English Polski
Tom 11, Nr 6 (2016)
Opis przypadku
Opublikowany online: 2016-09-15

dostęp otwarty

Wyświetlenia strony 1131
Wyświetlenia/pobrania artykułu 1666
Pobierz cytowanie

Eksport do Mediów Społecznościowych

Eksport do Mediów Społecznościowych

_08_FC_PK_Letek

PRACA KAZUISTYCZNA

The random placement of pacing lead in the left ventricle in a patient with patent foramen ovale

Przypadkowe umieszczenie elektrody stymulatora w lewej komorze serca u pacjenta z przetrwałym otworem owalnym

Agnieszka Łętek1, Jacek Kurzawski1, Marcin Sadowski1, 2

1II Department of Cardiology of the Świętokrzyskie Cardiology Center in Kielce

2Faculty of Medicine and Health Sciences, Jan Kochanowski University, Kielce, Poland

Adres do korespondencji: lek. Agnieszka Łętek, II Klinika Kardiologii, Świętokrzyskie Centrum Kardiologii, ul. Grunwaldzka 45, 25–736 Kielce, e-mail: agaletek@gmail.com

Abstract

The random placement of pacing lead in the left ventricle (LV) is a rare and late diagnosed complication of the permanent heart stimulation. In most cases the intracardiac electrode moves through the interatrial septum or through the patent foramen ovale to the LV. On ECG the sets of QRS have morphology of right bundle branch block. The decisive survey used to identify this complication is echocardiography, especially transesophageal study. The malposition of the pacing electrode is associated with the risk of thromboembolic complications. The treatment consists of early lead removal. In case of lack of such a possibility the chronic anticoagulation treatment has to be started. The paper presents a case of incorrect position of pacing electrode in LV in a patient with PFO.

Key words: cardiac pacing, inadvertent left ventricular pacing, lead removal, anticoagulation therapy

Folia Cardiologica 2016; 11, 6: 535–538

Case report

A 74-year-old male was admitted to the Department of Cardiology, due to atrioventricular arrhythmia (second degree atrioventricular block 2: 1) with MAS syndrome. During hospital admission the patient was secured by a temporary transvenosus pacing electrode. The next day the DDDR pacemaker was implanted. After surgery, on ECG recording right bundle branch block (RBBB) configuration of paced complexes were present (Figure 1A).

71886.png

Figure 1A, B. Electrocardiography: right bundle branch block (RBBB) pattern (A); left bundle branch block (LBBB) pattern (B)

The chests X-ray in posterior-anterior (PA) and lateral projections were correct (Figure 2).

71928.png

Figure 2A, B. Chest X-ray in posterior-anterior (PA) and lateral projections

The transthoracic echocardiography showed wrong location of pacing lead in the left ventricle (pacing electrode passed from the right atrium by the foramen ovale into the left atrium and then through the mitral valve into the left ventricle of the side wall) (Figure 3).

71919.png

Figure 3A, B. The transthoracic echocardiography. The arrow indicates pacing electrode [1]; LA — left atrium; RA — right atrium; LV — left ventricle; RV — right ventricle; Ao — aorta; TV — tricuspid valve; MV — mitral valve

To confirm a malposition of the electrode and to exclude possible artifacts, the transesophageal echocardiography was done. It also showed an incorrect position of ventricular electrode (Figure 4), current patent foramen ovale (PFO) and a trace of the left-to-right leak through the interatrial septum around the passage of the lead (Figure 5).

Letek_4.jpg

Figure 4. Transesophageal echocardiography. Pacing lead passing from the right to the left atrium (LA) and through the mitral valve (MV) into the left ventricle (LV); Ao — aorta; AV — aortic valve

Letek_5.jpg

Figure 5. Transesophageal echocardiography. Visible trace of the interatrial septum leak

The next day of hospitalization repositioning ventricular electrode was done. After the procedure, the control ECG showed normal right ventricular stimulation with left bundle branch block (LBBB) (Figure 1B). The echocardiography showed the correct location of the pacemaker electrodes, there was no damage to the structures of the heart or leakage through the septum. Control stimulation parameters were correct.

Discussion

The pacemaker implantation is the most common surgery in the invasive cardiology. Doctors are more often faced with problems of this treatment due to the fact that the population of pacemaker patients increases.

The article presents a very rare case of complication of the pacemaker implantation, which is the accidental placement of the pacing electrode in the left ventricle. The incidence of this complication is not known [1]. Most often this occurs during ventricular lead implantation in patients with PFO or interatrial or interventricular septum defect [2, 3]. The literature also describes isolated cases of the penetration interatrial, the interventricular septum or the right ventricular apex by the electrode displaced by the pericardial sac of the left ventricle. It is also described as a case of puncture of the subclavian artery and the introduction of the pacing electrode through the aortic valve into the left ventricle [3]. The abnormal ventricular lead position in the left ventricle of the heart is usually diagnosed late, most frequently during accidentally performed echocardiography. Therefore, the incidence of this complication is estimated by some authors to 2% of patients with an implanted pacemaker, which may be underestimated [1].

The delay in the diagnosis of abnormal location of the pacing electrode is influenced by several factors. During pacemaker implantation only electrocardiogram from limb leads are routinely assessed. Fluoroscopic image performed during surgery may be incorrectly evaluated as normal, especially when it is analyzed only by the PA projection [3]. Furthermore, the proper stimulation parameters measured intraoperatively are not differentiating because they are similar both in stimulating the right and left ventricle [2, 3]. Therefore, careful analysis and observation after pacemaker implantation are very important in order to identify the earliest possible complications.

The basic diagnostic tool to rank the correct position of the pacing electrode is a 12-lead ECG. Proper stimulation of the right ventricle gives a complete left bundle branch block (LBBB), usually with a sinistrogram. In contrast, stimulation of the left ventricle results in the right bundle branch block (RBBB), usually with the dextrogram. However, presence of RBBB in ECG during ventricular pacing not always proves wrong position of the electrode [4]. Right bundle branch block may occur in the case of placing the electrode in the coronary sinus or in the left ventricle of the heart or may occur as a result of septal or right ventricular free wall perforation. There are a few cases of atypical right bundle branch block (so-called the pseudo RBBB — in ECG in lead I, aVL is typical LBBB, and in V1–V3 is typical RBBB) in case of normal right ventricular apical pacing or apical segment of the septum in right ventricle, which can cause early left ventricular pacing. Moreover, if the ECG recording is technically wrong it can be mistakenly interpreted as a RBBB; the displacement of electrodes V1–V3 by one intercostal space below can normalize the ECG [4, 5]. The shape of the QRS with morphology RBBB during right ventricular pacing can be distinguished from the shape of the QRS with morphology RBBB during left ventricular pacing according to the criteria described by Okmen et al.: left axis deviation to the left in the frontal plane between –30 and –90 degrees, the transition zone in V3 lead (R-wave amplitude equal to the S-wave amplitude), the lack of S-wave in lead I and the sets of qR or RS in V1 lead. Nevertheless, each case of RBBB in ECG requires an accurate assessment of the electrode location [6].

The next step is to perform chest X-ray in the PA and lateral projections. Routine chest X-ray projection (PA) very often does not reveal any abnormalities. The lateral projection can be also correct in many cases. The most accurate study to evaluate the pacing lead fixation and to trace its exact course in the heart is transthoracic echocardiography, and especially transesophageal echocardiography [5, 7, 8]. In each case of suspicions of incorrectly located pacing leads the echocardiography should be performed.

Rapid diagnosis of pacemaker electrodes placed accidentally in the left ventricle is very important, because it may be a source of arterial embolism manifested by transient ischemic attack (TIA) or stroke. Thromboembolic complications occur in about 40% of patients with an electrode in the left ventricle of the heart who didn’t receive anticoagulation treatment [5, 7, 8]. In addition, there may be other complications such as perforation of the mitral valve, mitral regurgitation, endocarditis, or the loss of stimulation.

If the malposition of pacing lead is detected early after surgery it is possible to remove the electrodes usually by transvenous access. Transvenous lead extractions are effectively accomplished in many hospital centers by qualified doctors and with various techniques. Surgical lead extraction should be taken into consideration, especially if the cardiac surgery has to be performed because of other indications or in case of serious complications such as pericardial perforation [5, 7, 8]. If the leaving of pacing electrode in the left ventricle is necessary, anticoagulants (acenocoumarol or warfarin) must be given with therapeutic range of international normalization ratio (INR) between 2.5 and 3.0. Lower levels of anticoagulation or the antiplatelet treatment alone are ineffective [8].

Conclusions

The accidental introduction of pacing lead into the left ventricle is a potentially dangerous complication, which is rare, but can be overlooked during the routine pacemaker check. 12-lead ECG is recommended after each heart pacemaker implantation. It is the simplest test method used to assess the position of the pacing electrode. RBBB seen during ventricular pacing should arouse suspicion of improper position of the electrode in the left ventricle. Echocardiography is the most important study for evaluating location of pacing lead, while the chest X-ray has a little diagnostic value. The treatment of an inadvertent left ventricular pacing has two options: lead removal or chronic anticoagulation therapy.

Conflict of interest(s)

None declared.

Streszczenie

Przypadkowe umieszczenie elektrody stymulatora w lewej komorze (LV) jest rzadkim i późno rozpoznawanym powikłaniem stałej stymulacji serca. Najczęściej elektroda wewnątrzsercowa przemieszcza się przez przegrodę międzyprzedsionkową lub przetrwały otwór owalny (PFO) do lewego przedsionka, a następnie przez zastawkę dwudzielną do LV. W zapisie EKG w przypadku stymulacji LV obecne są zespoły QRS o morfologii bloku prawej odnogi pęczka Hisa. Decydującym badaniem służącym do rozpoznania tego powikłania jest badanie echokardiograficzne, zwłaszcza przezprzełykowe. Nieprawidłowe położenie elektrody w LV wiąże się ze znacznym ryzykiem powikłań zakrzepowo-zatorowych. Leczenie polega na wczesnym usunięciu nieprawidłowo położonej elektrody z LV lub, w przypadku braku takiej możliwości, na włączeniu na stałe leczenia przeciwzakrzepowego. W pracy przedstawiono przypadek nieprawidłowego umiejscowienia elektrody stymulującej w LV u pacjenta z PFO.

Słowa kluczowe: stymulacja serca, nieprawidłowe umieszczenie elektrody, antykoagulacja

Folia Cardiologica 2016; 11, 6: 535–538

References

  1. 1. Maziarz A., Ząbek A., Małecka B., Lelakowski J. An inadvertent placement of pacing lead in the left ventricle — an analysis of three cases. Pol. Przegl. Kardiol. 2010; 12: 164–168.
  2. 2. Shmuely H., Erdman S., Strasberg B., Rosenfeld J.B. Seven years of left ventricular pacing due to malposition of pacing electrode. PACE 1992; 15: 369–372.
  3. 3. Ghani M., Thacur R.K., Boughner D. et al. Malposition of transvenous pacing lead in the left ventricle. PACE 1993; 16: 1800–1807.
  4. 4. Yung-Nien Y, Wei-Hsian Y, Mason Shing Y. Case report. Safe right bundle branch block pattern during permanent right ventricular pacing. J. Electrocardiol. 2003; 36: 67–71.
  5. 5. Zielińska M., Kaczmarek K., Koniarek W. et al. The significance of electrocardiogram in the estimation lead position in patients with permanent ventricular pacing. Wiad. Lek. 2006; 59: 580–584.
  6. 6. Okmen E., Erdinler I., Oguz E. et al. An electrocardiographic algorithm for determining the location of pacemaker electrode in patients with right bundle branch block configuration during permanent ventricular pacing. Angiology 2006; 57: 623–630.
  7. 7. Moszczyńska-Stulin J., Kübler G., Goździk A. et al. An accidental localisation of transvenous pacemaker lead in the left ventricle — an uncommon or rare-diagnosed complication? Folia Cardiol. 2003; 10: 231–236.
  8. 8. Van Gelder B.M., Bracke F.A., Oto A. et al. Diagnosis and management of inadvertently placed pacing and ICD leads in the left ventricle: A multicenter experience and review of the literature. PACE 2000; 23: 877–883.