Broken intraocular lens haptic during cataract surgery

Broken intraocular lens haptic during cataract surgery

Panagiota Mylova, Ioannis A. Mallias, Aikaterini Mouzaka, Anastasia Tassiopoulou

Laser Plus Eye, Nea Smyrni, Athens, Greece

ABSTRACT

This is a case report describing a planned routine cataract surgery that was complicated intraoperatively by breakage of the trailing haptic of the intraocular lens. The possibility of lens implantation despite the broken haptic is discussed, and the patient’s postoperative course one month after the surgery is described, in this report.

KEY WORDS: cataract surgery, intraocular lens

Ophthalmol J 2017; Vol. 2, No. 1, 33–34

INTRODUCTION

We describe a planned routine cataract surgery that was complicated intraoperatively by breakage of the trailing haptic of the intraocular lens. The possibility of lens implantation despite the broken haptic is discussed, and the patient’s postoperative course one month after the surgery.

CASE REPORT

An 83-year-old man with age-related cataract was admitted for routine cataract surgery of the right eye. Best corrected visual acuity (BCVA) of the right eye was 20/80, and BCVA of the left eye was 20/30. The surgery was performed by Dr. Mallias under topical anaesthesia. At the end of the uneventful procedure an attempt was made to implant an Envista (Bausch & Lomb, USA) Intraocular lens (IOL) through a 2.2 mm incision. The Envista is a hydrophobic acrylic IOL. During the implantation, the trailing haptic was stuck between the cartridge and the plunger of the IOL injector. The surgeon tried to free the trailing haptic with a Sinsky hook, and during manipulation the trailing haptic of the IOL was broken (Fig. 1).

Broken intraocular lens haptic during cataract surgery

FIGURE 1. Broken intraocular lens haptic

The surgeon decided to implant the IOL with the broken haptic in the capsular bag. Despite the fact that the haptic was broken, no IOL dislocation was noticed postoperatively. The capsulorrhexis was about 5 mm. The incision was made temporally and the IOL was implanted in three to nine o’clock orientation.

Patient’s BCVA was 20/25 on the first postoperative day, and the IOL was well centred in the capsular bag (Fig. 2).

Broken intraocular lens haptic during cataract surgery

FIGURE 2. Well centred intraocular lens

The patient was examined again one month postoperatively. Slit lamp examination with pharmacologic pupil dilation was performed, which showed that the IOL was still well centred inside the capsular bag (Fig. 3).

Broken intraocular lens haptic during cataract surgery

FIGURE 3. Well centred intraocular lens, one month after surgery

DISCUSSION

There are some cases [1] of fractured IOL during surgery, but to our knowledge there is none reporting IOL placement inside the capsular bag. In all papers that we found in the literature, the surgeon removed the IOL when one haptic was broken. In cases where the IOL has to be removed, the surgeon has to either fold [2] the IOL or chop [3] it in half and remove the two separate pieces with forceps. The purpose of this case report is to mention the fact that the IOL can remain well centred inside the capsular bag, even in cases where the one haptic is missing. Therefore, there is no need to replace the IOL.

The IOL haptic was stuck between the cartridge and the plunger of the injector probably due to misplacement of the IOL in the cartridge from the beginning, and not because of material malfunction [4]. At this point we would like to stress the fact that the viscoelastic was completely removed from the capsular bag at the end of the procedure, and this helped the anterior and posterior capsule to come in contact and help hold the IOL in place.

There are also cases in which the haptics spontaneously break inside the capsular bag, causing complications [5] that reduce the vision. Hopefully in this case, the one remaining haptic seems more than enough to maintain the IOL well centred inside the capsular bag. In other studies [6, 7], the IOL needs to be removed due to dislocation that consequently reduces the patient’s vision. What should be taken into consideration in spontaneous haptic fractures is the further research about the material biocompatibility [5].

REFERENCES

  1. Kirkpatrick JN, Cook SD. Broken intraocular lens during cataract surgery. Br J Ophthalmol. 1992; 76(8): 509.

  2. Lee MH, Webster DL. Intraocular lens exchange-removing the optic intact. Int J Ophthalmol. 2016; 9(6): 925–928, doi: 10.18240/ijo.2016.06.23, indexed in Pubmed: 27366699.

  3. Hayashi K, Ogawa S, Manabe SI, et al. A classification system of intraocular lens dislocation sites under operating microscopy, and the surgical techniques and outcomes of exchange surgery. Graefes Arch. Clin. Exp. Ophthalmol. 2016; 254(3): 505–513, doi: 10.1007/s00417-016-3273-6, indexed in Pubmed: 26816295.

  4. Kang H, Han KE, Kim Ti, et al. Spontaneous fracture of an implanted posterior chamber polyimide intraocular lens haptic: a case report. Indian J Ophthalmol. 2014; 62(3): 348–350, doi: 10.4103/0301-4738.111195, indexed in Pubmed: 23619497.

  5. Stallings S, Werner L, Chayet A, et al. Intraocular polyimide intraocular lens haptic breakage long-term postoperatively. J Cataract Refract Surg. 2014; 40(2): 323–326, doi: 10.1016/j.jcrs.2013.11.024, indexed in Pubmed: 24461504.

  6. Caça I, Unlü K, Ari S, et al. Spontaneous fracture of an implanted posterior chamber intraocular lens. Eur J Ophthalmol. 2005; 15(4): 507–509, indexed in Pubmed: 16001387.

  7. Alasbali T. Risk of Breaking Haptic of Trifocal IOL and Its Management. Middle East Afr J Ophthalmol. 2016; 23(4): 321–322, doi: 10.4103/0974-9233.194090, indexed in Pubmed: 27994398.

CORRESPONDING AUTHOR:

Ioannis A. Mallias, MD, PhD, Mediterraneo Hospital, Glyfada, 3 Kallipoleos Street, Nea Smirni, 17123, Athens, Greece, tel.: 00302109320215, e-mail: johnmallias@yahoo.com

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

Publisher: VM Media Group sp. z o.o., Grupa Via Medica, 73 Świętokrzyska St., 80–180 Gdańsk

tel.:+48 58 310 94 94, faks:+48 58 320 94 60, e-mail: viamedica@viamedica.pl