Implantation of the Micra transcatheter pacing system: Single Polish center experience with the real costs of hospitalization analysis
Abstract
Background: The Micra transcatheter pacing system (TPS) is a miniaturized, single-chamber pacemaker system. Study reported herein is an initial experience with implantation of the Micra TPS.
Methods: The leadless pacemaker was implanted in 10 patients with standard indications for a permanent pacemaker implantation. All hospitalization costs were calculated for all patients.
Results: The mean age of the patients was 75 ± 7.1 years, 6 were men and 4 were women. Four patients had permanent atrial fibrillation as the basal rhythm and 6 patients had sinus rhythm. All patients had at least one relative contraindication that precluded the use of a traditional pacing system. Mean intraoperative ventricular sensing amplitude was 10.6 ± 5.4 mV, impedance 843 ± 185 ohms, and pacing threshold at 0.24 ms was 0.56 ± 0.23 V. At discharge, those values were 13.9 ± 5.6 mV, 667 ± 119 ohms and 0.47 ± 0.17, respectively. The mean duration of implantation procedure was 82 min, while mean fluoroscopy time was 3.5 min. Two patients developed hematoma at the groin puncture site post-implantation. In 1 case there was a need for erythrocyte mass transfusion and surgical intervention. Mean total time of hospitalization was 26 days and time from procedure to discharge 12 days. Average cost of hospitalization per 1 patient was 11,260.15 EUR minimal cost was 9,052.68 EUR, while maximal cost was 16,533.18 EUR.
Conclusions: Implantation of leadless pacemakers is feasible, safe and provides advantages over the conventional system. Hospitalization costs vary for individual patients in wide range.
Keywords: leadless pacemakerscomplicationsprocedure costhospitalization cost
References
- Udo EO, Zuithoff NPA, van Hemel NM, et al. Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm. 2012; 9(5): 728–735.
- Kirkfeldt RE, Johansen JB, Nohr EA, et al. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014; 35(18): 1186–1194.
- Ritter P, Duray GZ, Steinwender C, et al. Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study. Eur Heart J. 2015; 36(37): 2510–2519.
- Knops RE, Tjong FVY, Neuzil P, et al. Chronic performance of a leadless cardiac pacemaker: 1-year follow-up of the LEADLESS trial. J Am Coll Cardiol. 2015; 65(15): 1497–1504.
- El-Chami MF, Roberts PR, Kypta A, et al. How to Implant a Leadless Pacemaker With a Tine-Based Fixation. J Cardiovasc Electrophysiol. 2016; 27(12): 1495–1501.
- Sohail MR, Hussain S, Le KY, et al. Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. J Am Coll Cardiol. 2007; 49(18): 1851–1859.
- Baddour LM, Epstein AE, Erickson CC, et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation. 2010; 121(3): 458–477.
- Kypta A, Blessberger H, Kammler J, et al. Leadless cardiac pacemaker implantation after lead extraction in patients with severe device infection. J Cardiovasc Electrophysiol. 2016; 27(9): 1067–1071.
- Reynolds D, Duray G, Omar R, et al. A Leadless Intracardiac Transcatheter Pacing System. N Engl J Med. 2016; 374(6): 533–541.
- Roberts PR, Clementy N, Al Samadi F, et al. A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval Registry. Heart Rhythm. 2017; 14(9): 1375–1379.
- Reddy VY, Exner DV, Cantillon DJ, et al. Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker. N Engl J Med. 2015; 373(12): 1125–1135.
- Duray GZ, Ritter P, El-Chami M, et al. Long-term performance of a transcatheter pacing system: 12-Month results from the Micra Transcatheter Pacing Study. Heart Rhythm. 2017; 14(5): 702–709.