Introduction
Although bare metal stent (BMS) implantation has a lower restenosis rate than that of balloon angioplasty, the reduction of the restenosis rate by BMS is limited [1]. The use of durable polymer drug-eluting stents (DP-DES) has significantly decreased restenosis rates [2, 3]. However, DP-DES may induce stent thrombosis through local hypersensitivity reactions and inflammation caused by polymer persistence [4, 5]. To overcome this limitation, new-generation DES, including biodegradable polymer DES (BP-DES) and polymer-free DES, have been developed [6, 7].
This study evaluated the safety and efficacy of a novel polymer-free everolimus-eluting stent (EES) with a nitrogen-doped titanium dioxide (N-TiO2) film, comparing them with those of a commercial polymer-free sirolimus-eluting stent (SES). To the best of our knowledge, this is the first pre-clinical study comparing a polymer-free EES with an N-TiO2 film with another polymer-free stent design in a swine model with up to 6 months of follow-up.
Methods
The novel polymer-free EES with an N-TiO2 film was designed by Chonnam National University Hospital (Gwangju, Republic of Korea) and manufactured by Cell & Growth Factor Biotechnology (CG Bio Co., Ltd., Seoul, Republic of Korea). The stent strut based on cobalt-chromium was coated with N-TiO2 using a plasma-enhanced chemical vapor deposition process (Fig. 1) [8, 9]. The stent was covered with everolimus using an electrospinning technique [10]. This stent has been patented in the United States (patent numbers 09795987 and 10343184) and was approved for clinical trials by the Ministry of Food and Drug Safety in June 2020.
Characterization of the stent surface
Roughness of the stent surface
The surface roughness of the BMS, BMS coated with N-TiO2, polymer-free EES coated with N-TiO2, and polymer-free SES was assessed using atomic force microscopy (NX10; Park System, Suwon, Republic of Korea). Ra (arithmetic mean roughness), a commonly used roughness parameter, is measured using the average of the absolute values of height deviations from the mean line within the evaluation length. The higher the Ra value, the greater the difference between the peaks and valleys of the stent surface. The scan size was 10–20 μm.
Hydrophobicity of the stent surface
The contact angle test is a tool to evaluate the wettability of a surface. Hydrophobicity, defined as a contact angle over 90°, intensifies with larger contact angles. The degree of the dispersion of 2 μL of water on the stent surface was evaluated using a contact angle analyzer (Phoenix 300; SEO, Suwon, Republic of Korea). The hydrophobicity test was performed 10 times for each stent.
Study groups
A total of 30 pigs were assigned to 1 month (n = 10), 3 months (n = 10), and 6 months of follow-up (n = 10) from March to December 2018. One polymer-free EES with an N-TiO2 film (3.0 × 16 mm; CG Bio Co., Seoul, Republic of Korea) and polymer-free SES (Coroflex ISAR® 3.0 × 16 mm; B. Braun. Melsungen, Hessen, Germany) were randomly implanted into two different coronary arteries in 1 pig. Yorkshire × Landrace F1 crossbred neutered male pigs were followed up for 1 month (n = 10; 15-20 kg, 3 months of age, Chuwol grandparent farm in southwest Republic of Korea). Male minipigs (trade name micro-pig; T-type pig) were followed up for 3 months (n = 10) and 6 months (n = 10) (15–20 kg, 24 months of age, APURES in northwest Republic of Korea) (Fig. 2). Contrary to previous studies that used DP-DES for the control group, this study used the polymer-free SES [10, 11]. This study conformed to the Guide for the Care and Use of Laboratory Animals and was approved by the Chonnam National University Hospital Animal Ethics Committee based on the Animal Protection Act (Institutional Review Board approval number: CNU IACUC-H-2018-5).
Anesthesia for experimental animals
General anesthesia was induced using an intramuscular injection of 12 mg/kg of ketamine and 8 mg/kg of xylazine. Oxygen was supplied using an intubation tube, and saline was supplied through the ear veins. Then, local anesthesia with 2% lidocaine was sterilely induced at the center of the animal’s neck. The left carotid artery of the pig was cut, and a 7-Fr sheath was inserted. A dose of 200 IU/kg of heparin was administered as a single injection into the artery before advancing the guiding catheter.
Experimental process
Before stenting, coronary angiography (CAG) was performed to identify the coronary artery segment for the 3.0 × 16 mm stent implantation. Quantitative coronary analyses (QCA) were performed to evaluate the coronary lumen diameter and length. Then, the balloon was expanded for 30 s. The stent-to-vessel ratio ranged from 1.1:1 to 1.3:1. Optical coherence tomography (OCT) was performed to ensure optimal stent expansion and apposition, together with the absence of edge dissections. After stenting, the left carotid artery was ligated, and the skin of the neck was sutured. Acetylsalicylic acid (100 mg/day) and clopidogrel (75 mg/day) were administered 7 days before stenting and until the end of the study period. Because the healing response and neointimal growth pattern after stent implantation occur 5- to 6-times faster in pigs than in humans, 1-, 3-, and 6-month follow-up periods were used during this study to represent 1.5 to 3 years of follow-up for humans who receive DES [12]. The right carotid artery was used for performing CAG, QCA, and OCT during follow-up.
Histopathologic study
The experimental animals fasted the day before anesthesia. After follow-up CAG, 40 mL of potassium chloride was injected through a catheter sheath to induce euthanasia. The heart was removed through an incision in the left thoracic cavity. Perfusion fixation with 4% formalin was performed for more than 24 h at a perfusion pressure of 70 mmHg. The stented coronary artery was isolated, and three sections were prepared to estimate the coronary artery lesion. Samples were embedded in paraffin. To evaluate re-endothelialization of the stent, an immunohistochemical analysis was performed. The tissue samples were divided into 5-µm-thick sections using a rotary microtome. To reduce autofluorescence, 0.1% Sudan black B was used for 20 min. To prevent non-specific reactivity, 3% fetal bovine serum in phosphate-buffered saline was used for 60 min. An anti-mouse monoclonal cluster of differentiation 31 (CD31) antibody (1:100; Abcam, Cambridge, Massachusetts, USA) was used for immunohistochemistry. Streptavidin Alexa Fluor® 488-conjugated anti-mouse IgG (1:400; Invitrogen) was used as the secondary antibody for fluorescence microscopy.
Analysis
QCA evaluation
Two experienced examiners conducted the QCA. CAG was performed after the administration of nitroglycerin into the coronary artery. The analysis was conducted using a QCA system (CAAS QCA Workstation; Pie Medical Imaging, Maastricht, Netherlands). The reference vessel diameter, stenosis diameter in the target lesion, and minimal lumen diameter were measured. Late lumen loss (LLL) was defined as the difference between the minimal lumen diameter immediately after stent implantation and at the time of follow-up.
OCT evaluation
Optical coherence tomography with C7-XRTM (Light Imaging Inc., Westford, Massachusetts, USA) was used to observe the strut apposition during the stent implantation and the neointimal response in the coronary vessels during follow-up. A DragonflyTM OPTIS catheter (Abbott Vascular, Chicago, IL, USA) was used for the acquisition. The optical catheter was previously purged with contrast (Omnihexol 350; Korea United Pharm Inc., Seoul, Republic of Korea). A 6-Fr guiding catheter was used, and the guidewire was inserted into the coronary artery where the stent was located. The OCT probe was advanced along the guidewire and placed at the intended scanning location using C-arm guidance. A total of 30 mL of contrast was injected into the coronary artery at a pressure of 300 psi/s using an automatic injector, thus enabling OCT image acquisition using the non-occlusive technique. A 54-mm coronary segment was imaged at a pullback speed of 20 mm/s and a rotation speed of 100 frames/s.
Histopathologic evaluation
Experts at Chonnam National University Hospital and CG Bio Inc. conducted the histopathologic analysis. The external and internal elastic laminae, lumen, stent, and neointimal areas and stent strut thickness were measured. Damage to the vessel walls caused by stent overexpansion was assessed by the following scoring system: disruption of the internal elastic lamina, 1 point; disruption of the media, 2 points; and disruption of the external elastic lamina, 3 points. The inflammatory index was used to quantify the degree of inflammatory cell infiltration into the entire vessel. When the inflammatory cells were scattered, the inflammatory index score was 1. When they were concentrated around one-half of the struts, the score was 2. When they were concentrated around all struts, the score was 3. Fibrin deposition was scored from 0 to 3 and assessed for each stent section. The rate of re-endothelialization was calculated based on the percentage of the lumen area covered by endothelial cells.
Statistical analysis
All measurements are expressed as mean ± standard deviation. They were analyzed using unpaired t-test and one-way analysis of variance. Statistical analyses were conducted using SPSS software (version 25.0; SPSS Inc., Chicago, IL, USA). Statistical significance was set at p < 0.05.
Results
Stent surface roughness and hydrophobicity
The average surface roughness was 83.3 nm for the BMS and 407.2 nm after the plasma-enhanced chemical vapor deposition of N-TiO2. The surface of the EES with an N-TiO2 coating was rougher than that of the polymer-free SES (628.1 nm vs. 558.1 nm) (Fig. 1). The water contact angles were 124.3° ± 0.18° for the BMS, 117.0° ± ± 0.06° for the BMS with an N-TiO2 coating, 108.1° ± 0.18° for the EES with an N-TiO2 coating, and 123.8° ± 0.11° for the polymer-free SES. The hydrophobicity values of the stent surfaces were significantly different between the BMS and the BMS with an N-TiO2 coating (p < 0.001), between the BMS with an N-TiO2 coating and the EES with an N-TiO2 coating (p < 0.001), and between the EES with an N-TiO2 coating and the polymer-free SES (p < 0.001) (Fig. 3).
Coronary lumen diameters and LLL
The baseline coronary lumen diameters were not significantly different between the follow-up groups. Stent implantation was successful in all animals, with a thrombolysis in myocardial infarction flow grade of 3. After stent implantation, the coronary lumen diameters were not significantly different between the follow-up groups. On all follow-up QCA, the coronary lumen diameter (p = 0.732 at 1 month, p = 0.186 at 3 months, p = 0.314 at 6 months) and LLL (p = 0.520 at 1 month, p = 0.218 at 3 months, p = 0.449 at 6 months) were not different in the comparison groups (Table 1). No death occurred during the follow-up period.
N-TiO2 EES (n = 30) |
Polymer-free SES (n = 30) |
P |
|
1 month |
n = 10 |
n = 10 |
|
QCA: |
|||
Before stenting |
|||
Vessel diameter [mm] |
2.73 ± 0.24 |
2.73 ± 0.24 |
0.984 |
After stenting |
|||
Lumen diameter [mm] |
2.89 ± 0.23 |
2.80 ± 0.25 |
0.446 |
Stent-to-artery ratio |
1.06 ± 0.03 |
1.03 ± 0.07 |
0.232 |
Follow-up |
|||
Lumen diameter [mm] |
2.49 ± 0.32 |
2.43 ± 0.40 |
0.732 |
Late lumen loss [mm] |
0.38 ± 0.31 |
0.47 ± 0.21 |
0.520 |
Histologic findings: |
|||
Injury score |
1.01 ± 0.02 |
1.00 ± 0.02 |
0.637 |
Inflammatory score |
1.00 ± 0.00 |
1.02 ± 0.07 |
0.118 |
Fibrin score |
1.89 ± 0.45 |
2.53 ± 0.54 |
0.001 |
Stenosis area [%] |
30.5 ± 6.7 |
35.5 ± 10.4 |
0.003 |
Re-endothelialization rate [%] |
78.4 ± 14.7 |
66.2 ± 27.0 |
0.193 |
3 months |
n = 10 |
n = 10 |
|
QCA: |
|||
Before stenting |
|||
Vessel diameter [mm] |
2.77 ± 0.24 |
2.80 ± 0.26 |
0.738 |
After stenting |
|||
Lumen diameter [mm] |
2.95 ± 0.25 |
3.00 ± 0.22 |
0.625 |
Stent-to-artery ratio |
1.07 ± 0.03 |
1.08 ± 0.05 |
0.613 |
Follow-up |
|||
Lumen diameter [mm] |
2.50 ± 0.32 |
2.32 ± 0.28 |
0.186 |
Late lumen loss [mm] |
0.50 ± 0.37 |
0.73 ± 0.36 |
0.218 |
OCT: |
|||
Neointimal area [%] |
22.5 ± 11.4 |
32.1 ± 12.3 |
< 0.001 |
Neointimal volume [%] |
22.9 ± 11.6 |
29.6 ± 11.2 |
0.205 |
Histologic findings: |
|||
Injury score |
1.06 ± 0.10 |
1.05 ± 0.10 |
0.472 |
Inflammatory score |
1.04 ± 0.07 |
1.05 ± 0.09 |
0.871 |
Fibrin score |
0.22 ± 0.46 |
0.18 ± 0.43 |
0.620 |
Stenosis area [%] |
34.4 ± 10.2 |
39.7 ± 14.0 |
0.019 |
Re-endothelialization rate [%] |
81.3 ± 19.8 |
78.8 ± 20.6 |
0.732 |
6 months |
n = 10 |
n = 10 |
|
QCA: |
|||
Before stenting |
|||
Vessel diameter [mm] |
2.59 ± 0.20 |
2.69 ± 0.25 |
0.368 |
After stenting |
|||
Lumen diameter [mm] |
2.93 ± 0.19 |
2.90 ± 0.15 |
0.719 |
Stent-to-artery ratio |
1.13 ± 0.08 |
1.08 ± 0.06 |
0.119 |
Follow-up |
|||
Lumen diameter [mm] |
2.58 ± 0.28 |
2.41 ± 0.40 |
0.314 |
Late lumen loss [mm] |
0.40 ± 0.28 |
0.53 ± 0.45 |
0.449 |
OCT: |
|||
Neointimal area [%] |
27.9 ± 10.5 |
28.8 ± 13.4 |
0.524 |
Neointimal volume [%] |
27.6 ± 9.7 |
30.8 ± 12.9 |
0.531 |
Histologic findings: |
|||
Injury score |
1.07 ± 0.11 |
1.07 ± 0.14 |
0.797 |
Inflammatory score |
1.04 ± 0.23 |
1.02 ± 0.05 |
0.413 |
Fibrin score |
0.00 ± 0.00 |
0.02 ± 0.14 |
0.322 |
Stenosis area [%] |
45.5 ± 7.5 |
51.7 ± 12.3 |
0.002 |
Re-endothelialization rate [%] |
92.6 ± 11.2 |
85.8 ± 15.8 |
0.208 |
Neointimal area and volume
The percentage of neointimal area obstruction in the polymer-free SES was larger than that in the polymer-free EES with an N-TiO2 film at 3 months (32.1% ± 12.3% vs. 22.5% ± 11.4%, p < 0.001) but was not significant at 6 months of follow-up (28.8% ± 13.4% vs. 27.9% ± 10.5%, p = 0.524). No significant differences in the neointimal volume between the two stent groups at 3 and 6 months of follow-up were recorded (Fig. 4, Table 1).
Histopathologic outcomes
No significant differences in the injury and inflammatory scores between the two stent groups at 1 month were observed. The fibrin score with the polymer-free SES was significantly higher than that with the polymer-free EES with an N-TiO2 film (p = 0.001). The area of stenosis with the polymer-free SES was larger than that with the polymer-free EES with an N-TiO2 film (p = 0.003). The injury, inflammatory, and fibrin scores were not significantly different between the two stent groups at 3 months. The area of stenosis with the polymer-free SES was larger than that with the polymer-free EES with an N-TiO2 film (p = 0.019). At 6 months, the injury, inflammatory, and fibrin scores were not significantly different between the two stent groups. The area of stenosis with the polymer-free SES was larger than that with the polymer-free EES with an N-TiO2 film (p = 0.002) (Figs. 4, 5, Table 1). The rates of re-endothelialization were not significantly different between the two stent groups at 1, 3, or 6 months (Fig. 6, Table 1).
Discussion
The main findings of this study can be summarized as follows: A polymer-free EES with an N-TiO2 film showed borderline superior efficacy than that of a polymer-free SES for up to 6 months in a swine study, as indicated by the smaller neointimal area obstruction on OCT at 3 months and the smaller stenosis area on QCA at 6 months of follow-up. Although the polymer-free EES with an N-TiO2 film resulted in a lower fibrin score than that of the polymer-free SES at 1 month, all other parameters of inflammation, re-endothelialization, and safety were not significantly different.
The use of DP-DES decreased the rate of in-stent restenosis from 15–30% to less than 15% after 12 months [5]. However, DP-DES have a high probability of developing very late stent thrombosis. The remaining stent material, including the polymer, induced chronic inflammation and a hypersensitivity reaction, which may cause very late stent thrombosis. Although the polymer-free DES was developed to improve these problems, it has not shown meaningful outcomes compared with those of the DP-DES. In the Intracoronary Stenting and Angiographic Results: Test Efficacy of Three Limus-Eluting Stents (ISAR-TEST) 2 and ISAR-TEST 5 trials, polymer-free DES had similar death, myocardial infarction, and revascularization rates (including definite or probable stent thrombosis) compared to those of DP-DES [13–15]. In contrast, fluoropolymer-coated stents demonstrated outstanding results in in vitro tests. Fluoropolymers bind tightly with albumin and passivate the surface to protect against thrombogenic materials [16]. The platelet adherence and inflammatory cell density were lower with the fluoropolymer-coated EES compared with the polymer-free biolimus A9-coated stent in ex vivo swine arteriovenous shunt model experiments [17]. To overcome the limitations of the previous polymer-free DES, the polymer-free EES with N-TiO2 films used differentiated coating technologies and drugs.
Several technologies have been used to coat the polymer-free DES with drugs, including use of microporous surfaces, direct coating, crystallization of the drug, and inorganic porous coatings. These coating technologies affect the drug loading, drug-eluting rate, and platelet activation [18]. The stent platform of the polymer-free SES used in the control group was sandblasted to create a microporous surface that allows drug deposition. The novel polymer-free EES used specific coating technologies, such as an inorganic porous coating with an N-TiO2 film. A plasma-enhanced chemical vapor deposition process with N-TiO2 creates a coarse surface on the stent for the drug. One study showed that the smooth surface of the stent had less thrombogenicity and induced less neointimal hyperplasia [19]. However, in the study by Dibra et al. [20], rough stent surfaces did not increase LLL and restenosis rates compared with those of smooth stent surfaces in patients with coronary artery disease who underwent percutaneous coronary intervention. Another study showed that microscopic parallel grooves improved the endothelialization compared with that of smooth surfaces [21]. The process of N-TiO2 coating also results in an improved hydrophobicity at the stent surface (Fig. 3). The hydrophilic surface tends to improve the endothelial healing and reduce the platelet adhesion. The ultra-hydrophilic surface-treated BMS showed less intimal hyperplasia compared with that of the second-generation DES in a swine coronary model [22]. During inflammation, monocytes are more likely to adhere to hydrophobic surfaces than to hydrophilic surfaces [23].
Drugs (such as limus analogs and paclitaxel) may be used to reduce neointimal hyperplasia, which is related to restenosis, by suppressing the activation of smooth muscle and inflammatory cells. However, re-endothelization may also be inhibited, consequently increasing the stent thrombosis. Use of the EES resulted in less neointimal hyperplasia, less fibrin deposition, and more re-endothelialization compared with use of the SES in a rabbit model of iliac artery stenting [24]. Although similar clinical outcomes were reported for the EES, a zotarolimus-eluting stent, and SES in the BIO-RESORT trial [25], the EES had a lower risk of very late stent thrombosis and better outcomes than the SES in the Scandinavian Organization for Randomized Trials with Clinical Outcome IV trial [26]. Because of the influences of the coating technology and drugs used, the percentage of the area of stenosis during 6 months of follow-up and the fibrin score, which reflects the state of arterial healing, at 1 month after the stent placement were lower in the polymer-free EES with an N-TiO2 film than in the polymer-free SES. Similar inflammatory scores and re-endothelialization rates between the two groups throughout the follow-up period can also be influenced by these factors.
The novel polymer-free EES with an N-TiO2 film showed good biocompatibility and favorable 1-month outcomes in previous swine coronary models [10, 11]. In the preclinical study of Cho et al. [27], the polymer-free EES with an N-TiO2 film also demonstrated similar 6-month OCT and histopathologic findings compared with those of fluoropolymer-coated stents. Although a significant difference was not observed between the two stent types, the polymer-free EES with an N-TiO2 film had a fibrin score of 0 at 6 months (0 vs. 0.07 ± ٠.١١, p = 0.180). Therefore, reasonable mid- to long-term outcomes regarding restenosis and stent thrombosis are expected based on the results of our study. The polymer-free DES with an N-TiO2 film can be an alternative to DP-DES and other polymer-free DES.
Limitations of the study
However, this study has some limitations. First, the stent was implanted in a healthy swine coronary artery that had no significant atherosclerosis. Second, the type of pigs used during the 1-month follow-up period was different from the type of pigs used during the other follow-up periods. Finally, the follow-up period was limited to 6 months. Further research is warranted to obtain data regarding the longer-term safety and efficacy.
Conclusions
The domestic polymer-free EES with an N-TiO2 film outperformed the safety and efficacy of the polymer-free SES in experimental and pre-clinical 6-month follow-up tests in many aspects. The novel polymer-free EES with an N-TiO2 film may be used effectively in clinical practice.
Acknowledgments
This study was supported by a grant from the Korea Health Technology R&D Project (HI13C1527), Ministry of Health and Welfare, Republic of Korea. It was also supported by funds from Chonnam National University Hospital Biomedical Research Institute (BCRI-21054) and the Ministry of Science and ICT; the Ministry of Trade, Industry, and Energy; the Ministry of Health and Welfare; and the Ministry of Food and Drug Safety (1711138916, KMDF_PR_20200901_0280), Republic of Korea. This article underwent English language editing by Editage.