The effect of an antibacterial envelope on cardiac implantable device-related infection — A real-world analysis from a tertiary center
Abstract
Background: Infections related to cardiac implantable electronic devices (CIED) are associated with significant morbidity and mortality. Antibiotic-eluting envelopes have been introduced as a technology to prevent CIED infections. The aim of this study was to evaluate the effectiveness of the antibacterial envelope in the real-world population of a tertiary center.
Methods: This cohort study includes consecutively enrolled patients undergoing a device procedure from 01/2014 to 12/2020 at the University Hospital in Zurich. During period A (01/2014-12/2019) antibacterial envelopes were not used, whereas during period B (01/2020-12/2020) antibacterial envelopes were used in all device interventions. Follow-up was conducted by assessing all available patient records from patient visits and hospitalization.
Results: 1757 patients (male 70.5%, mean age 67.1 ± 16 years), were analyzed during a follow-up of 24 months. In 302 patients (17.2%) an antibacterial envelope was used. The overall occurrence of a device infection was low (n = 15, 0.85%). Factors that were associated with the incidence of an infection were not undergoing a primary implantation procedure (p = 0.024) and a CRT-P/D intervention (p = 0.022). There was no difference in the rate of infection between patients in whom a bacterial envelope was implanted vs. those in whom it was not used (0.6 vs. 0.9%, p = 0.693).
Conclusion: In a contemporary cohort of consecutive, unselected patients undergoing a device intervention at a large tertiary care center, the rate of device infection was low and not significantly different with vs. without the use of an antibacterial envelope. The data have important practical as well as economic implications for physicians performing such procedures.
Keywords: antibacterial envelopecardiac implantable electronic device
References
- Tarakji KG, Mittal S, Kennergren C, et al. Worldwide Randomized Antibiotic EnveloPe Infection PrevenTion Trial (WRAP-IT). Am Heart J. 2016; 180: 12–21.
- Wilkoff B, Boriani G, Mittal S, et al. Impact of Cardiac Implantable Electronic Device Infection. Circulation: Arrhythmia and Electrophysiology. 2020; 13(5).
- Boriani G, Kennergren C, Tarakji KG, et al. Cost-Effectiveness Analyses of an Absorbable Antibacterial Envelope for Use in Patients at Increased Risk of Cardiac Implantable Electronic Device Infection in Germany, Italy, and England. Value Health. 2021; 24(7): 930–938.
- Sohail MR, Henrikson CA, Braid-Forbes MJo, et al. Mortality and cost associated with cardiovascular implantable electronic device infections. Arch Intern Med. 2011; 171(20): 1821–1828.
- Uslan DZ, Sohail MR, St Sauver JL, et al. Permanent pacemaker and implantable cardioverter defibrillator infection: a population-based study. Arch Intern Med. 2007; 167(7): 669–675.
- Tarakji KG, Wilkoff BL, Tarakji KG, et al. WRAP-IT Investigators. Antibacterial Envelope to Prevent Cardiac Implantable Device Infection. N Engl J Med. 2019; 380(20): 1895–1905.
- Blomström-Lundqvist C, Traykov V, Erba PA, et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections-endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID) and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Europace. 2020; 22(4): 515–549.
- Polyzos KA, Konstantelias AA, Falagas ME. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace. 2015; 17(5): 767–777.
- Essebag V, Verma A, Healey JS, et al. BRUISE CONTROL Investigators. Clinically Significant Pocket Hematoma Increases Long-Term Risk of Device Infection: BRUISE CONTROL INFECTION Study. J Am Coll Cardiol. 2016; 67(11): 1300–1308.
- Tarakji KG, Korantzopoulos P, Philippon F, et al. Infectious consequences of hematoma from cardiac implantable electronic device procedures and the role of the antibiotic envelope: A WRAP-IT trial analysis. Heart Rhythm. 2021; 18(12): 2080–2086.
- Birnie DH, Wang J, Alings M, et al. Risk Factors for Infections Involving Cardiac Implanted Electronic Devices. J Am Coll Cardiol. 2019; 74(23): 2845–2854.
- Kleemann T, Becker T, Strauss M, et al. Prevalence of bacterial colonization of generator pockets in implantable cardioverter defibrillator patients without signs of infection undergoing generator replacement or lead revision. Europace. 2010; 12(1): 58–63.
- Olsen T, Jørgensen OD, Nielsen JC, et al. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982-2018). Eur Heart J. 2019; 40(23): 1862–1869.