Vol 80, No 3 (2022)
Clinical vignette
Published online: 2022-02-02

open access

Page views 4957
Article views/downloads 433
Get Citation

Connect on Social Media

Connect on Social Media

_27_HTML__KP_03_2022__Plaskota__Transcatheter
  • „ Clinical vignette

Transcatheter tricuspid valve-in-valve replacement due to severe bioprosthesis dysfunction in a patient with endocardial leads

Karolina Plaskota1, Marcin Demkow2, Ilona Michałowska3, Maciej Sterliński4, Mirosław Kowalski1, Piotr Hoffman1

1Department of Adult Congenital Heart Disease, National Institute of Cardiology, Warszawa, Poland

2Department of Coronary and Structural Heart Disease, National Institute of Cardiology, Warszawa, Poland

3Department of Radiology, National Institute of Cardiology, Warszawa, Poland

41st Department of Arrhythmia, National Institute of Cardiology, Warszawa, Poland

Correspondence to:

Karolina Plaskota, MD, PhD,

Department of Adult Congenital Heart Disease,

National Institute of Cardiology,

Alpejska 42, 04–462 Warszawa, Poland,

phone: +48 22 343 44 00,

e-mail: kplaskota@ikard.pl

Copyright by the Author(s), 2022

DOI: 10.33963/KP.a2022.0028

Received: January 4, 2022

Accepted: January 30, 2022

Early publication date: February 2, 2022

The safety and efficacy of transcatheter valve-in-valve implantation have led to the rapid growth of this treatment strategy especially in high surgical risk patients with aortic valve prosthetic dysfunction [1]. Although less proven, transcatheter tricuspid valve-in-valve replacement has also good short-term outcomes. Performing such a procedure in patients with endocardial leads poses an additional risk. Potential concerns include the impact of the pacemaker lead on the valve deployment, the presence and the extent of the perivalvular leak along with the pacemaker lead, and the risk of jailed lead dysfunction during the procedure and in the long-term follow-up [2–5].

A 59-year-old female patient with Ebstein’s anomaly, after surgical tricuspid valve replacement with a 31-mm Carpentier-Edwards bioprosthesis due to severe tricuspid regurgitation in 1987, after dual-chamber pacemaker implantation due to symptomatic second-degree atrioventricular heart block in 2003 was regularly followed-up in our outpatient clinic. From the beginning of 2021, she gradually started to present heart failure symptoms (exertional dyspnea with nausea and lower extremity edema) ultimately in class III according to the New York Heart Association classification. Laboratory tests revealed increased serum N-terminal prohormone for brain natriuretic peptide concentration of 1526 pg/ml (in comparison to 669 pg/ml a year before). Transthoracic echocardiographic examination demonstrated tricuspid bioprosthesis dysfunction with a mean transvalvular gradient of 14.5 mm Hg and a small perivalvular leak in the lateral part of the tricuspid ring, the enlarged right ventricle with a preserved systolic function, and severely enlarged right atrium (Figure 1A, B).

5206.png

Figure 1. A. Transthoracic echocardiography. Four-chamber apical zoom view. Degenerated leaflets of the tricuspid bioprosthesis. B. Transthoracic echocardiography. Four-chamber apical view. Color Doppler flow. Turbulent flow through degenerated leaflets of the tricuspid bioprosthesis. C. Cardiac computed tomography. Multiplanar reconstruction, short-axis view. Bioprosthetic tricuspid valve degeneration (the arrow) thickening and calcifications of the valve leaflets. D. Cardiac computed tomography after reimplantation of the tricuspid valve. Multiplanar reconstruction. Tricuspid valve bioprosthesis (the black arrow), pacemaker lead (the white arrow). E. Transthoracic echocardiography after the valve-in-valve procedure. Four-chamber apical zoom view. Correct position of the tricuspid valve-in-valve bioprosthesis. F. Transthoracic echocardiography after the valve-in-valve procedure. Substernal long-axis view. Color Doppler flow. Mild perivalvular leak in the lateral part of the tricuspid ring in the surrounding of the ventricular lead

Cardiac computed tomography confirmed the diagnosis (Figure 1C). Taking into consideration prior sternotomy, benign thrombocytopenia, and her new affiliation with Jehovah’s witness group, the Heart Team decided to refer her for percutaneous valve-in-valve implantation with right ventricle lead jailing. Pre-procedural pacemaker interrogation showed 88% pacing with underlying sinus rhythm with second to third-degree atrioventricular block resulting in the heart rate of 4065 bpm, dual-chamber pacing mode, and correct right ventricle lead threshold. The patient underwent tricuspid valvuloplasty with a 25-mm Edwards balloon followed by valve-in-valve 29-mm Sapien 3 bio­prosthesis implantation via the right femoral approach. The procedure was complicated by a femoral vein bleeding with an extensive hematoma and a hemoglobin concentration decrease not requiring blood transfusion. There were no changes in the device function in the post-procedural interrogation. Multimodality imaging using transthoracic echocardiography and cardiac computed tomography revealed correct position and good function of the bioprosthesis with a mean transvalvular gradient of 3.4 mm Hg and a mild perivalvular leak in the lateral part of the tricuspid ring in the surrounding of the ventricular lead (Figure 1DF). At a three-month follow-up, the patient was free from heart failure symptoms, and her hemoglobin concentration was stable at the level of 13.3 g/dl.

Transcatheter tricuspid valve-in-valve implantation is a safe alternative to redo surgical tricuspid valve replacement, especially in high surgical risk patients. Reported data show that jailing of the right ventricle lead does not cause a severe perivalvular leak or pacemaker dysfunction in the short-term follow-up. Yet, more studies are required to evaluate long-term outcomes.

Article information

Conflict of interest: None declared.

Open access: This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.

References

  1. 1. Sanon S, Cabalka AK, Babaliaros V, et al. Transcatheter tricuspid valve-in-valve and valve-in-ring implantation for degenerated surgical prosthesis. JACC Cardiovasc Interv. 2019; 12(15): 14031412, doi: 10.1016/j.jcin.2019.05.029, indexed in Pubmed: 31326424.
  2. 2. Dannenberg V, Donà C, Koschutnik M, et al. Transcatheter treatment by valve-in-valve and valve-in-ring implantation for prosthetic tricuspid valve dysfunction. Wien Klin Wochenschr. 2021; 133(15-16): 780785, doi: 10.1007/s00508-021-01842-x, indexed in Pubmed: 33791869.
  3. 3. Eleid MF, Asirvatham SJ, Cabalka AK, et al. Transcatheter tricuspid valve-in-valve in patients with transvalvular device leads. Catheter Cardiovasc Interv. 2016; 87(4): E160E165, doi: 10.1002/ccd.25990, indexed in Pubmed: 25946058.
  4. 4. Demir OM, Regazzoli D, Mangieri A, et al. Transcatheter tricuspid valve replacement: principles and design. Front Cardiovasc Med. 2018; 5: 129, doi: 10.3389/fcvm.2018.00129, indexed in Pubmed: 30283790.
  5. 5. Loyalka P, Nascimbene A, Metz B, et al. Transcatheter tricuspid valve-in-valve replacement with an Edwards Sapien 3 valve. Tex Heart Inst J. 2017; 44(3): 209213, doi: 10.14503/THIJ-15-5639, indexed in Pubmed: 28761403.



Polish Heart Journal (Kardiologia Polska)