Is the diagnostic function of pacemakers a reliable source of information about ventricular arrhythmias?

Magdalena Kumor, Rafał Baranowski, Edward Koźluk, Franciszek Walczak
DOI: 10.5603/cj.21337
·
Cardiol J 2010;17(5):495-502.

open access

Vol 17, No 5 (2010)
Original articles
Submitted: 2013-01-14
Published online: 2010-09-23

Abstract


Background: The aim of this study was to evaluate the reliability of pacemaker diagnostic function in diagnosing ventricular arrhythmias.
Methods: We compared the occurrence of ventricular ectopic beats in 51 simultaneous 24-hour electrocardiogram (ECG) recordings and pacemaker event counters printouts. The diagnostic function of a pacemaker allowed also for a qualitative assessment in 38 patients. In these cases, the occurrence of complex forms of ventricular arrhythmias was cross-checked for accelerated ventricular rhythms together with ventricular tachycardia, and triplets and couplets. The detection of at least one type of complex ventricular form of arrhythmia, diagnosed by both methods, was considered as an agreement between the methods.
Results: The results of ventricular ectopic beat counts differed significantly between the methods. In three (6%) patients, the results were consistent; in 20 (39%) the pacemaker underestimated results; in 28 (55%) they were overestimated. When more liberal criteria of agreement were applied, clinically significant differences were observed in 24 (47%) patients; in seven (29%) patients the count made by the pacemaker was lowered; and in 17 (71%) it was overestimated. Ventricular tachycardias were recorded in 24-hour ECG in eight patients. In three, they were identified by the pacemaker diagnostic function. In five, the pacemaker did not recognize tachycardia (because of its frequency being below 120/min). In nine, tachycardia was recognized falsely. The sensitivity in ventricular tachycardia diagnosis by pacemaker diagnostic function was 38%, specificity - 70%, the value of a positive result - 25%, negative - 81%.
Conclusions: The evaluation of ventricular arrhythmias by pacemaker cannot serve as the only reliable diagnostic method of arrhythmias. The presence of a large number of sequences that may correspond to ventricular arrhythmia or failure to sense, should result in verification via 24-hour ECG monitoring. (Cardiol J 2010; 17, 5: 495-502)

Abstract


Background: The aim of this study was to evaluate the reliability of pacemaker diagnostic function in diagnosing ventricular arrhythmias.
Methods: We compared the occurrence of ventricular ectopic beats in 51 simultaneous 24-hour electrocardiogram (ECG) recordings and pacemaker event counters printouts. The diagnostic function of a pacemaker allowed also for a qualitative assessment in 38 patients. In these cases, the occurrence of complex forms of ventricular arrhythmias was cross-checked for accelerated ventricular rhythms together with ventricular tachycardia, and triplets and couplets. The detection of at least one type of complex ventricular form of arrhythmia, diagnosed by both methods, was considered as an agreement between the methods.
Results: The results of ventricular ectopic beat counts differed significantly between the methods. In three (6%) patients, the results were consistent; in 20 (39%) the pacemaker underestimated results; in 28 (55%) they were overestimated. When more liberal criteria of agreement were applied, clinically significant differences were observed in 24 (47%) patients; in seven (29%) patients the count made by the pacemaker was lowered; and in 17 (71%) it was overestimated. Ventricular tachycardias were recorded in 24-hour ECG in eight patients. In three, they were identified by the pacemaker diagnostic function. In five, the pacemaker did not recognize tachycardia (because of its frequency being below 120/min). In nine, tachycardia was recognized falsely. The sensitivity in ventricular tachycardia diagnosis by pacemaker diagnostic function was 38%, specificity - 70%, the value of a positive result - 25%, negative - 81%.
Conclusions: The evaluation of ventricular arrhythmias by pacemaker cannot serve as the only reliable diagnostic method of arrhythmias. The presence of a large number of sequences that may correspond to ventricular arrhythmia or failure to sense, should result in verification via 24-hour ECG monitoring. (Cardiol J 2010; 17, 5: 495-502)
Get Citation

Keywords

pacemaker diagnostic functions; ventricular arrhythmias; 24-hour Holter ECG monitoring

About this article
Title

Is the diagnostic function of pacemakers a reliable source of information about ventricular arrhythmias?

Journal

Cardiology Journal

Issue

Vol 17, No 5 (2010)

Pages

495-502

Published online

2010-09-23

Page views

509

Article views/downloads

1174

DOI

10.5603/cj.21337

Bibliographic record

Cardiol J 2010;17(5):495-502.

Keywords

pacemaker diagnostic functions
ventricular arrhythmias
24-hour Holter ECG monitoring

Authors

Magdalena Kumor
Rafał Baranowski
Edward Koźluk
Franciszek Walczak

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., Grupa Via Medica, ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: viamedica@viamedica.pl