Clinical vignette

Impella-supported high-risk percutaneous coronary intervention complicated by a stuck pump and somersault in the aorta

Sylwia IwańczykWłodzimierz SkorupskiPatrycja WoźniakAnna Smukowska-GoryniaMarek GrygierMaciej Lesiak
1st Department of Cardiology, Poznan University of Medical Sciences, Poznań, Poland

Correspondence to:

Sylwia Iwańczyk, MD,

1st Department of Cardiology,

Poznan University of Medical Sciences,

Długa 1/2,, 61–848 Poznań, Poland,

phone: +48 61 854 92 22,

e-mail: syl.iwanczyk@gmail.com

Copyright by the Author(s), 2022

DOI: 10.33963/KP.a2022.0210

Received: July 5, 2022

Accepted: August 1, 2022

Early publication date: September 10, 2022

A 65-year-old man with non-ST-segment elevation myocardial infarction and three-vessel disease was admitted for further invasive treatment. His concomitant diseases included chronic heart failure, hypertension, hypercholesterolemia, and type 2 diabetes mellitus. Echocardiography showed hypokinesis of the inferior and anterior walls with reduced left ventricular ejection fraction of 35%. Coronary angiography showed a chronically occluded right coronary artery and critical left

main (LM) bifurcation stenosis (Medina 0,1,1) (Supplementary material, Videos S1S3). The local Heart Team considered the best therapeutic approach. Due to high operative risk (SYNTAX score, 47.0 points; EuroSCORE II, 2.65%) and coronary arteries incompatible with grafting, the patient was disqualified from coronary artery bypass grafting and qualified for high-risk rescue percutaneous coronary intervention (PCI) with the use of Impella CP (Abiomed, Denver, CO, US) support.

The Impella CP was successfully introduced into the left ventricle (LV) via the right femoral approach. The catheter needed to be repositioned twice before being placed in the optimal position (Figure 1A). When removing the guide wire, the operator noticed that the wire was trapped inside the lumen of the catheter. Following the applied traction, the Impella and trapped guidewire was moved back to the aortic arch and the descending aorta, with simultaneous flexion just distal to the pump motor and further down the catheter (Figure 1B). To straighten the kinked Impella catheter, we used a 6 F 1220 mm Atrieve Vascular Snare™ through the right radial approach. After several attempts, the distal part of the pump was captured by the vascular snare (Figure 1C). The kinks were straightened by gently pulling both the snare and Impella catheter, and the pump was retrieved (Supplementary material, Video S4). Due to concerns about the efficiency of Impella, we decided to continue the procedure without protection, allowing the use of the new device as a bailout.

PCI was performed using the 14 F Impella CP sheath with the EBU 3.5 Guide Catheter (7 F). Because of severe calcifications, we performed a rotablation of the LM along with the proximal and middle segments of the left descending artery (LAD). Two drug-eluting stents (DESs), 3.0 × 12 mm and 2.5 × 28 mm, were successfully implanted in the LAD. LM bifurcation was treated with 3.5 × 34 mm DES implantation by the provisional technique. The procedure was completed with the proximal optimization technique (POT) with a 5.0 mm NC balloon. The optimal final angiographic result was obtained and confirmed on the IVUS (Figure 1DE). The patient remained hemodynamically stable throughout the procedure.

Figure 1. A. Angiographic view of the Impella CP placed in the left ventricle. B. Angiographic view of the Impella and guidewire displaced in the descending aorta, with flexion just behind the pump and further down the catheter (white arrows). C. Angiographic view of the distal part of the pump captured by the vascular snare (red arrow). D The final angiographic result. E. Final IVUS cross-sections of the left main (LM) and anterior descending artery (LAD) with a well-expanded stent

Complex PCI in a patient with poor LV function is one of the indications for the use of Impella CP [1, 2]. In the presented case, the procedure was successfully performed without hemodynamic support, mainly due to the operators’ experience and only moderate LV dysfunction.

A possible cause of problems with retrieving the guidewire may be its distal kink and damage that occurred during Impella’s repositioning, hence we suggest using a new guidewire in such cases. The presented case shows also that snaring can be effectively and safely used to capture and reposition or remove kinked Impella catheters [3].

Supplementary material

Supplementary material is available at https://journals.viamedica.pl/kardiologia_polska.

Article information

Conflict of interest: None declared.

Funding: None.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.