Application of Monte Carlo calculations for validation of a treatment planning system in high dose rate brachytherapy

Alireza Naseri, Asghar Mesbahi
DOI: 10.1016/j.rpor.2009.12.003
·
Rep Pract Oncol Radiother 2009;14(6):200-204.

open access

Vol 14, No 6 (2009)
Published online: 2009-11-01
Submitted:

Abstract

Aim

The accuracy of treatment planning systems is of vital importance in treatment outcomes in brachytherapy. In the current study the accuracy of dose calculations of a high dose rate (HDR) brachytherapy treatment planning system (TPS) was validated using the Monte Carlo method.

Materials and methods

Three 60Co sources of the GZP6 afterloading brachytherapy system were modelled using MCNP4C Monte Carlo (MC) code. The dose distribution around all the sources was calculated by MC and a dedicated treatment planning system. The results of both methods were compared.

Results

There was good agreement (<2%) between TPS and MC calculated dose distributions except at a point near the sources (<1[[ce:hsp sp="0.25"/]]cm) and beyond the tip of the sources.

Conclusions

Our study confirmed the accuracy of TPS calculated dose distributions for clinical use in HDR brachytherapy.

Abstract

Aim

The accuracy of treatment planning systems is of vital importance in treatment outcomes in brachytherapy. In the current study the accuracy of dose calculations of a high dose rate (HDR) brachytherapy treatment planning system (TPS) was validated using the Monte Carlo method.

Materials and methods

Three 60Co sources of the GZP6 afterloading brachytherapy system were modelled using MCNP4C Monte Carlo (MC) code. The dose distribution around all the sources was calculated by MC and a dedicated treatment planning system. The results of both methods were compared.

Results

There was good agreement (<2%) between TPS and MC calculated dose distributions except at a point near the sources (<1[[ce:hsp sp="0.25"/]]cm) and beyond the tip of the sources.

Conclusions

Our study confirmed the accuracy of TPS calculated dose distributions for clinical use in HDR brachytherapy.

Get Citation

Keywords

Brachytherapy; High dose rate

About this article
Title

Application of Monte Carlo calculations for validation of a treatment planning system in high dose rate brachytherapy

Journal

Reports of Practical Oncology and Radiotherapy

Issue

Vol 14, No 6 (2009)

Pages

200-204

Published online

2009-11-01

DOI

10.1016/j.rpor.2009.12.003

Bibliographic record

Rep Pract Oncol Radiother 2009;14(6):200-204.

Keywords

Brachytherapy
High dose rate

Authors

Alireza Naseri
Asghar Mesbahi

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: journals@viamedica.pl