open access

Vol 3, No 2 (2014)
Research paper
Submitted: 2014-04-17
Accepted: 2014-04-23
Published online: 2014-05-14
Get Citation

Analytical evaluation of the Contour® Plus glucose meter

Maria Kapusta, Krystyna Słowińska-Solnica, Jan Skupień, Maciej T. Małecki, Bogdan Solnica
Diabetologia Kliniczna 2014;3(2):62-68.

open access

Vol 3, No 2 (2014)
Original articles
Submitted: 2014-04-17
Accepted: 2014-04-23
Published online: 2014-05-14

Abstract

Introduction. Glucose meters are widely used for self­-monitoring of blood glucose considered an integral part of the treatment of diabetes. In this study we evaluated the analytical performance of the Contour® Plus glucose meter utilizing glucose dehydrogenase reaction and amperometric measurement technique, intended for use in self-monitoring of blood glucose as well as for point-of-care testing. Material and methods. The evaluation of precision and test strips lot dependent variability was based on a series of measurements of glucose concentration in venous blood collected on EDTA. With the use of the evaluated meter, glucose concentrations were measured with 208 fresh venous blood samples. As a comparator hexokinase method on the Maxmat PLII analyzer was used. The hematocrit effect was assessed by measuring glucose concentrations in blood samples with hematocrit modified by adding or removing a de­fined volume of plasma. Results. Results of glucose concentration measure­ments using both methods correlated with each other (r = 0.99, P < 0.0001). The within-run imprecision coefficients of variation for glucose concentrations of ~72 mg/dl and ~222 mg/dl amounted to 2.40% and 2.37%, respectively. The between-run coefficient of variation for glucose concentration of ~123 mg/dl was 2.14%. Lot dependent relative differences between the results were equal to 0.88% and 1.16%. Glucose meter error calculated for the entire range of glucose con­centrations was 3.7%. The error grid analysis yielded 99.5% of the results within zone A. The meter meets the accuracy requirements of the ISO 15197:2013. Su­pplementing a blood sample in the test strip did not affect the accuracy. Hematocrit dependent change of glucose concentration amounted to –0.126 mg/dL (0.073%) per 1% increase in hematocrit. Conclusions. The evaluated glucose meter meets the current requirements of analytical quality and suitable for use in self-monitoring of blood glucose as well as in the hospital and outpatient setting.

Abstract

Introduction. Glucose meters are widely used for self­-monitoring of blood glucose considered an integral part of the treatment of diabetes. In this study we evaluated the analytical performance of the Contour® Plus glucose meter utilizing glucose dehydrogenase reaction and amperometric measurement technique, intended for use in self-monitoring of blood glucose as well as for point-of-care testing. Material and methods. The evaluation of precision and test strips lot dependent variability was based on a series of measurements of glucose concentration in venous blood collected on EDTA. With the use of the evaluated meter, glucose concentrations were measured with 208 fresh venous blood samples. As a comparator hexokinase method on the Maxmat PLII analyzer was used. The hematocrit effect was assessed by measuring glucose concentrations in blood samples with hematocrit modified by adding or removing a de­fined volume of plasma. Results. Results of glucose concentration measure­ments using both methods correlated with each other (r = 0.99, P < 0.0001). The within-run imprecision coefficients of variation for glucose concentrations of ~72 mg/dl and ~222 mg/dl amounted to 2.40% and 2.37%, respectively. The between-run coefficient of variation for glucose concentration of ~123 mg/dl was 2.14%. Lot dependent relative differences between the results were equal to 0.88% and 1.16%. Glucose meter error calculated for the entire range of glucose con­centrations was 3.7%. The error grid analysis yielded 99.5% of the results within zone A. The meter meets the accuracy requirements of the ISO 15197:2013. Su­pplementing a blood sample in the test strip did not affect the accuracy. Hematocrit dependent change of glucose concentration amounted to –0.126 mg/dL (0.073%) per 1% increase in hematocrit. Conclusions. The evaluated glucose meter meets the current requirements of analytical quality and suitable for use in self-monitoring of blood glucose as well as in the hospital and outpatient setting.

Get Citation

Keywords

diabetes mellitus, self-monitoring of blood glucose, glucose meter error

About this article
Title

Analytical evaluation of the Contour® Plus glucose meter

Journal

Clinical Diabetology

Issue

Vol 3, No 2 (2014)

Article type

Research paper

Pages

62-68

Published online

2014-05-14

Page views

1015

Article views/downloads

7073

Bibliographic record

Diabetologia Kliniczna 2014;3(2):62-68.

Keywords

diabetes mellitus
self-monitoring of blood glucose
glucose meter error

Authors

Maria Kapusta
Krystyna Słowińska-Solnica
Jan Skupień
Maciej T. Małecki
Bogdan Solnica

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

 

Wydawcą serwisu jest  "Via Medica sp. z o.o." sp.k., ul. Świętokrzyska 73, 80–180 Gdańsk

tel.:+48 58 320 94 94, faks:+48 58 320 94 60, e-mail:  viamedica@viamedica.pl