Vol 82, No 1 (2024)
Clinical vignette
Published online: 2023-10-25

open access

Page views 854
Article views/downloads 408
Get Citation

Connect on Social Media

Connect on Social Media

21__HTML__KP_01_2024__Goreczny___290
  • CLINICAL VIGNETTE

First Polish pediatric experience with percutaneous self-expandable pulmonary valve implantation

Sebastian Góreczny1, Judyta Szeliga1, Maksym Lazu1, Beata Załuska-Pitak1, Marc Gewillig2

1Department of Pediatric Cardiology, University Children’s Hospital, Jagiellonian University Medical College, Kraków, Poland

2Department of Pediatric Cardiology, University Hospitals Leuven, Leuven, Belgium

Correspondence to:

Sebastian Góreczny MD, PhD,

Department of Pediatric Cardiology,

University Children’s Hospital,

Faculty of Medicine,

Jagiellonian University Medical College,

Wielicka 265, 30–669 Kraków, Poland,

phone: +48 12 333 90 50,

e-mail: sebastian.goreczny@uj.edu.pl

Copyright by the Author(s), 2024

DOI: 10.33963/v.kp.97724

Received: September 15, 2023

Accepted: October 5, 2023

Early publication date: October 25, 2023

Percutaneous pulmonary valve implantation is an alternative way of restoring valve function to the right ventricular outflow tract [1]. More recently, self-expandable valves have been providing additional options, especially for patients with large outflow tracts [2, 3]. We present the first Polish experience with the Venus P-valve (Venus MedTech) in two pediatric patients. Both were born with tetralogy of Fallot with unusual coronary artery anatomy; both underwent surgical repair with a monocusp pulmonary homograft and presented with progressive pulmonary regurgitation, which was confirmed with non-invasive imaging (Supplementary material, Table S1). Virtual Reality models (VMersive) were created to present the anatomy and simulate valve size and position [4].

PATIENT A

In a 10-year-old girl (37 kg), the virtual model showed a 30 × 25 mm Venus P-valve as the most suitable (Figure 1A). After initial angiography and measurements, a 40 mm PTS-X sizing balloon (NuMed) was inflated to check the size and distensibility of the outflow tract. Diameters of the outflow tract could also allow a large balloon-expandable valve but, due to the close proximity and the course of the anomalous coronary artery, discouraged this option. Through a 24 Fr Dryseal sheath (Gore), a 30 × 25 mm Venus P-valve was deployed from the right pulmonary artery (Figure 1B). Control angiography showed proper expansion of the valve, with tapering of the distal flare on lateral imaging (Figure 1C and D). Oversizing of the valve could lead to infolding of one of the walls and significant regurgitation [5]. Although the latter was not observed, the distal segment was adapted with the PTS-X balloon (Figure 1E) to gain further expansion. A control angiogram confirmed the proper position and function of the valve and excluded coronary artery compression (Figure 1F). Pre-discharge and 6-month follow-up echocardiograms showed good function of the valve with trivial central regurgitation. ECG-Holter monitoring showed no arrhythmia.

Figure 1. Percutaneous Venus P-valve (Venus MedTech) implantation in a 10-year-old tetra­logy of Fallot patient with anomalous coronary artery (an additional left anterior descending artery from the right coronary artery) after patch repair in infancy and currently with significant pulmonary regurgitation. A. Virtual reality model processed from cardiac magnetic resonance scans with VMersive software (VR-Learning, Poland) to simulate a 30 mm diameter and 25 mm length of Venus P-valve. B. An angiogram during deployment of the distal flare in the proximal right pulmonary artery. C. Full expansion of the valve in cranial projection. D. The valve tapers towards the distal end (white arrows) in the lateral view. E. Adaptation of the distal flare of the valve with a 40 mm PTS-X (NuMed) balloon. F. The control aortography shows unobstructed coronary artery flow, including the additional left descending artery (black arrow) originating from the right coronary artery. The widened distal flare (white arrows) of the valve is seen as well

PATIENT B

In a 17-year-old boy (75 kg), the virtual model revealed a conical-shaped outflow measuring 35 mm proximally and 23 mm just before the bifurcation (Supplementary material, Figure S1A, Video S1). After an initial angiogram (Supplementary material, Figure S1B) and subsequent balloon (40 mm PTS-X) sizing with coronary compression exclusion (Supplementary material, Figure S1C), a 36 × 25 mm Venus P-valve was introduced through a 26 Fr Dryseal and positioned in the proximal left pulmonary artery (Supplementary material, Figure S1D). During the uncovering of the distal flare, the valve shifted below the bifurcation to the middle of the outflow tract. It was recaptured with the Dryseal sheath and once more deployed from the proximal left pulmonary artery with more push on the system. This enabled covering of the distal main pulmonary artery narrowing with the distal flare of the valve. The final angiogram confirmed the full expansion of the valve with unobstructed flow to the pulmonary arteries and a trace of regurgitation (Supplementary material, Figure S1E). Retrospectively, a deployment from the right pulmonary artery might have allowed positioning of the distal flare at the bifurcation, beyond the narrowing. Pre-discharge (Supplementary material, Figure S1F) and follow-up echocardiograms confirmed good valve function. ECG-Holter monitoring showed a slow irregular sinus rhythm.

Supplementary material

Supplementary material is available at https://journals.viamedica.pl/kardiologia_polska.

Article information

Acknowledgments: To Anna Grondalski from Pomeranian Medical University for editing the text.

Conflict of interest: MG is a proctor for Venus MedTech. Other authors declared no conflicts of interest.

Funding: Virtual Reality project is supported by the Jagiellonian University Medical College internal grant No. N41/DBS/001219.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, which allows downloading and sharing articles with others as long as t`hey credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.

REFERENCES

  1. Fiszer R, Dryżek P, Szkutnik M, et al. Immediate and long-term outcomes of percutaneous transcatheter pulmonary valve implantation. Cardiol J. 2017; 24(6): 604611, doi: 10.5603/CJ.a2017.0023, indexed in Pub­med: 28248409.
  2. Sivakumar K, Sagar P, Qureshi S, et al. Outcomes of Venus P-valve for dysfunctional right ventricular outflow tracts from Indian Venus P-valve database. Ann Pediatr Cardiol. 2021; 14(3): 281292, doi: 10.4103/apc.APC_175_20, indexed in Pubmed: 34667398.
  3. Morgan GJ, Sivakumar K, Promphan W, et al. Early clinical experience with the straight design of Venus P-valve™ in dysfunctional right ventricular outflow tracts. Catheter Cardiovasc Interv. 2020; 96(6): E653E659, doi: 10.1002/ccd.28819, indexed in Pubmed: 32096924.
  4. Szeliga J, Kołcz J, Piwowarczyk B, et al. Multimodality imaging and hybrid treatment of pulmonary artery stenosis in a patient with a high risk of airway compression. Kardiol Pol. 2023; 81(11): 11511152, doi: 10.33963/v.kp.97211, indexed in Pubmed: 37718587.
  5. Riahi M, Ang HL, Jones M, et al. Infolding of the Venus P-valve after transcatheter pulmonary valve implantation. Circ Cardiovasc Interv. 2018; 11(4): e005923, doi: 10.1161/CIRCINTERVENTIONS.117.005923, indexed in Pubmed: 29618579.