Early device migration to the pulmonary artery after percutaneous patent ductus arteriosus closure in a neonate

Karolina Andrzejczyk12, Maciej Chojnicki1, Julia Haponiuk13, Mariusz Steffens1, Konrad Paczkowski1, Katarzyna Gierat-Haponiuk14, Anna Romanowicz1, Marta Paśko-Majewska1, Wiktor Szymanowicz1, Ireneusz Haponiuk14
Folia Cardiologica 2018;13(5):481-484.

Abstract

Device displacement with pulmonary artery embolisation (PDA) is a rare complication after percutaneous closure of
a patent arterial duct. An 11-day old neonate with PDA was admitted to the Department of Paediatric Cardiac Surgery for
treatment. The newborn was qualified for a transcatheter closure of PDA. Device displacement to the pulmonary artery
was observed during the procedure. The patient was qualified for a surgical procedure of PDA closure, with removal of
the implant via midline sternotomy to provide an additional safeguard with extracorporeal circulation. The procedure and
early postoperative course were uneventful. The patient was discharged home in good general condition.
Interventional cardiology can provide effective minimally-invasive solutions for congenital heart disease treatment, but
there is always a risk of potentially life-threatening complications.

Article available in PDF format

View PDF (Polish) Download PDF file

References

  1. Philip R, Waller BR, Agrawal V, et al. Morphologic characterization of the patent ductus arteriosus in the premature infant and the choice of transcatheter occlusion device. Catheter Cardiovasc Interv. 2016; 87(2): 310–317.
  2. Delaney JW, Fletcher SE. Patent ductus arteriosus closure using the Amplatzer® vascular plug II for all anatomic variants. Catheter Cardiovasc Interv. 2013; 81(5): 820–824.
  3. Garay FJ, Aguirre D, Cárdenas L, et al. Use of the amplatzer vascular plug II device to occlude different types of patent ductus arteriosus in pediatric patients. J Interv Cardiol. 2015; 28(2): 198–204.
  4. Ramakrishnan S. Vascular plugs — a key companion to interventionists — 'just plug it'. Indian Heart J. 2015; 67(4): 399–405.
  5. Haponiuk I, Paczkowski K, Chojnicki M, et al. Iatrogenic obstruction of the aorta — a sequence of delayed, fatal complications after 'off-label' interventional persistent ductus arteriosus closure. Videosurgery Miniinv. 2016; 11(1): 44–48.
  6. Backes CH, Cheatham SL, Deyo GM, et al. Percutaneous Patent Ductus Arteriosus (PDA) closure in very preterm infants: feasibility and complications. J Am Heart Assoc. 2016; 5(2): 1–10.
  7. Krichenko A, Benson LN, Burrows P, et al. Angiographic classification of the isolated, persistently patent ductus arteriosus and implications for percutaneous catheter occlusion. Am J Cardiol. 1989; 63(12): 877–880.
  8. Pawelec-Wojtalik M, Masura J, Wojtalik M, et al. Treatment of the patent ductus arteriosus with the Amplatzer Duct Occluder. Folia Cardiol Excerpta. 2004; 11(4): 285–291.
  9. Białkowski J, Głowacki J, Zabal C, et al. Patent ductus arteriosus at low and high altitudes: anatomical and haemodynamic features and their implications for transcatheter closure. Kardiol Pol. 2011; 69(5): 431–436.
  10. Koltowski P, Patkowski P, Adamus J, et al. Thoracoscopic closure of patent ductus arteriosus in two premature infants weighing 700 and 800 g. Kardiochir Torakochir Pol. 2012; 9(4): 428–430.