Review

Infections of cardiac implantable electronic devices: Epidemiology, mechanisms, and preventive measures

Vassil Traykov1Krasimir Dzhinsov23Paweł T Matusik45
1Department of Invasive Electrophysiology, Clinic of Cardiology, Acibadem City Clinic University Hospital Tokuda, Sofia, Bulgaria
2Interventional Electrophysiology and Cardiac Pacing Unit, Interventional Cardiology Department, University Hospital “Sveti Georgi” — Plovdiv, Bulgaria
3Research Institute, Medical University — Plovdiv, Bulgaria
4Department of Electrocardiology, Institute of Cardiology, Faculty of Medicine, Jagiellonian University Medical College, Kraków, Poland
5Department of Electrocardiology, The John Paul II Hospital, Kraków, Poland

Correspondence to:

Vassil Traykov, MD, PhD,

Clinic of Cardiology,

Acibadem City Clinic Tokuda Hospital Sofia, Bulgaria,

phone: +35 98 79 430 365,

e-mail: vtraykov@yahoo.com

Copyright by the Author(s), 2023

DOI: 10.33963/v.kp.97249

Received: August 10, 2023

Accepted: September 3, 2023

Early publication date: September 3, 2023

Abstract
Cardiac implantable electronic device (CIED) infections represent a complication associated with high morbidity and mortality. Despite enormous efforts to prevent them, the rates of infections continue to rise out of proportion to the reported increase in CIED implantation rates. Following extensive research of various prevention strategies and new technologies, several organizations have issued recommendations and consensus papers covering this topic. Our narrative review aims to provide a summary of the existing preventive strategies put forward by the European Heart Rhythm Association consensus and European Society of Cardiology guidelines and introduce the most recent developments in the field, including optimized surgical site management and appropriate periprocedural antithrombotic drug use. It also provides an overview of epidemiology, mechanisms, risk factors, and risk stratification approaches. It focuses on the pre-, intra-, and postprocedural actions that should be taken to mitigate CIED infection risks. Future directions in the prevention of CIED infections have also been addressed.
Key words: cardiac implantable electronic device, defibrillator, epidemiology, pacemaker, infection, prevention, risk

Introduction

Following the first reports on cardiac pacemakers published in the late 1950s and the subsequent development of implantable cardioverter-defibrillators (ICD) in the 1980s, cardiac implantable electronic devices (CIEDs) have become the standard of care in managing cardiac rhythm and conduction disturbances. Published data show a constant increase in the numbers and complexity of CIED implantations worldwide [1]. This growth has been accompanied by an increasing rate of complications, especially with the wider introduction of cardiac resynchronization therapy pacemakers (CRT-P) and defibrillators (CRT-D) [2]. The rate of CIED infections has been shown to increase out of proportion to the reported rise in device implantation [1, 3]. The possible causes are the increasing CIED complexity, comorbidities, and longer life expectancy. On the other hand, CIED infections represent an essential factor for increased morbidity and mortality among CIED recipients [4]. From an economic perspective, CIED infection management puts a significant financial burden on healthcare systems due to additional treatment, prolonged hospital stays, and reinterventions [5–7].

Despite various preventive strategies to reduce CIED complications [8], reports show significant differences in their implementation [9]. Meticulous antisepsis and preoperative antibiotic prophylaxis are highly effective and recommended by various consensus papers and guidelines [8, 10]. New technologies, including subcutaneous ICDs and leadless pacemakers, also aid in the reduction in CIED infections. However, these apply only to selected patients. The role of antibiotic-eluting envelopes (AEEs) for effective CIED infection prevention has been demonstrated by randomized studies [11]. Moreover, advances in diagnostics, including use of procalcitonin in the recognition of device pocket infection, have been done recently [12].

This narrative review presents an overview of the epidemiology and mechanisms of CIED infections as well as the existing and developing strategies to prevent them. It highlights the strategies for risk stratification and focuses on the value of preprocedural, intraprocedural, and postprocedural measures and actions to prevent CIED infections.

Epidemiology, etiology, and mechanisms of CIED infections

Infections related to the CIED develop at a rate ranging from 1 to 7% depending on the type and complexity of the implantation [2, 6, 11, 13]. Previously reported data demonstrated a significant rise in the infection rates over time from 1.45% to 3.41%, with the highest increase for CRT-P/D devices [1]. Real-life data on infection rates contrast with results from randomized studies, which report much lower infection rates in the range of 0.6%–1.3% [4, 11, 13–15]. This could result, at least in part, from predominant participation of high-volume centers in randomized studies. The infection rate is highest early after the procedure (in the first 3 months) [16]. Infections are well-known to be associated with increased morbidity and mortality, especially in the case of systemic and delayed (312 months) localized infections [4, 16, 17]. This trend is preserved even after lead extraction (complete CIED removal) [4] and successful infection eradication [17].

CIED infections develop via two major mechanisms. The most common is local hardware (leads and pulse generator) contamination [18]. The introduction of normal skin flora might occur in the surgical wound during the implantation or later with the development of erosion. Contamination of the pocket leads to bacterial growth and subsequent (mostly early) pocket infection [11, 19, 20]. Later in its course, the infection may spread along the leads and eventually cause secondary systemic infection resulting in device-related endocarditis. In the second mechanism, remote infectious foci (e.g., from contaminated vascular catheters, surgical site infection, septic thrombophlebitis, etc.) causing bacteremia might result in direct lead seeding, which later may progress to systemic infection, while the device pocket remains unaffected.

Device infections are caused mainly by Gram-positive bacteria (70%–90% of the isolates). Some of them are normally non-pathogenic. These are usually coagulase-negative staphylococci (mainly Staphylococcus epidermidis). Staphylococcus aureus is also a commonly isolated microorganism responsible for pocket infection (especially in early cases) and also the most common isolate in bacteremia [21–26]. Almost half of all staphylococcal CIED infections have been reported to be caused by methicillin-resistant staphylococci [21]. Gram-negative bacteria are isolated in about 9% of the cases, while fungi are rare [25]. No causative microorganism is identified in about a third of the patients [11].

Risk assessment

Any preventive measure shows the highest benefit when directed to the population at the highest risk. Identifying risk factors and risk stratification play a central role in determining the CIED recipients for whom more aggressive preventive measures should be taken to reduce the infection rate.

Factors associated with higher CIED infection risk can be modifiable, with specific interventions addressing them able to mitigate the risk, or non-modifiable, determining persistently elevated risk of infection. Apart from that, risk factors can be grouped into patient-related, procedure-related, and device-related [8]. The magnitude of different risk factors is presented in Table 1.

Table 1. Major risk factors for cardiac implantable electronic device infections

Risk factors

Odds ratio

Patient-related factors

End-stage renal disease

8.73

Prior CIED infection

7.84

Fever before implantation

4.27

Immunosuppression

3.44

Renal failure (eGFR <30 ml/min/1.73 m2)a

1.45a–3.02

COPD

2.95

NYHA class ≥II

2.47

Skin disorder

2.46

Immunocompromised (therapy or disease-suppressing resistance to infection)

2.28a

Malignancy

2.23

Diabetes mellitus

2.08

Heparin bridging

1.87

Congestive heart failure

1.65

Oral anticoagulation

1.59

Device-related factors

Epicardial leads

8.09

Abdominal pocket

4.01

CRT

2.73a

Two or more leads

2.02

ICD

1.77a

Dual chamber device

1.45

Procedure-related factors

Reintervention <30 days

16.29

Procedure duration >1 hour

13.96

Hematoma

4.95-11.3b

Revision or upgrade

4.01a–6.46

Lead repositioning

6.37

Replacement

4.93

Two or more prior procedures

3.43a

Inexperienced operator

2.85

Temporary pacing

2.31

Single prior procedure

1.51a (P = 0.058)a

Patient-related risk factors

Some comorbidities are well-established risk factors. End-stage renal disease and renal insufficiency or failure are consistently reported as one of the most important risk factors [14, 27, 28]. Other conditions such as immunosuppression, chronic obstructive pulmonary disease (COPD), congestive heart failure, valvular heart disease (esp. prior valvular surgery), systemic autoimmune disorders, malignancy, diabetes, and skin disorders also carry a significant risk of CIED infection [15, 27–29]. Although often non-modifiable as risk factors, optimal management of these conditions (e.g., control of diabetes) has been shown to lower infection risk [30].

Younger age has been identified as a risk factor in the randomized Prevention of Arrhythmia Device Infection Trial (PADIT) population and by a recent large observational study [15, 29]. While a consistent explanation for this finding is lacking, the qualities of subcutaneous tissue at a younger age might predispose to more traumatization during implantation (esp. pocket creation) and subsequent higher predisposition to infection [31]. Conversely, a smaller observational study including only 1000 ICD and CRT recipients demonstrated a significantly higher risk of CIED infection (odds ratio [OR], 5.93; 95% confidence interval [CI], 1.7719.84) in patients older than 75 years [28]. This might be due to the inclusion of more frail and morbid patients in this series.

Fever before implantation is another well-established and major modifiable risk factor for CIED infection (OR, 5.34; 95% CI,1.00228.43) [8]. The administration of certain medications, such as corticosteroids and antithrombotic drugs, also represents a potentially modifiable patient-related risk factor. According to a recent analysis of randomized data, a history of atrial arrhythmia and the number of previous procedures were also associated with increased infection risk after secondary procedures [32]. In the same study, some geographical regions (outside North America and Europe) and lower body mass index were also associated with increased risk.

Device-related factors

Device-related factors mainly include system size and complexity. Implantation of complex systems, presence of at least two leads, and implantation of high-power devices are associated with increased infection risk [8, 27, 33–35] (Table 1). A sizeable real-life registry from Denmark reported significantly increased infection risk in patients with complex devices ICD (HR, 1.26, 95% CI, 1.091.47), CRT-P (HR, 1.68; 95% CI, 1.672.11), and CRT-D systems (HR, 2.22; 95% CI, 1.832.70) as compared to conventional antibradycardia devices [35]. Another, more recent analysis of data from the same registry reported that complex systems (CRT-P and CRT-D) are associated with increased risk for both pocket and systemic CIED infections, while ICD implantation portended a higher risk of systemic infection compared to implantation of antibradycardia pacemakers [15]. In the PADIT study population, implantation of CRT and ICD, as well as secondary procedures, were all associated with increased risk for CIED infection in a full prediction model (OR, 2.73; 95% CI, 1.724.31; OR, 1.77; 95% CI, 1.092.87 and OR, 4.01; 95% CI, 2.626.13, for CRT, ICD, and secondary procedures, respectively) [14]. Similar findings were reported in the WRAP-IT dataset as well [32].

Procedure-related risk factors

Previously published observational and randomized studies demonstrated that early reintervention (within 30 days) and lengthy procedure duration (> 1 hour) were associated with the highest risk of CIED infections [8, 27, 32–34]. Procedure duration is mainly affected by procedure complexity, patients’ anatomy, and operator skills and experience.

Postprocedural hematoma is another well-established risk factor that has been widely studied. The randomized BRUISE CONTROL INFECTION study, including 659 patients with CIED infection, demonstrated that the development of hematoma was associated with more than 7-fold increased risk of infection (hazard ratio [HR], 7.7; 95% CI, 2.920.5) within one-year follow-up [33]. Another recent analysis based on the WRAP-IT population (n = 6800 participants) demonstrated an 11-fold increase in CIED infection risk (HR, 11.3; 95% CI, 5.523.2) in the patients developing clinically significant hematoma [34].

Risk scores

Risk score systems for preprocedural risk assessment represent an essential tool for better risk stratification of low- and high-risk patients. They can not only facilitate clinical decision-making and patient counseling but also help healthcare systems and decision-makers be prepared for the scale of these severe complications. Mittal et al. [36] were among the first to develop a risk scoring system that included 7 clinical variables and 0 to 25 points (a higher number signifying higher risk). The infection risk increased significantly from the low-risk group (score 0-7, 1% infection rate) to the medium-risk group (score 814, 3.4% infection rate), and the high-risk group (score ≥15, 11.1% infection rate). Shariff et al. [37] also proposed a risk score including ten clinical and procedural variables. In a retrospective study, in patients who underwent de novo CIED implantation, Shariff score ≥4 was associated with more than three-fold increased risk of CIED infection RR 3.20 (1.2912.59) [38]. Another risk score designed by Kolek et al. included several clinical variables also known to be associated with CIED infection risk [39, 40]. The only risk score developed based on a dataset of a randomized trial is the PADIT risk score system [29]. It identified five independent predictors: Prior procedure(s) (P, 1 = 1 point, at least 2 = 4 points), Age (A, 6069 years = 1 point, <60 years = 2 points), Depressed estimated glomerular filtration rate (D, <30 ml/min = 1 point), Immunocompromised (I, 3 points), and Type of procedure (T, ICD = 2 points, CRT = 4 points, revision or upgrade = 5 points). The score, ranging from 0 to 15 points, was used to group patients into low (<1%, 0 to 4 points), intermediate (1%–3%, 5 to 6 points), and high (>3%, ≥7 points) risk groups with hospitalization rates due to CIED infection of 0.51%, 1.42%, and 3.41%, respectively. The PADIT risk score was validated externally in a large dataset of 54 042 procedures where each unit increase in the PADIT risk score was associated with a 28% increase in the infection risk [41]. Following PADIT risk score development, Boriani et al. [42] developed the RI-AIAC infection score based on registry data with 2675 patients. The RI-AIAC score is a 5-point scoring system, and the authors have identified several major clinical characteristics associated with increased CIED infection risk (especially type of procedure and diabetes). Interestingly, a score created to assess the risk of bleeding complications in CIED recipients the PACE DRAP score has also been shown to be helpful in CIED infection risk stratification [28]. It is important to note that none of these risk scores are entirely exhaustive. For instance, the most widely used PADIT risk score does not include important risk factors such as prior CIED infection, some comorbidities (e.g. malignancy), and concomitant antithrombotic therapy. Real-life studies have shown that previous CIED infection remains an important risk factor despite adjustment for the PADIT risk score [41].

Prevention strategies

Infections associated with CIEDs represent a significant challenge for healthcare providers and systems. Therefore, prevention is essential to reduce their incidence and diminish mortality and morbidity associated with them. In the case of CIED infection, preventive strategies include multiple measures at different time points during the management of these patients before, during, and immediately after the implantation [8].

Preprocedural measures

Patient selection and preprocedural patient-related factors

Careful patient selection and procedure timing are essential in CIED infection prevention. The risk-benefit ratio should always be considered individually before the procedure, with strict adherence to the recommendations [10]. For instance, a significant proportion (up to 50%) of patients might not need reimplantation of a new device following extraction for CIED infection [43-45]. Cardioneuroablation, as a new treatment modality for vasovagal syncope, is also likely to make implantation in some patients obsolete [46]. Careful consideration of temporal variation in the risk and postponing an implantation/reimplantation procedure to gain time to implement preventive measures play a central role in the decision-making process [47].

Preprocedural fever is a factor that necessitates postponing the procedure. As suggested by the available data, a reasonable afebrile period before undertaking the implantation procedure is at least 24 hours [48]. Isolated leucocytosis, without other clinical symptoms and signs (bacteremia, elevated inflammatory biomarkers) of ongoing infection, has not been associated with CIED infections and should not delay implantation [49]. Optimizing treatment and better control of comorbidities (e.g., better glycemic control in diabetic patients) is very important to minimize the risk of CIED infections.

Some studies have demonstrated the benefit of identifying S. aureus carriers by nasal swabs and subsequent decolonization with topical mupirocin and chlorhexidine skin wash to reduce healthcare-associated S. aureus infections [50]. Whether this strategy would prove beneficial in reducing CIED-related infections has not been specifically studied.

Implantation of temporary pacing leads should be avoided to reduce the risk of CIED infection. Alternative solutions, such as considering transcutaneous pacing in the most severe cases or administering medications to increase heart rate, should be sought and implemented. When needed, temporary transvenous pacing is better carried out via jugular/subclavian access rather than groin access, as this may be associated with lower infection risk. If possible, removing all central venous lines should be considered before CIED surgery [27]. In the case when vascular access was via a subclavian vein, the CIED should be implanted on the contralateral side. If that is not possible or feasible, it is always advisable to postpone the implantation after removing the central venous line.

If hair removal at the procedural site is needed, this should be done using electric clippers with a disposable head (not razors) [51]. Preprocedural skin wash with an antimicrobial agent is not routinely recommended due to diverging data from studies on other types of surgery and not specifically CIED implantation [8].

Antithrombotic therapy

Patients undergoing CIED implantation frequently need concomitant antiplatelet or anticoagulant therapy. As shown above, postoperative hematoma is a decisive risk factor for CIED infection; therefore, every effort should be made to minimize the risk of hematoma formation. One widely implemented strategy is uninterrupted vitamin K antagonist (VKA) therapy during implantation [52]. The Bridge or Continue Coumadin for Device Surgery Randomized Controlled Trial (BRUISE-CONTROL) demonstrated that uninterrupted VKA therapy, as compared to heparin bridging, resulted in fewer clinically significant device-pocket hematomas as compared to a strategy involving perioperative VKA interruption and bridging with heparin in patients with high thromboembolic risk (including patients after mechanical heart valve replacement) 3.5 vs. 16% (RR, 0.19; 95% CI, 0.100.36) [53]. The randomized BRUISE CONTROL-2 study found no difference in the clinically significant device pocket hematoma incidence with continued direct oral anticoagulants (DOAC) therapy vs. DOAC interruption in patients with atrial fibrillation and CHA2DS2-VASc score ≥2 [54]. A combined analysis of these two trials also demonstrated similar bleeding and pocket hematoma outcomes between interrupted or continued DOAC therapy vs. uninterrupted VKA (OR, 0.86; 95% CI, 0.381.96) [55]. In patients with low thromboembolic risk, temporary withholding of oral anticoagulation for the implantation is a well-established strategy [8].

Concomitant single or dual antiplatelet therapy increases bleeding risk in CIED recipients [56, 57]. Analyses of randomized studies demonstrated that a clinically significant hematoma develops in 9.8% of the patients on concomitant antiplatelet therapy versus 4.3% in those without, corresponding to a doubling of the risk (OR, 1.97; 95% CI, 1.203.21) [55]. Therefore, recent guidelines and consensus documents recommend discontinuing antiplatelet therapy (especially P2Y12 inhibitors) for at least 5 days before the procedure, if possible [8, 10]. In patients on dual antiplatelet therapy following percutaneous coronary intervention (PCI), discontinuation of one of the antiplatelet agents (usually P2Y12 inhibitor) for 37 days before the procedure is recommended based on thromboembolic and bleeding risk [10].

Antibiotic prophylaxis

Previous studies have demonstrated a significant and considerable reduction in the incidence of CIED infection with preprocedural intravenous antibiotic prophylaxis [58, 59]. In the randomized trial of de Oliveira et al., preprocedural administration of 1 g cefazolin was associated with a significant reduction in infection rates as compared to placebo (0.63% vs. 3.28%; RR, 0.19; P = 0.016) [59]. Current guidelines recommend such a strategy as the standard of care [8, 10]. Antibiotics should protect against S. aureus as the most common causative organism in acute infections. Randomized trials have used flucloxacillin (12 g) and first-generation cephalosporins e.g., cefazolin (12 g) [14, 59]. In cases of allergy to beta-lactams, the recommended choice is vancomycin (15 mg/kg) [8]. Antibiotics against methicillin-resistant S. aureus (MRSA) are not used routinely and could be considered based on local MRSA prevalence and patient risk. Antibiotic administration should be completed within one hour before the skin incision to ensure adequate antibiotic tissue levels.

Alternative systems and approaches in high-risk patients

The development of technology has brought in leadless pacemakers and subcutaneous ICD as an alternative to conventional transvenous systems. As these devices have no or only minimal intravascular components, they are expected to be associated with lower risk of infection. The absence of a pocket in leadless pacemakers eliminates the risk of CIED pocket infection although hematogenous seeding might still be possible. Extensive observational studies report a significantly lower infection rate with this new technology, but results from randomized studies are lacking [60]. Leadless pacemakers may also be associated with reduced risk of infection in patients after transvenous lead extraction [61]. High costs and lack of reimbursement are among the factors limiting their use in clinical practice. However, when considering total costs for the management of patients with recurrent CIED infections, leadless pacemaker implantation seems to be financially justified, at least in some healthcare systems [62].

Subcutaneous ICDs (S-ICD) are a viable option for patients requiring protection from ventricular tachyarrhythmias and have no pacing or CRT indication. Results from the EFFORTLESS S-ICD registry showed that at five years of follow-up, the overall infection rate in S-ICD recipients was 2.4% with an erosion rate of 1.7% [63]. A recent secondary analysis of the PRAETORIAN trial demonstrated a significantly lower rate of systemic infections in S-ICD recipients than those receiving a transvenous ICD (0% vs. 1.2%) [64].

Implanting an epicardial system may also provide a solution in selected high-risk patients, particularly those in whom preserving venous access is crucial [65].

Other preprocedural measures

An appropriate environment in the operating room/catheterization laboratory where CIED implantations are carried out is essential. These facilities should meet all the standards applicable for other surgical procedures involving implants [8, 66]. The staff at the implantation facility should be trained to follow strict sterile techniques.

Procedure times should be minimized as the duration of the implantation is a well-established risk factor for CIED infection. Long procedures (>60 minutes) are associated with infectious complications [27]. Extensive real-life data have demonstrated that, compared to procedure durations up to 30 minutes, the risk of infections is 2.4-fold higher in procedures longer than 120 minutes [35]. Many factors have an impact on procedure duration. Among those are lack of appropriate staff training [67], certification of operators [68], and patient volume [69]. These are all organizational issues that should be best addressed before starting any activity i.e., before performing any procedures. However, procedural difficulties associated with patient-related factors, e.g. anatomical/structural abnormalities/changes or bleeding, also play a role in procedural duration.

Surgical technique and intraprocedural factors

Surgical preparation

Results from randomized trials demonstrated that skin antisepsis with a 2% alcoholic chlorhexidine solution was associated with a lower incidence of surgical site infections as compared to povidone-iodine (alcoholic or aqueous solution) [70]. It is also associated with a lower infection rate with intravascular catheter insertion [71]. Despite the lack of randomized data on CIED implantation, the use of alcoholic chlorhexidine is recommended [8]. To provide sufficient time for the antiseptic to exert its effect and to minimize fire hazards when using electrocautery, it should be left to dry completely before the incision is made. Many operators use iodophor-impregnated incisive drapes, but there are no data showing that they reduce infection rates [9].

Surgical technique

Good surgical techniques including minimizing operative tissue damage, meticulous hemostasis, and appropriate wound closure, are crucial elements in infection prevention during the CIED implantation procedure.

Gloves change

Many operators change their gloves initially during prepping and/or later before handling the device. This is usually done by removing the outer pair of gloves with double-gloving or re-scrubbing. Observational studies have shown a high rate of glove contamination during the implantation before handling the device [72]. As significant, randomized studies in the field are lacking, the practice of glove change has been recommended based on expert consensus. The use of non-powdered gloves is preferable because glove powder has been demonstrated to facilitate infection [73].

Hemostasis and prevention of hematoma

Adequate hemostasis is key in the prevention of hematoma formation. Minimizing trauma by respecting tissue architecture and ensuring good wound closure is extremely important. Electrocautery is widely implemented in most centers, but the use of a plasma electron avalanche knife has been shown to be associated with a reduced incidence of hematoma compared to electrocautery in high-risk patients [74]. Some observational studies advocate for the use of hemostatic agents such as tranexamic acid [75]. However, results are controversial, and therefore this strategy cannot be recommended as a standard practice until larger-scale studies demonstrate its unequivocal benefit and safety. Routine addition of epinephrine to the local anesthetic during the procedure is discouraged as one small randomized single-center study demonstrated a higher incidence of hematoma formation with this strategy [76]. Capsulectomy entails the removal of the fibrous capsule formed around the device during secondary procedures. The rationale behind this practice is that the fibrous capsule has been known to facilitate bacterial colonization and subsequent infection. A randomized study demonstrated that routine capsulectomy during secondary procedures results in more hemorrhagic complications (6.1% vs. 0.8%; P = 0.03) with no effect on the incidence of pocket infection (1.5% vs. 4.7%; P = 0.13) [77]. Therefore, performing capsulectomy on a routine basis is discouraged.

Pocket irrigation and local instillation of antibiotics and antiseptics

The PADIT trial demonstrated no difference in the infection rate with the application of incremental antibiotic strategy, including antibiotic pocket wash before skin closure along with postoperative cephalexin or cephadroxil as compared to preprocedural cefazolin infusion only [14]. The recent Randomized Stand-Alone Use of the Antimicrobial Envelope in High-Risk Cardiac Device Patients (ENVELOPE) trial showed no difference in infection rates in high-risk patients receiving chlorhexidine skin preparation, preprocedural antibiotics, and an AEE (control arm) compared to adding an antibiotic pocket wash and a 3-day course of postoperative antibiotics to the initial treatment [13]. Observational studies do not support performing routine povidone-iodine pocket irrigation to reduce infection rates [78]. Based on these data, local instillation with antibiotics or antiseptic solutions is not recommended [8]. However, gentamicin-impregnated collagen sponge use was associated with reduced CIED infections in a recent 10-year analysis with propensity score matching [79]. In all cases, vigorous pocket irrigation with saline should be done to remove debris and potential contaminants from the pocket during the implantation.

Antibiotic eluting envelopes

In their early versions, AEEs consisted of non-absorbable polypropylene mesh, but this design was associated with significant pocket fibrosis and was therefore abandoned. An antibacterial mesh envelope has been designed and marketed (TYRX™; Medtronic, Inc. Monmouth Junction, NJ, US). It is made of a synthetic mesh of glycolide, caprolactone, and trimethylene carbonate absorbed in the body over nine weeks. The mesh is coated with an absorbable polyacrylate polymer releasing minocycline and rifampin in the tissues over seven days. This antibiotic combination has been shown to have additive effects on resistant bacteria such as MRSA [80] and covers the whole spectrum of Staphylococcus spp. (81), as well as other species [82].

The randomized WRAP-IT trial assessed AEE benefits in patients undergoing device implantations. It included 6983 patients with high infection risk randomized to AEE vs. standard of care [82]. Major infections occurred in 0.7% of patients receiving TYRX™ vs. 1.2% in controls (HR, 0.60; 95% CI, 0.360.98) [11]. The positive outcome was entirely driven by the lower rate of pocket infections, which comprised 75% of all major events 0.4% vs. 1% in the control group (HR, 0.39; 95% CI, 0.210.72). The benefit of AEE was sustained during long-term follow-up [83]. A meta-analysis summarizing a major observational and randomized trials and a recent real-world study demonstrated similar findings [84, 85]. Further analyses of the WRAP-IT population showed a more than 11-fold higher risk of major CIED infection in patients with pocket hematoma and without the AEE [34]. In patients who received the AEE and later developed pocket hematoma, the risk was 82% lower (HR, 0.18; 95% CI, 0.040.85), and the infection rate was comparable to those without hematoma. A significant limitation of the study was the exclusion of patients at very high risk (e.g., those with pocket intervention in the previous 365 days, patients on dialysis, on chronic immunosuppressive therapy, or those with previous CIED infection within 12 months), which probably explains the lower than expected infection rate. The cost-effectiveness of this device, especially in high-risk patients, has been demonstrated in many healthcare systems [86–89].

Another available absorbable CIED envelope is made of a decellularized and non-crosslinked extracellular matrix produced from porcine intestinal submucosa [90]. This envelope does not possess antibiotic-eluting properties per se but can be impregnated with gentamycin before implantation [90, 91]. Before recommending this envelope for routine clinical use, results from ongoing randomized trials are awaited.

Wound closure

Adequate wound closure is of paramount importance to prevent pocket infections. Closure in layers has been shown to reduce the risk of dehiscence [92]. Various suture materials, staples, or adhesives may be used for wound closure. However, it is extremely important to ensure timely (within 714 days) removal of non-absorbable suture material. No firm data have demonstrated the impact of suture material on the infection risk, but consensus documents recommend the use of non-braided monofilament sutures for skin closure as they may be less prone to bacterial adhesion [8]. With absorbable sutures, care should be taken to avoid a “stitch abscess”, especially at the pole of the wound where the knot is located.

Postprocedural measures

Postprocedural antibiotic therapy

Postoperative antibiotic therapy is not recommended based on the results of the large PADIT trial. The trial tested the benefit of incremental perioperative antibiotic therapy to reduce CIED infections in a cluster cross-over design. In 19 603 patients (of whom 12 842 were high risk), the authors did not find a significant reduction in infections in the patients treated with an incremental regimen consisting of preprocedural cefazolin plus vancomycin, bacitracin pocket wash, and postoperative 2-day administration of oral cephalexin (OR, 0.77; 95% CI, 0.561.05) [14]. However, incremental antibiotic use compared to standard care was associated with a trend toward a 23% reduction in hospitalization for infection. This finding was not significant, at least in part due to the low infection rate during only 1-year follow-up in the PADIT trial [29]. Similarly, the recent ENVELOPE trial did not find an additional benefit of antibiotic pocket wash and a 3-day postoperative antibiotic course in addition to standard care and an AEE in high-risk CIED patients [13]. These results should be interpreted in light of the low incidence of CIED infections with peri- and post-operative antibiotic use (course of 5 days) in the long-term follow-up [93].

Wound care

Mechanical compression devices have also been designed to be applied after wound closure. Some of these devices have demonstrated benefit in reducing postoperative hematoma [94–96]. Pressure dressings may be used for 24 hours although their efficacy has not been demonstrated. In any case, a sterile dressing should be left on the wound for 210 days, and patients should be given instructions for wound care, i.e., changing the dressing only if impregnated with blood or wound secretions and not soaking the wound until completely healed [8].

Reintervention

Some procedure-related complications (lead dislodgement, hematoma, etc.) may require reintervention. Proper timing is crucial in these cases as the infection risk of repeat procedures is time-dependent and very high in early reinterventions [27, 48]. As shown by the Prospective Evaluation of Pacemaker Lead Endocarditis (PEOPLE) study, reinterventions before hospital discharge are associated with 15-fold increased risk of CIED infection [48]. Apart from taking all the measures to avoid the need for repeat procedures (meticulous hemostasis, good lead fixation, etc.), careful consideration of the risks and benefits of early reintervention is extremely important.

The most important risk factors and major risk reduction strategies are summarized in Figure 1.

Figure 1. The most important risk factors of cardiovascular implantable electronic device infections along with the corresponding strategies to mitigate that risk. Colored circles reflect the magnitude of risk based on the results presented in Table 1

Future directions

There are several gaps in evidence related to CIED infection prevention that require further study to answer important questions in the field. More studies on nasal and/or skin treatment of bacterial decolonization to prevent CIED infections would be valuable, especially in high-risk patients. Randomized studies on skin preparation before CIED placement and use of adhesive incise drapes are eagerly awaited. Studies on the use of antiseptic/antimicrobial solutions (e.g. taurolidine) for pocket and hardware wash are ongoing (NCT05576194), but large, randomized trials are needed. Investigations onto different approaches expected to increase guideline-driven care for patients with CIED infections are ongoing (NCT05471973).

Conclusion

As a result of increasing device complexity and more prevalent comorbidities in patients undergoing CIED placement, the incidence of CIED infections has grown significantly. They are associated with high morbidity and mortality, as well as high healthcare costs for hospital stay, diagnostics, medical therapy, and interventional (or surgical) procedures. Therefore, identifying risk factors is crucial for implementing structured prevention measures and actions at the preprocedural, intraprocedural, and postprocedural levels, which could bring about a meaningful reduction in CIED infection incidence. These actions should target patients and procedures but also should be directed to the environment in the operating room, staff training, and institutional measures. Importantly, studying and incorporating new methods and technologies such as AEEs, leadless pacemakers, and S-ICDs is another action to be taken for more effective prevention of CIED infections.

Article information

Conflict of interest: VT declares receiving speaker fees and other honoraria from Boehringer Ingelheim, Astra Zeneca, Berlin Menarini, Abbott, Novartis, Bayer, Merck, Pfizer, and Biotronik. KD declares receiving speaker fees from Sandoz, Servier, Boehringer Ingelheim, Astra Zeneca, Berlin-Chemie/Menarini, Novartis, Bayer, Pfizer, and Medtronic. PTM received speech honorarium from Boehringer Ingelheim, Polish Cardiac Society 2018 Scientific Grant in cooperation with Berlin-Chemie/Menarini (sponsor of the grant: Berlin-Chemie/Menarini Poland LLC) and participated in educational activities which were supported by CIED manufacturers as well as Polpharma. None of the declared potential conflicts of interest are related to the current work.

Funding: PTM contribution was supported by the National Science Centre, Poland (grant number 2021/05/X/NZ5/01511), Jagiellonian University Medical College, the John Paul II Hospital, and the Ministry of Science and Higher Education stipend for outstanding young scientists.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, which allows downloading and sharing articles with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.

REFERENCES

  1. Joy PS, Kumar G, Poole JE, et al. Cardiac implantable electronic device infections: Who is at greatest risk? Heart Rhythm. 2017; 14(6): 839845, doi: 10.1016/j.hrthm.2017.03.019, indexed in Pubmed: 28315744.
  2. Greenspon AJ, Patel JD, Lau E, et al. 16-year trends in the infection burden for pacemakers and implantable cardioverter-defibrillators in the United States 1993 to 2008. J Am Coll Cardiol. 2011; 58(10): 10011006, doi: 10.1016/j.jacc.2011.04.033, indexed in Pubmed: 21867833.
  3. Voigt A, Shalaby A, Saba S. Rising rates of cardiac rhythm management device infections in the United States: 1996 through 2003. J Am Coll Cardiol. 2006; 48(3): 590591, doi: 10.1016/j.jacc.2006.05.016, indexed in Pubmed: 16875991.
  4. Lee DH, Gracely EJ, Aleem SY, et al. Differences of mortality rates between pocket and nonpocket cardiovascular implantable electronic device infections. Pacing Clin Electrophysiol. 2015; 38(12): 14561463, doi: 10.1111/pace.12748, indexed in Pubmed: 26351247.
  5. Sridhar AR, Lavu M, Yarlagadda V, et al. Cardiac implantable electronic device-related infection and extraction trends in the U.S. Pacing Clin Electrophysiol. 2017; 40(3): 286293, doi: 10.1111/pace.13009, indexed in Pubmed: 28084622.
  6. Clémenty N, Carion PL, Léotoing Lde, et al. Infections and associated costs following cardiovascular implantable electronic device implantations: a nationwide cohort study. Europace. 2018; 20(12): 19741980, doi: 10.1093/europace/eux387, indexed in Pubmed: 29672690.
  7. Romanek J, Farkowski M, Bukowski H, et al. The cost of CIED infectious complications treatment in Poland from the perspective of Polish hospitals. Kardiol Pol. 2022; 80(9): 919925, doi: 10.33963/KP.a2022.0144, indexed in Pubmed: 35698969.
  8. Blomström-Lundqvist C, Traykov V, Erba PA, et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections-endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID) and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Europace. 2020; 22(4): 515549, doi: 10.1093/europace/euz246, indexed in Pubmed: 31702000.
  9. Traykov V, Bongiorni MG, Boriani G, et al. Clinical practice and implementation of guidelines for the prevention, diagnosis and management of cardiac implantable electronic device infections: results of a worldwide survey under the auspices of the European Heart Rhythm Association. Europace. 2019; 21(8): 12701279, doi: 10.1093/europace/euz137, indexed in Pubmed: 31209483.
  10. Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2021; 42(35): 34273520, doi: 10.1093/eurheartj/ehab364, indexed in Pubmed: 34455430.
  11. Tarakji KG, Mittal S, Kennergren C, et al. Antibacterial Envelope to Prevent Cardiac Implantable Device Infection. N Engl J Med. 2019; 380(20): 18951905, doi: 10.1056/NEJMoa1901111, indexed in Pubmed: 30883056.
  12. Knoll K, O’Connor M, Chouchane A, et al. A prospective case-control validation of procalcitonin as a biomarker diagnosing pacemaker and implantable cardioverter-defibrillator pocket infection. Kardiol Pol. 2022; 80(7-8): 782791, doi: 10.33963/KP.a2022.0124, indexed in Pubmed: 35521718.
  13. Ellis CR, Greenspon AJ, Andriulli JA, et al. Randomized trial of stand-alone use of the antimicrobial envelope in high-risk cardiac device patients. Circ Arrhythm Electrophysiol. 2023; 16(5): e011740, doi: 10.1161/CIRCEP.122.011740, indexed in Pubmed: 36960716.
  14. Krahn AD, Longtin Y, Philippon F, et al. Prevention of arrhythmia device infection trial: the PADIT trial. J Am Coll Cardiol. 2018; 72(24): 30983109, doi: 10.1016/j.jacc.2018.09.068, indexed in Pubmed: 30545448.
  15. Olsen T, Jørgensen OD, Nielsen JC, et al. Risk factors for cardiac implantable electronic device infections: a nationwide Danish study. Eur Heart J. 2022; 43(47): 49464956, doi: 10.1093/eurheartj/ehac576, indexed in Pubmed: 36263789.
  16. Han HC, Wang J, Birnie DH, et al. Association of the timing and extent of cardiac implantable electronic device infections with mortality. JAMA Cardiol. 2023; 8(5): 484491, doi: 10.1001/jamacardio.2023.0467, indexed in Pubmed: 37017943.
  17. Rizwan Sohail M, Henrikson CA, Jo Braid-Forbes M, et al. Increased long-term mortality in patients with cardiovascular implantable electronic device infections. Pacing Clin Electrophysiol. 2015; 38(2): 231239, doi: 10.1111/pace.12518, indexed in Pubmed: 25243484.
  18. Da Costa A, Lelièvre H, Kirkorian G, et al. Role of the preaxillary flora in pacemaker infections: a prospective study. Circulation. 1998; 97(18): 17911795, doi: 10.1161/01.cir.97.18.1791, indexed in Pubmed: 9603533.
  19. Uslan DZ, Sohail MR, St Sauver JL, et al. Permanent pacemaker and implantable cardioverter defibrillator infection: a population-based study. Arch Intern Med. 2007; 167(7): 669675, doi: 10.1001/archinte.167.7.669, indexed in Pubmed: 17420425.
  20. Polewczyk A, Jacheć W, Polewczyk M, et al. Early, delayed and late cardiac implantable electronic device infections: do the timing of onset and pathogens matter? J Clin Med. 2022; 11(14), doi: 10.3390/jcm11143929, indexed in Pubmed: 35887692.
  21. Hussein AA, Baghdy Y, Wazni OM, et al. Microbiology of Cardiac Implantable Electronic Device Infections. JACC Clin Electrophysiol. 2016; 2(4): 498505, doi: 10.1016/j.jacep.2016.01.019, indexed in Pubmed: 29759872.
  22. Bongiorni MG, Tascini C, Tagliaferri E, et al. Microbiology of cardiac implantable electronic device infections. Europace. 2012; 14(9): 13341339, doi: 10.1093/europace/eus044, indexed in Pubmed: 22399202.
  23. Wang R, Li X, Wang Qi, et al. Microbiological characteristics and clinical features of cardiac implantable electronic device infections at a tertiary hospital in China. Front Microbiol. 2017; 8: 360, doi: 10.3389/fmicb.2017.00360, indexed in Pubmed: 28321212.
  24. Sandoe JAT, Barlow G, Chambers JB, et al. Guidelines for the diagnosis, prevention and management of implantable cardiac electronic device infection. Report of a joint Working Party project on behalf of the British Society for Antimicrobial Chemotherapy (BSAC, host organization), British Heart Rhythm Society (BHRS), British Cardiovascular Society (BCS), British Heart Valve Society (BHVS) and British Society for Echocardiography (BSE). J Antimicrob Chemother. 2015; 70(2): 325359, doi: 10.1093/jac/dku383, indexed in Pubmed: 25355810.
  25. Sohail MR, Uslan DZ, Khan AH, et al. Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. J Am Coll Cardiol. 2007; 49(18): 18511859, doi: 10.1016/j.jacc.2007.01.072, indexed in Pubmed: 17481444.
  26. Longtin Y, Gervais P, Birnie DH, et al. Impact of choice of prophylaxis on the microbiology of cardiac implantable electronic device infections: insights from the prevention of arrhythmia device infection trial (PADIT). Open Forum Infect Dis. 2021; 8(11): ofab513, doi: 10.1093/ofid/ofab513, indexed in Pubmed: 34859113.
  27. Polyzos KA, Konstantelias AA, Falagas ME. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace. 2015; 17(5): 767777, doi: 10.1093/europace/euv053, indexed in Pubmed: 25926473.
  28. Sławek-Szmyt S, Araszkiewicz A, Grygier M, et al. Predictors of Long-Term Infections After Cardiac Implantable Electronic Device Surgery - Utility of Novel PADIT and PACE DRAP Scores. Circ J. 2020; 84(10): 17541763, doi: 10.1253/circj.CJ-20-0305, indexed in Pubmed: 32893235.
  29. Birnie DH, Wang J, Alings M, et al. Risk factors for infections involving cardiac implanted electronic devices. J Am Coll Cardiol. 2019; 74(23): 28452854, doi: 10.1016/j.jacc.2019.09.060, indexed in Pubmed: 31806127.
  30. de Vries FEE, Gans SL, Solomkin JS, et al. Meta-analysis of lower perioperative blood glucose target levels for reduction of surgical-site infection. Br J Surg. 2017; 104(2): e95e9e105, doi: 10.1002/bjs.10424, indexed in Pubmed: 27901264.
  31. Mahtani K, Maclean E, Schilling RJ. Prevention and Management of Cardiac Implantable Electronic Device Infections: State-of-the-Art and Future Directions. Heart Lung Circ. 2022; 31(11): 14821492, doi: 10.1016/j.hlc.2022.06.690, indexed in Pubmed: 35989213.
  32. Tarakji KG, Krahn AD, Poole JE, et al. Risk Factors for CIED Infection After Secondary Procedures: Insights From the WRAP-IT Trial. JACC Clin Electrophysiol. 2022; 8(1): 101111, doi: 10.1016/j.jacep.2021.08.009, indexed in Pubmed: 34600848.
  33. Essebag V, Verma A, Healey JS, et al. Clinically significant pocket hematoma increases long-term risk of device infection: BRUISE CONTROL INFECTION study. J Am Coll Cardiol. 2016; 67(11): 13001308, doi: 10.1016/j.jacc.2016.01.009, indexed in Pubmed: 26988951.
  34. Tarakji KG, Korantzopoulos P, Philippon F, et al. Infectious consequences of hematoma from cardiac implantable electronic device procedures and the role of the antibiotic envelope: A WRAP-IT trial analysis. Heart Rhythm. 2021; 18(12): 20802086, doi: 10.1016/j.hrthm.2021.07.011, indexed in Pubmed: 34280568.
  35. Olsen T, Jørgensen OD, Nielsen JC, et al. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982-2018). Eur Heart J. 2019; 40(23): 18621869, doi: 10.1093/eurheartj/ehz316, indexed in Pubmed: 31155647.
  36. Mittal S, Shaw RE, Michel K, et al. Cardiac implantable electronic device infections: incidence, risk factors, and the effect of the AigisRx antibacterial envelope. Heart Rhythm. 2014; 11(4): 595601, doi: 10.1016/j.hrthm.2013.12.013, indexed in Pubmed: 24333543.
  37. Shariff N, Eby E, Adelstein E, et al. Health and economic outcomes associated with use of an antimicrobial envelope as a standard of care for cardiac implantable electronic device implantation. J Cardiovasc Electrophysiol. 2015; 26(7): 783789, doi: 10.1111/jce.12684, indexed in Pubmed: 25845917.
  38. Balla C, Brieda A, Righetto A, et al. Predictors of infection after “de novo” cardiac electronic device implantation. Eur J Intern Med. 2020; 77: 7378, doi: 10.1016/j.ejim.2020.02.030, indexed in Pubmed: 32127301.
  39. Kolek MJ, Dresen WF, Wells QS, et al. Use of an antibacterial envelope is associated with reduced cardiac implantable electronic device infections in high-risk patients. Pacing Clin Electrophysiol. 2013; 36(3): 354361, doi: 10.1111/pace.12063, indexed in Pubmed: 23252988.
  40. Kolek MJ, Patel NJ, Clair WK, et al. Efficacy of a bio-absorbable antibacterial envelope to prevent cardiac implantable electronic device infections in high-risk subjects. J Cardiovasc Electrophysiol. 2015; 26(10): 11111116, doi: 10.1111/jce.12768, indexed in Pubmed: 26222980.
  41. Ahmed FZ, Blomström-Lundqvist C, Bloom H, et al. Use of healthcare claims to validate the Prevention of Arrhythmia Device Infection Trial cardiac implantable electronic device infection risk score. Europace. 2021; 23(9): 14461455, doi: 10.1093/europace/euab028, indexed in Pubmed: 33755136.
  42. Boriani G, Proietti M, Bertini M, et al. Incidence and predictors of infections and all-cause death in patients with cardiac implantable electronic devices: the italian nationwide RI-AIAC registry. J Pers Med. 2022; 12(1), doi: 10.3390/jpm12010091, indexed in Pubmed: 35055406.
  43. De Schouwer K, Vanhove R, Garweg C, et al. Re-implantation after extraction of a cardiac implantable electronic device. Acta Cardiol. 2020; 75(6): 505513, doi: 10.1080/00015385.2019.1620997, indexed in Pubmed: 31145671.
  44. Al-Hijji MA, Killu AM, Yousefian O, et al. Outcomes of lead extraction without subsequent device reimplantation. Europace. 2017; 19(9): 15271534, doi: 10.1093/europace/euw184, indexed in Pubmed: 27707785.
  45. Döring M, Hienzsch L, Ebert M, et al. Extraction of infected cardiac implantable electronic devices and the need for subsequent re-implantation. Int J Cardiol. 2020; 309: 8491, doi: 10.1016/j.ijcard.2019.12.044, indexed in Pubmed: 31973885.
  46. Piotrowski R, Baran J, Sikorska A, et al. Cardioneuroablation for reflex syncope: efficacy and effects on autonomic cardiac regulation-a prospective randomized trial. JACC Clin Electrophysiol. 2023; 9(1): 8595, doi: 10.1016/j.jacep.2022.08.011, indexed in Pubmed: 36114133.
  47. Maille B, Koutbi L, Resseguier N, et al. Seasonal variations in cardiac implantable electronic device infections. Heart Vessels. 2019; 34(5): 824831, doi: 10.1007/s00380-018-1292-4, indexed in Pubmed: 30415372.
  48. Klug D, Balde M, Pavin D, et al. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation. 2007; 116(12): 13491355, doi: 10.1161/CIRCULATIONAHA.106.678664, indexed in Pubmed: 17724263.
  49. Kumar DS, Tompkins CM, Veenhuyzen GD, et al. Significance of leukocytosis prior to cardiac device implantation. Pacing Clin Electrophysiol. 2018; 41(9): 11971200, doi: 10.1111/pace.13443, indexed in Pubmed: 29987911.
  50. Bode LGM, Kluytmans JA, Wertheim HFL, et al. Preventing surgical-site infections in nasal carriers of Staphylococcus aureus. N Engl J Med. 2010; 362(1): 917, doi: 10.1056/NEJMoa0808939, indexed in Pubmed: 20054045.
  51. Tanner J, Melen K, Tanner J, et al. Preoperative hair removal to reduce surgical site infection. Cochrane Database Syst Rev. 2011; 8(11): CD004122, doi: 10.1002/14651858.CD004122.pub4, indexed in Pubmed: 22071812.
  52. Du L, Zhang Y, Wang W, et al. Perioperative anticoagulation management in patients on chronic oral anticoagulant therapy undergoing cardiac devices implantation: a meta-analysis. Pacing Clin Electrophysiol. 2014; 37(11): 15731586, doi: 10.1111/pace.12517, indexed in Pubmed: 25234639.
  53. Birnie DH, Healey JS, Wells GA, et al. Pacemaker or defibrillator surgery without interruption of anticoagulation. N Engl J Med. 2013; 368(22): 20842093, doi: 10.1056/NEJMoa1302946, indexed in Pubmed: 23659733.
  54. Birnie DH, Healey JS, Wells GA, et al. Continued vs. interrupted direct oral anticoagulants at the time of device surgery, in patients with moderate to high risk of arterial thrombo-embolic events (BRUISE CONTROL-2). Eur Heart J. 2018; 39(44): 39733979, doi: 10.1093/eurheartj/ehy413, indexed in Pubmed: 30462279.
  55. Essebag V, Healey JS, Joza J, et al. Effect of direct oral anticoagulants, warfarin, and antiplatelet agents on risk of device pocket hematoma: combined analysis of BRUISE CONTROL 1 and 2. Circ Arrhythm Electrophysiol. 2019; 12(10): e007545, doi: 10.1161/CIRCEP.119.007545, indexed in Pubmed: 31610718.
  56. Fensman SK, Grove EL, Johansen JB, et al. Predictors of pocket hematoma after cardiac implantable electronic device surgery: A nationwide cohort study. J Arrhythm. 2022; 38(5): 748755, doi: 10.1002/joa3.12769, indexed in Pubmed: 36237873.
  57. Kutinsky IB, Jarandilla R, Jewett M, et al. Risk of hematoma complications after device implant in the clopidogrel era. Circ Arrhythm Electrophysiol. 2010; 3(4): 312318, doi: 10.1161/CIRCEP.109.917625, indexed in Pubmed: 20558847.
  58. Da Costa A, Kirkorian G, Cucherat M, et al. Antibiotic prophylaxis for permanent pacemaker implantation: a meta-analysis. Circulation. 1998; 97(18): 17961801, doi: 10.1161/01.cir.97.18.1796, indexed in Pubmed: 9603534.
  59. de Oliveira JC, Martinelli M, Nishioka SA, et al. Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators: results of a large, prospective, randomized, double-blinded, placebo-controlled trial. Circ Arrhythm Electrophysiol. 2009; 2(1): 2934, doi: 10.1161/CIRCEP.108.795906, indexed in Pubmed: 19808441.
  60. Crossley GH, Piccini JP, Longacre C, et al. Leadless versus transvenous single-chamber ventricular pacemakers: 3 year follow-up of the Micra CED study. J Cardiovasc Electrophysiol. 2023; 34(4): 10151023, doi: 10.1111/jce.15863, indexed in Pubmed: 36807378.
  61. El-Chami MF, Johansen JB, Zaidi A, et al. Leadless pacemaker implant in patients with pre-existing infections: Results from the Micra postapproval registry. J Cardiovasc Electrophysiol. 2019; 30(4): 569574, doi: 10.1111/jce.13851, indexed in Pubmed: 30661279.
  62. Syska P, Farkowski MM, Raulinajtys-Grzybek M, et al. How to decrease the cost of pacemaker infection treatment by adopting seemingly costly innovation? A budget impact analysis of a leadless pacemaker implantation. Kardiol Pol. 2022; 80(12): 12601262, doi: 10.33963/KP.a2022.0284, indexed in Pubmed: 36573599.
  63. Boersma L, Barr C, Knops R, et al. Implant and Midterm Outcomes of the Subcutaneous Implantable Cardioverter-Defibrillator Registry: The EFFORTLESS Study. J Am Coll Cardiol. 2017; 70(7): 830841, doi: 10.1016/j.jacc.2017.06.040, indexed in Pubmed: 28797351.
  64. Knops RE, Pepplinkhuizen S, Delnoy PP, et al. Device-related complications in subcutaneous versus transvenous ICD: a secondary analysis of the PRAETORIAN trial. Eur Heart J. 2022; 43(47): 48724883, doi: 10.1093/eurheartj/ehac496, indexed in Pubmed: 36030464.
  65. Asif A, Carrillo R, Garisto JD, et al. Epicardial cardiac rhythm devices for dialysis patients: minimizing the risk of infection and preserving central veins. Semin Dial. 2012; 25(1): 8894, doi: 10.1111/j.1525-139X.2010.00757.x, indexed in Pubmed: 20807391.
  66. Haines DE, Beheiry S, Akar JG, et al. Heart Rythm Society expert consensus statement on electrophysiology laboratory standards: process, protocols, equipment, personnel, and safety. Heart Rhythm. 2014; 11(8): e951, doi: 10.1016/j.hrthm.2014.03.042, indexed in Pubmed: 24814989.
  67. Al-Khatib SM, Greiner MA, Peterson ED, et al. Patient and implanting physician factors associated with mortality and complications after implantable cardioverter-defibrillator implantation, 2002-2005. Circ Arrhythm Electrophysiol. 2008; 1(4): 240249, doi: 10.1161/CIRCEP.108.777888, indexed in Pubmed: 19169382.
  68. Curtis JP, Luebbert JJ, Wang Y, et al. Association of physician certification and outcomes among patients receiving an implantable cardioverter-defibrillator. JAMA. 2009; 301(16): 16611670, doi: 10.1001/jama.2009.547, indexed in Pubmed: 19383957.
  69. Lin YS, Hung SP, Chen PR, et al. Risk factors influencing complications of cardiac implantable electronic device implantation: infection, pneumothorax and heart perforation: a nationwide population-based cohort study. Medicine (Baltimore). 2014; 93(27): e213, doi: 10.1097/MD.0000000000000213, indexed in Pubmed: 25501080.
  70. Darouiche RO, Wall MJ, Itani KMF, et al. Chlorhexidine-Alcohol versus povidone-iodine for surgical-site antisepsis. N Engl J Med. 2010; 362(1): 1826, doi: 10.1056/NEJMoa0810988, indexed in Pubmed: 20054046.
  71. Mimoz O, Lucet JC, Kerforne T, et al. Skin antisepsis with chlorhexidine-alcohol versus povidone iodine-alcohol, with and without skin scrubbing, for prevention of intravascular-catheter-related infection (CLEAN): an open-label, multicentre, randomised, controlled, two-by-two factorial trial. Lancet. 2015; 386(10008): 20692077, doi: 10.1016/S0140-6736(15)00244-5, indexed in Pubmed: 26388532.
  72. Kozon I, Riahi S, Lundbye-Christensen S, et al. Risk factors of cardiac device infection: Glove contamination during device procedures. Am J Infect Control. 2017; 45(8): 866871, doi: 10.1016/j.ajic.2017.03.019, indexed in Pubmed: 28526307.
  73. Suding P, Nguyen T, Gordon I, et al. Glove powder increases Staphylococcus aureus abscess rate in a rat model. Surg Infect (Larchmt). 2010; 11(2): 133135, doi: 10.1089/sur.2008.055, indexed in Pubmed: 19622028.
  74. Lind A, Ahsan M, Kaya E, et al. Early Pacemaker Implantation after Transcatheter Aortic Valve Replacement: Impact of PlasmaBlade™ for Prevention of Device-Associated Bleeding Complications. Medicina (Kaunas). 2021; 57(12), doi: 10.3390/medicina57121331, indexed in Pubmed: 34946276.
  75. Beton O, Saricam E, Kaya H, et al. Bleeding complications during cardiac electronic device implantation in patients receiving antithrombotic therapy: is there any value of local tranexamic acid? BMC Cardiovasc Disord. 2016; 16: 73, doi: 10.1186/s12872-016-0251-1, indexed in Pubmed: 27105588.
  76. Ilov N, Ilov N, Nechepurenko A, et al. Arguments to apply epinephrine for pocket hematoma reduction. The MAITRE study. J Atr Fibrillation. 2016; 9(1): 1391, doi: 10.4022/jafib.1391, indexed in Pubmed: 27909508.
  77. Lakkireddy D, Pillarisetti J, Atkins D, et al. IMpact of pocKet rEvision on the rate of InfecTion and other CompLications in patients rEquiring pocket mAnipulation for generator replacement and/or lead replacement or revisioN (MAKE IT CLEAN): A prospective randomized study. Heart Rhythm. 2015; 12(5): 950956, doi: 10.1016/j.hrthm.2015.01.035, indexed in Pubmed: 25625720.
  78. Lakkireddy D, Valasareddi S, Ryschon K, et al. The impact of povidone-iodine pocket irrigation use on pacemaker and defibrillator infections. Pacing Clin Electrophysiol. 2005; 28(8): 789794, doi: 10.1111/j.1540-8159.2005.00173.x, indexed in Pubmed: 16105006.
  79. Matteucci A, Bonanni M, Massaro G, et al. Treatment with gentamicin-impregnated collagen sponges in reducing infection of implantable cardiac devices: 10-year analysis with propensity score matching. Rev Port Cardiol. 2023; 42(8): 711717, doi: 10.1016/j.repc.2023.01.023, indexed in Pubmed: 37085085.
  80. Bowker KE, Noel AR, Macgowan AP. Pharmacodynamics of minocycline against Staphylococcus aureus in an in vitro pharmacokinetic model. Antimicrob Agents Chemother. 2008; 52(12): 43704373, doi: 10.1128/AAC.00922-07, indexed in Pubmed: 18519719.
  81. Gleva MJ, Poole JE. Prevention of cardiac implantable electronic device infections: update and evaluation of the potential role for capsulectomy or the antibiotic pouch. J Atr Fibrillation. 2017; 9(5): 1540, doi: 10.4022/jafib.1540, indexed in Pubmed: 29250274.
  82. Tarakji KG, Krahn AD, Poole JE, et al. Worldwide randomized antibiotic envelope infection prevention trial (WRAP-IT). Am Heart J. 2016; 180(7): 1221, doi: 10.1016/j.ahj.2016.06.010, indexed in Pubmed: 27659878.
  83. Mittal S, Wilkoff BL, Kennergren C, et al. The World-wide Randomized Antibiotic Envelope Infection Prevention (WRAP-IT) trial: Long-term follow-up. Heart Rhythm. 2020; 17(7): 11151122, doi: 10.1016/j.hrthm.2020.02.011, indexed in Pubmed: 32087357.
  84. Asbeutah AA, Salem MH, Asbeutah SA, et al. The role of an antibiotic envelope in the prevention of major cardiac implantable electronic device infections: A systematic review and meta-analysis. Medicine (Baltimore). 2020; 99(26): e20834, doi: 10.1097/MD.0000000000020834, indexed in Pubmed: 32590773.
  85. Ziacchi M, Biffi M, Iacopino S, et al. REducing INFectiOns thRough Cardiac device Envelope: insight from real world data. The REINFORCE Project. Europace. 2023 [Epub ahead of print], doi: 10.1093/europace/euad224, indexed in Pubmed: 37490930.
  86. Frausing MH, Johansen JB, Afonso D, et al. Cost-effectiveness of an antibacterial envelope for infection prevention in patients undergoing cardiac resynchronization therapy reoperations in Denmark. Europace. 2023; 25(6), doi: 10.1093/europace/euad159, indexed in Pubmed: 37345858.
  87. Boriani G, Kennergren C, Tarakji KG, et al. Cost-Effectiveness analyses of an absorbable antibacterial envelope for use in patients at increased risk of cardiac implantable electronic device infection in Germany, Italy, and England. Value Health. 2021; 24(7): 930938, doi: 10.1016/j.jval.2020.12.021, indexed in Pubmed: 34243836.
  88. Kay G, Eby EL, Brown B, et al. Cost-effectiveness of TYRX absorbable antibacterial envelope for prevention of cardiovascular implantable electronic device infection. J Med Econ. 2018; 21(3): 294300, doi: 10.1080/13696998.2017.1409227, indexed in Pubmed: 29171319.
  89. Wilkoff BL, Boriani G, Mittal S, et al. Cost-Effectiveness of an antibacterial envelope for cardiac implantable electronic device infection prevention in the US healthcare system from the WRAP-IT trial. Circ Arrhythm Electrophysiol. 2020; 13(10): e008503, doi: 10.1161/CIRCEP.120.008503, indexed in Pubmed: 32915063.
  90. Xiang K, Catanzaro JN, Elayi C, et al. Antibiotic-Eluting envelopes to prevent cardiac-implantable electronic device infection: past, present, and future. Cureus. 2021; 13(2): e13088, doi: 10.7759/cureus.13088, indexed in Pubmed: 33728111.
  91. Deering TF, Catanzaro JN, Woodard DA. Physician antibiotic hydration preferences for biologic antibacterial envelopes during cardiac implantable device procedures. Front Cardiovasc Med. 2022; 9: 1006091, doi: 10.3389/fcvm.2022.1006091, indexed in Pubmed: 36620632.
  92. Grubb BP, Welch M, Karabin B, et al. Initial experience with a technique for wound closure after cardiac device implantation designed to reduce infection and minimize tissue scar formation. Am J Ther. 2012; 19(2): 8891, doi: 10.1097/MJT.0b013e3181ebd344, indexed in Pubmed: 20720483.
  93. Senaratne JM, Wijesundera J, Chhetri U, et al. Reduced incidence of CIED infections with peri- and post-operative antibiotic use in CRT-P/D and ICD procedures. Medicine (Baltimore). 2022; 101(40): e30944, doi: 10.1097/MD.0000000000030944, indexed in Pubmed: 36221436.
  94. Turagam MK, Nagarajan DV, Bartus K, et al. Use of a pocket compression device for the prevention and treatment of pocket hematoma after pacemaker and defibrillator implantation (STOP-HEMATOMA-I). J Interv Card Electrophysiol. 2017; 49(2): 197204, doi: 10.1007/s10840-017-0235-9, indexed in Pubmed: 28413854.
  95. Awada H, Geller JC, Brunelli M, et al. Pocket related complications following cardiac electronic device implantation in patients receiving anticoagulation and/or dual antiplatelet therapy: prospective evaluation of different preventive strategies. J Interv Card Electrophysiol. 2019; 54(3): 247255, doi: 10.1007/s10840-018-0488-y, indexed in Pubmed: 30460588.
  96. Fei YP, Wang L, Zhu CY, et al. Effect of a novel pocket compression device on hematomas following cardiac electronic device implantation in patients receiving direct oral anticoagulants. Front Cardiovasc Med. 2022; 9: 817453, doi: 10.3389/fcvm.2022.817453, indexed in Pubmed: 35282349.