Nuclear Medicine Review 1/2015-Preliminary assessment of inter- and intraobserver reproducibility, and normative values of renal mean transit time (MTT) and parenchymal transit time (PTT) for 99mTc-etylenodicysteine

Original

Preliminary assessment of inter- and intraobserver reproducibility, and normative values of renal mean transit time (MTT) and parenchymal transit time (PTT) for 99mTc-etylenodicysteine

Paweł Cichocki1, Marian Surma1, Wojciech Woźnicki1, Małgorzata Bieńkiewicz2, Anna Płachcińska2, Jacek Kuśmierek1

1Department of Nuclear Medicine, Medical University of Łódź, Poland

2Department of Quality Control and Radiological Protection, Medical University of Łódź, Poland

[Received 31 X 2014; Accepted 07 I 2015]

Abstract

BACKGROUND: The clinical significance of MTT and PTT, determined by deconvolution of renographic curves, is arguable. Their usefulness in diagnosis of obstructive uro- and nephropathy, renovascular hypertension and monitoring of transplanted kidneys is pointed out, but susceptibility of deconvolution methods to errors resulting from “statistical noise” is also stressed. So far there are no reports on normative MTT values for 99mTc-EC, although such values were already determined for 131I-OIH, 99mTc-DTPA and 99mTc-MAG3. The aim of this study is an assessment of inter- and intraobserver reproducibility of MTT and PTT for 99mTc-EC, and determination of normative values for these parameters.

MATERIALS AND METHODS: 31 patients (17 women and 14 men aged 19-75, average 44 years) referred for dynamic renal scintigraphy with: unilateral flow impairment (11), unilateral nephrolithiasis (2), control after unilateral lithotripsy (4), moderate hypertension (demographically with > 99% probability of primary hypertension) (4), suspected cirrhosis of one kidney (3), future kidney donors (3), control after abdominal injuries (3), incontinence (1).

42 functionally efficient kidneys were included in the study. Criteria for recognition of a kidney as functionally efficient were:

  • no earlier history of renal disease, signs of renal damage in basic blood and urine tests, or abnormalities in ultrasonography;
  • normal result of dynamic renal scintigraphy (in terms of sequential images and renographic curve).

MTT and PTT values were determined independently by two operators, using a matrix method for deconvolution of renographic curves.

RESULTS: Differences between mean MTT and PTT from two studies by one operator were insignificant and those values were closely correlated (r = 0.99 and r = 0.97, respectively). Differences of values obtained by both operators were practically insignificant for MTT (r = 0.93), and significant for PTT (r = 0.81 and p < 0.001). These differences do not disqualify that processing method. The upper limits of normative values of MTT and PTT were based on the results from first study performed by more experienced operator – 200 s and 170 s, respectively.

CONCLUSIONS: The procedure of processing dynamic renal scintigraphy used in this study is reproducible. Normative values of MTT and PTT for 99mTc-EC were established as 200 s and 170 s, respectively. An attempt to optimize and standardize the technique of determining parenchymal ROI in a matrix deconvolution method, followed by an evaluation of clinical usefulness of these parameters in the diagnosis of chosen renal function impairments would be a logical continuation of this initial research.

KEY words: adult, humans, radioisotope renography, reproducibility of results, radiopharmaceuticals, technetium Tc 99m-ethylenedicysteine

Nuclear Med Rev 2015; 18, 1: 29-34

Background

Complexity of the renal function makes it difficult to cover its whole range with a single diagnostic procedure. Scintigraphic methods are limited to the assessment of the two arbitrary renal functions. Those are: “uptake function”, that is the ability of a kidney to absorb a specific radiopharmaceutical from the bloodstream and “transit function”, that is the ability to transport an absorbed tracer through the kidney parenchyma to the urinary tract. Dynamic renal scintigraphy, used for many years in the diagnosis of the condition of kidneys, allows the assessment of both of these functions using radiopharmaceuticals such as 131I-OIH, 99mTc-DTPA, 99mTc-MAG3 and the most recently introduced – 99mTc-EC.

Dynamic renal scintigraphy allows the evaluation of these functions based on a qualitative assessment of sequential scintigraphic images and the shape of the renographic curves. Additionally, basic quantitative parameters are also calculated – the relative participation of each kidney in their overall function (split function) and times: of achieving maximum activity of the radiopharmaceutical in the kidney – T and decrease of that activity to half of the maximum – T1/2. Some other quantitative parameters are also used, for example: ratio of activity in the kidney in the 20th minute to maximum activity (R20/max), or to activity in the 3rd minute (R20/3), or the ratio of the maximum activity to activity in the 890th second (ER – excretion ratio) [1-3]. For a more thorough analysis of the function of each kidney and urinary tract, parametric images of renal clearance function [4-6], ureteral peristalsis [7] and others [8-10] are also generated. Moreover, after processing of the scintigraphic data, additional parameters concerning the transport of the radiopharmaceutical through the entire kidney – MTT (Mean Transit Time) and its parenchyma – PTT (Parenchymal Transit Time), can be obtained.

MTT and PTT can have an important role in the diagnosis of obstructive uro- and nephropathy [11, 12], renovascular hypertension [12-14] and in monitoring the function of the transplanted kidney [15, 16]. The most common methods used to determine the values of these parameters are based on the deconvolution of renographic curves and blood activity curve. Some authors, however, point out an uncertainty of the time parameters calculated by deconvolution, caused by a susceptibility of the impulse response function to the noise from the stochastic nature of radioactive decay and radiation detection [17, 18]. For that reason, Rutland et al. [19] and Samal et al. [20] presented methods for determining MTT and PTT not relying on deconvolution of renographic curves. According to Fleming’s comparative analysis of these parameters determined by means of deconvolution and a method based on Rutland’s theory their similar usefulness for the calculation of renal transit time parameters has been shown [21]. On the other hand, the method proposed by Samal et al. [20] based on factor analysis, requires the acquisition of sequential images in short time intervals, which in turn requires administering the patient higher activity of the radiopharmaceutical. Therefore, MTT and PTT are most frequently measured with deconvolution methods based on the matrix algorithm or Fourier transformation [21-23].

It is expected that MTT and PTT determination techniques, as well as other research methods, meet the requirements of inter- and intraobserver reproducibility. Therefore, in this study, these values for the procedure based on the matrix deconvolution algorithm were evaluated.

The upper limit of normative values of MTT and PTT is also considerably significant (lower limit is not taken into account, since shorter transit time through kidneys is not important from a clinical point of view). Such values were determined for 131I-OIH [24], 99mTc-DTPA [25-27] and for 99mTc-MAG3 [3, 28], both based on groups of healthy volunteers as well as patients with kidneys considered as functionally efficient. So far there are no reliable normative values of MTT and PTT for 99mTc-EC. Gupta et al. [29] only determined the value of PTT, but just on a group of five healthy volunteers.

Among nephrothropic radiopharmaceuticals, 99mTc-EC has favourable pharmacological properties (a high extraction rate), which make it especially useful and applicable for dynamic renal scintigraphy [30-32]. However, different renal extraction mechanism of 99mTc-EC than of 99mTc-DTPA and 99mTc-MAG3, and different proportions of its filtration and secretion than of 123/131I-OIH, make it necessary to determine separate normative values of MTT and PTT for these radiopharmaceuticals.

The purpose of this study is to evaluate the inter- and intraobserver reproducibility of the procedure, and determine normative values of MTT and PTT for 99mTc-EC.

Materials and methods

Dynamic renal scintigraphies performed at our department in years 2011-2013 were analysed retrospectively. Kidneys were considered functionally efficient based on the following criteria:

  • patients had routine urinalysis within normal limits and blood levels of urea and creatinine within normal ranges according to the reference values provided by the laboratory;
  • they had no history of any diseases of the urinary tract or the selected kidney;
  • in ultrasound, selected kidneys were typically located and had normal shapes and sizes. There were no signs of dilated pelvis or calices, cysts, cortical defects, or other morphological abnormalities;
  • the result of dynamic renal scintigraphy of the selected kidney in terms of sequential images and renographic curve, assessed independently by two physicians, was normal (that includes the split function of the kidney > 45%, Tmax < 300 s and T1/2 < 600 s).

Based on the above criteria, 42 kidneys regarded as functionally efficient were selected in 31 patients – 17 women and 14 men aged 19-75 years (44 years on average).

Those patients were referred for dynamic renal scintigraphy with:

  • suspected unilateral outflow impairment (all confirmed impairments were minor and did not require a diuretic test – 11 patients) (only contralateral kidney was selected);
  • suspected unilateral nephrolithiasis (2 patients) (only contralateral kidney was selected);
  • control after unilateral lithotripsy (no sooner than 1 year after therapy – 4 patients) (only contralateral kidney was selected);
  • moderate, freshly diagnosed hypertension, demographically with > 99% probability of primary hypertension (study without captopril – 4 patients) (both kidneys were selected);
  • suspected cirrhosis of one of the kidneys (excluded after renal scintigraphy – 3 patients) (only contralateral kidney was selected);
  • qualification of future kidney donors (3 patients) (both kidneys were selected);
  • control after abdominal injuries (any damage to the kidneys was excluded by morphological and scintigraphic evaluation – 3 patients) (both kidneys were selected);
  • incontinence (1 patient) (both kidneys were selected).

In all patients, dynamic renal scintigraphy was carried out according to the standard protocol without a diuretic test. Each patient approx. 1/2 h before the test was recommended to drink 500 ml of water and empty the bladder immediately before testing. The study was performed on cameras: Mediso Nucline AP Infinia Hawkeye 2 and Infinia Hawkeye 4, equipped with low-energy, general purpose collimators (LEGP). During the test, a patient was placed in a supine position. Gamma camera detector was positioned under the patient with the field of view including both the kidneys and the heart. The study was started at the time of intravenous administration of 111 MBq (3 mCi) of 99mTc-EC, collecting 60 twenty-second images in matrix 64 x 64 during 20 minutes.

Scintigraphic data processing was performed by drawing regions of interest (ROI) including the heart, the entire kidney, the renal parenchyma (possibly excluding the pelvis and calices) and area between the kidneys (blood background). The ROI for the whole kidney was determined by isocontour at 20% of the maximum counts on the summed images recorded between 2nd and 3rd minute of the study. Other ROIs were drawn manually: for heart – on the first registered image, for renal parenchyma – on summed images from between 14th and 15th minute of the study (avoiding calices and pelvis visible in these images). Then, time-activity curves were drawn for each ROI. Curves were smoothed with a three-point filter with 1-2-1 weight coefficients. Blood background curve was subtracted from the curves for the whole kidney and renal parenchyma (after correction due to ROI size). From these curves, MTT and PTT were determined using the matrix algorithm deconvolution.

In order to assess intra- and interobserver reproducibility of determined values of MTT and PTT, scintigraphic data were processed independently by two operators:

  • more experienced (A) – twice, with time interval of about two weeks, resulting in two pairs of MTT and PTT for each kidney (MTTA1 and PTTA1, and MTTA2 and PTTA2);
  • less experienced (B) – once (MTTB and PTTB).

To establish normative values of determined parameters, the value obtained by the operator A in the first study was taken.

Statistical analysis: Assessments of the intra- and interobserver reproducibility were based on a comparison of the mean values of the studied variables (MTT and PTT) obtained by the operator A in the two studies (MTTA1 with MTTA2 and PTTA1 with PTTA2) – for in traobserver reproducibility, or obtained from the first study by the operator A and the study by the operator B (MTTA1 with MTTB and PTTA1 with PTTB) – for interobserver reproducibility. For this purpose Student’s t-test for pairs was applied (normality of distributions was verified with Shapiro-Wilk’s test). In addition to statistical significance, practical significance of observed differences was also assessed, using Cohen’s d-coefficient, which is a measure of the effect size, that is – the difference between two mean values of MTT or PTT from different studies divided by a standard deviation for the data. {Formula: d = (M1 – M2)/SDpooled, where M1 is the mean value of MTTA1 or PTTA1, M2 is the mean value of MTTA2 or PTTA2 for intraobserver, and MTTB or PTTB for interobserver reproducibility, respectively; while SDpooled is the standard deviation for the samples (SDpooled = √[(SD12+sD22)/2])}.

It is assumed that d > 0.8 indicates significant differences, 0.5-0.8 – medium, 0.2-0.5 – small, and < 0.2 means that the differences are practically trivial. As an additional measurement of assessment reliability, linear correlation coefficients were applied, similarly for the results obtained in the two studies of the same operator in case of intraobserver reproducibility, and the results obtained by two independent operators for interobserver reproducibility.

Statistical analysis was performed with Statistica 10.0 software.

Results

Basic parameters of distributions of MTT and PTT values obtained by each of the operators are summarized in Table 1. There were no statistically significant differences between mean values of MTT and PTT from two studies by operator A (respectively p = 0.87 and p = 0.36). The mean difference between the results did not exceed 1s. Cohen’s d-coefficients for MTT and PTT were d = 0.03 and d = 0.05, respectively.

Table 1. Values of basic parameters describing distribution of results acquired by both operators

Study Min. Max. Mean St. dev. Mean + 2sd Compared pairs of results Statistics
p d
MTTA1 Ill 214 149 26 201
MTTA2 105 218 148 26 200 MTTA1 vs MTTA2 0.87 0.03
MTTB 97 208 145 25 195 MTTAl vs MTTB 0.02 0.16
PTTA1 80 l7l 128 2l 169 PTTA1 vs PTTA2 0.36 0.05
PTTA2 8l 176 129 2l 170 PTTA1 vs PTTB < 0.001 0.77
PTTb 86 150 ll4 15 145

On the other hand, MTT and PTT obtained by operators A and B differed significantly (p = 0.02 and p < 0.001, respectively). The mean difference between the results was 4s for MTT and 14s for PTT. Cohens d-coefficients for MTT and PTT amounted to d = 0.16 and d = 0.77, respectively.

Linear correlation coefficients between MTT and PTT values from two studies by operator A were r = 0.99 (Figure 1A) and r = 0.97 (Figure 1B), respectively, while for the pairs of values for MTT and PTT obtained by two operators – r = 0.93 (Figure 1C) and r = 0.81 (Figure 1D), respectively.

Figure 1. Correlations between values obtained by the same operator in two studies for (A) MTT and (B) PTT – intraobserver reproducibility; and correlations between values obtained by both operators; (C) MTT and (D) PTT – interobserver reproducibility

Normative values for MTT and PTT for normal kidneys were determined based on the first study performed by operator A. Their upper limits were determined as mean values plus two standard deviations (rounded), that is for MTT (149 + 2 x 26) s ≈ 200 s, and for PTT (128 + 2 x 21) s ≈ 170 s.

Discussion

Lack of significant differences between the mean values of MTT and PTT, and their close correlation in studies performed by the same operator (Figure 1A and 1B) give grounds to consider the processing procedure of dynamic renal scintigraphy used in this work to have high intraobserver reproducibility.

In case of studies performed by two operators, there were statistically significant differences between the mean values of the results. However, for MTT, these differences had no practical significance (d < 0.2). Therefore, our method can be considered to have high interobserver reproducibility in this field, which is confirmed by a close correlation between pairs of MTT values obtained by both operators (Figure 1C).

The results were different in case of PTT values obtained by both operators. Their mean values show statistically significant differences with medium size of effect (d = 0.77) and the correlation results were not as strict as in the case of MTT, although they were still high (r = 0.81, Figure 1D). This suggests that interobserver reproducibility is slightly lower in this field. These differences, however, do not disqualify the processing methods of dynamic renal scintigraphy.

High reproducibility of the results obtained by both operators for the MTT and the differences between the results obtained by these operators for PTT, can be explained by different methods of determining regions of interest for the whole kidney and its parenchyma (Figure 2). Whole kidney ROIs (for determining MTT) were drawn with isocontour with a constant value of 20% of the maximum number of counts in the kidney. This method of determining ROIs is strictly defined and reproducible. Only small, ischemic or cystic kidneys may cause some problems. On the other hand, parenchymal ROIs (for determining PTT) were drawn manually, so subjective factors could have a significant impact on their shape. Depending on the size of the selected area, it can for example contain smaller or larger share of calices, and that can change the value of determined parameter. Further efforts are necessary to minimize the impact of subjective elements in determination of parenchymal ROIs, by either standardization or automatization of the methods of selecting the ROI. Similar solution was also suggested by Bergmann et al. [33].

Figure 2A. ROI for MTT, selected with isocontour with constant value of 20% maximum counts; B. ROI for PTT, selected manually

Because of the fact that the results of MTT and PTT had high inter- and intraobserver reproducibility (even in case of PTT), they can be used as a basis for determ ining the upper lim its of normative values for these parameters for normally functioning kidneys. Taking into account the values obtained by a more experienced operator in the first study, the upper limit of normal range was assumed to be 200 s for MTT and 170 s for PTT

In Table 2, mean values of normal MTT and PTT obtained in our work were compared with those obtained by different authors for other radiopharmaceuticals in kidneys considered functionally efficient. It is apparent that mean values of MTT obtained in this work are lower than values obtained both by Piepsz et al. [25] and Rajabi et al. [17] for 99mTc-DTPA, as well as those calculated by Russell et al. [3] and Gonzales et al. [28] for 99mTc-MAG3. On the other hand, they are comparable with values reported by Kenny et al. [24] for 131I-OIH. The mean values of PTT are also lower than those found in literature for 99mTc-DTPA [17] and 99mTc-MAG3 [3, 28]. It is also notable that Gupta et al. [29] obtained surprisingly high normative values of PTT for 99mTc-EC (in 5 healthy volunteers determined PTT was from 125 s to 206 s, on average – 175 s). Differences between MTT and PTT obtained in this work for 99mTc-EC and values of these parameters for other radiopharmaceuticals may result both from methodological factors and different mechanisms of their extraction in nephrone. 99mTc-DTPA is a subject only to glomerular filtration. 99mTc-MAG3 is also excreted in only one mechanism – tubular secretion. 131I-OIH on the other hand, is a subject to both glomerular filtration and tubular secretion, just like 99mTc-EC, which can explain similar MTT values obtained for these radiopharmaceuticals.

Table 2. Mean normal values of MTT and PTT for OIH, DTPA and MAG3 reported by different authors

Authors Year Number of patients RF MTT ± SD PTT ± SD
This publication 2014 31 EC 149 ± 26 128 ± 21
Kenny et al. [24] 1975 19 OIH 134 ± 16
Diffy et al. [27] 1976 14 DTPA 180 ± 30
Piepsz et al. [25] 1982 33 DTPA 216 ± 64
Piepsz et al. [25] 1982 26 DTPA 201 ± 43
Rajabi et al. [17] 1998 37 DTPA 237 ± 50 199 ± 35
Gonzales et al. [28] 1994 25 MAG3 174 ± 27 148 ± 22
Russel et al. [3] 1995 14 MAG3 193 ± 58 157 ± 44

Conclusions

Obtained results justify the opinion that the procedure of processing dynamic renal scintigraphy studies used in this research to determine the values of MTT and PTT gives reproducible results. Normative values of transit times for 99mTc-EC through the whole kidney and its parenchyma were determined as 200 s and 170 s, respectively. An attempt to optimize and standardize the technique of determining parenchymal ROI in a matrix deconvolution method, followed by an evaluation of clinical usefulness of these parameters in the diagnostics of chosen renal function impairments would be a logical continuation of this initial research.

References

  1. Taylor A, Nally JV. Clinical applications of renal scintigraphy. Am J Roentgenol 1995; 164: 31-41.
  2. Kempi V, Sutton DG. Estimating the diagnostic yields resulting from renography and deconvolution parameters: a logistic regression analysis. J Nucl Med 1995; 36: 147-152.
  3. Russell CD, Japanwalla M, Khan S, Scott JW, Dubovsky EV. Techniques for measuring renal transit time. Eur J Nucl Med. 1995; 22: 1372-1378.
  4. Frieske I, Surma MJ, Pietrzak-Stelmasial E et al. Assessment of kidney parenchyma function by means of parametric imaging in patients after Extracorporeal Shock Wave Lithotripsy (ESWL). Eur J Nucl Med Mol Imaging 2008; 36: 367 (Abs).
  5. Frieske I, Pietrzak-Stelmasiak E, Bieńkiewicz M, Surma MJ, Kuśmierek J. Conventional and parametric kidney scintigrams – reproducibility of semiquantitative image evaluation. Nucl Med Rev Cent East Eur 2008; 11: 22-25.
  6. Frieske I, Surma MJ, Rogozińska-Zawiślak A et al. Parametric clearance kidney scintigrams; diagnostic potential in diabetes. Nucl Med Rev Cent East Eur 2007; 10: 16-20.
  7. Lepej J, Kliment J, Horák V, Buchanec J, Marosová A, Belákova S. A new approach in radionuclide imaging to ureteric peristalsis using 99Tcm-MAG 3 and condensed images. Nucl Med Commun 1991; 12: 397-407.
  8. Croft BY Functional Imaging. In: Esser PD (ed). Functional Mapping of Organ Systems and Other Computer Topics. Society of Nuclear Medicine, New York; 1981: 1-12.
  9. Szabo Z, Kutkuhn B, Georgescu G, Mecklenbeck W, Suatmadji A, Vosberg H. Parametrische Darstellung der Nierenfunktion mit 99mTc-Merkaptoaze- tyltriglyzin (MAG3). Nucl-Med 1989; 28: 73-83.
  10. Oppenheim BE, Appledorn CR. Parameters for functional renal imaging. In: Esser PD (ed). Functional Mapping of Organ Systems and Other Computer Topics. Society of Nuclear Medicine, New York 1981: 39-55.
  11. Britton K, Nimmon CC, Whitfield HN, Hendry WF, Wickham JEA. Obstructive nephropathy: successful evaluation with radionuclides. Lancet 1979; 313: 905-907.
  12. Durand E, Blaufox MD, Britton KE et al. International Scientific Committee of Radionuclides in Nephrourology (ISCORN) consensus on renal transit time measurements. Semin Nucl Med 2008; 38: 82-102.
  13. Gruenewald SM, Collins LT. Renovascular hypertension: quantitative renography as a screening test. Radiology 1983; 149: 287-291.
  14. Dondi M, Monetti N, Fanti S et al. Use of technetium-99m-MAG3 for renal scintigraphy after angiotensin-converting enzyme inhibition. J Nucl Med 1991; 32: 424-428.
  15. Bajén MT, Puchal R, González A et al. MAG3 renogram deconvolution in kidney transplantation: utility of the measurement of initial tracer uptake. J Nucl Med 1997; 38: 1295-1299.
  16. Nankivell BJ, Cohn DA, Spicer ST Evans SG, Chapman JR, Gruenewald SM. Diagnosis of kidney transplant obstruction using Mag3 diuretic renography. Clin Transplant. 2001; 15: 11-18.
  17. Rajabi H, Pant G. Renal transit times using a modified method of deconvolution. Iran J Nucl Med. 2003; 18: 37–3.
  18. Ham H. Is renography suitable for deconvolution analysis? J Nucl Med 1996; 37: 403-404.
  19. Rutland MD. Mean transit times without deconvolution. Nucl Med Commun 1981; 2: 337-344.
  20. Sámal M, Nimmon C. Relative renal uptake and transit time measurements using functional factor images and fuzzy regions of interest. Eur J Nucl Med 1998; 25: 48-54.
  21. Fleming JS, Kemp PM. A comparison of deconvolution and the Patlak-Rut- land plot in renography analysis. J Nucl Med 1999; 40: 1503-1507.
  22. Knesaurek K, Spaventi S. Comparison of three deconvolution techniques in renography. Eur J Nucl Med 1984; 9: 254-256.
  23. Durand E, Blaufox M, Britton K. Appendix of the ISCORN consensus on renal transit times measurements. Semin Nucl Med 2008; 1-14.
  24. Kenny RW, Ackery DM, Fleming JS, Goddard BA, Grant RW. Deconvolution analysis of the scintillation camera renogram. Br J Radiol 1975; 48: 481-486.
  25. Piepsz A, Ham HR, Erbsmann F, et al. A co-operative study on the clinical value of dynamic renal scanning with deconvolution analysis. Br J Radiol 1982; 55: 419-433.
  26. Rutland MD. A comprehensive analysis of renal DTPA studies. I. Theory and normal values. Nucl Med Commun 1985; 6: 11-20.
  27. Diffey BL, Hall FM, Corfield JR. The 99mTc-DTPA dynamic renal scan with deconvolution analysis. J Nucl Med 1976; 17: 352-355.
  28. Gonzalez A, Puchal R, Bajen M. 99Tcm-MAG3 renogram deconvolution in normal subjects and in normal functioning kidney grafts. Nucl Med Commun 1994; 15: 680-684.
  29. Gupta NK, Bomanji JB, Waddington W et al. Technetium-99m-L,L-ethylen- edicysteine scintigraphy in patients with renal disorders. Eur J Nucl Med 1995; 22: 617-624.
  30. Surma MJ, Wiewióra J, Liniecki J. Usefulness of 99Tcm-N,N’-ethylene-1-di- cysteine complex for dynamic kidney investigations. Nucl Med Commun 1994; 15: 628-635.
  31. Surma MJ, Wiewiora J, Szadkowska A, Liniecki J. Pharmacokinetic characteristics of 99mTc-Ethylene-l-dicysteine (99mTc-EC). Nucl Med Rev Cent East Eur 1999; 2: 20-27.
  32. Surma MJ, Wiewiora J, Liniecki J. Pharmacokinetics of 99Tcm-ethylenedi- cysteine (99Tcm-EC). Nucl Med Commun 1998; 19: 514 (Abs).
  33. Bergmann H, Dworak E, König B. Improved automatic separation of renal parenchyma and pelvis in dynamic renal scintigraphy using fuzzy regions of interest. Eur J Nucl Med 1999; 26: 837-843.

Correspondence to:
Paweł Cichocki
ul. Czechosłowacka 8/10, 92-216 Łódź
Phone: 42 675-72-90 Fax: 42 675-72-85
E-mail: pawelcichocki.zmn@gmail.com

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.