Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Ewa Fabiszewska1, Katarzyna Pasicz, Iwona Grabska
·
Nowotwory. Journal of Oncology 2017;67(2):115-120.
Affiliations
  1. Department of Medical Physics, Maria Skłodowska-Curie Memorial Cancer Centre and Institute of Oncology, Warszawa, Poland

open access

Vol 67, No 2 (2017)
Original article
Published online: 2017-10-16

Abstract

Between 2007–2014, there have been considerable changes in the use of different types of detector systems at mammography facilities undertaking screening programs in Poland. The use of screen-film systems (termed SF systems) has largely decreased and been replaced by either computed radiography (termed CR systems) or digital radiography systems (termed DR systems); this inevitably affecting mean glandular dose values. The study aim was to evaluate changes of mean glandular dose values for a dose exposure of 4.5 cm (MGD4,5cmPMMA) achieved by using different image detectors. The study consisted of analysing 1499 protocols for quality control tests carried out by medical physicists at 16 Regional Coordination Offices in 2007, 2011, 2012 and 2013. The mean CR system values were higher than for SF systems; by 25% in 2011, by 26% in 2012, and by 28% in 2013. In subsequent years, the MGD4,5cmPMMA mean values for DR systems were higher than for SF systems by respectively 15%, 4% and 5%. Also in the subsequent years, the MGD4,5cmPMMA mean values for CR systems were higher than for DR systems by respectively 13%, 23% and 24%. In all cases, the MGD4,5cmPMMA mean values were within the acceptable level of 2.5 mGy. The MGD4,5cmPMMA mean values were different, depending on the detection system used for mammography screening in Poland  

Abstract

Between 2007–2014, there have been considerable changes in the use of different types of detector systems at mammography facilities undertaking screening programs in Poland. The use of screen-film systems (termed SF systems) has largely decreased and been replaced by either computed radiography (termed CR systems) or digital radiography systems (termed DR systems); this inevitably affecting mean glandular dose values. The study aim was to evaluate changes of mean glandular dose values for a dose exposure of 4.5 cm (MGD4,5cmPMMA) achieved by using different image detectors. The study consisted of analysing 1499 protocols for quality control tests carried out by medical physicists at 16 Regional Coordination Offices in 2007, 2011, 2012 and 2013. The mean CR system values were higher than for SF systems; by 25% in 2011, by 26% in 2012, and by 28% in 2013. In subsequent years, the MGD4,5cmPMMA mean values for DR systems were higher than for SF systems by respectively 15%, 4% and 5%. Also in the subsequent years, the MGD4,5cmPMMA mean values for CR systems were higher than for DR systems by respectively 13%, 23% and 24%. In all cases, the MGD4,5cmPMMA mean values were within the acceptable level of 2.5 mGy. The MGD4,5cmPMMA mean values were different, depending on the detection system used for mammography screening in Poland  

Get Citation

Keywords

mean glandular dose, mammography screening, quality control

About this article
Title

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Journal

Nowotwory. Journal of Oncology

Issue

Vol 67, No 2 (2017)

Article type

Research paper (original)

Pages

115-120

Published online

2017-10-16

Page views

698

Article views/downloads

759

DOI

10.5603/NJO.2017.0018

Bibliographic record

Nowotwory. Journal of Oncology 2017;67(2):115-120.

Keywords

mean glandular dose
mammography screening
quality control

Authors

Ewa Fabiszewska
Katarzyna Pasicz
Iwona Grabska

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

Wydawcą serwisu jest VM Media Group sp. z o.o., ul. Świętokrzyska 73, 80–180 Gdańsk

tel.:+48 58 320 94 94, faks:+48 58 320 94 60, e-mail: viamedica@viamedica.pl