Two-year longitudinal evaluation of a second-generation thin-strut sirolimus-eluting bioresorbable coronary scaffold with hybrid cell design in porcine coronary arteries
Abstract
Background: The first commercially available bioresorbable scaffold (BRS) had a strut thickness of 156 microns. As such, it had the potential for delivery challenges and higher thrombogenicity. The aim herein, is to evaluate biomechanical performance, pharmacokinetics and vascular healing of a novel thin strut (100 μm) sirolimus eluting BRS (MeRes-100, Meril Life Sciences, Gujarat, India) against the once clinically used BRS (Absorb BVS, Abbott, Santa Clara, CA) in porcine coronary arteries.
Methods: Following device implantation, angiographic and optical coherence tomography (OCT) evaluation were performed at 45, 90, 180 days, 1 year and 2 years. Histological evaluation was performed at 30, 90 and 180 days.
Results: At 2 years, both lumen (MeRes-100 7.07 ± 1.82 mm2 vs. Absorb BVS 7.57 ± 1.39 mm2, p = NS) and scaffold areas (MeRes-100 9.73 ± 1.80 mm2 vs. Absorb BVS 9.67 ± 1.25 mm2, p = NS) were comparable between tested and control scaffolds. Also, the late lumen area gain at 2 years was similar in both groups tested (MeRes-100 1.03 ± 1.98 mm2 vs. Absorb BVS 0.85 ± 1.56 mm2, p = NS). Histologic examination up to 6 months showed comparable healing and inflammation profiles for both devices.
Conclusions: The novel sirolimus-eluting BRS with thinner struts and hybrid cell design showed similar biomechanical durability and equivalent inhibition of neointimal proliferation when compared to the first-ever Absorb BVS up to 2 years in normal porcine coronary arteries.
Keywords: optical coherence tomographybioresobable scaffoldshistology
References
- Onuma Y, Ormiston J, Serruys PW. Bioresorbable scaffold technologies. Circ J. 2011; 75(3): 509–520.
- Ellis SG, Kereiakes DJ, Metzger DC, et al. Everolimus-Eluting Bioresorbable Scaffolds for Coronary Artery Disease. N Engl J Med. 2015; 373(20): 1905–1915.
- Serruys PW, Chevalier B, Dudek D, et al. A bioresorbable everolimus-eluting scaffold versus a metallic everolimus-eluting stent for ischaemic heart disease caused by de-novo native coronary artery lesions (ABSORB II): an interim 1-year analysis of clinical and procedural secondary outcomes from a randomised controlled trial. Lancet. 2015; 385(9962): 43–54.
- Serruys PW, Chevalier B, Sotomi Y, et al. Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trial. Lancet. 2016; 388(10059): 2479–2491.
- Kereiakes DJ, Ellis SG, Metzger C, et al. ABSORB III Investigators. 3-Year Clinical Outcomes With Everolimus-Eluting Bioresorbable Coronary Scaffolds: The ABSORB III Trial. J Am Coll Cardiol. 2017; 70(23): 2852–2862.
- Otsuka F, Cheng Qi, Yahagi K, et al. Acute Thrombogenicity of a Durable Polymer Everolimus-Eluting Stent Relative to Contemporary Drug-Eluting Stents With Biodegradable Polymer Coatings Assessed Ex Vivo in a Swine Shunt Model. JACC Cardiovasc Interv. 2015; 8(9): 1248–1260.
- Koskinas KC, Chatzizisis YS, Antoniadis AP, et al. Role of endothelial shear stress in stent restenosis and thrombosis: pathophysiologic mechanisms and implications for clinical translation. J Am Coll Cardiol. 2012; 59(15): 1337–1349.
- Azzalini L. L'Allier PL. Bioresorbable vascular scaffold thrombosis in an all-comer patient population: single-center experience. J Invasive Cardiol. 2015; 27: 85–92.
- Onuma Y, Serruys PW, Gomez J, et al. Comparison of in vivo acute stent recoil between the bioresorbable everolimus-eluting coronary scaffolds (revision 1.0 and 1.1) and the metallic everolimus-eluting stent. Catheter Cardiovasc Interv. 2011; 78(1): 3–12.
- Cheng Y, Gasior P, Shibuya M, et al. Comparative characterization of biomechanical behavior and healing profile of a novel ultra-high-molecular-weight amorphous poly-l-lactic acid sirolimus-eluting bioresorbable coronary scaffold. Circ Cardiovasc Interv. 2016; 9(10).
- Gomez-Lara J, Brugaletta S, Diletti R, et al. A comparative assessment by optical coherence tomography of the performance of the first and second generation of the everolimus-eluting bioresorbable vascular scaffolds. Eur Heart J. 2011; 32(3): 294–304.
- Strandberg E, Zeltinger J, Schulz DG, et al. Late positive remodeling and late lumen gain contribute to vascular restoration by a non-drug eluting bioresorbable scaffold: a four-year intravascular ultrasound study in normal porcine coronary arteries. Circ Cardiovasc Interv. 2012; 5(1): 39–46.
- Yazdani SK, Pacheco E, Nakano M, et al. Vascular, downstream, and pharmacokinetic responses to treatment with a low dose drug-coated balloon in a swine femoral artery model. Catheter Cardiovasc Interv. 2014; 83(1): 132–140.
- Stone GW, Gao R, Kimura T, et al. 1-year outcomes with the Absorb bioresorbable scaffold in patients with coronary artery disease: a patient-level, pooled meta-analysis. Lancet. 2016; 387(10025): 1277–1289.
- Tenekecioglu E, Torii R, Bourantas C, et al. Preclinical assessment of the endothelial shear stress in porcine-based models following implantation of two different bioresorbable scaffolds: effect of scaffold design on the local haemodynamic micro-environment. EuroIntervention. 2016; 12(10): 1296.
- Gogas BD, King SB, Timmins LH, et al. Biomechanical assessment of fully bioresorbable devices. JACC Cardiovasc Interv. 2013; 6(7): 760–761.
- Chatzizisis YS, Coskun AU, Jonas M, et al. Role of endothelial shear stress in the natural history of coronary atherosclerosis and vascular remodeling: molecular, cellular, and vascular behavior. J Am Coll Cardiol. 2007; 49(25): 2379–2393.
- Koppara T, Cheng Qi, Yahagi K, et al. Thrombogenicity and early vascular healing response in metallic biodegradable polymer-based and fully bioabsorbable drug-eluting stents. Circ Cardiovasc Interv. 2015; 8(6): e002427.
- Kang J, Han JK, Yang HM, et al. Bioresorbable vascular scaffolds: are we facing a time of crisis or one of breakthrough? Circ J. 2017; 81(8): 1065–1074.
- Suzuki T, Kopia G, Hayashi S, et al. Stent-based delivery of sirolimus reduces neointimal formation in a porcine coronary model. Circulation. 2001; 104(10): 1188–1193.
- Vetrovec GW, Rizik D, Williard C, et al. Sirolimus PK trial: a pharmacokinetic study of the sirolimus-eluting Bx velocity stent in patients with de novo coronary lesions. Catheter Cardiovasc Interv. 2006; 67(1): 32–37.
- Seth A, Onuma Y, Costa R, et al. First-in-human evaluation of a novel poly-L-lactide based sirolimus-eluting bioresorbable vascular scaffold for the treatment of de novo native coronary artery lesions: MeRes-1 trial. EuroIntervention. 2017; 13(4): 415–423.