ARTYKUŁ ORYGINALNY / ORYGINAL ARTICLE |
Cardiac pacing in 21 patients with Emery-Dreifuss muscular dystrophy: a single-centre study with a 39-year follow-up
Roman Steckiewicz1, Przemysław Stolarz1, Elżbieta Świętoń1, Agnieszka Madej-Pilarczyk2, Marcin Grabowski1, Michał Marchel1, Marian Pieniak1, Krzysztof J. Filipiak1, Irena Hausmanowa-Petrusewicz2, Grzegorz Opolski1
1Department of Cardiology, Medical University of Warsaw, Warsaw, Poland
2Neuromuscular Unit, Mossakowski Medical Research Centre, Polish Academy of Sciences, Warsaw, Poland
Address for correspondence:
Roman Steckiewicz, MD, PhD, Department of Cardiology, Medical University of Warsaw, ul. Banacha 1A, 02–097 Warszawa, Poland, tel: +48 22 599 29 58,
e-mail: r.steckiewicz@pro.onet.pl
Received: 28.05.2015 Accepted: 22.09.2015 Available as AoP: 12.11.2015
Abstract |
Background: Emery-Dreifuss muscular dystrophy (EDMD) is a genetic condition associated with cardiac arrhythmias. The patients typically develop early, asymptomatic bradyarrhythmia, which may lead to sudden death, preventable with a cardiac implantable electronic device (CIED). EDMD may be characterised by atrial electrical silence. Intra-operative electrophysiological evaluation of the myocardium helps ultimately determine the true nature of the disorder and select an appropriate CIED. |
Aim: To analyse permanent electrotherapy procedures in EDMD patients: atrial pacing limitations that stem from the electrophysiological properties of the myocardium and long-term follow-up of implanted devices. |
Methods: A total of 21 EDMD patients (mean age 29 ± 9 years) with a CIED implanted (1976–2014) due to bradyarrhythmia were included in the study. The implantation procedures and factors determining the CIED type selection were analysed. |
Results: CIEDs were implanted in five women and in 16 men with EDMD types 1 and 2 (mean follow-up: 11 ± 8 years). Intra-operatively assessed atrial electrophysiology resulted in changing the planned CIED type during the procedure in three men with EDMD type 1. Eventually, we implanted: eight DDD, one VDD, 11 VVI, and one CD-DR device, with four of the patients’ devices switched later from DDD to VVI mode in response to electrophysiological changes in the atria. |
Conclusions: Intra-operative assessment of atrial electrophysiological properties resulted in changing the planned DDD mode for VVI in 19% of patients with EDMD type 1. Progression of the underlying disease over a 39-year follow-up resulted in a later change of the initially selected pacing mode from DDD to VVI in 40% of cases. |
Key words: Emery-Dreifuss muscular dystrophy, cardiac arrhythmias, conduction abnormalities, cardiac pacing |
Kardiol Pol 2016; 74, 6: 576–583 |
INTRODUCTION
Emery-Dreifuss muscular dystrophy (EDMD) presents with neurologic symptoms and cardiac problems that may be life-threatening as well as difficult to diagnose. Neurologically, the disease manifests as humeroperoneal muscle weakness and atrophy, joint contractures, and spinal stiffness. These factors determine the phenotype. EDMD is associated with structural and/or functional deregulation of genes encoding proteins that create the nuclear envelope [1, 2]. Although EDMD patients have similar phenotypes, the genes responsible for the disease may be different. EDMD type 1 patients have a mutation in emerin (EMD) located on the X chromosome (Xq28) [3]. EDMD 1 is an X-linked recessive disease. Women do not have skeletal muscle symptoms; however, 20% may have cardiac symptoms. EDMD type 2 patients have mutations in the LMNA gene, which encodes lamin A and C. It is located on chromosome 1 (1q22) [4]. However, in almost 60% of EDMD patients, mutations are found in other genes that encode nuclear proteins.
The clinical presentation of EDMD is exceptional because even mild pathology of the skeletal muscles is accompanied by cardiac pathology, which may result in sudden cardiac death (SCD) [5]. The first cardiologic symptoms are observed during the second to fourth decades of life. Symptoms can be mild or even absent, which seems to be the reason behind SCD in EDMD patients [6]. In the beginning of disease development, the atria are affected, while the size of the left ventricle remains unaffected for a long period of time. The first symptoms include conduction block, bradyarrhythmia, and/or tachyarrhythmia. Early diagnosis and implantation of a permanent pacemaker protects against SCD, but it does not mitigate disease progression [7, 8]. The heart can be affected prior to the muscular system. Ventricular arrhythmias are the main cause of SCD [9]. An implantable cardioverter-defibrillator (ICD) should be implanted for primary prevention of these arrhythmias [10–12].
Fatty and fibrous infiltrates can be found in the myocardium and the cardiac conduction system. Morphologically, this results in electrical inhomogeneity. Inhomogeneous degeneration affects various cardiac areas, including the atria and ventricles [5, 13]. The EDMD-associated changes that develop over time in the nervous and cardiovascular systems have both direct and indirect effects on preliminary indications for, and the ultimate implantation of, the specific permanent electrotherapy device.
We present the results of cardiac implantable electronic device (CIED) placement in EDMD patients throughout a 39-year follow-up. The procedures were performed by one operator (RS). Most papers describe only a few cases. Rarely, larger groups of patients are presented [14, 15].
The aim of our study was to analyse permanent electrotherapy in EDMD patients. We focused on: (i) atrial pacing limitations due to electrophysiological properties of the myocardium and (ii) long-term follow-up of implanted devices.
METHODS
Patients
Twenty-one patients with EDMD type 1 or 2 (mean age at the time of procedure was 29 ± 9 years) were hospitalised between December 29, 1976 and December 31, 2013 for elective CIED implantation. The follow-up periods lasted from six months to 36 years (mean: 11 ± 8 years).
All patients showed a complete EDMD phenotype (Table 1). EDMD type 1 (EMD gene) GenBank accession no. NM000117. EDMD type 2 (LMNA gene) GenBank accession no. NM170707.
Table 1. Emery-Dreifuss muscular dystrophy (EMMD) patient characteristics in chronological order of procedures. Genetic mutations and cardiac arrhythmias recorded prior to the procedure
Pt. No. |
Sex |
Follow-up (months since procedure) |
Dystrophy type |
Form: FF/SF |
Gene mutation |
Age during procedure |
Rhythm before/during procedure |
AVB/WP during procedure |
PACs |
PVCs according to L-W |
CIED (planned) implemented/→ retained |
1 |
M |
1976/432 † 01.2013 |
EDMD 1 |
FF |
c.256C>T, p.Gln86X |
22 |
CAF |
+ |
II |
VVI |
|
2 |
M |
1996/210 |
EDMD 1 |
FF |
c.187+1G>A |
34 |
SR/SAB |
III |
PAF |
0/I |
VVI |
3 |
M |
1996/209 |
EDMD 1 |
FF |
c.153delC, p.Ser52AlafsX13 |
21 |
Atrial standstill |
+ (junctional rhythm) |
0/I |
(DDD)VVI |
|
4 |
M |
1996/193 † 12.2012 |
EDMD 1 |
FF |
c.153delC, p.Ser52AlafsX13 |
28 |
SR/SAB |
I |
Isolated |
0/1 |
VVI |
5 |
M |
1997/198 |
EDMD 1 |
FF |
c.153delC, p.Ser52AlafsX13 |
34 |
CAF |
+ |
0/I |
VVI |
|
6 |
M |
2000/102 † 01.2009 |
EDMD 1 |
SF |
c.397C>T, p.Gln133X |
25 |
CAF |
+ |
II |
VVI |
|
7 |
M |
2000/162 |
EDMD 1 |
FF |
c.153delC, p.Ser52AlafsX13 |
25 |
SR < 40/min |
SVT |
0/I |
DDD → VVI |
|
8 |
M |
2001/152 |
EDMD 1 |
FF |
c.153delC, p.Ser52AlafsX13 |
32 |
SR/SAB |
I/II |
Isolated |
II |
DDD → VVI |
9 |
M |
2001/146 |
EDMD 1 |
SF |
c.153delC, p.Ser52AlafsX13 |
18 |
CAF |
+ |
II |
VVI |
|
10 |
M |
2001/145 |
EDMD 1 |
FF |
c.1A>G |
24 |
SR < 34/min |
WP < 130 |
Isolated |
0/I |
DDD |
11 |
F |
2002/50 † 2006 |
EDMD 2 |
FF |
c.1357C>T, p.Arg453Trp |
41 |
CAF |
+ |
0/I |
VVI |
|
12 |
M |
2002/135 |
EDMD 1 |
FF |
c.450dup, p.Glu151GlyfsX59 |
18 |
SR/SAB |
I/II/III |
Isolated |
0/I |
DDD → VVI |
13 |
F |
2002/133 |
EDMD 2 |
SF |
c.1072G>A, p.Glu358Lys |
25 |
SR |
I/II |
Isolated |
0/I |
VDD → VVI |
14 |
F |
2003/10 † 2004 |
EDMD 2 |
FF |
c.788T>C, p.Leu263Pro |
28 |
CAFL |
+ |
0/I |
VVI |
|
15 |
M |
2005/97 |
EDMD 1 |
FF |
c.399+1G>C |
16 |
SR/SAB |
WP < 130 |
0 |
DDD |
|
16 |
M |
2007/78 |
EDMD 1 |
SF |
c.417_418dup, p.Leu140PhefsX98 |
18 |
SR |
SAB WP < 130 |
PAF |
0/I |
DDD |
17 |
M |
2008/67 |
EDMD 1 |
FF |
c.187+1G>A |
35 |
Atrial standstill |
+ (junctional rhythm) |
0/I |
(DDD)VVI |
|
18 |
F |
2008/66 |
EDMD 1 |
FF |
c.187+1G>A |
44 |
SR |
II |
Isolated |
0/I |
DDD |
19 |
M |
2009/57 |
EDMD 1 |
SF |
c.1 A>G |
25 |
Atrial standstill |
+ (junctional rhythm) |
0/1 |
(DDD)VVI |
|
20 |
F |
2011/28 |
EDMD 2 |
FF |
c.1355T>C, p.Val452Ala |
52 |
SR |
I/II |
Isolated |
II |
DDD |
21 |
M |
2013/6 |
EDMD 2 |
SF |
c.1357C>T, p.Arg453Trp |
35 |
SR |
I WP < 130 |
IVb |
ICD-DR |
|
EDMD type 1 (gene EMD) GenBank accession no. NM000117; EDMD type 2 (gen LMNA) GenBank accession no. NM170707; †Date of death; AVB — atrioventricular block; CAF — chronic atrial fibrillation; CAFL — chronic atrial flutter; CIED — cardiac implantable electronic device; F — female; FF — familial form; L-W — Lown’s grade of premature ventricular contractions; M — male; PACs — premature atrial contractions; PAF — paroxysmal atrial fibrillation; PVCs — premature ventricular contraction/complexes; SAB — sinoatrial block; SF — sporadic form; SR — sinus rhythm; SVT — supraventricular tachycardia; WP — +Wenckebach point (< 130) |
Twelve-lead electrocardiogram (ECG) at rest and/or 24-h Holter monitoring in these patients revealed significant dysfunction of the cardiac conduction system, which was an indication for CIED.
Device implantation procedure and follow-up
Transvenous implantations of CIEDs were performed as usual: in the subclavicular region, with local anaesthesia. Leads were inserted through the cephalic and/or axillary or subclavian vein. As EDMD patients need to turn to one side in order to assume an upright position and in the process tend to use the arm less affected by the disease, the side of device implantation was selected individually, based on where it would least inhibit the patient’s mobility.
Leads were placed under fluoroscopy. Specific electrical measurements were analysed with the appropriate device. Lead placement depended on the parameters obtained during a pacing test, lead tip stabilisation (passive or active), and electrophysiological properties of adjacent tissue recorded during the procedure.
Atrial leads were placed in the right atrial appendage. Ventricular leads were placed in the right ventricle at the site with optimal pacing parameters: the lowest pacing threshold and a voltage amplitude that ensured proper functioning of the device; this was intended to avoid areas of myocardial inhomogeneity.
Patients with preserved spontaneous sinus rhythm underwent atrioventricular node conduction assessment after atrial lead placement but prior to any ventricular lead insertion. In patients with suspected atrial electrical silence, atrial stimulation and pacing parameters were assessed in various areas of the heart.
After discharge, patients were followed-up in the outpatient clinic. We interrogated the devices and, when needed, modified the pacing mode.
RESULTS
We included patients undergoing their first implantation procedure (six females, mean age: 38 ± 11 years; 15 males, mean age: 25 ± 7 years; p = 0.016). The mean follow-up period was 10.6 ± 7.9 years. Sixteen patients had EDMD type 1 (mean age: 26 ± 8 years), and five patients had EDMD type 2 (mean age: 36 ± 11 years); EDMD type 1:EDMD type 2, NS (p = 0.068).
All of the patients (those with sinus rhythm as well as those with atrial fibrillation [AF]) had various degrees of atrioventricular conduction block. In patients with ECG tracings at rest showing normal sinus rhythm and normal atrioventricular conduction, intraprocedural atrial stimulation at an increased rate revealed nodal dysfunction (Table 1). Nearly 30% of ECG tracings showed intraventricular conduction anomalies: left anterior fascicular block or incomplete right bundle branch block.
Mild or even asymptomatic bradyarrhythmia was common, and it was independent of the rhythm shown in ECG tracings (sinus rhythm, junctional rhythm, or chronic AF). We observed bradyarrhythmia at a rate of 20 bpm in 24-h Holter monitoring. Nonetheless, they did not result in symptomatic central nervous system ischaemia. No patient had fully symptomatic Morgagni-Adams-Stokes syndrome prior to device implantation. Patient symptoms included irregular heartbeat and palpitations. There were episodes of paroxysmal AF (Table 1).
Table 1 presents the devices we had planned to implant, the devices we actually implanted, and the associated cardiac arrhythmias registered prior to, or during, the implantation procedure. We planned to implant a VVI device in six patients with atrioventricular conduction abnormalities, AF, atrial flutter, or atrial tachycardia insufficiently controlled with medication. Fifteen patients had atrioventricular conduction abnormalities in ECG and 24-h Holter monitoring. Two patients had atrial electrical silence. After periprocedural Wenckebach point analysis, four patients with sinus rhythm had atrioventricular conduction ratio of 1:1, at the rate not faster than 130 impulses per minute (mean 120 imp/min). In these four patients we implanted a device for atrioventricular pacing. One patient developed AF during the procedure. After a failed attempt to manage the arrhythmia with drugs, we implanted a VVI pacemaker. Because of atrial electrophysiological properties noted during the procedure in three patients, we implanted a VVI pacemaker instead of a DDD device that we had planned to implant (Table 1). Because of abnormalities at the atrioventricular node, we did not implant any AAI pacemakers. The device pocket was located in the left subclavian region in nine patients (in the right subclavian region in 12 patients). The pocket was situated over the pectoral muscle.
Presentation of cases
The indication for pacemaker implantation in a 21-year-old patient (PA, no. 3), was a complete atrioventricular block with ventricular rhythm of 36 bpm. During the procedure, performed on July 15, 1996, 100-impulse/min pacing with a lead placed in the right atrial appendage resulted in a significant delay (190 ms) between the pacing spike and atrial excitation (Fig. 1).
Figure 1. Delay (190 ms for pacing rhythm 100 imp/min) between pacing spike and atrial excitation
A 35-year-old patient (DG, no. 17) received a pacemaker due to bradycardia (38/min from the atrioventricular node). The procedure was performed on May 15, 2008. Due to lack of excitability of the atrial myocardium, we changed the primary device type that had been planned for implantation. Lack of excitability was consistent in all of the tested places in the right atrial endocardium and veins leading to the coronary sinus; it also did not depend on the scale of electrical impulse.
The indication for the procedure in a 25-year-old patient (DR, no. 19), was atrioventricular node rhythm (42/min). A VVI pacemaker was implanted on March 30, 2009. Due to lack of excitability of the right atrium in response to electrical impulses, no lead was placed in this heart chamber. The scale of electrical impulse had no effect on cardiac excitation.
During the procedure in a 41-year-old patient (KG, no. 11), we observed actual electrophysiological properties of the atrium (Fig. 2).
Figure 2. Patient no. 11, 12-lead electrocardiogram at rest. Initially, we interpreted it as a nodal rhythm of 42/min and sinus arrest or complete atrioventricular block. No atrial electrical activity
We found low-amplitude atrial flutter and excluded atrial electrical silence suggested by earlier ECG and Holter findings (Fig. 3).
Figure 3. The same patient during an electrophysiological study. Figure is from the lead in the right atrium; atrial flutter
During lead implantation (screw-in, bipolar) in patient DG (no. 17), prior to final lead placement, we found cardiac areas of a markedly increased pacing threshold. An magnetic resonance imaging scan showed an area of delayed enhancement in the right ventricle, suggesting fibrosis.
During follow-up, three patients had an additional operation due to pacemaker dysfunction (no planned pacemaker replacement).
One month after the implantation procedure in a 21-year-old patient, PA (no. 3), unsuccessful pacing was observed in the ventricular lead (tined, unipolar) placed in the right ventricular apex (pacing threshold > 4.8 V, 0.5 ms; initially: 0.9 V, 0.5 ms). Because of a suspected exit block, glucocorticoids were prescribed, which proved ineffective. A new lead (tined, bipolar) was implanted in a new place, where the pacing threshold was also increased (1.9 V, 0.5 ms).
Patient DA (no. 1) received a new VVI mode pacemaker 20 years after the first procedure. The indication for the second procedure was lead failure and no intrinsic rhythm.
An examination conducted three months after the procedure in patient DA (no. 2) revealed no pacing due to Twiddler’s syndrome. A chest radiogram showed spiral torsion of the lead in the pocket and right ventricle. The unipolar lead in the adapter area was completely broken. We replaced the adapter and inserted the pacemaker into a new pocket, achieving adequate pacing.
Pacemaker interrogations in the rest of the patients showed parameter changes (including pacing threshold values) to be consistent with those normally occurring during and after post-implantation healing. The generated pulse amplitude (voltage) in patients with a preserved intrinsic rhythm with a rate above the minimal rate programmed into the device did not change in the follow-up period in an outpatient setting.
Six of the 10 patients received a DDD pacemaker. In four out of these patients the pacemaker mode was changed to VVI 3–7 years (mean 5 ± 1.6 years) after the first procedure. The indication for that change was chronic AF. One male patient with EDMD type 2 received an ICD-DR because of unsustained ventricular tachycardia, clinical presentation, and genetic background. This decision was made according to the SCD primary prevention guidelines.
Five patients died (three had EDMD type 1, two had EDMD type 2). Autopsies were not performed. Three patients (no. 1, 4, and 6) had SCD; one had ischaemic stroke (no. 11); and one had advanced heart failure (no. 14). These patients had been paced for a mean of 13.2 ± 24.8 years (range: 1–36 years).
DISCUSSION
EDMD is a congenital myopathy. The life-threatening aspect of the condition, however, is not due to the slowly progressive neurological anomalies, but to cardiac arrhythmias, which depend on disease stage [7, 15–17].
These bradyarrhythmias are mildly symptomatic or even asymptomatic. This is not related to the aetiology of bradyarrhythmia, instead it seems to be a result of patient adaptation to slower and slower rhythm during significantly decreased physical activity compared with that of their peers. The asymptomatic character of bradyarrhythmia was an important obstacle in obtaining patient consent for the procedure. Supraventricular arrhythmias were present in patients with EDMD type 1 and 2 [18]. Most of the patients had chronic AF or atrial flutter. Moreover, in some of the patients treated with atrioventricular pacing and antiarrhythmic medications, AF progressed to its chronic form. In some patients with EDMD, cardiac pacing indicated based on ECG tracings may not be possible to implement. In some of these patients, electrical instability, which facilitates AF, might be present with decreased cardiac tissue excitability. The mechanism and grade of atrial failure influences the electrophysiological properties observed during the procedure and affects the steps of the procedure [10, 15].
Electrical atrial silence is typical for EDMD. Patients with that electrophysiological condition require a VVI pacemaker implantation or reprogramming of a pre-existing CIED into a VVI mode. Electrical atrial silence heralds a decline in atrial haemodynamic function [18, 19]. Electrical silence shown in a surface ECG trace needs to be verified during the procedure, as was done in this case.
At our institution, during three implantations in patients with EDMD type 1, unfavourable atrial electrophysiology precluded the initially planned atrial pacing. One of those patients developed a considerable atrial spike-to-P-wave delay. This may have been associated with hyperpolarisation of excitable tissue because of degeneration, which led to increased conduction time. It seems that the other two patients had similar electrophysiological mechanisms, but more advanced and with complete failure to conduct pacemaker-generated impulses.
Most of our patients were in their second to fourth decade of life, which is consistent with other publications [4, 9]. We observed conduction abnormalities, as well as an increased chance of supraventricular arrhythmias in people in that age range [20]. We triggered supraventricular arrhythmias during atrial lead placement in DDD pacemaker implantation procedures. Then, we observed short episodes of AF or atrial flutter, which resolved with no treatment. However, AF triggered in one patient during the procedure could not be interrupted with pharmacological treatment. Therefore, we implanted a VVI pacemaker.
Because of the progression from paroxysmal to chronic AF, we changed the pacing mode to VVI in four patients during follow-up. To our knowledge, the data presented here constitute the greatest number of EDMD patients who received a DDD pacemaker and were followed up for one of the longest periods. There is no published data in this field that can be compared with our results. For example, a study by Boriani et al. [18] included 18 patients; 10 out of these patients received pacemakers: nine VVI/VVIR and one DDDR. The median follow-up after pacemaker implantation was 12 years. Pacemaker dependency was recorded in all cases at the end of follow-up. Five (50%) implanted patients developed atrial standstill after the development of paroxysmal AF or atrial flutter and one patient developed permanent AF after pacemaker implantation.
The atria and ventricles are involved in heterogenic degeneration of the myocardium in which muscular tissue is replaced by fibrotic and adipose tissue [5, 10, 13]. In all patients except one leads were placed in areas with proper pacing parameters despite the possibility of local right ventricular degeneration.
The type and progression of arrhythmia were similar in all of our patients. However, what we could actually observe in individual patients during follow-up were the various stages of arrhythmia progression. Thanks to these observations, we managed to identify the chronology of electrophysiological states of the atria: 1) isolated supraventricular beats, 2) sustained arrhythmia evident in surface ECG, 3) sustained arrhythmia, detectable only during periprocedural intra-atrial assessment of electrophysiological properties in patients with surface ECG tracings suggesting atrial electrical silence, and, eventually, 4) true atrial electrical silence.
In summary, our study shows limitations of pacemaker implantation caused by atrial electrical silence and long-term changes observed during follow-up. Despite generally similar outcomes, we observed differences in terms of arrhythmia onset and detectability via surface ECG. In light of this observation, our research appears to be one of the few studies presenting subsequent stages of electrophysiological changes in the heart, particularly in the atria. These changes include early development of atrial rhythm abnormalities, which can be diagnosed using ECG. Some of the atrial rhythm abnormalities were diagnosed during periprocedural electrophysiological studies when the surface ECG tracing suggested electrical silence. In some cases, there was true electrical silence. We also presented genetic mutations in patients who underwent electrotherapy.
Limitations of the study
Over a period of almost 40-year experience with patients with EDMD at our centre, the indications for CIED have changed. In some patients with an implanted pacemaker, the pacing mode should be modified according to current American College of Cardiology/European Heart Rhythm Association guidelines. However, progression of cardiovascular abnormalities is not as inconvenient for EDMD patients as the neurological aspect of their condition. It seems to be the reason for rejection of device adjustment by EDMD patients.
Device materials and design have changed over the study period. In the beginning, passive leads determined the pacing site in the right ventricular apex. Screw-in leads overcame this problem, and now it is feasible to pace from other sites.
In patients with DDD-mode pacing systems, we also assessed other right atrium sites besides the auricle, and the obtained results were similar. However, even having screw-in leads at our disposal, we still ended up placing them in the auricle because our earlier observations indicated good long-term pacing results from that location in other EDMD patients.
A retrospective analysis of electrophysiological measurements performed today vs. results of earlier studies does not allow for drawing unequivocal conclusions in this area, due to differences in the devices and methods used to collect data.
CONCLUSIONS
Because of the electrophysiological atrial properties measured during the procedure, we had to implant VVI instead of the previously planned DDD pacemakers in 19% of patients with EDMD type 1. Because of the progression of the underlying disease in the atria and rhythm disturbances during a 38-year follow-up, we had to change the double-chamber (atrial and ventricular) pacing to ventricular pacing in 40% of cases. Atrioventricular blocks of various degrees were seen in all patients.
This work was supported by the Medical University of Warsaw and statutory funds from the Mossakowski Medical Research Centre, Polish Academy of Sciences.
Conflict of interest: none declared
References
- 1. Dauer WT, Worman HJ. The nuclear envelope as a signaling node in development and disease. Dev Cell, 2009; 17: 626–638. doi: 10.1016/j.devcel.2009.10.016.
- 2. Helbling-Leclerc A, Bonne G, Schwartz K. Emery-Dreifuss muscular dystrophy. Eur J Hum Genet, 2002; 10: 157–161.
- 3. Bione S, Maestrini E, Rivella S et al. Identification of a novel X-linked gene responsible for Emery-Dreifuss muscular dystrophy. Nat Genet, 1994; 8: 323–327.
- 4. Bonne G, Di Barletta MR, Varnous S et al. Mutations in the gene encoding lamin A/C cause autosomal dominant Emery-Dreifuss muscular dystrophy. Nat Genet, 1999; 21: 285–288.
- 5. Fishbein MC, Siegel RJ, Thompson CE, Hopkins LC. Sudden death of a carrier of X-linked Emery-Dreifuss muscular dystrophy. Ann Intern Med, 1993; 119: 900–905.
- 6. Rakovec P, Zidar J, Sinkovec M et al. Cardiac involvement in Emery-Dreifuss muscular dystrophy: role of a diagnostic pacemaker. Pacing Clin Electrophysiol, 1995; 18: 1721–1724.
- 7. Ishikawa K, Mimuro M, Tanaka T. Ventricular arrhythmia in X-linked Emery-Dreifuss muscular dystrophy: a lesson from an autopsy case. Intern Med, 2011; 50: 459–462.
- 8. Yoshioka M, Saida K, Itagaki Y, Kamiya T. Follow up study of cardiac involvement in Emery-Dreifuss muscular dystrophy. Arch Dis Child, 1989; 64: 713–715.
- 9. Sanna T, Dello Russo A, Toniolo D et al. Cardiac features of Emery-Dreifuss muscular dystrophy caused by lamin A/C gene mutations. Eur Heart J, 2003; 24: 2227–2236.
- 10. Meune C, Van Berlo JH, Anselme F et al. Primary prevention of sudden death in patients with lamin A/C gene mutations. N Engl J Med, 2006; 354: 209–210.
- 11. McMurray JJ, Adamopoulos S, Anker SD et al. ESC Committee for Practice Guidelines. ESC guidelines for the diagnosis and treatment of acute and chronic heart failure 2012: the task force for the diagnosis and treatment of acute and chronic heart failure 2012 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association (HFA) of the ESC. Eur Heart J, 2012; 33: 1787–1847. doi: 10.1093/eurjhf/hfs105.
- 12. Brignole M, Auricchio A, Baron-Esquivias G et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Europace, 2013; 15: 1070–1118. doi: 10.1093/europace/eut206.
- 13. van Tintelen JP, Tio RA, Kerstjens-Frederikse WS et al. Severe myocardial fibrosis caused by a deletion of the 5’ end of the lamin A/C gene. J Am Coll Cardiol, 2007; 49: 2430–2439.
- 14. Parmar MS, Parmar KS. Emery-Dreifuss humeroperoneal muscular dystrophy: cardiac manifestations. Can J Cardiol, 2012; 28: 516. doi: 10.1016/j.cjca.2012.01.012.
- 15. Buckley AE, Dean J, Mahy IR. Cardiac involvement in Emery Dreifuss muscular dystrophy: a case series. Heart, 1999; 82: 105–108.
- 16. Golzio PG, Chiribiri A, Gaita F. ‘Unexpected’ sudden death avoided by implantable cardioverter defibrillator in Emery Dreifuss patient. Europace, 2007; 9: 1158–1160. doi: 10.1093/europace/eum236.
- 17. Gayathri N, Taly AB, Sinha S et al. Emery Dreifuss muscular dystrophy: a clinico-pathological study. Neurol India, 2006; 54: 197–199.
- 18. Boriani G, Gallina M, Merlini L et al. Clinical relevance of atrial fibrillation/flutter, stroke, pacemaker implant, and heart failure in Emery-Dreifuss muscular dystrophy: a long-term longitudinal study. Stroke, 2003; 34: 901–908. doi: 10.1161/01.STR.0000064322.47667.49.
- 19. Marshall TM, Huckell VF. Atrial paralysis in a patient with Emery-Dreifuss muscular dystrophy. Pacing Clin Electrophysiol, 1992; 15: 135–140. doi: 10.1111/j.1540-8159.1992.tb03056.x.
- 20. Karst ML, Herron KJ, Olson TM. X-linked nonsyndromic sinus node dysfunction and atrial fibrillation caused by emerin mutation. J Cardiovasc Electrophysiol, 2008; 19: 510–515. doi: 10.1111/j.1540-8167.2007.01081.x.
Cite this article as: Steckiewicz R, Stolarz P, Świętoń E et al. Cardiac pacing in 21 patients with Emery-Dreifuss muscular dystrophy: a single-centre study with a 39-year follow-up. Kardiol Pol, 2016; 74: 576–583. doi: 10.5603/KP.a2015.0218. |