A 73-year-old male was admitted to the Cardiology Ward from the Emergency Department after an episode of syncope for further evaluation and treatment. The patient had a history of hypertension and recurrent presyncope episodes. On admission, he was in stable condition with no significant findings on physical examination. Electrocardiography (ECG) recorded after the syncope showed sinus bradycardia with junctional escape rhythm and a heart rate of 38 beats per minute. There were no other significant findings on 24-hour Holter monitoring. The patient was diagnosed with symptomatic sinus node dysfunction and scheduled for dual-chamber pacemaker implantation the next day. The procedure was carried out routinely with a 52 cm right atrial lead placed in the right atrial appendage with a capture threshold of 1 V at 0.4 ms, sensing at 3 mV, and impedance of 628 Ω, and a 58 cm ventricular lead placed in the right ventricular outflow tract with a capture threshold of 0.5 V at 0.4 ms, sensing at >12 mV and impedance of 640 Ω. Lead placement was confirmed with two independent fluoroscopy angles. Within minutes after placement of the ventricular lead, the patient developed chest pain and became hypotensive with a sudden drop in blood pressure to 60/30 mm Hg. Pneumothorax and cardiac tamponade were ruled out using fluoroscopy in 30° left anterior oblique view and intraprocedural transthoracic echocardiography. The patient’s status quickly improved after administration of inotropes and intravenous fluids and the symptoms subsided; the procedure was completed without further complications. After obtaining a post-procedure 12-lead ECG, ST-segment elevation myocardial infarction was confirmed (Figure 1A). Emergency coronary angiography showed critical stenosis of the left anterior descending artery with compromised distal flow, of the diagonal branch, and significant stenosis of the circumflex artery. Angioplasty of the left anterior descending artery and the diagonal branch was performed using two drug-eluting stents; the patient was scheduled for elective circumflex artery angioplasty before discharge.
Due to the complicated pacemaker implantation course and the ECG suggesting epicardial pacing, suspicion of direct injury to the coronary artery was raised. After careful re-evaluation of coronary angiography, a direct left anterior descending artery perforation by the ventricular lead helix was confirmed (Figure 1B, D; Supplementary material, Video S1–S3). The lead was replaced without further complications, and scheduled angioplasty was carried out as planned. No significant pericardial effusion was present on control echocardiography following the procedure. After follow-up, the patient was discharged home.
Myocardial infarction due to coronary artery injury is a rare complication of cardiac electronic device implantation [1]. There were only several recorded cases of left anterior descending artery injury caused by a ventricular lead [2–5]. In the presented case, the patient developed symptoms of acute myocardial infarction due to inadvertent malposition of the ventricular lead in the right ventricle anteroseptal junction leading to coronary artery perforation by the lead helix, and he required urgent revascularization and lead replacement. Routine use of 12-lead ECG and right anterior oblique (RAO) imaging during implantation should be considered to avoid this high-risk pseudo-septal position and potential complications. Utilization of novel imaging algorithms such as iterative metal artifact reduction for computed tomography might be helpful for better evaluation of the perforated leads [6].
Supplementary material
Supplementary material is available at https://journals.viamedica.pl/kardiologia_polska.
Article information
Conflict of interest: None declared.
Funding: None.
Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, which allows downloading and sharing articles with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.