Progressive non-ischaemic cardiomyopathy resulting in T-wave oversensing and inappropriate shocks: case report

Bahri Köroğlu, Umut Celikyurt, Aysen Agacdiken , Ahmet Vural
DOI: 10.5603/FC.2018.0009
·
Folia Cardiologica 2018;13(1):51-54.

open access

Vol 13, No 1 (2018)
Case Reports
Published online: 2018-03-22

Abstract

T-wave oversensing is one of the causes of inappropriate shocks. It results from various conditions including progressive cardiomyopathy, diminution of R-wave amplitude and relative or dynamic gain in the T-wave amplitude, electrolyte abnormalities, injury current-related increase in the T-wave voltage. We report a 59-year-old patient who presented with implantable cardioverter defibrillator shocks after 5 year implantation with decreased R-wave amplitude and T-wave changes on 12-lead electrocardiography.

Abstract

T-wave oversensing is one of the causes of inappropriate shocks. It results from various conditions including progressive cardiomyopathy, diminution of R-wave amplitude and relative or dynamic gain in the T-wave amplitude, electrolyte abnormalities, injury current-related increase in the T-wave voltage. We report a 59-year-old patient who presented with implantable cardioverter defibrillator shocks after 5 year implantation with decreased R-wave amplitude and T-wave changes on 12-lead electrocardiography.

Get Citation

Keywords

progressive non-ischaemic cardiomyopathy, T-wave oversensing, inappropriate shocks

About this article
Title

Progressive non-ischaemic cardiomyopathy resulting in T-wave oversensing and inappropriate shocks: case report

Journal

Folia Cardiologica

Issue

Vol 13, No 1 (2018)

Pages

51-54

Published online

2018-03-22

DOI

10.5603/FC.2018.0009

Bibliographic record

Folia Cardiologica 2018;13(1):51-54.

Keywords

progressive non-ischaemic cardiomyopathy
T-wave oversensing
inappropriate shocks

Authors

Bahri Köroğlu
Umut Celikyurt
Aysen Agacdiken
Ahmet Vural

References (4)
  1. Jimenez-Lopez J, Arias MA, Akerstrom F, et al. A single shock during an episode of ventricular tachycardia: is the device working properly? Indian Pacing Electrophysiol J. 2012; 12(5): 226–228.
  2. Srivathsan K, Scott LR, Altemose GT. T-wave oversensing and inappropriate shocks: a case report. Europace. 2008; 10(5): 552–555.
  3. Rasania SP, Mountantonakis S, Patel VV. Inappropriate ICD shocks caused by T-wave oversensing due to acute alcohol intoxication. Pacing Clin Electrophysiol. 2012; 35(9): e267–e271.
  4. Watanabe H, Chinushi M, Izumi D, et al. Decrease in amplitude of intracardiac ventricular electrogram and inappropriate therapy in patients with an implantable cardioverter defibrillator. Int Heart J. 2006; 47(3): 363–370.

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

 