English Polski
Tom 11, Nr 3 (2016)
Opis przypadku
Opublikowany online: 2016-07-13

dostęp otwarty

Wyświetlenia strony 576
Wyświetlenia/pobrania artykułu 1636
Pobierz cytowanie

Eksport do Mediów Społecznościowych

Eksport do Mediów Społecznościowych

Acute dislocation of fully deployed stent after use of non-compliant balloon: an enigma

Santosh Kumar Sinha, Mahmadulla Razi, Ramesh Thakur, Mukesh Jitendra Jha, Amit Goel, Vikas Mishra, Ashutosh Kumar
Folia Cardiologica 2016;11(3):222-225.

Streszczenie

Stent embolism is an established but rare complication of percutaneous coronary intervention, usually encountered when an undeployed stent unintentionally dislocates from the balloon. Published literature regarding incidence or clinical outcomes of embolism of fully deployed coronary stents is sparse. Here we report an unusual case of a 41-year-old male who had dislocation of a fully deployed stent into the distal part of left anterior descending artery following post dilatation by non-compliant balloon during percutaneous coronary intervention.