A broken balloon catheter entrapped in coronary artery due to its rupture during angioplasty of a chronic total occlusion: a rare complication of angioplasty managed conservatively

Uwięźnięcie uszkodzonego cewnika balonowego w tętnicy wieńcowej spowodowane jego pęknięciem w trakcie angioplastyki przewlekłej całkowitej zmiany: rzadkie powikłanie angioplastyki leczone zachowawczo

Celal Kilit1, Afşin Parspur2, Taner Şen2, Mehmet Ali Astarcıoğlu2

1Department of Cardiology, Faculty of Medicine, Dumlupınar University, Kütahya, Turkey

2Department of Cardiology, Dumlupınar University, Evliya Çelebi Training and Research Hospital, Kütahya, Turkey

Address for correspondence: Dr. Celal Kilit, Dumlupınar University, Evliya Çelebi Training and Research Hospital, İstiklal, Okmeydanı Cd., 43050 Kütahya, Turkey, tel.: +90 274 742 3166, e-mail: ckilit@hotmail.com

Abstract

We present a case of a 72-year-old patient who underwent percutaneous transluminal coronary angioplasty due to a chronic total occlusion of left anterior descending artery. The balloon catheter ruptured and entrapped in the chronic lesion and had broken while attempting to remove from the lesion. End portion of the balloon catheter retained in the lesion. We tried various extraction methods but were unsuccessful. Because of the patient’s high surgical risk, we left the broken part of balloon catheter in lesion and fixed it to the coronary wall by stenting.

Key words: percutaneous coronary intervention, balloon angioplasty, complications, coronary device entrapment

Folia Cardiologica 2016; 11, 3: 219–221

Introduction

Entrapped and broken percutaneous transluminal coronary angioplasty (PTCA) balloon catheter is a rare but serious complication of intervention cardiology. It requires retrieval of the catheter to prevent further complications. This report describes a rare situation in which a ruptured and entrapped balloon with broken catheter was caught in an incompletely dilated chronic total occlusion (CTO) during a percutaneous coronary intervention (PCI) procedure in a 72 year old woman.

Case report

A 72-year-old female patient was underwent PCI due to a CTO of left anterior descending artery (LAD) (Figure 1).

Kilit_1.jpg

Figure 1. Coronary angiography with chronic total occlusion in left anterior descending artery

The lesion was successfully crossed with a chronic occlusion guide wire with the help of Tornus penetration catheter (Asahi Kasei Medical, Japan). We changed the chronic occlusion guide wire with a floppy wire. A 1.0 × 20 mm CTO balloon catheter was passed to inflate the lesion. After this, in order to facilitate the stent crossing, lesion dilated sequentially from distal to proximal with a 2.5 × 20 mm non-compliant balloon. Balloon ruptured during the inflation of the balloon. The balloon catheter entrapped in the lesion and it could not be retrieved from the LAD. The tip of the balloon catheter broke and remained in the lesion while trying to get back by the operator. The patient developed angina, bradycardia and hypotension. We crossed the lesion with a second guide wire. The tip of the balloon could not be retrieved by snare. The lesion was dilated with another non-compliant balloon for the release of the trapped broken balloon. We tried to capture the broken balloon by snare again, but have not succeeded. Because of the high surgical risk of the patient, the surgical removal of the broken balloon was not considered. It was decided to leave the broken part of balloon catheter in lesion and fix it to the coronary wall by stenting. A drug-eluting stent was passed through the lesion to cover the broken part of the balloon catheter (Figure 2).

Kilit_2.jpg

Figure 2. The cranial coronary angiographic image shows markers of drug eluting stent (black arrows) and distal marker of broken balloon catheter in left anterior descending coronary artery (white arrow)

Due to suboptimal stent expansion, post-dilation was made with a non-compliant balloon and patient had a good angiographic outcome (Figure 3, 4). Patient was followed for 24 hours in intensive care unit and she had no complication.

Kilit_3.jpg

Figure 3. The caudal right anterior oblique angiographic image shows suboptimal stent expansion and distal marker of broken balloon catheter in left anterior descending coronary artery (white arrow)

Kilit_4.jpg

Figure 4. The cranial coronary angiographic image shows optimal stent expansion in left anterior descending coronary artery after post-dilatation

Conclusions

PCI for CTO of coronary arteries is one of the most technically challenging areas for the interventional cardiologist with lower procedural success rates and higher complication rates compared with those for the non-occluded coronary arteries or acutely-occluded arteries [1]. Despite advances in device technology, the management of resistant, calcific lesions remains one of the greatest challenges in successful CTO intervention. One of the established techniques to modify calcific lesions is the use of high-pressure non-compliant balloon dilation. The nature of the fibro-calcific plaque reduces vessel distensibility, making it resistant to aggressive balloon dilation. Therefore, even at low pressures, balloon rupture and entrapment can be observed in calcified lesions. If equipment is retained or entrapped within a coronary artery, this can lead to myocardial ischemia, infarction, and lethal arrhythmias [2, 3]. Management of these cases is dictated by the hemodynamics; therefore, the immediate removal of the broken piece of the angioplasty catheter is imperative when the patient becomes hemodynamically unstable or there is evidence of an acute coronary event. There are reports in which guidewires or components of PTCA lodged in coronary arteries have managed conservatively [4, 5]. Various percutaneous retrieval methods have already been described in the literature: loop snare, dormia basket, double or triple guidewire technique, hooked guidewire, Fogarty technique, jailed-wire technique, and balloon inflation technique [3, 6–10]. Although a variety of techniques have been introduced in these situations, most of them have been time and cost consuming methods. In addition, none of these techniques have emerged as the best retrieval modality. In cases where the broken piece cannot be removed and impaired the patient’s hemodynamics, stenting broken piece in coronary artery without inhibit of coronary circulation, may be the most appropriate approach.

Complications of retained or entrapped equipment in the coronary system are still encountered during angioplasty procedures. In cases where the broken piece cannot be remove, fixing it to the coronary wall with stenting, can be considered as a management option.

Conflict of interest(s)

The authors declare no conflict of interest.

Streszczenie

Przedstawiono przypadek 72-letniego chorego poddanego przezskórnej śródnaczyniowej angioplastyce wieńcowej z powodu przewlekłej całkowitej okluzji przedniej lewej tętnicy zstępującej. Cewnik balonowy pękł i uwiązł w przewlekłej zmianie, a następnie oderwał się w czasie próby jego wyciągnięcia. Końcowa część cewnika pozostała w zmianie. Autorom nie udało się jej usunąć mimo stosowania różnych metod. Ze względu na wysokie ryzyko operacyjne u pacjenta pozostawiono oderwany fragment cewnika w zmianie i naprawiono ścianę naczynia za pomocą stentu.

Słowa kluczowe: przezskórna interwencja wieńcowa, angioplastyka balonowa, powikłania, uwięźnięcie urządzenia w tętnicy wieńcowej

Folia Cardiologica 2016; 11, 3: 219–221

References

  1. 1. Fefer P., Knudtson M.L., Cheema A.N. et al. Current perspectives on coronary chronic total occlusions: the Canadian Multicenter Chronic Total Occlusions Registry. J. Am. Coll. Cardiol. 2012; 59: 991997.
  2. 2. Chang W.T., Chen J.Y., Li Y.H. et al. A two-case series of entrapment of a ruptured balloon in the coronary artery: avoidable complications and nonsurgical management. J. Formos. Med. Assoc. 2015; 114: 1135–1139.
  3. 3. Girish M.P., Gupta M.D., Mittal A. Percutaneous retrieval of entrapped partially inflated broken coronary angioplasty balloon by modified Fogarty technique. J. Invasive Cardiol. 2011; 23: E173–E176.
  4. 4. Sen T., Aksu T., Parspur A., Kilit C. Broken guidewire during primary percutaneous coronary intervention. Anadolu Kardiyol. Derg. 2012; 12: E7–E8.
  5. 5. Sianos G., Papafaklis M.I. Septal wire entrapment during recanalization of a chronic total occlusion with the retrograde approach. Hellenic J. Cardiol. 2011; 52: 7983.
  6. 6. Collins N., Horlick E., Dzavik V. Triple wire technique for removal of fractured angioplasty guidewire. J. Invasive Cardiol. 2007; 19: E230–E234.
  7. 7. Hung C.L., Tsai C.T., Hou C.J.Y. Percutaneous transcatheter retrieval of retained balloon catheter in distal tortuous coronary artery: a modified double-helix approach. Catheter. Cardiovasc. Interv. 2004; 62: 471–475.
  8. 8. Sato A., Abe D., Aonuma K. A successful retrieval of entrapped guidewire with a novel “jailed-wire technique”. Int. J. Cardiol. 2015; 192: 18.
  9. 9. Cho Y., Pak S., Kim J. et al. Rescuing entrapped guidewire using a Tornus catheter. Circ. J. 2007; 71: 1326–1327.
  10. 10. Brilakis E.S., Best P.J., Elesber A.A. et al. Incidence, retrieval methods, and outcomes of stent loss during percutaneous coronary intervention: a large single-center experience. Catheter. Cardiovasc. Interv. 2005; 66: 333–340.