Vol 80, No 12 (2022)
Original article
Published online: 2022-09-02

open access

Page views 3343
Article views/downloads 424
Get Citation

Connect on Social Media

Connect on Social Media

Eligibility of patients with temporary paced rhythm for a subcutaneous implantable cardioverter-defibrillator

Barbara Opielowska-Nowak1, Maciej Kempa1, Szymon Budrejko1, Grzegorz Sławiński1, Grzegorz Raczak1
Pubmed: 36052564
Kardiol Pol 2022;80(12):1231-1237.

Abstract

Background: A concomitant use of a pacemaker and a subcutaneous implantable cardioverter-de-fibrillator (S-ICD) may be required in some patients.
Aims: Our study aimed to evaluate the influence of permanent cardiac pacing on the morphology of the QRS complex in the context of S-ICD screening.
Methods: One hundred patients with cardiac electronic implantable devices (CIEDs) were included in whom S-ICD screening could be performed both during intrinsic and paced rhythm.
Results: The positive result of screening during spontaneous rhythm for at least one vector (in both supine and standing positions) was obtained in 80% and for 2 vectors in 59% of patients. Positive screening during paced rhythm for at least one vector was recorded in 36% of patients (78% right ventricular and 22% biventricular pacing) and for 2 vectors in 15% of patients (93% right ventricular and 7% biventricular pacing). At least one vector acceptable during both types of rhythm and in both positions was recorded in 23% of patients and at least 2 vectors in 8% of patients.
Conclusions: The use of S-ICD in patients with paced ventricular rhythm is associated with a serious risk of inappropriate sensing due to different QRS morphology during intrinsic and paced rhythm, and it is particularly high in patients in whom periods of spontaneous rhythm interchange with periods of ventricular pacing. That risk has been hardly acknowledged in available reports, but according to our data, it is significant, and therefore it should be considered during S-ICD screening.

Original article

Eligibility of patients with temporary paced rhythm for a subcutaneous implantable cardioverter-defibrillator

Barbara Opielowska-NowakMaciej KempaSzymon BudrejkoGrzegorz SławińskiGrzegorz Raczak
Department of Cardiology and Electrotherapy, Medical University of Gdansk, Gdańsk, Poland

Correspondence to:

Maciej Kempa, MD, PhD,

Department of Cardiology and Electrotherapy,

Medical University of Gdansk,

Smoluchowskiego 17, 80–214 Gdańsk, Poland,

phone: +48 58 584 47 60,

e-mail: kempa@gumed.edu.pl

Copyright by the Author(s), 2022

DOI: 10.33963/KP.a2022.0205

Received: March 24, 2022

Accepted: August 30, 2022

Early publication date: September 2, 2022

ABSTRACT
Background: A concomitant use of a pacemaker and a subcutaneous implantable cardioverter-defibrillator (S-ICD) may be required in some patients.
Aims: Our study aimed to evaluate the influence of permanent cardiac pacing on the morphology of the QRS complex in the context of S-ICD screening.
Methods: One hundred patients with cardiac electronic implantable devices (CIEDs) were included in whom S-ICD screening could be performed both during intrinsic and paced rhythm.
Results: The positive result of screening during spontaneous rhythm for at least one vector (in both supine and standing positions) was obtained in 80% and for 2 vectors in 59% of patients. Positive screening during paced rhythm for at least one vector was recorded in 36% of patients (78% right ventricular and 22% biventricular pacing) and for 2 vectors in 15% of patients (93% right ventricular and 7% biventricular pacing). At least one vector acceptable during both types of rhythm and in both positions was recorded in 23% of patients and at least 2 vectors in 8% of patients.
Conclusions: The use of S-ICD in patients with paced ventricular rhythm is associated with a serious risk of inappropriate sensing due to different QRS morphology during intrinsic and paced rhythm, and it is particularly high in patients in whom periods of spontaneous rhythm interchange with periods of ventricular pacing. That risk has been hardly acknowledged in available reports, but according to our data, it is significant, and therefore it should be considered during S-ICD screening.
Key words: cardiac pacing, implantable cardioverter-defibrillator, subcutaneous implantable cardioverter-defibrillator, sudden cardiac arrest, sudden cardiac death

WHAT’S NEW?

In patients with paced ventricular rhythm the risk of inappropriate sensing by a coexisting subcutaneous implantable cardioverter-defibrillator is high, especially if periods of spontaneous rhythm and ventricular pacing interchange constantly. Following the results of our study we postulate that the issue is significant, it has been underestimated in the available reports, and it definitely should be considered during subcutaneous implantable cardioverter-defibrillator screening.

INTRODUCTION

A subcutaneous implantable cardioverter-defibrillator (S-ICD) has recently become a recognized method of treatment used for prevention of sudden cardiac death [1]. The S-ICD system detects ventricular arrhythmias based on the analysis of one of three available electrocardiogram (ECG) vectors. Those vectors are recorded between either the upper or lower sensing ring on the lead located along the left margin of the sternum and the device can or between those two rings without involvement of the device can. Sensing signals of appropriate quality are required for accurate operation of the system. Therefore, a patient qualified for implantation has to undergo a dedicated ECG test (the so-called ECG screening) that is intended to ensure the correct detection of the cardiac rhythm by the device. Screening is performed with the use of specialized software provided by the manufacturer. It relies on the automated analysis of the ECG signal recorded from the surface of the patient’s chest, using vectors similar to the predicted location of the lead and can of the S-ICD system. At least one of the three available vectors should be acceptable to allow for implantation of S-ICD, but some authors require at least two acceptable vectors to consider screening positive.

The S-ICD system has become widely used in current clinical practice although its use is still limited by the inability to perform permanent cardiac pacing and cardiac resynchronization therapy. As a consequence, the need for permanent cardiac pacing is one of contraindications for S-ICD therapy [1]. But the indication for pacing may develop later even if it was not present at the time of S-ICD implantation. In such a situation, the change of QRS morphology during paced rhythm may potentially lead to inappropriate sensing of cardiac activity by the S-ICD system. A similar problem occurs in a patient with an implanted pacemaker who develops indications for an implantable cardioverter-defibrillator (ICD). If the patient does not give their consent to implantation of a transvenous ICD or there is no possibility to perform such a procedure (that requires transvenous extraction of the existing right ventricular pacing lead with all its risks), S-ICD implantation may become one of the options. But then, again, the transient changes of the QRS morphology during spontaneous and paced rhythm may lead to inappropriate sensing and inadequate interventions. And last but not least, infective complications may require extraction of the transvenous ICD system and implantation of an epicardial pacemaker in pacing-dependent patients, which ceases protection against ventricular arrhythmias. The S-ICD system implanted in addition to the epicardial pacemaker might be a solution in such a complex case. At least one sensing vector acceptable both during intrinsic and paced rhythm should be confirmed before the decision is made to use S-ICD together with a pacemaker. Notably, it has to be the same vector for both types of rhythm as the S-ICD system cannot adjust the sensing vector automatically to the changing rhythm and QRS morphology when the intrinsic and paced rhythm are constantly interchanging. The sensing vector can be altered only by a physician during the follow-up procedure. Our study aimed to evaluate the influence of permanent cardiac pacing on the morphology of the QRS complex in the context of S-ICD screening and on the possibility of concomitant use of S-ICD and a permanent pacemaker.

METHODS

The study was designed to include 100 consecutive patients hospitalized in the Department of Cardiology and Electrotherapy of the Medical University of Gdańsk, Poland who had just undergone implantation of cardiac electronic implantable devices (CIEDs) due to sick sinus syndrome, atrial fibrillation with bradycardia, or heart failure. We collected data regarding demographical parameters, rates of concomitant diseases, left ventricular ejection fraction (based on echocardiography), cardiac rhythm and pacing mode of the implanted device, and registered standard ECG for assessment of cardiac rhythm and measurement of standard electrocardiographic parameters. Data collection was performed between July and December 2021. We included patients in whom it was possible to record both spontaneous rhythm with intrinsic conduction to the ventricles and paced ventricular rhythm forced by the implanted device in any mode of ventricular pacing (DDD, VVI, or biventricular). Patients with an advanced atrioventricular block and ventricular escape rhythm (or no escape rhythm at all) were not qualified for the study. The eligibility screening for S-ICD was performed with the Boston Scientific programmer and EMBLEM™ automated screening tool software within 5 days from implantation of CIED. The ECG signal was recorded for 3 vectors: primary (the proximal pole on the lead [on the left margin of the xiphoid process] to the device can [in the position of ECG lead V6]), secondary (the distal pole on the lead [14 cm above the proximal pole, on the left margin of the sternomanubrium junction] to the device can), and alternate (the distal pole to the proximal pole on the lead). Only the standard set of vectors and typical positioning were performed in our study. Body surface ECG was recorded for those 3 vectors in supine and standing positions both during intrinsic rhythm and ventricular pacing. To record the intrinsic rhythm, the implanted device was set to VVI mode with a basic rate of 30 bpm. To force ventricular pacing, the device was set to DDD or VVI mode with a basic rate of 10 bpm more than the intrinsic rhythm, and in the case of DDD or cardiac resynchronization therapy (CRT) devices with the atrioventricular delay time short enough to force ventricular pacing (interventricular delay was set to 0 ms in all CRT devices). Since experts hold different views on the definition of positive screening (one or two passing vectors), both those situations were analyzed, as is stated and underlined in this article at every occurrence of that issue. We analyzed the percentages of positive screening during intrinsic and paced rhythm and then by groups divided according to the mode of pacing (right ventricular pacing in comparison with biventricular pacing). And finally, we planned a comparison of patients with positive screening during spontaneous rhythm (at least 2 vectors acceptable, both supine and standing positions) divided into subgroups with either negative or positive results of screening during paced rhythm. That analysis was intended to include demographical variables (age, sex) and clinical variables (heart rate, electrocardiographic parameters, left ventricular ejection fraction, mode of pacing, and rates of concomitant diseases). The study design was approved by the Ethical Board at the Medical University of Gdańsk, Poland.

Statistical analysis

Continuous variables were presented as mean and standard deviation or median and interquartile range in case of non-normal distribution. Categorical parameters were presented as numbers and percentages. The normality of distribution was tested with the Shapiro-Wilk test. The χ2 test and Student’s t-test or the Mann-Whitney U test (depending on the analysis of distribution and variance) were used to compare the groups, as appropriate for a given variable. A P value below 0.05 was considered statistically significant. Data management and statistical analysis were performed with Microsoft Excel and Statistica 13.1 software (TIBCO Software, Palo Alto, CA, US).

RESULTS

One hundred consecutive patients with a pacemaker or ICD were included in the study group. Clinical data of the patients are summarized in Table 1. All the right ventricular leads were in the apical position.

Table 1. Clinical data of patients in the study group

Total number, n (%)

100 (100)

Male, n (%)

65 (65)

Age, mean (SD), range, years

73 (12), 28–94

Coronary artery disease, n (%)

42 (42)

Non-ischemic cardiomyopathy, n (%)

19 (19)

Heart failure, n (%)

62 (62)

LVEF, median (IQR), %

40 (27–55)

Sinus rhythm, n (%)

69 (69)

Atrial fibrillation, n (%)

31 (31)

Type of the implanted device, n (%)

VVI pacemaker

8 (8)

DDD pacemaker

43 (43)

ICD single-chamber

14 (14)

ICD dual-chamber

6 (6)

CRT pacemaker

5 (5)

CRT defibrillator

24 (24)

Screening during spontaneous rhythm (implanted device inactive)

Data regarding the number of vectors acceptable for S-ICD implantation are presented in Figures 1 and 2. The positive result of screening, if at least one acceptable vector was required (both in supine and standing positions), was eventually obtained in 80 patients (80%) and if two positive vectors were required in 59 patients (59%).

Screening during paced ventricular rhythm forced by the implanted device

Data on the number of vectors acceptable for S-ICD implantation are presented in Figures 1 and 2.

Figure 1. Percentages of patients with different numbers of vectors acceptable in S-ICD screening during spontaneous and paced rhythm
Abbreviation: S-ICD, subcutaneous implantable cardioverter-defibri­llator
Figure 2. Percentages of vectors acceptable in S-ICD screening during spontaneous and paced rhythm
Abbreviation: see Figure 1

The positive result of screening if at least one acceptable vector was required (both in supine and standing positions) was eventually obtained in 36 patients (36%). In that subgroup, 28 patients (78%) had right ventricular (RV) pacing and 8 (22%) biventricular (BiV) pacing. But 8 of those 36 patients (22%) had none of the vectors acceptable during spontaneous rhythm. If two positive vectors were required, the positive result of screening was achieved in 15 patients (15%). In that subgroup, RV pacing was present in 14 cases (93%) and BiV pacing in one case (7%).

In the group of 100 patients, we obtained at least one acceptable sensing vector during both spontaneous and paced rhythm only in 28 cases (28%). Furthermore, in 5 patients, it was not the same vector for those two types of rhythm. Eventually, only in 23 cases of the initial 100 (23%), we managed to find at least one vector acceptable during both types of rhythm (the same vector in both situations) and in both body positions (5 patients with BiV and 18 with RV pacing). If 2 vectors were required for positive screening (the same 2 vectors for both rhythms and both body positions), the final positive results of screening were obtained only in 8 patients (all with RV pacing).

In the context of pacing modality, 29 patients (29%) had BiV pacing, and 71 (71%) RV pacing. If one acceptable vector was enough, 5 patients of 29 with BiV pacing (17.2%) and 18 of 71 with RV pacing (25.4%) could be considered to have a positive result of screening (P = 0.38 for the difference). If two vectors were required, none of the patients with BiV pacing (0%) and 8 with RV pacing (11%) could be considered positive (P = 0.06). Therefore, the type of pacing did not influence the chance of having a positive result of screening with either one or two vectors required.

While analyzing which of the vectors were positive during spontaneous and paced rhythm, we found that it was predominantly the primary vector in both situations (65% and 21%, respectively, see Figure 2).

Then we analyzed the variance among patients with positive screening according to the criteria used typically in our department (two passing vectors) during spontaneous rhythm, dividing them into subgroups with the negative (group 1) or positive (group 2) results of screening during paced rhythm. Variables in that analysis included age, sex, left ventricular ejection fraction (LVEF), underlying cardiac disease, history of chronic heart failure, and electrocardiographic measurements (the width of paced QRS complex and intrinsic QRS complex, the increment of QRS width with pacing, the QT interval, and the presence of right or left bundle branch block). Only LVEF was significantly different between those subgroups, and it was 40% (25%–50%) in group 1 and 55.5% (50%–60%) in group 2. The results are presented in Table 2.

Table 2. Comparison of patients with positive screening results during spontaneous rhythm (at least 2 vectors acceptable, both supine and standing positions) divided into subgroups with either the negative (group 1) or positive (group 2) results of screening during paced rhythm

Variable

Group 1

(n = 51)

Group 2

(n = 8)

P-value

Age, median (IQR), years

73 (68–82)

74.5 (71.5–80)

0.89

Male sex, n (%)

30 (58.82)

5 (62.5)

0.84

Heart rate, median (IQR), bpm

70 (61–82)

62 (52–70.5)

0.08

Intrinsic QRS, median (IQR), ms

109 (90–160)

128.5 (100–150)

0.63

Paced QRS, median (IQR), ms

160 (150–170)

175 (150.5–189.5)

0.13

Difference between paced and intrinsic QRS width, median (IQR), ms

54 (0–72)

48.5 (31.5–63.5)

0.93

QT, mean (SD), ms

424.39 (56.06)

422.75 (46.19)

0.96

QTc, mean (SD), ms

455.86 (48.61)

424.75 (34.59)

0.06

LVEF, median (IQR), %

40 (25–50)

55.5 (50–60)

0.01

CRT, n (%)

20 (39.22)

0 (0)

0.03

AF, n (%)

17 (33.33)

2 (25)

0.64

RBBB, n (%)

4 (7.84)

2 (25)

0.14

LBBB, n (%)

13 (25.49)

0 (0)

0.11

IVCD, n (%)

5 (9.8)

2 (25)

0.28

ICM, n (%)

23 (45.1)

1 (12)

0.08

HA, n (%)

37 (72.55)

6 (75)

0.89

DCM, n (%)

13 (25.49)

0 (0)

0.11

CHF, n (%)

33 (64.71)

3 (37.5)

0.14

DISCUSSION

The inability to provide permanent cardiac pacing is one of the major limitations of the S-ICD system. The predicted rate of the need for pacemaker implantation was found to be between 2% and 6.8% per year of follow-up in numerous studies of patients with transvenous ICDs, and the rate of the need for CRT between 0.6% and 0.8% per year [2–5], but it should not be directly extrapolated to populations of potential S-ICD recipients. The reported risk of developing indications for permanent cardiac pacing in real-life populations of patients with implanted S-ICD systems is lower. In the analysis of early cumulative results of the EFFORTLESS and IDE studies, the need for permanent cardiac pacing occurred only in 2 of 889 patients during 22 months of follow-up [6]. In another report from Germany, the low risk of such a scenario was confirmed. In 28 patients, no need for pacemaker implantation was reported during follow-up until S-ICD battery depletion [7]. Finally, in recently published results of the prolonged follow-up of the EFFORTLESS study population (median implant duration 5.1 years), the need for conversion from S-ICD to transvenous ICD due to indications for cardiac pacing occurred only in 13 of 984 patients [8].

Therefore, the risk that a patient with pre-existing S-ICD will need a permanent cardiac pacemaker is low, yet not negligible. The opposite scenario seems more probable, in which a patient with a pre-existing cardiac pacemaker develops heart failure with reduced LVEF and, therefore, an indication for ICD in primary prevention of sudden cardiac death. In a study by Khurshid et al., [9] a decrease in mean LVEF from 62.1% to 36.2% over a mean follow-up period of 3.3 years occurred in 19.5% of the study population.

Two solutions for such a problem are available. One is to upgrade the pacemaker to transvenous ICD or CRT-D, possibly after transvenous extraction of the right ventricular pacing lead. The second solution is to implant an S-ICD system as a companion to the existing pacemaker.

Data on the concomitant use of pacing systems and S-ICDs are limited although such a solution has been successfully used and reported. Reports are available of S-ICD systems co-existing with both transvenous and epicardial pacemakers [10–13]. Moreover, in isolated cases, the S-ICD system was used together with a leadless pacemaker [14–16]. On the other hand, several cases were reported where the implanted pacemaker changed QRS morphology to such an extent that continuation of S-ICD therapy was not possible [17]. Therefore, the concomitant use of pacemakers and S-ICD systems is associated with a significant risk of undesired interactions between those devices. Careful programming may reduce the risk of such interactions. When programming a pacemaker, it is recommended to use low pulse amplitudes with minimal safety margins, as well as to turn off the automatic threshold and automatic switch-of-polarity functions. In DDD and CRT devices, the upper rate limit should be set lower than half of the first therapy zone in S-ICD [18]. But that recommended programming algorithm cannot completely eliminate the problem of unacceptable QRS morphology change due to ventricular pacing, which may preclude appropriate QRS sensing by the S-ICD device.

To evaluate the significance of that phenomenon, we analyzed data acquired from 100 consecutive patients in whom the ventricular rhythm was forced in the form of RV or BiV pacing. In that group, screening in only 36 patients (36%; 28 RV, 8 BiV) was acceptable for at least one vector, and in 15 cases (15%) for at least 2 vectors (14 RV, 1 BiV). The analysis of the results of screening by the mode of pacing showed that in patients with BiV pacing at least one vector was acceptable in 8 of 29 patients (28%) and with RV pacing in 28 of 71 patients (39%). Those values are significantly lower than the rates reported by Ip et al. [19]. Those authors reported positive screening in 80% of patients with biventricular pacing and 46% of patients with RV pacing. That analysis was performed manually using the Boston Scientific screening templates while in our population, the automated screening was performed.

There are other reports on the influence of permanent cardiac pacing on the QRS morphology and the impact of that phenomenon on S-ICD screening. Those reports were based on populations of patients with CRT or His bundle pacing. In those reports, S-ICD screening was acceptable in most patients with cardiac pacing, contrary to our results. The rates of positive screening were 82% to 85% for BiV pacing and 90% for His bundle pacing [20–22]. But the authors of those publications analyzed only the results of screening in paced rhythm and did not include the possible temporary change of rhythm for the intrinsic one. Such an event may occur in a setting of transient atrioventricular conduction disturbances, supraventricular tachycardia, or ineffective ventricular pacing. In our study group, in 36 of 100 patients (36%), we could find at least one acceptable vector during ventricular pacing, but in 8 patients of those 36, none of the vectors was acceptable during intrinsic activation when ventricular pacing was switched off. Therefore, those patients, if equipped with an S-ICD, would be at risk of inappropriate interventions in the case of reoccurrence of the intrinsic rhythm. In the subsequent analysis of the remaining 28 patients, we noticed that in 5 of them the vectors acceptable during spontaneous activation and ventricular pacing were different. The S-ICD system cannot automatically change the sensing vector depending on the type of ventricular activation (spontaneous versus paced). Therefore, those 5 patients would also be at risk of inappropriate interventions. Only 23 patients (23%) could be eventually deemed as having the positive result of screening with minimal requirements, that is at least one vector acceptable both during spontaneous rhythm and ventricular pacing and in both supine and standing positions. Assuming that the reasonable number of acceptable vectors to guarantee long-term safety is 2 (the same vectors in both body positions and both types of cardiac rhythm), the number of patients meeting such restricted criteria was 8 (8%). The phenomenon of the QRS morphology change between spontaneous and paced rhythm and its influence on S-ICD screening, was reported by Giammaria et al. [23]. In the group of 48 patients with biventricular pacing, at least one vector was acceptable in 34 patients (71%). However when pacing was switched off, that number was reduced to 22 (46%) during intrinsic ventricular activation.

Limitations of the study

A limited number of patients included in the study and a relatively low number of CRT recipients (because we included consecutive patients undergoing implantation of CIEDs) resulted in small subgroups, which restricted statistical analysis (especially for comparisons of proportions). We did not analyze the relationship between the position of LV leads and the result of screening, as it would further subdivide those relatively small groups.

CONCLUSIONS

The use of S-ICD is associated with a serious risk of inappropriate sensing in patients with other cardiac implantable electronic devices in whom periods of spontaneous rhythm interchange with periods of ventricular pacing. If a pacemaker is implanted first, analysis of such a risk is possible and requires repeated screening in both paced and intrinsic rhythms. However if the S-ICD system is implanted first, it is very difficult to predict the possible risk of inappropriate sensing because screening for the paced rhythm cannot be performed before the pacemaker implantation.

Article information

Conflict of interest: None declared.

Funding: None.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.

REFERENCES

  1. Priori SG, Blomström-Lundqvist C, Mazzanti A, et al. 2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: The Task Force for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death of the European Society of Cardiology (ESC). Eur Heart J. 2015; 36(41): 27932867, doi: 10.1093/eurheartj/ehv316, indexed in Pubmed: 26320108.
  2. de Bie MK, Thijssen J, van Rees JB, et al. Suitability for subcutaneous defibrillator implantation: results based on data from routine clinical practice. Heart. 2013; 99(14): 10181023, doi: 10.1136/heartjnl-2012-303349, indexed in Pubmed: 23704324.
  3. Melles MC, Yap SC, Bhagwandien RE, et al. Frequency of need for antitachycardia or antibradycardia pacing or cardiac resynchronization therapy in patients with a single-chamber implantable cardioverter-defibrillator. Am J Cardiol. 2018; 122(12): 20682074, doi: 10.1016/j.amjcard.2018.08.059, indexed in Pubmed: 30340766.
  4. Grett M, Christ M, Trappe HJ. Predictors for need of antibradycardia and antitachycardia pacing after ICD implantation : Implications for the subcutaneous ICD [article in German]. Herzschrittmacherther Elektrophysiol. 2016; 27(4): 390395, doi: 10.1007/s00399-016-0460-4, indexed in Pubmed: 27738753.
  5. Kutyifa V, Rosero SZ, McNitt S, et al. Need for pacing in patients who qualify for an implantable cardioverter-defibrillator: Clinical implications for the subcutaneous ICD. Ann Noninvasive Electrocardiol. 2020; 25(4): e12744, doi: 10.1111/anec.12744, indexed in Pubmed: 31994819.
  6. Burke MC, Gold MR, Knight BP, et al. Safety and efficacy of the totally subcutaneous implantable defibrillator: 2-year results from a pooled analysis of the ide study and effortless registry. J Am Coll Cardiol. 2015; 65(16): 16051615, doi: 10.1016/j.jacc.2015.02.047, indexed in Pubmed: 25908064.
  7. Bettin M, Rath B, Ellermann C, et al. Follow-up of the first patients with a totally subcutaneous ICD in Germany from implantation till battery depletion. Clin Res Cardiol. 2019; 108(1): 1621, doi: 10.1007/s00392-018-1296-1, indexed in Pubmed: 29948287.
  8. Lambiase PD, Theuns DA, Murgatroyd F, et al. Subcutaneous implantable cardioverter-defibrillators: long-term results of the EFFORTLESS study. Eur Heart J. 2022; 43(21): 20372050, doi: 10.1093/eurheartj/ehab921, indexed in Pubmed: 35090007.
  9. Khurshid S, Liang JJ, Owens A, et al. Incidence and predictors of right ventricular pacing-induced cardiomyopathy. Heart Rhythm. 2014; 11(9): 16191625, doi: 10.1016/j.hrthm.2014.05.040, indexed in Pubmed: 24893122.
  10. Porterfield C, DiMarco JP, Mason PK. Effectiveness of implantation of a subcutaneous implantable cardioverter-defibrillator in a patient with complete heart block and a pacemaker. Am J Cardiol. 2015; 115(2): 276278, doi: 10.1016/j.amjcard.2014.10.036, indexed in Pubmed: 25465940.
  11. Kuschyk J, Stach K, Tülümen E, et al. Subcutaneous implantable cardioverter-defibrillator: First single-center experience with other cardiac implantable electronic devices. Heart Rhythm. 2015; 12(11): 22302238, doi: 10.1016/j.hrthm.2015.06.022, indexed in Pubmed: 26073595.
  12. Gemein C, Haj M, Schmitt J. Combining an subcutaneous ICD and a pacemaker with abdominal device location and bipolar epicardial left ventricular lead: first-in-man approach. Europace. 2016; 18(8): 1279, doi: 10.1093/europace/euv443, indexed in Pubmed: 26843577.
  13. Kossidas K, Kalman R, Follis WP, et al. Managing cross talk between a subcutaneous implantable cardioverter-defibrillator and a dual-chamber unipolar pacemaker system. HeartRhythm Case Rep. 2017; 3(12): 579583, doi: 10.1016/j.hrcr.2017.08.004, indexed in Pubmed: 29296580.
  14. Mondésert B, Dubuc M, Khairy P, et al. Combination of a leadless pacemaker and subcutaneous defibrillator: First in-human report. HeartRhythm Case Rep. 2015; 1(6): 469471, doi: 10.1016/j.hrcr.2015.07.009, indexed in Pubmed: 28491609.
  15. Kaczmarek K, Czarniak B, Jakubowski P, et al. Leadless pacemaker and subcutaneous implantable cardioverter-defibrillator therapy: the first use of a novel treatment option in Poland. Kardiol Pol. 2018; 76(6): 1026, doi: 10.5603/KP.2018.0126, indexed in Pubmed: 29905373.
  16. Enomoto Y, Hashimoto H, Ishii R, et al. Leadless pacemaker and subcutaneous implantable cardioverter defibrillator combination in a hemodialysis patient. Circ J. 2018; 82(12): 31083109, doi: 10.1253/circj.CJ-18-0263, indexed in Pubmed: 29877203.
  17. Kempa M, Budrejko S, Zienciuk-Krajka A, et al. Subcutaneous implantable cardioverter-defibrillators for the prevention of sudden cardiac death: 5-year single-center experience. Kardiol Pol. 2020; 78(5): 447450, doi: 10.33963/KP.15235, indexed in Pubmed: 32186353.
  18. Gauthey A, Calle S, Accinelli S, et al. His bundle pacing for newly acquired pacing needs in patients implanted with a subcutaneous implantable cardioverter defibrillator: A feasibility study based on the automated screening score and clinical cases. J Cardiovasc Electrophysiol. 2020; 31(7): 17931800, doi: 10.1111/jce.14566, indexed in Pubmed: 32412155.
  19. Ip JE, Wu MS, Kennel PJ, et al. Eligibility of pacemaker patients for subcutaneous implantable cardioverter defibrillators. J Cardiovasc Electrophysiol. 2017; 28(5): 544548, doi: 10.1111/jce.13182, indexed in Pubmed: 28185354.
  20. Moubarak G, Maison-Blanche P, Thomas O. Eligibility of cardiac resynchronization therapy patients for subcutaneous implantable cardioverter defibrillators. J Electrocardiol. 2018; 51(3): 531533, doi: 10.1016/j.jelectrocard.2017.12.005, indexed in Pubmed: 29273233.
  21. Kawabata M, Goya M, Takahashi Y, et al. Candidacy for a subcutaneous implantable cardioverter defibrillator in patients with cardiac resynchronization therapy. Int Heart J. 2018; 59(5): 951958, doi: 10.1536/ihj.17-550, indexed in Pubmed: 30101850.
  22. Gauthey A, Calle S, Accinelli S, et al. His bundle pacing for newly acquired pacing needs in patients implanted with a subcutaneous implantable cardioverter defibrillator: A feasibility study based on the automated screening score and clinical cases. J Cardiovasc Electrophysiol. 2020; 31(7): 17931800, doi: 10.1111/jce.14566, indexed in Pubmed: 32412155.
  23. Giammaria M, Lucciola MT, Amellone C, et al. Eligibility of cardiac resynchronization therapy patients for subcutaneous implantable cardioverter defibrillators. J Interv Card Electrophysiol. 2019; 54(1): 4954, doi: 10.1007/s10840-018-0437-9, indexed in Pubmed: 30187250.



Polish Heart Journal (Kardiologia Polska)