Irregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device

Athanasios G. Manolis, Spyros Kourouklis, Dimitrios Chatzis, Konstantinos Kouvelas, Zenon Kyriakides
Cardiol J 2008;15(3):281-283.

Abstract

A case of sustained monomorphic ventricular tachycardia underdetected by a single chamber implantable cardioverter defibrillator because of RR interval irregularity is presented. The programmed stability criterion is responsible for the underdetection. Special attention must be paid when it comes to programming this detection parameter. (Cardiol J 2008; 15: 281-283)