Vol 81, No 3 (2023)
Clinical vignette
Published online: 2023-01-10

open access

Page views 2410
Article views/downloads 484
Get Citation

Connect on Social Media

Connect on Social Media

Clinical vignette

Usefulness of metal artifact reduction algorithms from the pacemaker lead in diagnosis of late perforation: The right choice makes it easier to make a decision

Rafał Młynarski*12Michał Sosna13*Maciej Honkowicz13Michał Gibiński24Karolina Simionescu24Michał Rozmus5Krzysztof S Golba24
1Department of Electroradiology, School of Health Sciences, Medical University of Silesia, Katowice, Poland
2 Department of Electrocardiology, Upper Silesian Medical Center, Katowice, Poland
3Department of Radiology, Upper Silesian Medical Center, Katowice, Poland
4Department of Electrocardiology and Heart Failure, School of Health Sciences, Medical University of Silesia, Katowice, Poland
5Department of Radiodiagnostics and Interventional Radiology, Prof. K Gibiński Memorial University Clinical Center of the Medical University of Silesia, Katowice, Poland
*These authors contributed equally to this work

Correspondence to:

Rafał Młynarski, MD,

Department of Electroradiology,

School of Health Sciences,

Medical University of Silesia,

Ziołowa 45, 40–635 Katowice, Poland,

phone: +48 323 598 900,

e-mail: rmlynarski@sum.edu.pl

Copyright by the Author(s), 2023

DOI: 10.33963/KP.a2022.0176

Received: November 2, 2022

Accepted: November 26, 2022

Early publication date: January 10, 2023

Dislocations of lead(s) in cardiac implantable electronic devices (CIED) often result in cardiac tamponade [1, 2]. The risk of lead dislocation is higher in very elderly patients (≥80 years). Additionally, the risk of complications increases in the latest methods of cardiac pacing including left bundle branch area pacing, where the overall incidence of lead dislodgments is about 1% [3]. A combination of various imaging techniques allows for a proper diagnosis and treatment. The biggest problem is artifacts from the exposed metal parts of the lead(s).

We present a case of massive asymptomatic lead dislocation, after implantation of a pacing and sensing capabilities in both the atrium and the ventricle (DDD) in 2007 due to sinus node dysfunction. The patient (woman) was admitted to the hospital for asymptomatic dislocation of the ventricular lead with perforation of the left ventricular wall and displacement to the pleural cavity, which was found in a routine chest X-ray examination. The patient denied fainting. The implanted device had been programmed in a pacing and sensing capabilities in the atrium, rate response active (AAIR mode; base rate 50/min; atrial pacing rate 12%).

The chest X-ray showed an abnormal location of the right ventricular lead with its tip projected below the contour of the left diaphragm, at the level of the gastric air bubble in the left upper abdominal quadrant (possibly in the intraabdominal adipose tissue).

To support the decision about lead extraction, we performed a computed tomography scan using a Siemens Somatom Force scanner. The obtained images showed the course of the ventricular lead along the interventricular septum and perforation of the wall of the right ventricle near the apex of the left ventricle. Then the lead ran subpleurally in the pericardial fat tissue, and its tip was located in the supradiaphragmatic fat tissue of the left costophrenic angle, at the level of the costochondral junction of the left VII rib.

Due to numerous lead artifacts, iterative metal artifact reduction (IMAR) techniques were superimposed on the image in various subroutines, as shown in Figure 1 [4, 5]. In our opinion, IMAR gives promising results, but not with pacemaker preset. Dental presets seemed to be the best in this rare case.

Figure 1. Posteroanterior (A) and lateral (B) chest X-ray shows the right ventricular lead projected under the left diaphragm. Native (C) and IMAR-filtered (D, E) computed tomography images showing reduction of the area of metallic artifacts around the lead tip, with best results with dental preset (E)
Abbreviations: IMAR, iterative metal artifact reduction

On Holter electrocardiography, no atrioventricular conduction disturbances were recorded. After the Heart Team’s decision, the patient was qualified for further conservative treatment. The patient was in good general condition and was discharged home.

Article information

Conflict of interest: None declared.

Funding: The publication of the article was financed by the Medical University of Silesia, Katowice, Poland.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, which allows downloading and sharing articles with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.

REFERENCES

  1. Waddingham PH, Elliott J, Bates A, et al. Iatrogenic cardiac perforation due to pacemaker and defibrillator leads: a contemporary multicentre experience. Europace. 2022; 24(11): 18241833, doi: 10.1093/europace/euac105, indexed in Pubmed: 35894862.
  2. Vanezis AP, Prasad R, Andrews R. Pacemaker leads and cardiac perforation. JRSM Open. 2017; 8(3): 2054270416681432, doi: 10.1177/2054270416681432, indexed in Pubmed: 28321317.
  3. Jiang Z, Chen Y, Chen C, et al. Feasibility and safety of left bundle branch area pacing in very elderly patients (≥80 years). Kardiol Pol. 2022; 80(4): 452460, doi: 10.33963/KP.a2022.0048, indexed in Pubmed: 35167114.
  4. Martin O, Boos J, Aissa J, et al. Impact of different iterative metal artifact reduction (iMAR) algorithms on PET/CT attenuation correction after port implementation. Eur J Radiol. 2020; 129: 109065, doi: 10.1016/j.ejrad.2020.109065, indexed in Pubmed: 32485336.
  5. Bolstad K, Flatabø S, Aadnevik D, et al. Metal artifact reduction in CT, a phantom study: subjective and objective evaluation of four commercial metal artifact reduction algorithms when used on three different orthopedic metal implants. Acta Radiol. 2018; 59(9): 11101118, doi: 10.1177/0284185117751278, indexed in Pubmed: 29310445.