What’s new? This retrospective study suggests that the percutaneous microaxial blood pump, Impella, is safe and effective in the treatment of high-risk percutaneous coronary intervention (HR-PCI). The PROTECT IV trial aimed to determine the safety and efficacy of Impella use in HR-PCI patients. The risk profile and mortality in cardiogenic shock (CS) patients were higher than in other registries; therefore, it remains challenging to compare our results with previously published data. The potential benefits of Impella in CS should be further investigated. |
Introduction
The use of mechanical circulatory support (MCS) devices, developed to provide circulatory support in the setting of critical cardiogenic shock (CS) or end-stage heart failure (HF), has expanded to prophylactic short-term support during percutaneous cardiovascular procedures [1]. Joint efforts in biomedical engineering over the last 50 years have led to a shift from intracorporeal surgically-implanted MCS devices to the first extracorporeal percutaneous MCS devices, including the intra-aortic balloon pump (IABP) and percutaneous microaxial blood pump (Impella) (Abiomed, Danvers, MA, US) [2]. Impella can provide hemodynamic support by continuously pumping blood from the left ventricle into the ascending aorta [3]. According to the European Society of Cardiology (ESC) guidelines, Impella should be considered in CS as bridge-to-recovery, bridge-to-decision, or bridge-to-bridge therapy (class IIa recommendation) [4]. Elective use of Impella during high-risk percutaneous coronary intervention (HR-PCI) procedures, while not clearly endorsed by the ESC [5], is advocated by the American College of Cardiology to prevent hemodynamic deterioration in selected high-risk patients, especially those with multivessel disease (MVD), left main (LM) disease, disease of the last patent conduit, and severe left ventricular dysfunction (class IIb recommendation) [6]. Since data on the superiority of Impella over the IABP are conflicting [7, 8], studies that evaluate the efficacy, safety, and cost-effectiveness of Impella use in real-world settings are urgently needed. Given that large, randomized trials of hemodynamic support in patients with CS and undergoing HR-PCI are challenging to conduct, national and international registries are a crucial source of high-quality data that provide novel insights into the characteristics of patients treated with Impella, supporting the decision-making process. Hitherto, four registries that specifically focus on Impella devices have been conducted: the Impella Italian Registry (IMP-IT) and German Registry in Europe, Japanese Registry for Percutaneous Ventricular Assist Device (J-PVAD) in Asia, and Catheter-Based Ventricular Assist Devices (cVAD) Registry in the US [9–12]. Regarding differences in international clinical practice and the dynamic development of Impella hemodynamic technology, the national, multicenter, investigator-initiated IMPELLA-PL registry was developed to share the knowledge and clinical experiences collected since the implementation of Impella technology in Poland.
The main goal of the IMPELLA-PL registry was to (1) describe clinical characteristics of patients treated with Impella during HR-PCI and CS; (2) evaluate the efficacy and safety of Impella-assisted treatment according to the prespecified endpoint definitions; and (3) compare the results with other registries.
Methods
Design
IMPELLA-PL is a national, multicenter, retrospective registry conducted under auspices of the Polish Association of Cardiovascular Interventions [13]. The registry included consecutive treated with Impella for CS and HR-PCI in all Polish interventional cardiac centers which performed at least 3 interventions using Impella. IMPELLA-PL included consecutive patients treated with Impella for CS and HR-PCI. The subgroup of patients undergoing Impella-assisted revascularization included hemodynamically stable patients with severe coronary artery disease undergoing elective or urgent HR-PI after a Heart Team had determined that it was the appropriate therapeutic option. The subgroup of patients treated with Impella due to CS included those with ongoing CS refractory to the optimal medical management and conventional treatment measures, including volume loading and use of pressors and inotropes, with or without an IABP [13].
Clinical characteristics, procedural data, and outcomes for consecutive patients treated with Impella devices from 2014 until December 2021 were collected retrospectively in a password-protected database, with a 12-month follow-up data collected on the basis of in-hospital and ambulatory medical records.
Endpoints
The main efficacy endpoints were (1) in-hospital mortality, (2) 12-month mortality; and (3) 12-month major adverse cardiovascular and cerebrovascular events (MACCEs), including mortality, rehospitalization for HF, acute myocardial infarction (MI), repeat revascularization, stroke, left ventricular assist device (LVAD) implantation, and heart transplantation following hospital discharge. Data on efficacy and safety were collected as well, including cardiosurgical intervention, exacerbation of HF, MI, acute kidney injury (AKI) inflammatory complications, severe bleeding complications (per operator judgment and defined as type ≥3 according to the Bleeding Academic Research Consortium; BARC), and device-related complications. The prespecified endpoint definitions have been published previously [13].
Statistical analysis
Statistical analysis was performed by an independent statistician with IBM SPSS Statistics, version 24.0. Categorical variables were summarized using frequencies and proportions and compared using the χ2 test. Continuous data were expressed as means (standard deviations) or medians (interquartile ranges) and compared using a t-test or Mann–Whitney U test, depending on distribution. Statistical tests were two-sided, with a significance level of 0.05.
Results
Altogether, 308 patients were enrolled in the registry in 20 Polish centers, including 253 (82.1%) who received Impella support for HR-PCI and 55 (17.9%) who received it for CS. Trends in the use of Impella in Poland during the study period in patients presenting with CS (blue line), undergoing HR-PCI (green line), and total insertions (red line) are shown in Figure 1.
The study chart diagram is shown in Figure 2. Overall, the use of Impella increased steadily from 2014 to 2019 and exponentially from 2019 to 2021, with 4.6-fold higher Impella use in HR-PCI, compared to CS. Baseline characteristics and angiographic and procedural characteristics of patients treated with Impella for CS and HR-PCI are presented in Table 1 and Table 2, respectively. In-hospital and 12-month outcomes are reported in Table 3.
Cardiogenic shock (n = 55; 17.9) |
HR-PCI (n = 253; 82.1) |
|
Age |
63.0 (50.0–69.0) |
70.0 (64.0–78.0) |
Male sex, n (%) |
42 (76.4) |
221 (87.4) |
BMI, kg/m2 |
27.7 (24.7–31.1) |
27.1 (24.4–30.5) |
Clinical presentation |
||
Acute coronary syndrome, n (%) |
49 (89.1) |
135 (53.4) |
STEMI, n (%) |
40 (72.7) |
6 (4.4) |
NSTEMI, n (%) |
9 (16.4) |
108 (80) |
Unstable angina, n (%) |
0 (0.0) |
21 (15.6) |
Chronic coronary syndrome, n (%) |
0 (0.0) |
118 (46.6) |
Myocarditis, n (%) |
2 (3.6) |
0 (0.0) |
Risk factors |
||
Hypertension, n (%) |
26 (47.3) |
199 (78.7) |
Dyslipidemia, n (%) |
21 (38.2) |
198 (78.3) |
Diabetes mellitus, n (%) |
18 (32.7) |
118 (46.6) |
Prior MI, n (%) |
19 (35.5) |
132 (52.2) |
Previous PCI, n (%) |
13 (23.6) |
93 (36.8) |
Previous CABG, n (%) |
0 (0) |
27 (10.7) |
Atrial fibrillation, % |
10 (18.2) |
75 (29.6) |
Paroxysmal |
8 |
37 |
Permanent |
1 |
25 |
Persistent |
1 |
13 |
Chronic heart failure, n (%) |
53 (96.4) |
249 (98.4) |
Previous stroke, n (%) |
7 (12.7) |
24 (9.5) |
Previous TIA, n (%) |
3 (5.5) |
12 (4.7) |
Chronic kidney disease, n (%) |
18 (32.7) |
94 (37.2) |
Dialysis, n (%) |
1 (1.8) |
4 (1.6) |
COPD, n (%) |
3 (5.5) |
28 (11.5) |
PAD, n (%) |
7 (12.7) |
76 (30.0) |
EuroSCORE II, median (range) |
21.8 (12.4–37.6) |
5.1 (2.7–9.4) |
Cardiac arrest before admission, n (%) |
26 (47.3) |
9 (3.6) |
VF, n (%) |
16 (29.1) |
4 (1.6) |
VT, n (%) |
3 (5.5) |
2 (0.8) |
PEA, n (%) |
4 (7.3) |
2 (0.8) |
Asystole, n (%) |
5 (9.1) |
1 (0.4) |
ICED, n (%) |
3 (5.5) |
43 (17.0) |
Pacemaker, n (%) |
0 (0) |
10 (4.0) |
ICD, n (%) |
3 (5.5 ) |
28 (11.1) |
CRT, n (%) |
0 (0) |
12 (4.7) |
Laboratory investigations |
||
Hemoglobin, g/dl |
13.3 (2.4) |
13.0 (2.2) |
Platelets, x109/l |
244.9 (88.7) |
222.6 (90.9) |
Creatinine, mg/dl |
1.4 (1.4) |
1.4 (0.7) |
NT-proBNP, pg/ml |
8784 (9357) |
7918 (14132) |
Troponin, ng/ml |
387 (1348) |
467 (3636) |
pH |
7.3 (7.1–7.4) |
7.4 (7.4–7.5) |
Lactate, mmol/l |
7.4 (7.2–7.5) |
1.7 (1.3–4.4) |
Echocardiographic characteristics |
||
LVEDD, mm |
53.5 (48.0-59.5) |
60.0 (53.0-66.3) |
LA, mm |
44.0 (38.0-45.0) |
45.0 (42.0-50.0) |
LVEF, % |
22.5 (15.0-29.5) |
26.0 (20.0-37.0) |
RV dysfunction, n (%) |
12 (21.8) |
45 (17.8) |
Mitral regurgitation grade 3 or 4, n (%) |
6 (10.9) |
43 (17.0) |
Tricuspid regurgitation grade 3 or 4, n (%) |
7 (12.7) |
36 (14.2) |
Severe aortic stenosis, n (%) |
1 (1.8) |
3 (1.2) |
Cardiogenic shock (n = 55; 17.9) |
High-risk PCI (n = 253; 82.1) |
|
Angiographic characteristics, n (%) |
||
Coronary angiography performed, % |
50 (90.9) |
253 (100.0) |
Number of vessels with significant stenosis |
3 (1.0-4.0) |
3 (3.0-4.0) |
Severe calcifications |
15 (27.3)a |
140 (55.3) |
Chronic total occlusions |
14 (25.5)a |
137 (54.2) |
In-stent restenosis |
3 (5.5)a |
17 (6.7) |
In-stent thrombosis |
2 (3.6)a |
1 (0.4) |
Intravascular imaging |
11 (20.0)a |
104 (41.1) |
IVUS |
11 (20.0) |
102 (40.3) |
OCT |
0 (0) |
2 (0.8) |
Functional assessment |
0 (0) |
9 (3.6) |
Extent of the disease |
||
One-vessel |
8 (14.5) |
1 (0.4) |
Multi-vessel (except for LM) |
17 (30.9) |
61 (24.1) |
Multi-vessel (including LM) |
21 (38.2) |
161 (63.6) |
Missing data |
30 (11.9) |
9 (16.4) |
SYNTAX Score II |
38.5 (32.3–47.5) |
43 (32.4–55.0) |
Procedural characteristics, n (%) |
||
PCI performed |
46 (83.6) |
251 (99.2) |
Rotational atherectomy used |
5 (9.1) |
77 (30.4) |
All lesions successfully treated |
35 (63.6) |
210 (83.0) |
Vessel treated |
||
LM |
26 (47.3) |
175 (69.2) |
LAD |
34 (61.8) |
198 (78.3) |
Cx |
14 (25.5) |
140 (55.3) |
RCA |
11 (20.0) |
48 (19.0) |
Impella |
||
Use of Impella CP, n (%) |
54 (98.2) |
253 (100.0) |
Use of Impella 5.0, n (%) |
1 (1.8) |
0 (0.0) |
Timing of Impella placement |
||
Before PCI, n (%) |
29 (52.7) |
207 (81.8) |
During PCI, n (%) |
15 (27.3) |
44 (17.4) |
After PCI, n (%) Missing data, n (%) |
8 (14.5) 3 (5.5) |
0 (0) 2 (3.6) |
Explantation in catheterization lab, n (%) |
8 (14.5) |
237 (93.7) |
Time of insertion, min |
20.0 (15.0–31.0) |
25.0 (15.0–40.0) |
Duration of support, h |
45.0 (19.0–120.0) |
3.0 (2.0–73.0) |
Vascular access for Impella |
||
Right femoral artery, n (%) |
32 (58.2) |
138 (54.5) |
Left femoral artery, n (%) |
22 (40.0) |
101 (39.9) |
Right subclavian artery, n (%) |
1 (1.8) |
8 (3.2) |
Left subclavian artery, n (%) |
0 (0) |
6 (2.4) |
Ultrasound-guided puncture, n (%) |
18 (32.7) |
70 (27.7) |
Surgical access, n (%) |
1 (1.8) |
38 (15.0) |
Single access, n (%) |
4 (7.3) |
45 (17.8) |
Contralateral safety access, n (%) |
1 (1.8) |
22 (8.7) |
Other cardiopulmonary support |
||
Use of catecholamines, n (%) |
54 (98.2) |
47 (18.6) |
Use of levosimendan, n (%) |
15 (27.3) |
13 (5.1) |
Use of mechanical ventilation, n (%) |
44 (80.0) |
10 (4.0) |
Mechanical ventilation, hours |
43.0 (24.0–110.0) |
46.0 (7.75–75.0) |
Use of ECMO, n (%) |
7 (12.7) |
6 (2.4) |
Use of IABP, n (%) |
14 (25.5) |
5 (2.0) |
Use of other LVAD, n (%) |
12 (21.8) |
27 (10.7) |
Last available LVEF, n (%) |
27.7 (12.6) |
32.9 (12.4) |
In-hospital stay, days |
5.5 (2.0–15.0) |
11.0 (7.0–18.0) |
Intensive care stay, days |
3.5 (2.0–9.0) |
6.5 (2.3–30.8) |
Cardiogenic shock (n = 55; 17.9) |
High-risk PCI (n = 253; 82.1) |
|
In-hospital outcomes |
||
Mortality, n (%) |
42 (76.4) |
21 (8.3) |
Need for cardiosurgical intervention, n (%) |
4 (7.3) |
1 (0.4) |
Exacerbation of HF, n (%) |
35 (63.6) |
12 (4.7) |
Acute myocardial infarction, n (%) |
5 (9.1) |
11 (4.3) |
Acute kidney injury, n (%) |
34 (61.8) |
32 (12.6) |
Need for renal replacement therapy, n (%) |
18 (32.7) |
4 (1.6) |
Inflammatory complications, n (%) |
22 (40.0) |
35 (13.8) |
Any bleeding complications, n (%) |
25 (45.5) |
34 (13.4) |
Severe bleeding complications, n (%) |
19 (34.5) |
16 (6.3) |
BARC 3a |
6 (10.9) |
12 (4.7) |
BARC 3b |
7 (12.7) |
4 (1.6) |
BARC 3c |
0 (0.0) |
0 (0.0) |
BARC 5a |
4 (7.3) |
0 (0.0) |
BARC 5b |
2 (3.6) |
0 (0.0) |
RBC transfusion, n (%) |
22 (40) |
34 (13.4) |
Number of RBC units transfused |
4.5 (3–6.5) |
2 (2.0–2.0) |
Device-related complications, n (%) |
||
Any access site bleeding |
17 (30.9) |
37 (14.6) |
Limb ischemia |
7 (12.7) |
6 (2.4) |
Endovascular intervention |
3 (5.5) |
8 (3.2) |
Surgical intervention |
3 (5.5) |
8 (3.2) |
Hemolysis |
6 (10.9) |
4 (1.6) |
Aortic injury |
0 (0.0) |
1 (0.4) |
12-month outcomes, n (%) |
||
Mortality after discharge |
2 (3.6) |
25 (9.9) |
Rehospitalization for HF |
3 (5.5) |
25 (9.9) |
MI |
1 (1.8) |
3 (1.2) |
Repeat revascularization |
0 (0) |
8 (3.2) |
PCI |
0 (0) |
8 (3.2) |
CABG |
0 (0) |
0 (0) |
Stroke |
1 (1.8) |
4 (1.6) |
Permanent LVAD implantation |
1 (1.8) |
1 (0.4) |
Heart transplantation |
1 (1.8) |
3 (1.2) |
Number of MACCEs |
9 (16.3) |
69 (27.3) |
Number of patients that experienced MACCEs |
5 (9.1) |
57 (22.5) |
Total mortality |
44 (80.0) |
46 (18.2) |
In patients who received Impella before PCI |
23/29 (79.3) |
39/207 (18.8) |
In patients who received Impella during or after PCI |
19/23 (82.6) |
5/44 (11.4) |
Impella for cardiogenic shock
In terms of baseline characteristics (Table 1), the median age of patients presenting with CS was 63.0 years, and 76% were male. The main CS etiology was ST-segment elevation myocardial infarction (STEMI), followed by non-ST-segment elevation myocardial infarction (NSTEMI) and myocarditis. Over 30% of patients had a history of prior MI, and over 20%, had a history of prior PCI. Median left ventricular ejection fraction (LVEF) was 22.5%, and the median EuroScore II value was 21.8.
Coronary angiography was performed in 90.9% of patients (Table 2). The majority of patients presented with MVD, either with or without LM coronary artery stenosis. The median SYNTAX Score II was 38.5. In terms of procedural characteristics, emergent PCI was done in 83.6% of patients, including the LM coronary artery PCI in 47.3%. All lesions were successfully treated in 63.6%.
All patients were treated with Impella Cardiac Power (CP), except for one case of Impella 5.0 use (Table 2). Impella was inserted before PCI in 52.7% of patients, during PCI in 27.3% of patients and after PCI in 14.5%. It was explanted in the catheterization laboratory in 14.5% of patients. The median insertion time was 20 minutes, and the median duration of support was 45 hours. The most common vascular access sites for Impella were the right and left femoral arteries (Table 2). Single access for simultaneous mechanical support and PCI was used in fewer than 10% of patients.
Regarding other measures of cardiopulmonary support, nearly all patients received catecholamines, most required mechanical ventilation, nearly 30% received levosimendan, 25.5% received an IABP (11 before and 3 after Impella insertion), and 12.7% received extracorporeal membrane oxygenation (ECMO; 3 patients before and 4 after Impella insertion).
The Kaplan-Meier curve showing 12-month survival in the IMPELLA-PL registry is in Figure 3. In-hospital and 12-month outcomes are presented in Table 3. The in-hospital mortality rate was 76.4% (42 patients), and the total 12-month mortality rate was 80.0% (44 patients). Five patients (9.1%) experienced 12 MACCEs during the 12-month follow-up period, including 2 post-discharge deaths, 3 readmissions for HF, 1 MI, 1 stroke, 1 LVAD implantation, and 1 heart transplantation.
Acute kidney injury occurred in over 60% of patients, and 32.7% of them required dialysis. One in three patients experienced bleeding complications according BARC criteria. Device-related complications, including any access site bleeding, limb ischemia, and hemolysis, occurred in 30.9%, 12.7%, and 10.9% of patients, respectively.
Impella to protect HR-PCI
The median age of patients in the HR-PCI group was 70.0 years, and 87.4% were male (Table 1). 53.4% of patients underwent HR-PCI in the setting of chronic coronary syndrome and the remaining in the setting of acute coronary syndrome, mostly NSTEMI. More than 50% of patients had a history of prior MI, nearly 40% a history of previous PCI, and over 10% a history of previous coronary artery bypass grafting. Median LVEF was 26.0%, and the median EuroScore II value was 5.1.
In terms of angiographic characteristics (Table 2), over 60% of patients presented with MVD including the LM, followed by MVD except for the LM. Severe calcifications and chronic total occlusions were present in 55.3% and 54.2%, respectively. The median SYNTAX Score II was 43. PCI was performed in nearly all patients, including the LM coronary artery in nearly 69.2% and the left anterior descending artery in nearly 78.3%. All lesions were successfully treated in over 83%. PCI was performed via the Impella sheath in about 17.8% of patients.
All patients were treated with Impella CP (Table 2). Impella was inserted before PCI in 81.8%. It was removed directly after PCI in 93.7%. The median insertion time was 25.0 minutes, and the median duration of support was 3.0 hours. The most common vascular access for Impella was the right or left femoral artery (54.5% and 39.9% of patients, respectively). Alternative access was used in 14 patients (approximately 5%). Single access for simultaneous mechanical support and PCI was used in 17.8%.
Other cardiopulmonary support (Table 2) included catecholamines (18.6%), levosimendan (5.1%), mechanical ventilation (4.0%), ECMO (2.4%; 1 before and 5 after Impella insertion), and an IABP (2.0%; 4 before and 1 af- ter Impella insertion).
The in-hospital mortality rate was 8.3% (21 patients), and the total 12-month mortality rate was 18.2% (46 patients, Figure 3, Table 3). In a group of patients who were discharged from the hospital, 57 experienced 69 MACCEs during the 12-month follow-up, including 25 post-discharge deaths, 25 readmissions for HF, 3 MI, 8 repeated revascularizations, 4 strokes, 1 LVAD implantation, and 3 heart transplantations.
AKI occurred in 12.6% of patients, and about 1% of them required dialysis (Table 3). Severe bleeding complications according to the BARC definition were reported in 16 patients (6.3%). The rate of device-related complications including any access site bleeding, limb ischemia, hemolysis, and aortic injury was 14.6%, 2.4%, 1.6%, and 0.4%, respectively.
Discussion
The main findings of the IMPELLA-PL registry are that (1) the use of Impella devices for CS and HR-PCI has greatly increased since their introduction in Poland, with HR-PCI being the predominant indication, with more than 80% of patients receiving Impella with nearly exclusive use of Impella CP; (2) the baseline risk profile of CS patients was substantially higher than in other registries and associated with high mortality and complication rates; (3) the risk profile of HR-PCI patients, their mortality and complications rates were consistent with other registries.
IMPELLA-PL differs from other registries in terms of MCS indications and the Impella model used. First, regarding the indications, HR-PCI constituted over 80% of patients treated with Impella in Poland, whereas over 50% of patients in the Italian registry received Impella due to CS [9]. Other registries published the results for CS and HR-PCI patients separately and from different periods, precluding direct comparisons [9, 11, 12]. Second, in our registry, all CS and HR-PCI patients were treated with Impella CP, except for one CS patient, in whom Impella 5.0 was surgically implanted. In 2008, Impella 2.5 became the first approved Impella model and was the most used device in other registries (60%–96%), although Impella CP was rapidly adopted after its introduction in 2012 [9–12]. The 14 F Impella CP, with an average maximum flow of 3.7 l/min and a peak flow of 4.3 l/min, is designed to offer a higher level of support compared to the 12 F Impella 2.5 [14]. Although there have been no prospective studies comparing both pump models in terms of efficacy and safety, improved prognosis has been reported following the switch from Impella 2.5 to Impella CP in individual patients [15]. Nevertheless, the crude rates of all-cause mortality did not differ according to the type of Impella device used [10]. Still, one should remain cautious when comparing results of different retrospective registries with different endpoint definitions, and prospective studies are needed to research further the development of Impella technology.
Impella in cardiogenic shock
The baseline risk profile of CS patients in our registry was extremely high, with 70% of patients presenting with STEMI, 70% with severe three-vessel disease with or without concomitant left main disease, close to 50% with cardiac arrest before admission; all received catecholamines, 80% required mechanical ventilation, 25% a concomitant IABP and over 10% concomitant ECMO. In other registries, the rate of patients with prior cardiac arrest was lower (23%–24%) [10, 16] and initial ejection fraction was higher [10, 16, 17], suggesting that the baseline risk profile of CS patients in our registry was higher than in other registries. Consequently, the mortality and complication rates were also higher, with AKI, bleeding, and inflammatory complications being the most frequent (Table 4).
Cardiogenic shock |
||||
IMPELLA-PL n = 55 |
IMP-IT n = 229 |
J-PVAD n = 819 |
cVAD n = 154 |
|
Hemolysis, % |
10.9 |
20.5 |
11.2 |
10.3 |
AKI, % |
61.8 |
50.5 |
– |
18.1 |
Bleeding, % |
45.5a |
15.7 |
6.1 |
20.1 |
Inflammatory, % |
40.0 |
30.5 |
– |
12.9 |
Neurological, % |
1.8 |
6.6 |
1.6 |
1.9 |
HR-PCI |
||||
IMPELLA-PL n = 253 |
IMP-IT n = 177 |
German Registry n = 154 |
cVAD n = 637 |
|
Hemolysis, % |
1.6 |
0.5 |
– |
0.2 |
AKI, % |
12.6 |
13.0 |
– |
5.8 |
Bleeding, % |
13.4a |
5.1 |
4.5 |
11.0 |
Inflammatory, % |
13.8 |
4.1 |
– |
– |
Neurological, % |
1.6 |
2.0 |
0.0 |
0.0 |
Currently, MCS has a class IIa recommendation in the recent ESC guidelines for the treatment of cardiogenic shock, with no preference towards a specific MCS type [4]. Initially, it was suggested that Impella may have an advantage over IABPs in patients with MI complicated by CS [18]. Data from systematic reviews and registry-based analyses questioned these assumptions, suggesting no mortality benefit and even adverse effects in patients treated with Impella compared to IABPs [19, 20]. However, the randomized controlled studies included in these meta-analyses had variable definitions of cardiogenic shock, slow enrollment rates, high crossover between the randomization arms, and variable time of Impella treatment initiation. For example, recent analyses showed that the timing of Impella insertion is a key to clinical success, with pre-PCI Impella insertion associated with a substantial survival benefit, compared to insertion during or after PCI [21, 22], especially in women [23]. In our registry, the baseline risk was very high, so Impella was inserted before PCI in about 50% of patients and mostly used to escalate IABP or ECMO therapy, which explains the very unfavorable outcomes [10]. Due to the retrospective design, we did not have complete clinical variables to establish the Society of Cardiovascular Angiography & Interventions (SCAI) Shock Classification. We believe that one of the reasons for high mortality in CS patients was implementation of Impella therapy far too late (as indicated by the median lactate of 7.4 mmol/l) potentially due to initial reimbursement problems with Impella in Poland. We are planning to complete the missing clinical variables and perform a separate analysis in CS patients to better understand the potential reasons for such high mortality. Altogether, further studies are required for heart teams to navigate toward the optimal patient selection and timing of MCS initiation and answer the question of whether the survival benefit of Impella therapy in CS outweighs the risk of complications, compared with the standard of care.
Impella to protect HR-PCI
The prospective multicenter PROTECT I trial (n = 20) demonstrated that Impella 2.5 can be successfully used during HR-PCI [24]. In the intention-to-treat analysis of the randomized controlled PROTECT II trial, patients supported with Impella 2.5 (n = 226) had numerically improved outcomes at 90 days compared to the IABP (n = 226) (P = 0.147). In the per-protocol analysis, Impella was associated with fewer MACCEs than the IABP (P = 0.048) [25]. Subsequently, analysis of the prospective single-arm PROTECT III trial including HR-PCI patients supported with Impella 2.5 and Impella CP (n = 504) demonstrated more complete revascularization, lower bleeding rate, and improved 90-day clinical outcomes compared to the historic cohort of PROTECT II patients with mean LVEF of 23% [26]. The use of Impella was associated with over 75% lower risk of post-PCI AKI than expected in the current risk models, and lower risk of AKI than the use of veno-arterial ECMO, suggesting that Impella insertion might be a new protective strategy against AKI during HR-PCI [27, 28]. However, a retrospective study including 1680 patients found that HR-PCI was successfully performed in over 98% of patients without MCS support, with a mortality rate of only 1.6% 30 days post-procedure [29]. However, detailed data on the completeness of revascularization as well as long-term outcomes were not provided. In addition, a recent single-center analysis of patients undergoing complex high-risk PCIs performed with either an IABP or Impella showed similar outcomes in terms of MACCE and mortality rates for both devices [30]. Altogether, the optimal selection of patients who truly require MCS during HR-PCI and the selection of the most suitable device remains to be further investigated.
Limitations
Our study has several limitations. First, since this was a registry-based study, it was limited by the completeness of the available medical records and the lack of an independent event adjudication committee. Thus, both baseline characteristics and data on endpoints might be prone to under or overreporting bias despite prespecified definitions. Second, there was no control group of patients treated with IABPs, ECMO, or no MCS, precluding any comparison between Impella and other MCS types. Third, due to the adoption of Impella mostly in the HR-PCI patients in Poland, the absolute number of CS patients included in the registry was low (55 patients over 8 years, ~7 patients per year in the whole country), making the statistical power of the CS subgroup analysis low and not reflecting contemporary medical practice. Altogether, given the observational, retrospective study design, our findings are hypothesis-generating and should be interpreted with caution.
Conclusions
The use of Impella in CS was low, compared with the use of Impella in HR-PCI, with almost exclusive use of Impella CP. The risk profile and mortality in CS patients were higher than in other registries, and the potential benefits of Impella in CS remain to be further investigated. In contrast, Impella seems safe and effective during HR-PCI, in accordance with the results from previous registries.
Article information
Acknowledgments: JetPub Scientific Communications, LLC, supported by funding from Abiomed Europe GmbH, provided editorial assistance to the authors during preparation of this manuscript.
Conflict of interest: AP and AG received a research grant from Abiomed (grant no. 69829335). JS is an Impella proctor. TP received speaking honoraria from Abiomed and PulseCath. All other authors have nothing to disclose.
Funding: None.
Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, which allows downloading and sharing articles with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.