Intracardiac electrogram method of VV-delay optimization in biventricular pacemakers

Kinga Gościńska-Bis, Bogusław Grzegorzewski, Mario Migschitz, Jarosław Bis, Włodzimierz Kargul
DOI: 10.5603/cj.21710
·
Cardiol J 2007;14(3):305-310.

open access

Vol 14, No 3 (2007)
Case Reports
Submitted: 2013-01-14
Published online: 2007-03-30

Abstract

Ventricle to ventricle (VV) delay optimization can provide an additional benefit to cardiac resynchronization therapy, but the methods currently used for optimization are time consuming and operator-dependent. We present two cases of VV-delay optimization with the use of a new intracardiac electrogram method. (Cardiol J 2007; 14: 305-310)

Abstract

Ventricle to ventricle (VV) delay optimization can provide an additional benefit to cardiac resynchronization therapy, but the methods currently used for optimization are time consuming and operator-dependent. We present two cases of VV-delay optimization with the use of a new intracardiac electrogram method. (Cardiol J 2007; 14: 305-310)
Get Citation

Keywords

cardiac resynchronization therapy; heart failure

About this article
Title

Intracardiac electrogram method of VV-delay optimization in biventricular pacemakers

Journal

Cardiology Journal

Issue

Vol 14, No 3 (2007)

Pages

305-310

Published online

2007-03-30

Page views

578

Article views/downloads

1144

DOI

10.5603/cj.21710

Bibliographic record

Cardiol J 2007;14(3):305-310.

Keywords

cardiac resynchronization therapy
heart failure

Authors

Kinga Gościńska-Bis
Bogusław Grzegorzewski
Mario Migschitz
Jarosław Bis
Włodzimierz Kargul

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., Grupa Via Medica, ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: viamedica@viamedica.pl