Vol 82, No 6 (2024)
Original article
Published online: 2024-04-24

open access

Page views 529
Article views/downloads 377
Get Citation

Connect on Social Media

Connect on Social Media

New-onset atrial high-rate episodes in left bundle branch area pacing versus right ventricular pacing for patients with atrioventricular block

Shigeng Zhang1, Wen Yang2, Shengchan Wang1, Yandi Cheng1, Zhixin Jiang2, Xiujuan Zhou2, Qijun Shan2
Pubmed: 38712772
Pol Heart J 2024;82(6):632-639.

Abstract

Background: Left bundle branch area pacing (LBBAP) demonstrated beneficial effects on clinical outcomes. Comparative data on the risk of atrial high-rate episodes (AHREs) between LBBAP and right ventricular pacing (RVP) are lacking.
Aims: This study aimed to investigate whether LBBAP can reduce the risk of new-onset AHREs compared with RVP in patients with atrioventricular block (AVB).
Methods: We enrolled 175 consecutive AVB patients with no history of atrial fibrillation undergoing dual-chamber pacemaker implantation (LBBAP or RVP). Propensity score matching for baseline characteristics yielded 43 matched pairs. The primary outcome was new-onset AHREs detected on a scheduled device follow-up. Changes in echocardiographic measurements were also compared between the groups.
Results: New-onset AHREs occurred in 42 (24.0%) of all enrolled patients (follow-up 14.1 [7.5] months) and the incidence of new-onset AHREs in the LBBAP group was lower than in the RVP group (19.8% vs. 34.7%; P = 0.04). After propensity score matching, LBBAP still resulted in a lower incidence of new-onset AHREs (11.6% vs. 32.6%; P = 0.02), and a lower hazard ratio for new-onset AHREs compared with RVP (HR, 0.274; 95% CI, 0.113–0.692). At 1 year, LBBAP achieved preserved left ventricular ejection fraction (LVEF) (63.0 [3.2]% to 63.1 [3.1]%; P = 0.56), while RVP resulted in reduced LVEF (63.4 [4.9]% to 60.5 [7.3]%; P = 0.01]). Changes in LVEF were significantly different between the 2 groups (by 2.6% [0.2 to 5.0]%; P = 0.03).
Conclusion: LBBAP demonstrated a reduced risk of new-onset AHREs compared with RVP in patients with AVB.

ORIGINAL ARTICLE

New-onset atrial high-rate episodes in left bundle branch area pacing versus right ventricular pacing for patients with atrioventricular block

Shigeng Zhang1Wen Yang2Shengchan Wang1Yandi Cheng1Zhixin Jiang2Xiujuan Zhou2Qijun Shan2
11st Clinical Medical School of Nanjing Medical University, Nanjing, China
2Department of Cardiology, 1st Affiliated Hospital of Nanjing Medical University, Nanjing, China

Correspondence to:

Qijun Shan, MD, PhD, FHRS,

Department of Cardiology,

1st Affiliated Hospital of Nanjing Medical University,

Nanjing, 210029, China,

phone: (+86) 13 505 142 015,

e-mail: qjshan@njmu.edu.cn

Copyright by the Author(s), 2024

DOI: 10.33963/v.phj.100403

Received: January 6, 2024

Accepted: April 24, 2024

Early publication date: April 24, 2024

ABSTRACT
Background: Left bundle branch area pacing (LBBAP) demonstrated beneficial effects on clinical outcomes. Comparative data on the risk of atrial high-rate episodes (AHREs) between LBBAP and right ventricular pacing (RVP) are lacking.
Aims: This study aimed to investigate whether LBBAP can reduce the risk of new-onset AHREs compared with RVP in patients with atrioventricular block (AVB).
Methods: We enrolled 175 consecutive AVB patients with no history of atrial fibrillation undergoing dual-chamber pacemaker implantation (LBBAP or RVP). Propensity score matching for baseline characteristics yielded 43 matched pairs. The primary outcome was new-onset AHREs detected on a scheduled device follow-up. Changes in echocardiographic measurements were also compared between the groups.
Results: New-onset AHREs occurred in 42 (24.0%) of all enrolled patients (follow-up 14.1 [7.5] months) and the incidence of new-onset AHREs in the LBBAP group was lower than in the RVP group (19.8% vs. 34.7%; P = 0.04). After propensity score matching, LBBAP still resulted in a lower incidence of new-onset AHREs (11.6% vs. 32.6%; P = 0.02), and a lower hazard ratio for new-onset AHREs compared with RVP (HR, 0.274; 95% CI, 0.1130.692). At 1 year, LBBAP achieved preserved left ventricular ejection fraction (LVEF) (63.0 [3.2]% to 63.1 [3.1]%; P = 0.56), while RVP resulted in reduced LVEF (63.4 [4.9]% to 60.5 [7.3]%; P = 0.01]). Changes in LVEF were significantly different between the 2 groups (by 2.6% [0.2 to 5.0]%; P = 0.03).
Conclusion: LBBAP demonstrated a reduced risk of new-onset AHREs compared with RVP in patients with AVB.
Key words: atrial high-rate episodes, atrioventricular block, left bundle branch area pacing, right ventricular pacing

WHAT’S NEW?

Left bundle branch area pacing (LBBAP) demonstrated a lower incidence of new-onset atrial high-rate episodes (AHREs), and LBBAP was associated with a significantly reduced risk of new-onset AHREs by 73% compared with right ventricular pacing (RVP) in atrioventricular block patients. LBBAP showed stable left ventricular ejection fraction and decreased left atrial diameter compared with RVP at 1-year follow-up.

INTRODUCTION

Conventional right ventricular pacing (RVP) induces ventricular electro-mechanical dyssynchrony and is associated with increased risk of heart failure hospitalization, atrial fibrillation (AF), and mortality in patients with high pacing burden [1]. Conduction system pacing could achieve favorable ventricular synchrony [2], and both His-bundle pacing (HBP) and left bundle branch area pacing (LBBAP) have been shown to reduce the incidence of new-onset AF in bradycardia patients [3–5].

Since the presentation of AF may be asymptomatic and intracardiac electrograms records are not always available, previous studies with clinical AF as the primary outcome should be interpreted with caution. Moreover, the patient population in previous studies was heterogeneous because it included patients with sinus node dysfunction and atrioventricular block (AVB). Atrial high-rate episodes (AHREs) represent continuous detection of atrial tachycardias by implanted cardiac devices; they are strongly associated with clinical AF, elevated stroke risk, and long-term mortality outcomes [6–8]. The present study was conducted to explore the effect of LBBAP on new-onset AHREs compared with RVP in AVB patients.

METHODS

Study design and participants

This prospective observational study was conducted at the First Affiliated Hospital of Nanjing Medical University, Nanjing; and the Affiliated People’s Hospital of Jiangsu University. We enrolled consecutive AVB patients with estimated ventricular pacing percentage (VP) >20% who underwent de novo successful dual-chamber pacemaker implantation if they had no AF history between January 2019 and June 2022. The pacing strategies were determined by operators according to clinical practice and not driven by the study. Patients were excluded if they (1) were younger than 18 years old; (2) had left ventricular ejection fraction (LVEF) <50% at baseline; (3) had severe valvular disease, congenital heart disease, or hypertrophic cardio­myopathy; (4) had myocardial infarction or open heart surgery within the past 3 months; (5) had indications for cardiac resynchronization therapy or implantable cardioverter defibrillator; (6) had concomitant serious diseases such as malignancy; (7) had a previous history of hyperthyroidism; (8) could not regularly attend appointments at our centers. This study was approved by the institutional review board of both hospitals, and all patients provided written informed consent.

Procedures

LBBAP was performed using the Select Secure pacing lead (model 3830, 69 cm, Medtronic) delivered through a fixed curve sheath (C315HIS, Medtronic) as previously described [9]. A delivery sheath was inserted into the right ventricle over a long guide wire through the subclavian or axillary vein. The pacing lead was then advanced through the sheath to the tip of the catheter. To identify the potential screwing site, a His bundle electrogram was identified first, and the system was advanced 1.02.0 cm along an imaginary line between the His bundle and the right ventricular apex. Unipolar pacing was performed at an output of 2.0 v/0.4 ms before screwing, and the paced QRS complex in lead V1 always displayed “W” morphology. The lead was then screwed into the interventricular septum by clockwise rotations until the right bundle branch block morphology of the paced QRS complex presented in electrocardiogram lead V1. Pacing stimulus to left ventricular activation time (LVAT) was measured repeatedly at low (2.0 v/0.4 ms) and high (5.0 v/0.4 ms) outputs in lead V6. LBBAP was considered successful if the unipolar paced QRS morphology presented with a right bundle branch block pattern along with a demonstration of transition from nonselective to selective LBB/left ventricular septum during threshold testing, or shortest and constant LVAT at high and low output pacing (commonly ≤75 ms), or a sudden increase in LVAT >10 ms at reduced pacing outputs. During right ventricular pacing, the right ventricular lead was inserted in a standard manner into the right ventricular septum or apex (RVA) based on operator preference.

Device programming was different between the two pacing modalities. In the LBBAP group, for patients with complete AVB, the atrioventricular (AV) delay was set as 120/90 ms, whereas for patients with incomplete AVB, the AV delay was set 30 ms longer than the intrinsic AV interval if the patient had a normal intrinsic QRS complex, and 30 ms shorter than the intrinsic AV interval if the patient had baseline bundle branch block to possibly correct the electrical dyssynchronization. The automatic AV delay optimization algorithms were turned off. In the RVP group, however, the AV delay was set as 150/120 ms for patients with complete AVB and 30 ms longer than the intrinsic AV interval for patients with incomplete AVB, and the automa­tic AV delay optimization algorithms were routinely turned on to avoid unnecessary ventricular pacing for patients with intermittent AVB.

Data collection and follow-up

We collected baseline demographics, comorbidities, prior medication history, electrocardiogram, and echocardiographic parameters. Patients underwent follow-up at 3, 6, and 12 months and annually after implantation. Pacing parameters were routinely documented. Ventricular pacing burden was recorded at the end of follow-up and censored to an early date if the primary outcome was reached. Echocardiographic evaluations were conducted at baseline and 12 months after the procedure. Biplane Simpson’s method on two-dimensional transthoracic echocardiography was used to evaluate LVEF.

The primary outcome was new-onset AHREs detected on a scheduled device follow-up. If no AHREs occurred during follow-up, patients would be censored at the last follow-up or death; once patients suffered from AHREs, the subjects were censored immediately. All data and follow-up dates were censored after June 30, 2023. AHREs were defined as events with an atrial frequency of ≥175 bpm and a duration of ≥5 min detected by a pacemaker device [10]. All episodes of pacemaker-detected AHREs were documented and reviewed both by physicians and experts from the pacemaker manufacturing company.

Statistical analysis

Continuous data were presented as means (standard deviation), and categorical data were summarized as frequency (percentages). The χ2 or Fisher’s exact test was used to analyze categorical data. Student’s t-test was used to analyze continuous data. The KaplanMeier survival curve and log-rank test were employed to estimate cumulative event rates. The effect of individual variables on the risk of new-onset AHREs was investigated by using univariate Cox proportional hazard models applied to the whole study population. Baseline variables considered to be clinically relevant or univariate predictors with P <0.1 were entered into multivariable Cox proportional hazard models. Echocardiographic parameters were compared between groups with the analysis of covariance (ANCOVA), which took into account baseline values. To adjust for bias due to potential confounders, a propensity score was computed for eligible participants using binary logistic regression which incorporated pacing modality (LBBAP or RVP) as dependent variables and baseline variables including age, sex, heart failure, hypertension, coronary artery disease, diabetes, LVEF, left atrial diameter (LAD), and intrinsic QRS duration (QRSd) as independent variables. Then, patients were matched 1:1 with a caliper as 0.02. Analysis was performed using SPSS version 26.0 (IBM Corporation, Armonk, NY, US). Statistical significance was set at P <0.05, and all tests were 2-sided.

RESULTS

Baseline characteristics

A total of 175 consecutive patients were enrolled. LBBAP was successful in 126 patients while 49 patients underwent successful RVP (30 paced from the right ventricular septum and 19 from the right ventricular apex) (Figure 1).

Figure 1. Flowchart of the study population enrollment
Abbreviations: LBBAP, left bundle branch area pacing; LVEF, left ventricular ejection fraction; RVP, right ventricular pacing

In the unmatched cohort, several baseline characteristics (e.g., age, hypertension) differed significantly between the two groups. Since AVB was an inclusion criterion, the QRSd at baseline was slightly longer than normal values in both groups. Propensity score matching identified 43 pairs of patients with balanced baseline characteristics, which were used for the final analysis (Table 1).

Table 1. Baseline characteristics of the general patients and the propensity score-matched cohort

General population

Propensity score matched

LBBAP

(n = 126)

RVP

(n = 49)

P-value

LBBAP

(n = 43)

RVP

(n = 43)

P-value

Age, years

68.3 (13.6)

72.9 (10.0)

0.02

75.0 (10.6)

72.4 (10.0)

0.24

Male, n (%)

74 (58.7)

29 (59.2)

0.96

28 (65.1)

26 (60.5)

0.66

HF, n (%)

36 (28.6)

7 (14.3)

0.049

11 (25.6)

7 (16.3)

0.29

HTN, n (%)

76 (60.3)

38 (77.6)

0.03

36 (83.7)

33 (76.7)

0.42

CAD, n (%)

32 (25.4)

7 (14.3)

0.11

11 (25.6)

7 (16.3)

0.29

DM, n (%)

33 (26.2)

10 (20.4)

0.43

13 (30.2)

9 (20.9)

0.32

LVEF (%)

63.0 (3.2)

63.4 (4.9)

0.54

62.2 (2.5)

63.3 (5.2)

0.24

LAD, mm

38.1 (4.6)

37.8 (4.5)

0.71

38.6 (5.0)

38.1 (4.5)

0.64

LVEDD, mm

48.7 (4.7)

49.1 (4.3)

0.62

48.3 (5.2)

49.2 (4.3)

0.39

Intrinsic QRSd, ms

119.4 (30.3)

110.3 (26.6)

0.07

116.3 (26.2)

113.1 (27.2)

0.58

β blocker, n (%)

32 (25.4)

10 (20.4)

0.49

12 (27.9)

8 (18.6)

0.31

ACEI/ARB, n (%)

54 (42.9)

27 (55.1)

0.15

29 (67.4)

23 (53.5)

0.19

Electrophysiological and pacing characteristics

Compared with RVP, LBBAP showed a lower pacing thres­hold, similar R wave amplitude, and pacing impedance on implantation. During follow-up, the pacing threshold was comparable between the two groups, while better R wave amplitude and lower pacing impedance was observed in the LBBAP group. LBBAP showed a higher ventricular pacing percentage (VP%) (99.6 [1.0]% vs. 88.1 [20.9]%; P = 0.001), which may have been caused by turning off the automatic AV delay optimization algorithms in this group. Paced QRSd was narrower in the LBBAP group than in the RVP group (114.7 [12.2] ms vs. 167.1 [12.9] ms; P <0.001). Pacing characteristics are listed in Table 2.

Table 2. Pacing characteristics between LBBAP and RVP

LBBAP

(n = 43)

RVP

(n = 43)

P-value

Baseline

Sense, mV

13.1 (5.5)

11.5 (3.3)

0.13

Threshold, V/0.4 ms

0.51 (0.12)

0.84 (0.20)

<0.001

Impedance, Ω

825.2 (179.7)

839.8 (201.8)

0.73

Follow-up

Sense, mV

14.5 (3.9)

10.7 (3.9)

0.001

Threshold, V/0.4 ms

0.70 (0.19)

0.75 (0.23)

0.33

Impedance, Ω

478.4 (75.8)

553.5 (119.9)

0.001

Paced QRSd, ms

114.7 (12.2)

167.1 (12.9)

<0.001

VP, %

99.6 (1.0)

88.1 (20.9)

0.001

Primary outcomes

During a mean follow-up duration of 14.1 (7.5) months, new-onset AHREs occurred in 42 (24.0%) of all enrolled patients, and the incidence rate of new-onset AHREs in the LBBAP group was lower than in the RVP group (19.8% vs. 34.7%; P = 0.04). Table S1 (Supplementary material) presents the univariate analysis of baseline characteristics and potential predisposing factors for new-onset AHREs. In multivariable analysis, LBBAP was independently associated with lower risk of new-onset AHREs (HR, 0.368; 95% CI, 0.1830.738; P = 0.005), while age increased the risk of new-onset AHREs (Figure 2 and Figure 3).

Figure 2. Comparison of new-onset AHREs between LBBAP and RVP groups. A. General patients. B. Propensity score matched cohort
Abbreviations: AHREs, atrial high rate episodes; other see Figure 1
Figure 3. Multivariable Cox regression analysis of new-onset AHREs
Abbreviations: CI, confidence interval; HR, hazard ratio; other see Figure 1 and Table 1 and 2

In the matched cohort, the LBBAP group showed markedly longer follow-up duration than RVP: 16.0 (7.6) months and 11.8 (5.3) months, respectively (P = 0.004). Nevertheless, new-onset AHREs occurred in 5 patients in the LBBAP group (11.6%), and 14 patients in the RVP group (32.6%) (P = 0.02). Patients with LBBAP showed longer AHRE-free survival than RVP patients (HR, 0.274; 95% CI, 0.1130.692; P = 0.007) (Figure 2).

Echocardiographic measurements

Compared with baseline, LBBAP showed stable LVEF at 1-year follow-up. Conversely, the RVP group showed decreased LVEF at 1-year follow-up. Changes in LVEF were significantly different between the treatment groups. The ANCOVA treatment effect was 2.6% in favor of LBBAP (Table 3, Figure 4). The LAD showed a slight increase after 1 year of pacing in the LBBAP group and remained unchanged in the RVP group while there were no other significant differences between the groups. The left ventricular end-diastolic diameter was nearly unchanged in both groups and was not significantly different between the groups.

Table 3. Changes of echocardiographic measurements between LBBAP and RVP

LBBAP

RVP

ANCOVA effect

P-value

LVEF, %

Baseline

63.0 (3.2)

63.4 (4.9)

–2.6 (–5.0, –0.2)

0.03

1-year follow-up

63.1 (3.1)

60.5 (7.3)a

LAD, mm

Baseline

38.1 (4.6)

37.8 (4.5)

0.2 (–1.1, 1.4)

0.79

1-year follow-up

37.5 (4.2)#

37.6 (4.4)

LVEDD, mm

Baseline

48.7 (4.7)

49.1 (4.3)

1.0 (–0.3, 2.3)

0.12

1-year follow-up

47.5 (4.0)

48.5 (4.4)

Figure 4. Comparison of echocardiographic measurements between LBBAP and RVP groups
Abbreviations: see Figure 1 and Table 1

DISCUSSION

In this prospective observational study, we evaluated the impact of LBBAP and RVP on the incidence of new-onset AHREs in AVB patients and demonstrated a lower incidence of new-onset AHREs with LBBAP. LBBAP was associated with a 63% relative risk reduction compared with RVP. During 1-year follow-up, LBBAP achieved preserved LVEF, while RVP resulted in reduced LVEF.

Previous studies have demonstrated that long-term RVP is associated with increased risk of AF, heart failure hospitalization, and mortality [1, 11]. The risk of AF increased linearly with VP% in a dual-chamber pacemaker group ([HR, 1.36; 95% CI, 1.09 – 1.69] for each 25% increase in VP%) [1]. HBP, as the most physiological pacing modality, could preserve or improve ventricular synchrony and has been associated with reduced risk of AF. Among patients with no history of AF, HBP demonstrated a lower risk of new-onset AF (HR, 0.53; 95% CI, 0.280.99) [3], and HBP also decreased the risk of persistent/permanent AF (HR, 0.28; 95% CI, 0.160.48) [12]. LBBAP, as one of the physiological pacing modalities, has been shown to have an obvious beneficial effect on clinical outcomes compared with RVP [13, 14]. Similar to HBP, LBBAP has been reported to decrease the risk of new-onset AF, and this kind of effect seems to be more pronounced in patients with VP% ≥20% [4, 5].

Nowadays, AHREs can be continuously recorded by cardiac implantable electronic devices to document episodes of AF and other atrial tachyarrhythmia. With AHREs as the primary outcome, this study showed a significantly decreased incidence of new-onset AHREs compared with RVP in AVB patients. Previous studies have shown that AHREs could increase the risk of clinical AF, ischemic stroke, and mortality outcomes. In the meta-analysis by Mahajan et al. [6], patients with documented AHREs were 5.7 times more likely to have documented clinical AF during the follow-up period. An ancillary study of the Mode Selection Trial indicated that the risk of death or stroke was increased by a factor of 2.5 in patients who had at least one episode of AHREs >5 min [15]. AHREs are commonly encountered in pacemaker patients without previous AF history [6], and it is of great importance to decrease their incidence.

Factors that predispose patients to AHREs are not clear. In 2020, Witt et al. [16] found that left atrial emptying fraction and left atrial minimum volumes (LAmin) assessed by cardiac computed tomography were significantly associated with AHREs (HR, 0.95; 95% CI, 0.920.98, and HR, 1.02; 95% CI, 1.001.05). In the same year, Kishima et al. [17] found that left ventricular stiffness assessed by diastolic wall strain was associated with AHREs in patients with a dual-chamber pacemaker. They speculated that increased left ventricular stiffness augmented left ventricular filling pressure and further led to left atrial remodeling, which may then induce AHREs. Unfortunately, right ventricular pacing could result in left atrial remodeling and reduced atrial function, which may be related to elevated filling pressures and impairment of left ventricular systolic function [18]. This would explain why minimizing ventricular pacing could reduce AF incidence in patients with sinus node disease.

However, generally minimizing ventricular pacing is not practical for AVB patients, and physiological pacing modality may be the best option for such patients. As the most physiological pacing option for ventricular pacing, HBP resulted in a more physiological left ventricular electromechanical activation/relaxation and better left atrial function compared with RVP [19]. LBBAP has provided an alternative pacing site for lead implantation along the HisPurkinje system. Previous studies have demonstrated beneficial effects of LBBAP on cardiac function. LBBAP could preserve satisfactory left ventricular intraventricular synchrony and improve interventricular dyssynchrony compared with RVP [20]. Liu et al. [21] found increased left atrial myocardial elasticity and left atrial strain capacity with LBBAP. Our study also demonstrated improved LVEF and decreased LAD after LBBAP when compared with RVP.

As conduction system pacing, both LBBAP and HBP have shown beneficial effects on clinical outcomes compared with RVP. However, HBP has some limitations. Studies have reported a gradually increased capture threshold and sensing issues, and the success rates for HBP varied between 65% and 92% [22–24]. LBBAP has been used for years in clinical practice, and it has been evaluated as a safe and feasible pacing modality. Su et al. [25] demonstrated a success rate of 97.8% in patients undergoing LBBAP with stable thresholds during a mean follow-up of 18.6 months. Particularly, LBBAP is safe and effective in patients ≥80 years old [26], and as AHREs may increase with aging [27, 28], older people may benefit more from LBBAP. Additionally, HBP was associated with significantly higher risk of complications compared with LBBAP (8.6% vs. 1.3%; P = 0.04) mainly because of more lead-related complications, whereas LBBAP was associated with risk of complications similar to that of RVP (3.5% vs. 1.3%; P = 0.36) [29]. Thus, LBBAP may be more promising for patients with AVB.

Study limitations

This study had several limitations. It was a non-randomized controlled study with a relatively small sample size. Since there were significantly different baseline characteristics between the two study groups, we used propensity score matching to balance the cohorts; however, this resulted in discarding 50.9% of the original sample, which may have led to decreased statistical precision in results. Nevertheless, statistical analysis before and after propensity score matching demonstrated consistent results, strengthening our conclusions.

By using a cutoff of >6 minutes, the rate of false-positive AHREs was 17.3% [8]. These AHREs were subcategorized as noise, sensing of farfield R wave, or repetitive non-reentrant ventriculoatrial synchrony. Although each electrogram was reviewed by physicians, the accuracy of AHRE categorization could not be fully guaranteed.

CONCLUSIONS

The results of this observational study indicate that LBBAP was associated with lower risk of new-onset AHREs compared with conventional RVP in patients with a high burden of VP%. Randomized trials with larger sample sizes are needed to further confirm these findings.

Supplementary material

Supplementary material is available at https://journals.viamedica.pl/polish_heart_journal.

Article information

Conflict of interest: None declared.

Funding: None.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, which allows downloading and sharing articles with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at polishheartjournal@ptkardio.pl

REFERENCES

  1. Sweeney MO, Hellkamp AS, Ellenbogen KA, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. 2003; 107(23): 29322937, doi: 10.1161/01.CIR.0000072769.17295.B1, indexed in Pubmed: 12782566.
  2. Hou X, Qian Z, Wang Y, et al. Feasibility and cardiac synchrony of permanent left bundle branch pacing through the interventricular septum. Europace. 2019; 21(11): 16941702, doi: 10.1093/europace/euz188, indexed in Pubmed: 31322651.
  3. Ravi V, Beer D, Pietrasik GM, et al. Development of new-onset or progressive atrial fibrillation in patients with permanent HIS bundle pacing versus right ventricular pacing: Results from the RUSH HBP registry. J Am Heart Assoc. 2020; 9(22): e018478, doi: 10.1161/JAHA.120.018478, indexed in Pubmed: 33174509.
  4. Ravi V, Sharma PS, Patel NR, et al. New-onset atrial fibrillation in left bundle branch area pacing compared with right ventricular pacing. Circ Arrhythm Electrophysiol. 2022; 15(4): e010710, doi: 10.1161/CIRCEP.121.010710, indexed in Pubmed: 35333096.
  5. Zhu H, Li X, Wang Z, et al. New-onset atrial fibrillation following left bundle branch area pacing vs. right ventricular pacing: A two-centre prospective cohort study. Europace. 2023; 25(1): 121129, doi: 10.1093/europace/euac132, indexed in Pubmed: 35942552.
  6. Mahajan R, Perera T, Elliott AD, et al. Subclinical device-detected atrial fibrillation and stroke risk: a systematic review and meta-analysis. Eur Heart J. 2018; 39(16): 14071415, doi: 10.1093/eurheartj/ehx731, indexed in Pubmed: 29340587.
  7. Gonzalez M, Keating RJ, Markowitz SM, et al. Newly detected atrial high rate episodes predict long-term mortality outcomes in patients with permanent pacemakers. Heart Rhythm. 2014; 11(12): 22142221, doi: 10.1016/j.hrthm.2014.08.019, indexed in Pubmed: 25131667.
  8. Kaufman ES, Israel CW, Nair GM, et al. Positive predictive value of device-detected atrial high-rate episodes at different rates and durations: an analysis from ASSERT. Heart Rhythm. 2012; 9(8): 12411246, doi: 10.1016/j.hrthm.2012.03.017, indexed in Pubmed: 22440154.
  9. Huang W, Chen X, Su L, et al. A beginner’s guide to permanent left bundle branch pacing. Heart Rhythm. 2019; 16(12): 17911796, doi: 10.1016/j.hrthm.2019.06.016, indexed in Pubmed: 31233818.
  10. Hindricks G, Potpara T, Dagres N, et al. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J. 2021; 42(5): 373498, doi: 10.1093/eurheartj/ehaa612, indexed in Pubmed: 32860505.
  11. Udo EO, van Hemel NM, Zuithoff NPA, et al. Risk of heart failure- and cardiac death gradually increases with more right ventricular pacing. Int J Cardiol. 2015; 185: 95100, doi: 10.1016/j.ijcard.2015.03.053, indexed in Pubmed: 25804349.
  12. Pastore G, Zanon F, Baracca E, et al. The risk of atrial fibrillation during right ventricular pacing. Europace. 2016; 18(3): 353358, doi: 10.1093/europace/euv268, indexed in Pubmed: 26443444.
  13. Li X, Zhang J, Qiu C, et al. Clinical outcomes in patients with left bundle branch area pacing vs. right ventricular pacing for atrioventricular block. Front Cardiovasc Med. 2021; 8: 685253, doi: 10.3389/fcvm.2021.685253, indexed in Pubmed: 34307499.
  14. Sharma PS, Patel NR, Ravi V, et al. Clinical outcomes of left bundle branch area pacing compared to right ventricular pacing: Results from the Geisinger-Rush Conduction System Pacing Registry. Heart Rhythm. 2022; 19(1): 311, doi: 10.1016/j.hrthm.2021.08.033, indexed in Pub­med: 34481985.
  15. Glotzer T, Hellkamp A, Zimmerman J, et al. Atrial high rate episodes detected by pacemaker diagnostics predict death and stroke. Circulation. 2003; 107(12): 16141619, doi: 10.1161/01.cir.0000057981.70380.45.
  16. Witt CT, Kronborg MB, Sommer A, et al. Left atrial function determined by cardiac computed tomography predicts device-detected atrial high-rate episodes in patients treated with cardiac resynchronization therapy. J Comput Assist Tomogr. 2020; 44(5): 784789, doi: 10.1097/RCT.0000000000001038, indexed in Pubmed: 32558773.
  17. Kishima H, Mine T, Fukuhara E, et al. Left ventricular stiffness assessed by diastolic Wall strain predicts asymptomatic atrial high-rate episodes in patients with pacemaker implantation. J Cardiol. 2021; 77(2): 195200, doi: 10.1016/j.jjcc.2020.08.002, indexed in Pubmed: 32888831.
  18. Xie JM, Fang F, Zhang Q, et al. Left atrial remodeling and reduced atrial pump function after chronic right ventricular apical pacing in patients with preserved ejection fraction. Int J Cardiol. 2012; 157(3): 364369, doi: 10.1016/j.ijcard.2010.12.075, indexed in Pubmed: 21239072.
  19. Pastore G, Aggio S, Baracca E, et al. Hisian area and right ventricular apical pacing differently affect left atrial function: an intra-patients evaluation. Europace. 2014; 16(7): 10331039, doi: 10.1093/europace/eut436, indexed in Pubmed: 24473501.
  20. Cai B, Huang X, Li L, et al. Evaluation of cardiac synchrony in left bundle branch pacing: Insights from echocardiographic research. J Cardiovasc Electrophysiol. 2020; 31(2): 560569, doi: 10.1111/jce.14342, indexed in Pubmed: 31919928.
  21. Liu Q, Yang J, Bolun Z, et al. Comparison of cardiac function between left bundle branch pacing and right ventricular outflow tract septal pacing in the short-term: A registered controlled clinical trial. Int J Cardiol. 2021; 322: 7076, doi: 10.1016/j.ijcard.2020.08.048, indexed in Pubmed: 32860843.
  22. Abdelrahman M, Subzposh FA, Beer D, et al. Clinical outcomes of His bundle pacing compared to right ventricular pacing. J Am Coll Cardiol. 2018; 71(20): 23192330, doi: 10.1016/j.jacc.2018.02.048, indexed in Pubmed: 29535066.
  23. Sharma PS, Dandamudi G, Naperkowski A, et al. Permanent His-bundle pacing is feasible, safe, and superior to right ventricular pacing in routine clinical practice. Heart Rhythm. 2015; 12(2): 305312, doi: 10.1016/j.hrthm.2014.10.021, indexed in Pubmed: 25446158.
  24. Vijayaraman P, Naperkowski A, Subzposh FA, et al. Permanent His-bundle pacing: Long-term lead performance and clinical outcomes. Heart Rhythm. 2018; 15(5): 696702, doi: 10.1016/j.hrthm.2017.12.022, indexed in Pubmed: 29274474.
  25. Su L, Wang S, Wu S, et al. Long-term safety and feasibility of left bundle branch pacing in a large single-center study. Circ Arrhythm Electrophy­siol. 2021; 14(2): e009261, doi: 10.1161/CIRCEP.120.009261, indexed in Pubmed: 33426907.
  26. Jiang Z, Chen Y, Chen C, et al. Feasibility and safety of left bundle branch area pacing in very elderly patients (≥80 years). Kardiol Pol. 2022; 80(4): 452460, doi: 10.33963/KP.a2022.0048, indexed in Pub­med: 35167114.
  27. Pastori D, Miyazawa K, Li Y, et al. Inflammation and the risk of atrial high-rate episodes (AHREs) in patients with cardiac implantable electronic devices. Clin Res Cardiol. 2018; 107(9): 772777, doi: 10.1007/s00392-018-1244-0, indexed in Pubmed: 29667016.
  28. Chen JH, Chen GY, Zheng H, et al. Atrial high-rate event incidence and predictors in patients with permanent pacemaker implantation. Front Cardiovasc Med. 2021; 8: 728885, doi: 10.3389/fcvm.2021.728885, indexed in Pubmed: 34708084.
  29. Palmisano P, Ziacchi M, Dell’Era G, et al. Rate and nature of complications of conduction system pacing compared with right ventricular pacing: Results of a propensity score-matched analysis from a multicenter registry. Heart Rhythm. 2023; 20(7): 984991, doi: 10.1016/j.hrthm.2023.03.009, indexed in Pubmed: 36906165.