Case report
Twiddler syndrome – a rare complication of ICD implantation

Michał Waśniewski, Przemysław Mitkowski, Krystian Stanek, Romuald Ochotny, Andrzej Cieśliński
DOI: 10.33963/v.kp.80787
Kardiol Pol 2007;65(9):1094-1096.

Abstract

Twiddler syndrome is a rare complication after pacemaker or ICD implantation. We present a case of a man who experienced 7 inappropriate ICD shocks because of this complication. Treatment included total system removal and new device implantation on the right side