open access

Vol 1, No 1 (2016)
Original article
Published online: 2016-08-11
Get Citation

Inactivation of Clostridium sporogenes and Geobacillus stearothermophilus spores with the use of microwave and steam sterilizers and microwave oven

Krzysztof Skowron, Justyna Bauza-Kaszewska, Agnieszka Kaczmarek, Anna Budzyńska, Joanna Kwiecińska-Piróg, Adrian Reśliński, Eugenia Gospodarek-Komkowska
·
Medical Research Journal 2016;1(1):15-22.

open access

Vol 1, No 1 (2016)
ORIGINAL ARTICLES
Published online: 2016-08-11

Abstract

Introduction. Equipment for sterilization used in medical laboratories must be absolutely effective in eliminating microorganisms and their spores. It often directly influences human health, even life. The aim of the study was to compare the effectiveness of sterilization using the steam sterilizer ASV E, microwave sterilizer EnbioJet ML1, microwave sterilizer for baby bottles and breast pumps AVENT and microwave oven.

Materials and methods. Evaluation of the effectiveness of sterilization with the use of selected devices based on pressure-thermal and microwave-thermal methods was conducted, on the basis of elimination of G. stearothermophilus PCM 2104 and C. sporogenes IW 1306 spores.

Results. After using the steam sterilizer, 100% inactivation of spores of both species was noted. In the case of EnbioJet ML1 sterilizer, in the test containing 106 CFU × cm–3 G. stearothermophilus spores, 1.63 × 101 CFU × cm–3 survived. The baby bottles sterilizer proved less effective. While the microwave, in the case of tests with the highest spore content, provided their inactivation only at the level of more than 70.0%. The steam sterilizer and EnbioJet ML1 sterilizer were the most effective, whereas the latter ensured a very short time of high temperature effect, which has a favorable impact on the properties of sterilized products, for example compounds decomposing in high temperature.

Conclusion. Results of own, as well as other authors’ studies allow to confirm the large potential in the scope of using microwave radiation for the sterilization and disinfection of materials of various sensitivity to temperature.  

Abstract

Introduction. Equipment for sterilization used in medical laboratories must be absolutely effective in eliminating microorganisms and their spores. It often directly influences human health, even life. The aim of the study was to compare the effectiveness of sterilization using the steam sterilizer ASV E, microwave sterilizer EnbioJet ML1, microwave sterilizer for baby bottles and breast pumps AVENT and microwave oven.

Materials and methods. Evaluation of the effectiveness of sterilization with the use of selected devices based on pressure-thermal and microwave-thermal methods was conducted, on the basis of elimination of G. stearothermophilus PCM 2104 and C. sporogenes IW 1306 spores.

Results. After using the steam sterilizer, 100% inactivation of spores of both species was noted. In the case of EnbioJet ML1 sterilizer, in the test containing 106 CFU × cm–3 G. stearothermophilus spores, 1.63 × 101 CFU × cm–3 survived. The baby bottles sterilizer proved less effective. While the microwave, in the case of tests with the highest spore content, provided their inactivation only at the level of more than 70.0%. The steam sterilizer and EnbioJet ML1 sterilizer were the most effective, whereas the latter ensured a very short time of high temperature effect, which has a favorable impact on the properties of sterilized products, for example compounds decomposing in high temperature.

Conclusion. Results of own, as well as other authors’ studies allow to confirm the large potential in the scope of using microwave radiation for the sterilization and disinfection of materials of various sensitivity to temperature.  

Get Citation

Keywords

spores, Clostridium sporogenes, Geobacillus stearothermophilus, sterilization, microwave radiation

About this article
Title

Inactivation of Clostridium sporogenes and Geobacillus stearothermophilus spores with the use of microwave and steam sterilizers and microwave oven

Journal

Medical Research Journal

Issue

Vol 1, No 1 (2016)

Article type

Original article

Pages

15-22

Published online

2016-08-11

Page views

931

Article views/downloads

2318

DOI

10.5603/MRJ.2016.0003

Bibliographic record

Medical Research Journal 2016;1(1):15-22.

Keywords

spores
Clostridium sporogenes
Geobacillus stearothermophilus
sterilization
microwave radiation

Authors

Krzysztof Skowron
Justyna Bauza-Kaszewska
Agnieszka Kaczmarek
Anna Budzyńska
Joanna Kwiecińska-Piróg
Adrian Reśliński
Eugenia Gospodarek-Komkowska

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: viamedica@viamedica.pl