open access

Vol 15, No 3 (2008)
Case Reports
Submitted: 2013-01-14
Published online: 2008-04-14
Get Citation

Irregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device

Athanasios G. Manolis, Spyros Kourouklis, Dimitrios Chatzis, Konstantinos Kouvelas, Zenon Kyriakides
Cardiol J 2008;15(3):281-283.

open access

Vol 15, No 3 (2008)
Case Reports
Submitted: 2013-01-14
Published online: 2008-04-14

Abstract

A case of sustained monomorphic ventricular tachycardia underdetected by a single chamber implantable cardioverter defibrillator because of RR interval irregularity is presented. The programmed stability criterion is responsible for the underdetection. Special attention must be paid when it comes to programming this detection parameter. (Cardiol J 2008; 15: 281-283)

Abstract

A case of sustained monomorphic ventricular tachycardia underdetected by a single chamber implantable cardioverter defibrillator because of RR interval irregularity is presented. The programmed stability criterion is responsible for the underdetection. Special attention must be paid when it comes to programming this detection parameter. (Cardiol J 2008; 15: 281-283)
Get Citation

Keywords

irregular ventricular tachycardia; implantable cardioverter defibrillator; amiodarone

About this article
Title

Irregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device

Journal

Cardiology Journal

Issue

Vol 15, No 3 (2008)

Pages

281-283

Published online

2008-04-14

Page views

1005

Article views/downloads

1030

Bibliographic record

Cardiol J 2008;15(3):281-283.

Keywords

irregular ventricular tachycardia
implantable cardioverter defibrillator
amiodarone

Authors

Athanasios G. Manolis
Spyros Kourouklis
Dimitrios Chatzis
Konstantinos Kouvelas
Zenon Kyriakides

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.