Preliminary assessment of interand intraobserver reproducibility, and normative values of renal mean transit time (MTT) and parenchymal transit time (PTT) for 99mTc-etylenodicysteine

Paweł Cichocki, Marian Surma, Wojciech Woźnicki, Małgorzata Bieńkiewicz, Anna Płachcińska, Jacek Kuśmierek
DOI: 10.5603/NMR.2015.0007
·
Pubmed: 25633514
·
Nucl. Med. Rev 2015;18(1):29-34.

open access

Vol 18, No 1 (2015)
Original articles
Submitted: 2014-10-31
Accepted: 2014-12-19
Published online: 2015-01-29

Abstract

BACKGROUND: The clinical significance of MTT and PTT, determined by deconvolution of renographic curves, is arguable. Their usefulness in diagnosis of obstructive uro- and nephropathy, renovascular hypertension and monitoring of transplanted kidneys is pointed out, but susceptibility of deconvolution methods to errors resulting from “statistical noise” is also stressed. So far there are no reports on normative MTT values for 99mTc-EC, although such values were already determined for 131I-OIH, 99mTc-DTPA and 99mTc-MAG3. The aim of this study is an assessment of inter- and intraobserver reproducibility of MTT and PTT for 99mTc-EC, and determination of normative values for these parameters.

MATERIALS AND METHODS: 31 patients (17 women and 14 men aged 19–75, average 44 years) referred for dynamic renal scintigraphy with: unilateral flow impairment (11), unilateral nephrolithiasis (2), control after unilateral lithotripsy (4), moderate hypertension (demographically with > 99% probability of primary hypertension) (4), suspected cirrhosis of one kidney (3), future kidney donors (3), control after abdominal injuries (3), incontinence (1). 42 functionally efficient kidneys were included in the study. Criteria for recognition of a kidney as functionally efficient were: — no earlier history of renal disease, signs of renal damage in basic blood and urine tests, or abnormalities in ultrasonography; — normal result of dynamic renal scintigraphy (in terms of sequential images and renographic curve). MTT and PTT values were determined independently by two operators, using a matrix method for deconvolution of renographic curves.

RESULTS: Differences between mean MTT and PTT from two studies by one operator were insignificant and those values were closely correlated (r = 0.99 and r = 0.97, respectively). Differences of values obtained by both operators were practically insignificant for MTT (r = 0.93), and significant for PTT (r = 0.81 and p < 0.001). These differences do not disqualify that processing method. The upper limits of normative values of MTT and PTT were based on the results from first study performed by more experienced operator — 200 s and 170 s, respectively.

CONCLUSIONS: The procedure of processing dynamic renal scintigraphy used in this study is reproducible. Normative values of MTT and PTT for 99mTc-EC were established as 200 s and 170 s, respectively. An attempt to optimize and standardize the technique of determining parenchymal ROI in a matrix deconvolution method, followed by an evaluation of clinical usefulness of these parameters in the diagnosis of chosen renal function impairments would be a logical continuation of this initial research.

Abstract

BACKGROUND: The clinical significance of MTT and PTT, determined by deconvolution of renographic curves, is arguable. Their usefulness in diagnosis of obstructive uro- and nephropathy, renovascular hypertension and monitoring of transplanted kidneys is pointed out, but susceptibility of deconvolution methods to errors resulting from “statistical noise” is also stressed. So far there are no reports on normative MTT values for 99mTc-EC, although such values were already determined for 131I-OIH, 99mTc-DTPA and 99mTc-MAG3. The aim of this study is an assessment of inter- and intraobserver reproducibility of MTT and PTT for 99mTc-EC, and determination of normative values for these parameters.

MATERIALS AND METHODS: 31 patients (17 women and 14 men aged 19–75, average 44 years) referred for dynamic renal scintigraphy with: unilateral flow impairment (11), unilateral nephrolithiasis (2), control after unilateral lithotripsy (4), moderate hypertension (demographically with > 99% probability of primary hypertension) (4), suspected cirrhosis of one kidney (3), future kidney donors (3), control after abdominal injuries (3), incontinence (1). 42 functionally efficient kidneys were included in the study. Criteria for recognition of a kidney as functionally efficient were: — no earlier history of renal disease, signs of renal damage in basic blood and urine tests, or abnormalities in ultrasonography; — normal result of dynamic renal scintigraphy (in terms of sequential images and renographic curve). MTT and PTT values were determined independently by two operators, using a matrix method for deconvolution of renographic curves.

RESULTS: Differences between mean MTT and PTT from two studies by one operator were insignificant and those values were closely correlated (r = 0.99 and r = 0.97, respectively). Differences of values obtained by both operators were practically insignificant for MTT (r = 0.93), and significant for PTT (r = 0.81 and p < 0.001). These differences do not disqualify that processing method. The upper limits of normative values of MTT and PTT were based on the results from first study performed by more experienced operator — 200 s and 170 s, respectively.

CONCLUSIONS: The procedure of processing dynamic renal scintigraphy used in this study is reproducible. Normative values of MTT and PTT for 99mTc-EC were established as 200 s and 170 s, respectively. An attempt to optimize and standardize the technique of determining parenchymal ROI in a matrix deconvolution method, followed by an evaluation of clinical usefulness of these parameters in the diagnosis of chosen renal function impairments would be a logical continuation of this initial research.

Get Citation

Keywords

adult, humans, radioisotope renography, reproducibility of results, radiopharmaceuticals, technetium Tc 99m-ethylenedicysteine

About this article
Title

Preliminary assessment of interand intraobserver reproducibility, and normative values of renal mean transit time (MTT) and parenchymal transit time (PTT) for 99mTc-etylenodicysteine

Journal

Nuclear Medicine Review

Issue

Vol 18, No 1 (2015)

Article type

Research paper

Pages

29-34

Published online

2015-01-29

Page views

1631

Article views/downloads

2256

DOI

10.5603/NMR.2015.0007

Pubmed

25633514

Bibliographic record

Nucl. Med. Rev 2015;18(1):29-34.

Keywords

adult
humans
radioisotope renography
reproducibility of results
radiopharmaceuticals
technetium Tc 99m-ethylenedicysteine

Authors

Paweł Cichocki
Marian Surma
Wojciech Woźnicki
Małgorzata Bieńkiewicz
Anna Płachcińska
Jacek Kuśmierek

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., Świętokrzyska 73 street, 80–180 Gdańsk, Poland

phone: +48 58 320 94 94, fax: +48 58 320 94 60, e-mail: viamedica@viamedica.pl