A Monte Carlo study for optimizing the detector of SPECT imaging using a XCAT human phantom

Mohammad Khoshakhlagh, Jalil Pirayesh Islamian, Seyyed Mohammad Abedi, Babak Mahmoudian, Masoud Shayesteh Azar
·
Pubmed: 28198517
·
Nucl. Med. Rev 2017;20(1):10-14.

open access

Vol 20, No 1 (2017)
Original articles
Submitted: 2016-02-19
Accepted: 2016-11-22
Published online: 2017-01-31

Abstract

BACKGROUND: Acquiring a high quality image has assigned an important concern for obtaining accurate diagnosis in nuclear medicine. Detector is a critical component of Single Photon Emission Computed Tomography (SPECT) imaging system for giving accurate information from exact pattern of radionuclide distribution in the target organ. The images are strongly affected by the attenuation, scattering, and response of the detector. The conventional detector is mainly made from sodium iodide activated by thallium [NaI(Tl)] in nuclear medicine imaging. The aim of the study. This study has planned to introduce a suitable for an optimized SPECT imaging. SIMIND Monte Carlo program was utilized for simulating a SPECT imaging system with a NaI(Tl) detector, and a low-energy high-resolution (LEHR) collimator.

MATERIAL AND METHODS: The Planar and SPECT scans of a 99mTc point source and also an extended Cardiac-Torso (XCAT) computerized phantom with the experiment and simulated systems were prepared. After verification and validation of the simulated system, the similar scans of the phantoms were compared from the point of view of image quality for 7 scintillator crystals including: NaI(Tl), BGO, YAG:Ce, YAP:Ce, LuAG:Ce, LaBr3 and CZT. The parameters of energy and spatial resolution, and sensitivity of the systems were compared. Images were analyzed quantitatively by SSIM algorithm with Zhou Wang and Rouse/Hemami methods, and also qualitatively by two nuclear medicine specialists.

RESULTS: Energy resolutions of the mentioned crystals obtained were: 9.864, 9.8545, 10.229, 10.221, 10.230, 10.131and10.223 percentage for 99mTc photopeak 140 Kev, respectively. Finally, SSIM indexes for the related phantom images were calculated to 0.794, 0.738, 0.735, 0.607, 0.760 and 0.811 compared to the NaI(Tl) acquired images, respectively. Medical diagnosis of the SPECT images of the phantom showed that the system with BGO crystal potentially provides a better detectability for hot and cold lesions in the liver of XCAT phantom.

CONCLUSIONS: The results showed that BGO crystal has a high sensitivity and resolution, and also provides a better lesion detectability from the point of view of image quality on XCAT phantom.

Abstract

BACKGROUND: Acquiring a high quality image has assigned an important concern for obtaining accurate diagnosis in nuclear medicine. Detector is a critical component of Single Photon Emission Computed Tomography (SPECT) imaging system for giving accurate information from exact pattern of radionuclide distribution in the target organ. The images are strongly affected by the attenuation, scattering, and response of the detector. The conventional detector is mainly made from sodium iodide activated by thallium [NaI(Tl)] in nuclear medicine imaging. The aim of the study. This study has planned to introduce a suitable for an optimized SPECT imaging. SIMIND Monte Carlo program was utilized for simulating a SPECT imaging system with a NaI(Tl) detector, and a low-energy high-resolution (LEHR) collimator.

MATERIAL AND METHODS: The Planar and SPECT scans of a 99mTc point source and also an extended Cardiac-Torso (XCAT) computerized phantom with the experiment and simulated systems were prepared. After verification and validation of the simulated system, the similar scans of the phantoms were compared from the point of view of image quality for 7 scintillator crystals including: NaI(Tl), BGO, YAG:Ce, YAP:Ce, LuAG:Ce, LaBr3 and CZT. The parameters of energy and spatial resolution, and sensitivity of the systems were compared. Images were analyzed quantitatively by SSIM algorithm with Zhou Wang and Rouse/Hemami methods, and also qualitatively by two nuclear medicine specialists.

RESULTS: Energy resolutions of the mentioned crystals obtained were: 9.864, 9.8545, 10.229, 10.221, 10.230, 10.131and10.223 percentage for 99mTc photopeak 140 Kev, respectively. Finally, SSIM indexes for the related phantom images were calculated to 0.794, 0.738, 0.735, 0.607, 0.760 and 0.811 compared to the NaI(Tl) acquired images, respectively. Medical diagnosis of the SPECT images of the phantom showed that the system with BGO crystal potentially provides a better detectability for hot and cold lesions in the liver of XCAT phantom.

CONCLUSIONS: The results showed that BGO crystal has a high sensitivity and resolution, and also provides a better lesion detectability from the point of view of image quality on XCAT phantom.

Get Citation

Keywords

gamma detector, SIMIND simulation, SSIM algorithm, XCAT phantom

About this article
Title

A Monte Carlo study for optimizing the detector of SPECT imaging using a XCAT human phantom

Journal

Nuclear Medicine Review

Issue

Vol 20, No 1 (2017)

Article type

Research paper

Pages

10-14

Published online

2017-01-31

Page views

1980

Article views/downloads

1244

DOI

10.5603/NMR.2017.0001

Pubmed

28198517

Bibliographic record

Nucl. Med. Rev 2017;20(1):10-14.

Keywords

gamma detector
SIMIND simulation
SSIM algorithm
XCAT phantom

Authors

Mohammad Khoshakhlagh
Jalil Pirayesh Islamian
Seyyed Mohammad Abedi
Babak Mahmoudian
Masoud Shayesteh Azar

References (26)
  1. Liu B, Shi C. Development of medical scintillator. J Chin Sci Bull. 2002; 47(13): 1057–1063.
  2. Melcher CL. Perspectives on the future development of new scintillators. Nucl Instrum Methods Phys Res A. 2005; 537(1-2): 6–14.
  3. Crystals.saint-gobain.com [homepage on the Internet]. BGO Bismuth germanates scintillation material. http://www.crystals.saint-gobain.com/uploadedFiles/SG-Crystals/Documents/BGO%20data%20sheet.pdf (Last accessed 2016 Oct 13).
  4. Rahmim A, Zaidi H. PET versus SPECT: strengths, limitations and challenges. Nucl Med Commun. 2008; 29(3): 193–207.
  5. Moszynski M, Kapusta M, Zalipska J, et al. Low energy γ-rays scintillation detection with large area avalanche photodiodes. IEEE Trans Nucl Sci. 1999; 46(4): 880–885.
  6. Chewpraditkul W, Swiderski L, Moszynski M, et al. Scintillation Properties of LuAG:Ce, YAG:Ce and LYSO:Ce Crystals for Gamma-Ray Detection. IEEE Transactions on Nuclear Science. 2009; 56(6): 3800–3805.
  7. Saizu MA, Cata-Danil G. Lanthanum Bromide Scintillation Detector for Gamma Spectrometry Applied in Internal Radioactive Contamination Measurements. UPB Sci Bull, Series A. 2011; 73(3): 119–126.
  8. Gehealthcare.com [homepage on the Internet]. CZT Technology: Fundamentals and Applications. http://www3.gehealthcare.com/en/products/categories/nuclear_medicine/cardiac_imaging/discovery_nm570c. (Last accessed 2016 Oct 13).
  9. Levin CS. New Imaging Technologies to Enhance the Molecular Sensitivity of Positron Emission Tomography. Proceedings of the IEEE. 2008; 96(3): 439–467.
  10. Eijk C. Inorganic scintillators in medical imaging. Phys Med Biol. 2002; 47(8): R85–R106.
  11. Dorenbos P. Light output and energy resolution of Ce3+-doped scintillators. Nucl Instrum Methods Phys Res A. 2002; 486(1-2): 208–213.
  12. Motta D, Schönert S. Optical properties of bialkali photocathodes. Nucl Instrum Methods Phys Res A. 2005; 539(1-2): 217–235.
  13. Data sheet of Siemens E.CAM Dual Head gamma camera. 1992. p.1-12.
  14. Vittori F, Notaristefani Fde, Malatesta T. Crystals and light collection in nuclear medicine. Nuclear Physics B - Proceedings Supplements. 1999; 78(1-3): 616–621.
  15. De Vries DJ, Moore SC, Zimmerman RE, et al. Development and validation of a Monte Carlo simulation of photon transport in an Anger camera. IEEE Trans Med Imaging. 1990; 9(4): 430–438.
  16. SIMIND Monte Carlo program. http://www.msf.lu.se/forskning/the-simind-monte-carlo-program (Last accessed 2016).
  17. Groch MW, Erwin WD. Single-photon emission computed tomography in the year 2001: instrumentation and quality control. J Nucl Med Technol. 2001; 29(1): 12–18.
  18. Islamian JP, Toossi MT, Momennezhad M, et al. Simulation of a Quality Control Jaszczak Phantom with SIMIND Monte Carlo and Adding the Phantom as an Accessory to the Program. Iran J Med Phys. 2012; 9(2): 135–140.
  19. Segars WP, Sturgeon G, Mendonca S, et al. 4D XCAT phantom for multimodality imaging research. Med Phys. 2010; 37(9): 4902–4915.
  20. Islamian JP, Toossi MT, Momennezhad M, et al. Monte Carlo study of the effect of collimator thickness on T-99m source response in single photon emission computed tomography. World J Nucl Med. 2012; 11(2): 70–74.
  21. Islamian JP, Toossi MB, Momennezhad M, et al. Monte Carlo Study of the Effect of Backscatter Material Thickness on 99mTc Source Response in Single Photon Emission Computed Tomography. Iran J Med Phys. 2013; 10(1): 69–77.
  22. Rouse D, Hemami S. Analyzing the role of visual structure in the recognition of natural image content with multi-scale SSIM. Human Vision and Electronic Imaging XIII. 2008.
  23. Wang Z, Bovik AC, Sheikh HR, et al. Image quality assessment: from error visibility to structural similarity. IEEE Trans Image Process. 2004; 13(4): 600–612.
  24. Khoshakhlagh M, Islamian JP, Abedi M, et al. A Study on Determination of an Optimized Detector for Single Photon Emission Computed Tomography. World J Nucl Med. 2016; 15(1): 12–17.
  25. Toossi MB, Islamian PJ, Momennezhad M, Ljungberg M, Naseri S. SIMIND Monte Carlo simulation of a single photon emission CT. J Med Phys 2010; 35(1): 42–7.
  26. Derenzo S, Budinger T, Huesman R, et al. Imaging Properties of a Positron Tomograph with 280 Bgo Crystals. IEEE Trans Nucl Sci. 1981; 28(1): 81–89.

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., Świętokrzyska 73 street, 80–180 Gdańsk, Poland

phone: +48 58 320 94 94, fax: +48 58 320 94 60, e-mail: viamedica@viamedica.pl