Introduction
Breast cancer is the most common cancer in women. Early invasive breast cancer is limited to the breast and ipsilateral axilla. The primary treatment of early invasive breast cancer is usually surgical excision followed frequently by risk adapted adjuvant treatments, including radiation therapy, hormone therapy, biological therapies, or chemotherapy. Whatever treatment is used, the reduction of chronic toxicity in patients with breast cancer becomes of the paramount importance due to the continuous improvement of the vital prognosis.
Radiation therapy (RT) to the left breast usually involves incidental irradiation of the heart and left lung. Cardiac irradiation increases the risk of radiation-induced cardiac alterations, including major coronary events, proportional to dose with a relative increase of 7.4% (95% confidence interval, 2.9 to 14.5) per Gy of mean heart dose (MHD) received [1]. There is no cardiac dose threshold below which radiation therapy is safe, so the cardiac dose should be kept as low as possible [2–4]. It is still unknown which dosimetric parameter is best related to specific cardiac toxicities, whether it is the MHD, the maximum doses (Dmax) to the coronary arteries, or the dose distribution in a particular cardiac substructure. In addition, RT can cause lung lesions, such as reduction of lung function, radiological alterations, radiation pneumonitis, fibrosis and radiation induced tumors [5]. The combination of lung and heart damage could cooperate for a net cardiorespiratory functional deterioration after irradiation mainly in the long term.
Deep inspiration breath-hold (DIBH) is a method of limiting radiation exposure to the lung and heart in patients receiving radiation therapy to the left breast [4]. The basis of the technique simply involves the movement of organs at risk (OARs) away from the treatment area by stopping respiratory movement at the optimal stage of the respiratory cycle. Numerous studies have shown that the DIBH technique allows for heart preservation during the RT of the left breast, reducing the MHD by 26.2–75.0% compared to standard free-breathing (FB) and leaving the absolute MHD between 0.7–5.0 Gy [6, 7]. The use of modern RT techniques with dose intensity modulation of the beam (IMRT) combined with DIBH can achieve even greater reductions in MHD [8]. Left lung dose reductions have also been observed with the DIBH technique [6, 7]. Other additional advantages as derived from the elimination of respiratory movement during RT with the DIBH technique have been described such as the reduction of the margins necessary to compensate for the movement of the target volume (the breast or chest wall) and the greater precision in the administration of the planned dose [9].
Quantifying radiation dose to cardiac substructures is important for research on the etiology and prevention of complications following RT [10]; however, segmentation of substructures is challenging. Cardiac atlases based on computed tomography (CT) images have been developed for the radiation therapy field, resulting in a variable number of substructures [11–19]. Deep-learning, machine-learning, knowledge-based, and artificial-intelligence methods can help reducing the variability inherent to the manual contouring of so many substructures [20].
The main purposes of this work are to quantify the magnitude of DIBH dosimetric benefits and identify which cardiac substructures experience a significant dose reduction in patients with diagnosis of early left-sided breast cancer who were treated with RT under DIBH conditions.
Materials and methods
Patients
Forty consecutive patients with left-sided breast cancer were included if they were in a good performance status [Eastern Cooperative Oncology Group (ECOG) 0–1], they were able to maintain DIBH for more than 20 seconds, and able to follow the visual instructions of the respiratory synchronization system. Conservative treatments as well as post-mastectomy treatments were included, with or without RT to the axillary lymph node levels 1–4.
Deep inspiration breath hold technique
The spirometric breath-hold system SDX® (Dyn’R, Toulouse, France) consists of blocking voluntarily the patient’s breathing in inspiration, during acquisition, and irradiation. The DIBH technique involved patient training immediately prior to CT simulation. The training or preparation phase is required to define a comfortable degree of deep inspiration for the patient. This level usually corresponds to 85% of the patient’s maximum inspiration with a 10% tolerated variability.
Immobilization and image acquisition
All patients were simulated and treated on a breast-board in a supine position with both hands above the head. Two CT acquisitions were performed for each patient: the first CT acquisition was performed with the DIBH device and a final acquisition was performed in FB. An injection of contrast agent was performed during DIBH acquisition to facilitate delineation of target volumes.
Contouring
Target volumes were defined according to International Commission on Radiation Units and Measurements general recommendations, following the European Society for Radiotherapy and Oncology (ESTRO) specific guidelines for elective radiation therapy of early stage breast cancer [21]. The defined OARs were the left and right lung, total lung volume, the heart and the heart substructures according Duane’s cardiac contouring atlas [16]. From the 15 cardiac segments included in Duane’s atlas, the following structures were contoured for this study: left anterior descendant (LAD) artery, whole heart, left and right ventricles, apical, anterior, septal, lateral, and inferior left ventricle (LV) segments.
Treatment planning and dosimetric data extraction
For each patient, tangential fields and sub-fields plans on the FB and DIBH scans were developed using segmental IMRT. The aim of the plans was to cover ≥ 95% of the planning target volume (PTV) with ≥ 95 % of the prescription dose. Two fractionation schedules were used: 50 Gy/25 fractions in 5 weeks (Conv) and 40 Gy/15 fractions hypofractionated schedule within 3 weeks (Hypo). Dose volume histograms (DVH) were exported from Eclipse (Varian Medical Systems) in text format. Automatic data extraction from DVH was scripted in MATLAB® (The MathWorks Inc.). Relevant dosimetric parameters from targets and OARs were tabulated for posterior statistical analysis.
Treatment administration
The daily reproducibility was assessed by electronic portal acquisitions according to a No-Action Level (NAL) protocol.
Endpoints and statistical analysis
To investigate the dose homogeneity of target coverage, the PTV median dose (D50%) and PTV V95% were calculated. The relative change when using DIBH as opposed to FB was compared for standard defined parameters: MHD, mean ventricles dose, V10 and V20 in the heart and ventricles, mean dose and V20 for several heart subvolumes, Dmax, D2% and V20 for LAD artery, and total lung and left lung V20, V10 and V5.
To test if an individual group of data (from FB or DIBH) comes from a standard normal distribution, the Kolmogorov-Smirnov (K-S) test was conducted. The Lilliefors test, a less restrictive test of normality, was also conducted. The significance level is established at 0.05 in both cases.
To evaluate FB vs. DIBH (paired data populations), non-parametric tests for continuous variables were used. The selected test for comparison was non-parametric Wilcoxon Signed Rank Test. Assuming data from the differences between the two observations comes from a symmetric and continuous distribution around its median value, the null hypotheses tests if this distribution has zero median. The significance level chosen is 0.05.
Results
Normality tests (Kolmogorov-Smirnoff and Lillie fors tests) showed that structure volumes followed a normal distribution but dose and dose-volume histogram (HDV) derived parameters did not. Therefore, mean values were used to describe the organ volumes and median values were used to describe the dose related parameters.
Patient characteristics
The patient and tumor characteristics are displayed in Table 1.
Characteristic |
Category |
Mean ± SD* |
n (%)** |
Age* |
52.8 ± 8.1 |
|
|
BMI* |
25.2 ± 4.6 |
|
|
Tumor location** |
Inner quadrants |
|
10 (25%) |
Outer quadrants |
|
21 (52.5%) |
|
Periareolar and interquadrants |
|
9 (22.5%) |
|
Stage T** |
cis |
|
2 (5.0%) |
pT1a |
|
1 (2.5%) |
|
pT1b |
|
11 (27.5%) |
|
pT1c1 |
|
17 (42.5%) |
|
pT22 |
|
9 (22.5%) |
|
Stage N** |
pN0 |
|
33 (82.5%) |
pNmic |
|
4 (10,0%) |
|
pN1 |
|
1 (2.5%) |
|
pN2a |
|
1 (2.5%) |
|
pNx |
|
1 (2.5%) |
|
Stage group** |
cis |
|
2 (5.0%) |
IA |
|
24 (60.0%) |
|
IB |
|
4 (10.0%) |
|
IIA |
|
8 -20.0%) |
|
IIB |
|
1 (2.5%) |
|
IIIA |
|
1 (2.5%) |
|
Supraclavicular field** |
Yes |
|
2 (5.0%) |
No |
|
38 (95%) |
|
Fractionation** |
Conv |
|
3 (7.5%) |
Hypo |
|
37 (92.5%) |
Target volumes and doses
The clinical target volume (CTV) and PTV volumes under FB or DIBH conditions were similar: 733 ± 337 mL vs. 729 ± 336 mL (p = 0.815) for CTVs and 917 ± 385 mL vs. 913 ± 385 mL (p = 0.861) for PTVs. PTV dose coverage was similar disrespect of respiratory conditions. For hypofractionated regime, median PTV-D50% was 41.1 Gy under FB conditions versus 41.0 under DIBH conditions, p = 0.665. For conventional fractionation regime, median D50% was 51.3 under FB conditions versus 51.4 under DIBH conditions, p = 1.0. Regarding PTV-V95% , the median values were almost identical: 97% for FB and 96.9% for DIBH conditions.
Lung volumes and doses
The mean total lung and left lung volumes under FB versus DIBH conditions were 2954 ± 578 mL vs. 4592 ± 740 mL (p < 0.001) and 1368 ± 311 mL vs. 2141 ± 372 mL (p < 0.001), respectively. Median total lung mean dose and median total lung V20 values under FB versus DIBH conditions were 3.5 Gy versus 3.3 Gy (p = 0.06) and 5.9% vs. 5.2% (p = 0.33). Three dose levels of 20Gy, 10 Gy, and 5 Gy were stablished to determine the fractional volumes V20, V10 and V5 for the left lung. The median values under FB versus DIBH conditions were, for V20 13.1% vs. 11–4% (p = 0.008); for V10, 18.6% vs. 17.1% (p = 0.003); and for V5, 28.2% vs. 26.2% (p < 0.001). All lung volumes and doses are described in Table 2.
Structure |
Parameter |
DIBH |
Range |
FB |
Range |
p |
Lungs |
Dmean [Gy] |
3.3 ± 0.7 |
5.3–2.0 |
3.7 ± 0.9 |
6.8–2.1 |
0.06 |
Dmedian [Gy] |
3.3 |
3.5 |
||||
Mean V20 (%) |
5.6 ± 1.6 |
9.5–2.9 |
6.1 ± 2.1 |
12.7–2.6 |
0.33 |
|
Median V20 (%) |
5.2 |
5.9 |
||||
Mean Δ-Volume [cm3] |
+1637.8 ± 555.3 |
3389.3–664.1 |
< 0.001 |
|||
Median Δ-Volume [cm3] |
+1632.8 |
|||||
Left lung |
Mean V20 (%) |
12.1 ± 3.4 |
20.4–6.2 |
13.3 ± 4.8 |
27.6–7.2 |
|
Median V20 (%) |
11.4 |
13.1 |
0.008 |
|||
Mean V10 (%) |
17.7 ± 3.9 |
28.3–9.9 |
19.0 ± 6.0 |
37.3–12.4 |
|
|
Median V10 (%) |
17.1 |
18.6 |
0.003 |
|||
Mean V5 (%) |
27.2 ± 5.1 |
38.7–16.6 |
29.2 ± 8.1 |
49.1–19.9 |
|
|
Median V5 (%) |
26.8 |
28.2 |
< 0.001 |
|||
Mean Δ-Volume [cm3] |
+783.5 ± 286.4 |
1,705.2–217.6 |
< 0.001 |
|||
Median Δ-Volume [cm3] |
+772.0 |
Left anterior descending artery doses
The radiation dose to the LAD artery was reduced under DIBH conditions: median Dmax, median D2%, and median V20% were 39.0 Gy, 37.2 Gy, and 44.2 % for FB conditions versus 27.6 Gy, 21.2 Gy, and 3.1% for DIBH conditions (p < 0.001 for all comparisons) (Tab. 3).
|
Parameter |
DIBH |
Range |
FB |
Range |
p |
LAD artery |
Dmáx [Gy] |
24.4 ± 14.2 (27.6) |
50.5–3.1 |
36.2 ± 11.1 (39.0) |
51.4–6.0 |
< 0.001 |
D2% [Gy] |
21.0 ± 14.2 (21.2) |
49.4–2.7 |
33.4 ± 12.1 (37.9) |
50.7–4.9 |
< 0.001 |
|
V20 (%) |
17.1 ± 25.8 (25.8) |
93.2–0.0 |
42.3 ± 30.9 (44.2) |
95.7–0.0 |
< 0.001 |
|
Heart |
Dmean [Gy] |
1.9 ± 1.0 (1.7) |
4.5–0.6 |
3.1 ± 1.4 (2.9) |
7.7–0.8 |
< 0.001 |
V20 (%) |
1.2 ± 1.8 (0.5) |
7.4–0.0 |
3.2 ± 2.8 (2.3) |
13.8–0.0 |
< 0.001 |
|
V10 (%) |
2.1 ± 2.6 (1.1) |
11.1–0.0 |
4.9 ± 3.8 (3.9) |
19.8–0.0 |
< 0.001 |
|
LV |
Dmean [Gy] |
2.5 ± 1.5 (2.1) |
7.6–0.9 |
4.5 ± 2.4 (3.8) |
12.7–1.2 |
< 0.001 |
V20 (%) |
1.7 ± 3.1 (0.3) |
7.4–0.0 |
5.3 ± 5.4 (4.5) |
24.6–0.0 |
< 0.001 |
|
Right ventricle |
Dmean [Gy] |
1.4 ± 0.6 (1.2) |
3.4–0.6 |
2.1 ± 1.2 (1.7) |
5.6–0.9 |
< 0.001 |
V20 (%) |
0.1 ± 0.5 (0.0) |
3.3–0.0 |
0.7 ± 1.5 (0.0) |
6.3–0.0 |
– |
|
LV Apical segment |
Dmean [Gy] |
5.5 ± 4.7 (4.0) |
21.7–1.5 |
13.0 ± 8.0 (11.3) |
38.0–2.3 |
< 0.001 |
V20 (%) |
6.9 ± 13.2 (0.3) |
56.0 - 0.0 |
25.1 ± 23.8 (17.5) |
94.9–0.0 |
< 0.001 |
|
LV Anterior segment |
Dmean [Gy] |
4.6 ± 4.2 (3.7) |
20.9–1.2 |
7.9 ± 7.1 (5.9) |
36.0–1.9 |
< 0.001 |
V20 (%) |
3.6 ± 13.0 (0.0) |
65.2–0.0 |
10.5 ± 19.3 (0.9) |
85.9–0.0 |
< 0.001 |
|
LV Septal segment |
Dmean [Gy] |
1.6 ± 0.9 (1.4) |
3.9–0.6 |
2.4 ± 1.4 (2.0) |
7.7–0.9 |
< 0.001 |
V20 (%) |
0.0 ± 0.1 (0.0) |
0.5–0.0 |
0.3 ± 1.6 (0.0) |
9.9–0.0 |
– |
|
LV Lateral segment |
Dmean [Gy] |
1.9 ± 1.2 (1.6) |
6.2–0.8 |
3.2 ± 2.6 (2.5) |
15.4–0.8 |
< 0.001 |
V20 (%) |
0.0 ± 0.2 (0.0) |
1.0–0.0 |
1.2 ± 5.3 (0.0) |
30.9–0.0 |
– |
|
LV Inferior segment |
Dmean [Gy] |
0.8 ± 0.7 (0.6) |
3,6–0.3 |
1.3 ± 1.0 (0.9) |
4.4–0.5 |
< 0.001 |
V20 (%) |
0.0 ± 0.0 (0.0) |
0.0–0.0 |
0.0 ± 0.0 (0.0) |
0.0–0.0 |
– |
Heart volumes and doses
Mean whole heart volume resulted in slightly smaller DIBH (475.7 ± 104.7 mL. vs. 503.3 ± 129.2 mL; p = 0.04). The median MHD was 2.9 Gy under FB conditions versus 1.7 Gy under DIBH conditions (p < 0.001) (Tab. 2). Dose-volume parameters corroborate the heart dose preservation with DIBH: Median V10 and V20 under FB were 3.9% and 2.3%, versus 1.1% and 0.5% under DIBH conditions (p ≤ 0.001 for both comparisons). Figure 1 shows a demonstrative case highlighting the dosimetric benefit obtained from the DIBH technique.
Median Dmean to the LVs were 2.1 Gy for DIBH and 3.8 Gy for FB conditions (p < 0.001). Respective values for the right ventricles were 1.2 Gy under DIBH and 1.7 Gy for FB (p < 0.001). Median V20 for LVs was 0.3% under DIBH conditions, and 4.5% for FB conditions (p < 0.001). Median right ventricles V20 were 0.0% under both respiratory conditions.
LV segments doses
Median Dmean and V20 values for LV apical, anterior, septal, lateral, and inferior LV segments under DIBH and FB conditions are displayed in Table 3. Median doses were significantly reduced under DIBH conditions for every segment. V20 values were negligible in lateral, septal and posterior segments whatever the respiratory conditions, but for apical and anterior segments the V20 values were significantly reduced under DIBH condition.
Discussion
Two recent meta-analyses compared the heart dose, left lung dose, LAD coronary artery dose, and target coverage in 1019–3599 patients from 12–41 observational studies, respectively [6, 7]. Different respiratory monitoring devices were used in those studies: a belt in one study, real time surface tracking system in two studies, real time position management system in four studies, and active-breathing coordinator in five studies [6]. The combined analysis showed that there was no significant difference in target volumes coverage (Dmean and V95) between the two groups [6]. Using the methods described in the present series, no difference in target coverage was observed.
The lung volume expansion under DIBH conditions is substantial, observing a mean volume difference of 1,638 mL for the whole lung, and 773 mL for the left lung. This volume expansion conditioned a significant decrease in the fraction of irradiated lung volume: a relative mean reduction of 9% for V20, and 7% for V10 and V5. The values analyzed in the meta-analysis (left lung Dmean, V20, V10 and V5) were quite similar to the ones observed in the present study [6, 7]. Seven studies reported left-lung V20 values under DIBH conditions, resulting in a weighted mean of 11.6%, similar to the 12.1% observed in the present series [6]. For the V5 value, we observed 27.2% while a weighted mean value of 23.4% from six studies was reported in the Lai J. et al. meta-analysis [6]. The lung beneficial effect derived from the DIBH technique is not only related to total lung volume increment, but also to reduction in tissue density and, therefore, reduction of lung tissue mass in the treated volume underestimated through the dose-volume relationships the real sparing of lung parenchyma.
Few DIBH studies have reported a comparison between treatment techniques and lung dosimetric parameters. Zhao-Feng et al. [25] reported ipsilateral lung V20 d = –2.11% with 3DCRT and d = –1.29% with IMRT. As the patients in the present series were treated with IMRT, it is no surprising to obtain only small improvements for ipsilateral lung dosimetric parameters.
The coronary arteries integrity is essential to maintain the cardiac function. The artery more heavily damaged after breast irradiation is the LAD coronary artery [22]. According QUANTEC guidelines, the risk of radiation induced cardiac death at 10 years appears to be very low if MHD is < 3.3 Gy and maximum LAD dose is < 45.4 Gy [3]. Therefore, dose preservation on this vessel is of the highest importance. Under DIBH conditions, the radiation dose was dramatically diminished on LAD coronary artery in comparison with FB conditions, with relative reductions in Dmax, Dquasi-max, and V20 of 29%, 44%, and 93%, respectively. The Lai J. et al. meta-analysis showed a significant reduction in LAD coronary artery Dmean and Dmax [6]. Four studies included in the metanalysis reported mean LAD Dmax between 29.8–41.9 Gy under FB conditions versus 15.5–21.96 Gy under DIBH conditions. Corresponding figures in the present series were 36.2 Gy for FB and 24.4 Gy for DIBH. The data from 27 studies involving 2146 patients in Lu Y. et al. metanalysis demonstrated that the LAD dose (Dmean and Dmax) of the DIBH group was significantly lower than that of the FB group [7].
The heart volume difference was small (27 mL, about 5%), but statistically significant. It could be related to physiological factors or it could be just a contouring artefact. The DIBH heart compression effect has been reported previously [6]. As the dosimetric results are expressed in relative values, the small absolute volume difference does not deserve special consideration in the analysis. Dose to the heart as a whole organ and to the LV were significantly reduced under DIBH conditions. A reduction of 1.2 Gy in MHD, as observed in the present study, would represent a risk reduction of 9% for major coronary events according Darby’s rule [1]. Based on more recent evidence, where the authors showed an increment of cumulative incidence of acute coronary events by 16.5% per Gy within 9 years of RT, a risk reduction of 20% for acute coronary events could be expected [22]. A recent study has shown that LAD-Dmean and MHD strongly correlated with coronary damage on computed tomographic angiography, with a 21% higher incidence of disease in the LAD per Gy for LAD-Dmean and a 95% higher incidence of disease in the LAD per Gy for MHD [23]. Therefore, a reduction of 1.2 Gy in MHD, as observed in the present study, would be followed by a substantial risk reduction in radiologically visible coronary disease.
It is under investigation which dose-distribution parameters, other than MHD and LAD dose, are good predictors for heart toxicity. The LV is the most exposed cardiac chamber during left-sided breast cancer 3D radiation therapy, with mean Dmean of 4.8 Gy according to Naimi et al. [10]. Similar Dmean value was obtained in the present study with IMRT under free-breathing conditions (mean Dmean of 4.5 Gy), but it was reduced to 2.5 Gy under DIBH conditions.
Zureick et al. have recently shown that increased LAD Dmean, LAD Dmax, and MHD were associated with increased risk of any cardiac event and a major cardiac event [24]. Van der Bogaard et al. showed that the volume of the LV receiving 5 Gy (LV–V5 Gy) was the most important prognostic dose-volume parameter, but only the four cardiac cavities were contoured in their study [26]. They chose the LV–V5 parameter because heart-V5 had been widely used in many other recent reports. The typical radiotherapy heart isodose map shows a very heterogeneous dose deposits in the LV, being the anterior and apical segments, the ones that receive higher doses. The subsegmental dosimetry in the present series showed that, in FB conditions, only the LV apex and anterior segments received a median dose larger than 5 Gy. Precisely, those segments were the zones where DIBH reduced more intensly the radiation dose.
Conclusion
Demonstration of reduced doses to the left lung and cardiac structures with voluntary DIBH under spirometric control justifies the adoption of this technique as a standard method for left sided breast adjuvant irradiation because of its excellent dosimetric performance. For optimal heart sparing, LAD, LV and its apical/anterior segments should be considered as separate organs at risk.
Conflict of interest
Authors declare no conflict of interests.
Funding
Gerencia Regional de Salud de Castilla y León, GRS 1871/A/18.