longer meeting the reimplantation criteria and mortality, Kaplan-Meier survival curves were plotted, whose course was assessed using the log-rank test. Kaplan-Meier log-rank survival analysis and Cox regression survival analysis considered the time from last TLE to the end of the follow-up period and were performed on the basis of data from 3464 patients. A P-value less than 0.05 was considered statistically significant. Statistical analysis was performed with Statistica version 13.3 (TIBCO Software Inc.).

Approval of the Bioethics Committee

All patients gave their informed written consent to undergo TLE, and we used anonymous data from their medical records. The telephone follow-up was approved by the Bioethics Committee at the Regional Chamber of Physicians in Lublin no. 288/2018/KB/VII. The study was carried out in accordance with the ethical standards of the 1964 Declaration of Helsinki.

Results

The study group consisted of 3646 patients (61.6% men), aged 596 years, the median age: 69.0 (60.077.0), who underwent lead extraction between 2006 and 2021. Indications for TLE included infectious (17% of patients) and non-infectious complications, such as dysfunction or dislocation of the leads, need to upgrade the system, and loss of indications for device therapy (Table 1).

Patients from group 1 were the youngest age group at first CIED implantation (37.1 years), and most often had unknown underlying heart disease. In 23.3% of patients, the main indication for system removal was infection.

Patients from the second group were characterized by the oldest age at TLE (72 years) and there were many women in this group (54%). In 32.0% of patients, the main indication for system removal was infection.

Group 3 (33 patients) was relatively young, both during system implantation and extraction (45.7 and 52.8 years, respectively); they rarely had ischemic heart disease (IHD)(24.2%), more frequently had normal LVEF (over 50% in 50% of patients), and a low score according to the Charlson co-morbidity index.

The best general health status was observed in patients from groups 1 and 3; it was significantly better than in the control group. Group 2, in spite of older age, presented, indeed, better LVEF than controls (Table 1).

Table 1. Clinical characteristics of the study patients

Unnecessary pacemaker, χNSR

Unnecessary pacemaker, AF, NHR

Unnecessary ICD/CRT, improvement in EF, NHR

All non-reimplanted patients

Control group Removal of necessary CIED with immediate or delayed reimplantation

Group number

1

2

3

4

5

χ2 test/

/Mann-Whitney

“U” test

1 vs. 5

χ2 test/

/Mann-Whitney

“U” test

2 vs. 5

χ2 test/

/Mann-Whitney

“U” test

3 vs. 5

χ2 test/

/Mann-Whitney

“U” test

4 vs. 5

Number of patients, (%)

N = 86 (2.4%)

N = 50 (1.4%)

N = 33 (0.9%)

N = 169 (4.6%)

N = 3477 (95.4%)

Patient age during TLE, years

50.5 (35.5–64.5)

P <0.001

72.0 (65.0–77.0)

P = 0.03

53.0 (36.0–67.0)

P <0.001

60.5 (41.0–71.0)

P <0.001

69.0 (60.0–77.0)

Patient age at first implantation, years

37.1 (20.5–49.5)

P <0.001

62.0 (55.0–68.0)

P = 0.27

45.0 (33.0–61.0)

P <0.001

46.5 (29.6–61.0)

P <0.001

61.0 (51.0–70.0)

Sex, female

35 (40.7)

P = 0.65

27 (54.0)

P = 0.02

8 (24.2)

P = 0.02

70 (41.4)

P = 0.46

1331 (38.3)

Underlying heart disease: ischemic heart disease

11 (12.8)

P <0.001

36 (72.0)

P = 0.04

8 (24.2)

P <0.001

55 (32.5)

P <0.001

1974 (56.8)

Underlying heart disease: valvular heart diseases, cardiomyopathies, post inflammatory

4 (4.7)

P = 0.01

3 (6.0)

P = 0.09

18 (54.6)

P <0.001

25 (14.8)

P = 0.7

553 (15.9)

Underlying heart disease: conduit, unknown, channelopathies, neurocardiogenic, surgical, post ablation

71 (82.6)

P <0.001

11 (22.0)

P = 0.51

7 (21.2)

P = 0.55

89 (52.7)

P <0.001

947 (27.2)

NYHA class III & IV

2 (2.3)

P <0.001

4 (8.0)

P = 0.2

0 (0.0)

P = 0.03

6 (3.6)

P <0.001

541 (15.6)

Congestive heart failure (symptomatic presently)

0 (0.0)

P <0.001

9 (18.0)

P = 1.0

4 (12.1)

P = 0.44

13 (7.7)

P <0.001

659 (19.0)

LVEF average, %

62.0 (60.0–65.0)

P <0.001

55.0 (45.0–60.0)

P = 0.11

52.0 (45.0–60.0)

P = 0.04

60.0 (54.0–65.0)

P <0.001

53.0 (35.0–60.0)

Charlson comorbidity index, points

1 (0–3)

P <0.001

4 (3–6)

P = 0.58

1 (0–3)

P <0.01

2 (0–4)

P <0.001

4 (3–6)

Main indications for TLE — (primary / predominant)

Infective endocarditis with or without pocket infection

14 (16.3)

P = 0.26

10 (20.0)

P = 0.87

5 (15.2)

P = 0.47

29 (17.2)

P = 0.17

764 (21.5)

Local (isolated) pocket infection

6 (7.0)

P = 0.45

6 (12.0)

P = 0.79

3 (9.1)

P = 0.98

15 (8.9)

P = 0.78

342 (9.8)

Mechanical lead damage (electric failure)

16 (18.6)

P = 0.09

3 (6.0)

P < 0.001

11 (33.3)

P = 0.57

30 (17.8)

P < 0.01

953 (27.4)

Lead dysfunction (exit/entry block, dislodgement, perforation, extracardiac pacing)

8 (9.3)

P = 0.046

3 (6.0)

P = 0.12

1 (3.0)

P = 0.06

12 (7.1)

P < 0.01

808 (23.2)

Other non-infective indicationsa

42 (48.8)

P <0.001

28 (56.0)

P <0.001

13 (39.4)

P <0.01

83 (49.1)

P <0.001

610 (17.5)

A decision to discontinue CIED therapy was made in 169 (4.6%) patients of the entire analyzed population (group 4), and it was made significantly more often in patients with pacemaker (136 patients, 5.3%) than in patients with ICD (28 [3.5%]) or CRT-D5 (1.9%) (Supplementary material, Table S1).

Independent prognostic factors for loss of indications for CIED reimplantation in the group of patients with pacemakers were: higher LVEF (odds ratio [OR], 1.03; P <0.001), presence of AF (OR, 3.8; P <0.001), older age (by one year) during first CIED implantation (OR, 0.97; P <0.001), and higher New York Heart Association (NYHA) class (OR, 0.61; P <0.01). The only predictive parameter for loss of indications for ICD/CRT-D reimplantation was an increase in LVEF (OR, 1.06; P <0.001) (Supplementary material, Table S2).

Implant duration was significantly longer in groups 1 and 2 than in the control group (Table 2).

Table 2. Risk factors for major complications and procedure complexity related to the device, history of pacing, and the procedure itself

Unnecessary pacemaker, NSR

Unnecessary pacemaker, AF, NHR

Unnecessary ICD/CRT, improvement in LVEF, NHR

All non-reimplanted patients

Control group Removal of necessary CIED with immediate or delayed reimplantation

Group number

1

2

3

4

5

χ2 test/

/Mann-Whitney

“U” test

1 vs. 5

χ2 test/

/Mann-Whitney

“U” test

2 vs.5

χ2 test/

/Mann-Whitney

“U” test

3 vs.5

χ2 test/

/Mann-Whitney

“U” test

4 vs.5

Number of patients, %

N = 86 (2.4%)

N = 50 (1.4%)

N = 33 (0.9%)

N = 169 (4.6%)

N = 3477 (95.4%)

Types of CIED systems

PM (A,V,D)

86 (100.0)

50 (100.0)

0 (0.00)

136 (80.5)

2448 (70.4)

ICD (V,D)

0 (0.00)

0 (0.00)

28 (84.9)

28 (16.57)

778 (22.4)

CRT-D

0 (0.00)

0 (0.00)

5 (15.2)

5 (3.0)

251 (7.2)

Number of procedures before lead extraction, n

2 (1–2)

P = 0.08

1 (1–2)

P = 0.21

2 (1–2)

P = 0.29

2 (1–2)

P = 0.94

2 (1–2)

Time since last CIED procedure (any), months

75.0 (34.5–115.5)

P <0.001

72.0 (34.0–93.0)

P <0.001

42.0 (13.0–76.0)

P = 1.0

67.0 (33.0–97.0)

P = 0.001

38.0 (16.0–70.0)

TLE-related potential risk factors for major complications and procedure complexity

Oldest extracted lead per patient, months

135.0 (82.4–213.6)

P <0.001

125.0 (79.0–157.0)

P <0.01

85.0 (42.0–132)

P = 0.37

122.0 (74.5–185.5)

P < 0.001

84.0 (43.0–22.9)

Cumulative dwell times of extracted leads per patient, years

18.1 (10.3–31.6)

P <0.001

19.0 (12.5–28.0)

P <0.001

10.1 (4.33–14.2)

P = 0.32

16.4 (9.9–28.0)

P = 0.001

9.25 (4.7–18.0)

SAFeTY-TLE probability score of MC risk, %

1.3 (0.4–3.2)

P <0.001

1.8 (0.5–3.7)

P = 0.06

0.33 (0.2–0.6)

P = 0.07

1.2 (0.4–2.6)

P = 0.001

0.5 (0.3–1.8)

TLE complexity

Procedure duration (sheath-to-sheath time), minutes

11.5 (8.0–25.0)

P <0.001

11.0 (8.0–19.0)

P = 0.04

9.0 (4.0–18.0)

P = 0.55

11.0 (8.0–21.0)

P <0.001

9.0 (4.0–12.0)

Average time of single lead extraction (sheath-to-sheath / number of extracted leads), minutes

6.5 (4.0–15.0)

P <0.001

5.0 (4.0–10.0)

P <0.59

5.0 (4.0–12.0)

P = 0.04

6.0 (4.0–12.5)

P <0.001

4.5 (4.0–9.0)

Unexpected procedure difficulties during TLE (any) [18]

28 (32.6)

P <0.01

9 (18.0)

P = 0.9

6 (18.2)

P = 0.8

43 (25.4)

P = 0.21

687 (19.8)

Lead-to-lead binding

11 (12.8)

P = 0.03

5 (10.0)

P = 0.45

1 (3.0)

P = 0.68

17 (10.1)

P = 0.08

220 (6.3)

Break of extracted lead

13 (15.1)

P < 0.01

3 (6.0)

P = 0.81

1 (3.3)

P = 0.76

17 (10.1)

P = 0.03

201 (5.8)

Two or unexpected procedure difficulties [18]

13 (15.12)

P = 0.001

3 (6.00)

P = 0.724

2 (6.1)

P = 0.88

18 (10.7)

P <0.001

139 (4.0)

Use of additional tools

Evolution (old and new) or TightRail

6 (6.98)

P = 0.001

0 (0.00)

P = 0.849

1 (3.0)

P = 0.93

7 (4.1)

P < 0.01

46 (1.3)

Use of advanced tools (metal sheaths, lasso- basket catheters/snares)

22 (24.2)

P <0.001

6 (10.7)

P = 0.79

3 (8.82)

P = 0.59

21 (12.4)

P = 0.94

409 (11.9)

The oldest leads were extracted from patients from group 1. That resulted in a higher risk of major TLE complications, which was predicted using the SAFeTY-TLE calculator in the groups with unnecessary pacemakers. In about 50% of all non-reimplanted patients, verification of indications took place during the third or next CIED-related procedure, which explains long or very long implant duration and subsequent slightly worse TLE effectiveness and increased complication rates. Procedure duration was longer in all patients with redundant CIEDs compared to the control group. Procedure difficulty and complexity (expressed as the number of technical problems, mutual lead-to-lead connections, with strong scars or breaks of extracted lead) were significantly more frequent in patients from group 1. Its resulted from the longest implant duration and youngest patient’s age. The necessity to utilize second-line tools (Evolution or TightRail, metal sheath, lasso catheter/snare) confirms the most difficult, and most complicated lead extraction in the group of patients with superfluous pacemakers and NSR (Table 2).

After TLE, the median follow-up of the whole study group was 1584 days. Occurrence of any major complications, necessity for urgent rescue cardiac surgery, tricuspid valve damage during TLE, partial radiographic success (lead remnants), and lack of complete procedural success were significantly more frequent in group 1. However, there were no procedure-related deaths (intra or post-procedural) in the non-reimplanted groups of patients. Analysis showed that periprocedural and 30-day mortality was similar and that there was no death in the non-reimplanted groups in this period. However, mortality rates during all-time follow-up at > 1 year and >3 years following TLE were significantly higher in reimplanted patients (Table 3, Figure 1).

Table 3. TLE major complications, outcomes and long-term mortality following lead extraction

Unnecessary pacemaker, NSR

Unnecessary pacemaker, AF, NHR

Unnecessary ICD/CRT, improvement in EF, NHR

All non-reimplanted patients

Control group Removal of necessary CIED with immediate or delayed reimplantation

Group numer

1

2

3

4

5

χ٢ test /

Mann-Whitney

“U” test

1 vs. 5

χ٢ test /

Mann-Whitney

“U” test

2 vs. 5

χ٢ test /

Mann-Whitney

“U” test

3 vs. 5

χ٢ test /

Mann-Whitney

“U” test

4 vs. 5

Number of patients, %

N = 86

(2.4%)

N = 50

(1.4%)

N = 33

(0.9%)

N = 169

(4.6%)

N = 3477 (95.4%)

Major complications (any)

9 (10.5)

P <0.001

1 (2.0)

P = 0.67

1 (3.0)

P = 0.2

11 (6.5)

P <0.001

62 (1.8)

Hemopericardium

8 (9.3)

P <0.001

1 (2.0)

P = 0.42

1 (3.0)

P = 0.81

10 (59.2)

P <0.001

37 (1.1)

Tricuspid valve damage during TLE (severe)

3 (3.5)

P <0.01

0 (0.0)

P = 0.54

0 (0.0)

P = 0.39

3 (1.8)

P = 0.1

17 (0.5)

Rescue cardiac surgery

9 (10.5)

P <0.001

1 (2.0)

P = 0.96

1 (3.0)

P = 0.73

11 (6.5)

P <0.001

32 (0.9)

Procedure-related death (intra-, post-procedural)

0 (0.0)

P = 0.89

0 (0.0)

P = 0.59

0 (0.0)

P = 0.39

0 (0.0)

P = 0.74

6 (0.2)

Partial radiographic success (a tip or <4 cm lead fragment remained)

9 (10.5)

P <0.01

1 (2.0)

P = 0.81

1 (3.0)

P = 0.78

11 (6.5)

P = 0.09

132 (3.8)

Complete procedural success

75 (87.2)

P <0.001

49 (98.0)

P = 0.6

32 (97.0)

P = 1.0

156 (92.3)

P = 0.09

3318 (95.4)

Survival after TLE during 1584 (718.0–2823) (1–5519) days of follow-up.

Log-rank P for all models: P <0.001

Length of the follow-up: survivors, days

1689

(752.0–2330)

1761

(1185–2494)

1213

(850.0–1629)

1577

(784.0–2389)

1913

(961.0–3145)

Length of the follow-up: death, days

595.0

(158.0–1116)

1228

(595.0–2592)

1081

(702.0–1894)

1081

(585.0–2147)

1112

(374.0–2072)

1-month mortality after TLE 1–30 days

0 (0.0)

P = 0.45

0 (0.0)

P = 0.74

0 (0.0)

P = 0.97

0 (0.0)

P = 0.18

56 (1.6)

Alive during all follow-up

81 (94.2)

P <0.001

39 (78.0)

P = 0.11

30 (90.9)

P <0.01

150 (88.8)

P <0.001

2304 (66.3)

Death during all follow-up

5 (95.8)

P <0.001

11 (22.0)

P = 0.11

3 (9.1)

P <0.01

19 (11.2)

P <0.001

1173 (33.7)

Figure 1. Survival after transvenous lead extraction according to continuation or discontinuation of cardiac implantable electronic device therapy for different reasons

In non-reimplanted patients, 19 died during follow-up. In the subgroup with removed ICD systems, there were two deaths up to 3 years and one death at >3 years, whereas in the subgroup with removed CRT-D systems, no patient died (Supplementary material, Table S3).

The mortality rate during long-term follow-up was significantly lower in non-reimplanted patients (11.2% vs. 33.7%,). The risk factors for death were infections, older age, low LVEF, IHD, and congestive heart failure.

Regression analysis confirmed that the risk of death during follow-up increased with patient age (by 5.0% for each year), infectious indications (by 49.4%), higher NYHA class (by 31.8% per one class), presence of diabetes (by 32.6%), renal dysfunction (by 85.2%), presence of CRT before TLE (by 9.4%). The loss of indications for electrotherapy lowered the risk of death by 35.0% (Supplementary material, Table S4).

Long-term telephone follow-up was performed for this analysis. In 86.4% of cases, a new system was not implanted, 4.7% received a new device, and we lost contact with 8.9% (all survived) (Supplementary material, Table S5).

Discussion

In clinical practice, it may appear that sometimes, after many years, in various clinical situations, some patients with CIEDs no longer meet the treatment criteria. Device therapy is no longer needed after lead extraction for various reasons [5–15]. Moreover, the guidelines for lead extraction strongly recommend detailed reassessment of indications for a new CIED implantation [1–3]. Discontinuation of CIED therapy was described in 19.3%–45.0% of patients with infectious indications [5–7], 14.0%–38.8% of patients with non-infectious indications for TLE [13–15], and 14.8%–35.0% with mixed reasons [8–12]. However, most reports show a high rate of discharge of patients who did not receive a new device with a recommendation to continue antibiotic therapy and follow-up [5–12]. In our study, we analyzed the reasons for the decision to resign from reimplantation in patients undergoing TLE mainly for non-infectious indications. The overall percentage of patients who did not undergo CIED reimplantation after transvenous lead extraction was only 4.64% (4.00% during long-term follow-up). The patients did not undergo subsequent device reimplantation mainly because of the return of stable NSR and normal AV conduction in 2.36% of patients, conversion of sinus node dysfunction (SND) to permanent AF with NHR in 1.37%, a significant improvement in LVEF, and no ventricular arrhythmia during long-term follow-up in 0.91% of patients. There are many potentially reversible causes of SND or AV conduction disturbances over a long or very long period [21–23]. Especially in young patients, sinus arrest or paroxysmal AV block may be a sign of different forms of neurocardiogenic syncope (vagal reflex only) without heart disease [24–27]. Reports of various forms of neurocardiogenic syncope do not present a long-term natural history of the disorder [27–30]. Our personal observations seem to indicate that in patients with neurocardiogenic syncope and pacemakers, the severity of vagal reflex decreases over time and the percentage of atrial or ventricular pacing becomes low. Considering potential lead-related complications (infectious or lead dysfunction) that can occur over the next decades, and difficulties with extraction of very old leads, we decided to remove unnecessary CIEDs without reimplantation in patients with higher life expectancy (most patients in group 1). Such decisions were based on device memory analysis and the results of the “minimum pacing output test” and “switch off test” with Holter monitoring. We believe that it is better to remove unnecessary CIEDs in the case of battery exhaustion than to reimplant the device and put off removal to a future time.

The most important aspect of the decision to resign from CIED reimplantation after TLE is assessment of rhythm stability and, above all, long-term survival. There are several reports on long-term survival among non-reimplanted patients [9–12, 14], and nobody has demonstrated a relationship between death and lack of CIED. In our study, the rates of periprocedural and 30-day mortality were similar, and there were no deaths in non-reimplanted patients in this period, but mortality at >1 year and >3 years following lead extraction was, paradoxically, higher among reimplanted patients. This can be accounted for by the fact that risk factors for mortality during follow-up were mainly infection, older age, IHD, lower LVEF, and congestive heart failure, while patients from groups without reimplantation were often younger and had fewer comorbidities.

The issue of discontinued ICD therapy is less known. Pediatric guidelines recommend reconsideration of indications for ICD for each CIED-related intervention [31]. The authors of the guidelines refer to the extensive study by Kini et al. [32], which showed that approximately 25% of patients receiving ICDs as primary prevention may no longer meet the indications for an ICD at the time of generator replacement. In other reports, a thorough analysis of non-implanted ICD patients is omitted [33] or patients with pacemaker and ICD are discussed as one group. Döring showed that more than one-third of patients undergoing TLE do not need reimplantation due to loss of indications [5]. Al-Hijji [14] pointed out that 14% of patients who had device extraction did not undergo reimplantation mainly because they no longer met CIED indications. The high mortality in these patients is related to device complications and comorbid conditions, whereas mortality (as reported by Döring) associated with arrhythmia is rare. Little is known about discontinuing CRT in responders as this therapy should be continued for as long as possible [34]. Recently, a report by D’Angelo et al. [35] was published, in which the authors indicate that mortality among non-reimplanted patients (with CRT and ICD systems removed, which constituted about 70% of the group) was even lower than in patients who received CIEDs again as indicated immediately after TLE or after recovery.

Study limitations

In many cases, especially in young patients, it was very difficult to establish the underlying heart disease and causes of rhythm disturbances. We showed that Holter findings of sinus node dysfunction or recurrent AV conduction disturbances may disappear within the service life of the device, but we cannot be sure about regression of, for example, vasovagal syndrome or another form of neurocardiogenic syncope and differentiate it retrospectively from inflammatory or drug-induced arrhythmia.

Conclusions

In the presented TLE population CIED reimplantation was not necessary, due to the disappearance of indications in 4.6% of patients during short-term follow-up. In about 70% of them, a decision was made during TLE performed due to infective or non-infective indications, but in 30 %, the decision was taken when patients were referred for TLE due to battery depletion and with intention not to reimplant their device. The most important predictive parameters for discontinuation of CIED therapy are younger age of patients during first CIED implantation, lower NYHA class, an increase in LVEF, and occurrence of AF (pacemaker group). Patients in whom indications for device therapy have disappeared had better general and cardiac status health (excluding patients with permanent AF) and had longer implant duration (excluding ICD/CET carriers). These factors affects procedure complicity, effectiveness, and complications, but not short, mid, and long-term mortality, which is strongly dependent on infective indications for TLE and general health status.

In about 50% of all non-reimplanted patients, verification of the necessity for continuation of CIED therapy takes place during the third or next CIED-related procedure, which explains long or very long implant duration and subsequent slightly worse TLE effectiveness and increased complication rates. Therefore, such verification should be performed in all CIED carriers, not only after lead extraction but before planned unit replacement due to battery depletion.

Supplementary material

Supplementary material is available at https://journals.viamedica.pl/kardiologia_polska.

Article information

Conflict of interest: None declared.

Funding: None.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, which allows downloading and sharing articles with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.

REFERENCES

  1. Wilkoff BL, Love CJ, Byrd CL, et al. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA). Heart Rhythm. 2009; 6(7): 10851104, doi: 10.1016/j.hrthm.2009.05.020, indexed in Pubmed: 19560098.
  2. Kusumoto FM, Schoenfeld MH, Wilkoff BL, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017; 14(12): e503e551, doi: 10.1016/j.hrthm.2017.09.001, indexed in Pubmed: 28919379.
  3. Bongiorni MG, Burri H, Deharo JC, et al. 2018 EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRS. Europace. 2018; 20(7): 1217, doi: 10.1093/europace/euy050, indexed in Pubmed: 29566158.
  4. Burri H, Starck C, Auricchio A, et al. EHRA expert consensus statement and practical guide on optimal implantation technique for conventional pacemakers and implantable cardioverter-defibrillators: endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), and the Latin-American Heart Rhythm Society (LAHRS). EP Europace. 2021; 23(7): 9831008, doi: 10.1093/europace/euaa367.
  5. Döring M, Hienzsch L, Ebert M, et al. Extraction of infected cardiac implantable electronic devices and the need for subsequent re-implantation. Int J Cardiol. 2020; 309: 8491, doi: 10.1016/j.ijcard.2019.12.044, indexed in Pubmed: 31973885.
  6. Sohail MR, Uslan DZ, Khan AH, et al. Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. J Am Coll Cardiol. 2007; 49(18): 18511859, doi: 10.1016/j.jacc.2007.01.072, indexed in Pubmed: 17481444.
  7. Kim DH, Tate J, Dresen WF, et al. Cardiac implanted electronic device-related infective endocarditis: clinical features, management, and outcomes of 80 consecutive patients. Pacing Clin Electrophysiol. 2014; 37(8): 978985, doi: 10.1111/pace.12452, indexed in Pubmed: 25060820.
  8. De Schouwer K, Vanhove R, Garweg C, et al. Re-implantation after extraction of a cardiac implantable electronic device. Acta Cardiol. 2020; 75(6): 505513, doi: 10.1080/00015385.2019.1620997, indexed in Pubmed: 31145671.
  9. Zsigmond EJ, Miklos M, Vida A, et al. Reimplantation and long-term mortality after transvenous lead extraction in a high-risk, single-center cohort. J Interv Card Electrophysiol. 2023; 66(4): 847855, doi: 10.1007/s10840-021-00974-4, indexed in Pubmed: 33723694.
  10. D’Angelo G, Zweiker D, Fierro N, et al. Check the need-prevalence and outcome after transvenous cardiac implantable electric device extraction without reimplantation. J Clin Med. 2021; 10(18), doi: 10.3390/jcm10184043, indexed in Pubmed: 34575152.
  11. Merchant FM, Levy MR, Kelli HM, et al. Predictors of long-term survival following transvenous extraction of defibrillator leads. Pacing Clin Electrophysiol. 2015; 38(11): 12971303, doi: 10.1111/pace.12733, indexed in Pubmed: 26256667.
  12. Gomes S, Cranney G, Bennett M, et al. Long-Term outcomes following transvenous lead extraction. Pacing Clin Electrophysiol. 2016; 39(4): 345351, doi: 10.1111/pace.12812, indexed in Pubmed: 26768807.
  13. Lewis K, Dai M, Patton KK, et al. Lead extraction for reduction of chronic pain related to cardiovascular implantable electronic device. Europace. 2019; 21(5): 781786, doi: 10.1093/europace/euy320, indexed in Pubmed: 30698694.
  14. Al-Hijji MA, Killu AM, Yousefian O, et al. Outcomes of lead extraction without subsequent device reimplantation. Europace. 2017; 19(9): 15271534, doi: 10.1093/europace/euw184, indexed in Pubmed: 27707785.
  15. Iskos D, Lurie KG, Sakaguchi S, et al. Termination of implantable pace­maker therapy: experience in five patients. Ann Intern Med. 1997; 126(10): 787790, doi: 10.7326/0003-4819-126-10-199705150-00007, indexed in Pubmed: 9148652.
  16. Kutarski A, Czajkowski M, Pietura R, et al. Effectiveness, safety, and long-term outcomes of non-powered mechanical sheaths for transvenous lead extraction. Europace. 2018; 20(8): 13241333, doi: 10.1093/europace/eux218, indexed in Pubmed: 29016781.
  17. Tułecki Ł, Polewczyk A, Jacheć W, et al. A study of major and minor complications of 1500 transvenous lead extraction procedures performed with optimal safety at two high-volume referral centers. Int J Environ Res Public Health. 2021; 18(19), doi: 10.3390/ijerph181910416, indexed in Pubmed: 34639716.
  18. Kutarski A, Jacheć W, Nowosielecka D, et al. Unexpected procedure difficulties increasing the complexity of transvenous lead extraction: the single centre experience with 3721 procedures. J Clin Med. 2023; 12(8), doi: 10.3390/jcm12082811, indexed in Pubmed: 37109149.
  19. Kosior J, Jacheć W, Polewczyk A, et al. A study of major and minor complications of 1500 transvenous lead extraction procedures performed with optimal safety at two high-volume referral centers. Int J Environ Res Public Health. 2021; 18(19): 242251, doi: 10.3390/ijerph181910416, indexed in Pubmed: 34639716.
  20. Jacheć W, Polewczyk A, Polewczyk M, et al. Transvenous lead extraction safety score for risk stratification and proper patient selection for removal procedures using mechanical tools. J Clin Med. 2020; 9(2), doi: 10.3390/jcm9020361, indexed in Pubmed: 32013032.
  21. Menozzi C, Brignole M, Alboni P, et al. The natural course of untreated sick sinus syndrome and identification of the variables predictive of unfavorable outcome. Am J Cardiol. 1998; 82(10): 12051209, doi: 10.1016/s0002-9149(98)00605-5, indexed in Pubmed: 9832095.
  22. Dideriksen JR, Christiansen MK, Johansen JB, et al. Long-term outcomes in young patients with atrioventricular block of unknown aetiology. Eur Heart J. 2021; 42(21): 20602068, doi: 10.1093/eurheartj/ehab060, indexed in Pubmed: 33599276.
  23. Osmonov D, Erdinler I, Ozcan KS, et al. Management of patients with drug-induced atrioventricular block. Pacing Clin Electrophysiol. 2012; 35(7): 804810, doi: 10.1111/j.1540-8159.2012.03410.x, indexed in Pubmed: 22530749.
  24. Palmisano P, Dell’Era G, Pellegrino PL, et al. Causes of syncopal recurrences in patients treated with permanent pacing for bradyarrhythmic syncope: Findings from the SYNCOPACED registry. Heart Rhythm. 2021; 18(5): 770777, doi: 10.1016/j.hrthm.2021.01.010, indexed in Pubmed: 33465512.
  25. Aksu T, Mandrola J, Raj SR. Permanent pacing for recurrent vasovagal syncope: New answers or just more questions? J Electrocardiol. 2021; 65: 8890, doi: 10.1016/j.jelectrocard.2021.01.010, indexed in Pubmed: 33581411.
  26. Maloney JD, Jaeger FJ, Rizo-Patron C, et al. The role of pacing for the management of neurally mediated syncope: carotid sinus syndrome and vasovagal syncope. Am Heart J. 1994; 127(4 Pt 2): 10301037, doi: 10.1016/0002-8703(94)90083-3, indexed in Pubmed: 8160577.
  27. Gopinathannair R, Salgado BC, Olshansky B. Pacing for vasovagal syncope. Arrhythm Electrophysiol Rev. 2018; 7(2): 95102, doi: 10.15420/aer.2018.22.2, indexed in Pubmed: 29967681.
  28. Akella K, Olshansky B, Lakkireddy D, et al. Pacing therapies for vasovagal syncope. J Atr Fibrillation. 2020; 13(1): 2406, doi: 10.4022/jafib.2406, indexed in Pubmed: 33024506.
  29. Dupliakov DV, Golovina GA, Sysuenkova EV, et al. Clinical features and history of vasovagal syncope [in Russian]. Kardiologiia. 2012; 52(5): 4855, indexed in Pubmed: 22839586.
  30. Biffi M, Boriani G, Bronzetti G, et al. Neurocardiogenic syncope in selected pediatric patients--natural history during long-term follow-up and effect of prophylactic pharmacological therapy. Cardiovasc Drugs Ther. 2001; 15(2): 161167, doi: 10.1023/a:1011179014084, indexed in Pubmed: 11669410.
  31. Shah M, Silka M, Silva J, et al. 2021 PACES expert consensus statement on the indications and management of cardiovascular implantable electronic devices in pediatric patients. Cardiology in the Young. 2021; 31(11): 17381769, doi: 10.1017/s1047951121003413.
  32. Kini V, Soufi MK, Deo R, et al. Appropriateness of primary prevention implantable cardioverter-defibrillators at the time of generator replacement: are indications still met? J Am Coll Cardiol. 2014; 63(22): 23882394, doi: 10.1016/j.jacc.2014.03.025, indexed in Pubmed: 24727249.
  33. Sterliński M, Przybylski A, Gepner K, et al. Over 10 years with an implantable cardioverter-defibrillator - a long term follow-up of 60 patients. Kardiol Pol. 2010; 68(9): 10231029, indexed in Pubmed: 20859894.
  34. Hoshiba Y, Sugimoto A, Doi S, et al. Discontinuation of cardiac resynchronization therapy for heart failure due to dilated cardiomyopathy in a 61-year-old female “-super-responder” with return of a reduced left ventricular ejection fraction to normal. Am J Case Rep. 2020; 21: e926704, doi: 10.12659/AJCR.926704, indexed in Pubmed: 32986689.
  35. D’Angelo G, Zweiker D, Fierro N, et al. Check the need-prevalence and outcome after transvenous cardiac implantable electric device extraction without reimplantation. J Clin Med. 2021; 10(18), doi: 10.3390/jcm10184043, indexed in Pubmed: 34575152.