Vol 76, No 8 (2018)
Original articles
Published online: 2018-04-06

open access

Page views 2667
Article views/downloads 1058
Get Citation

Connect on Social Media

Connect on Social Media

Lead-related complications after DDD pacemaker implantation

Maciej Dębski, Mateusz Ulman, Andrzej Ząbek, Krzysztof Boczar, Kazimierz Haberka, Marcin Kuniewicz, Jacek Lelakowski, Barbara Małecka
Kardiol Pol 2018;76(8):1224-1231.

Abstract

Background: Pacing leads remain the weakest link in pacemaker systems despite advances in manufacturing technology.

Aim: The aim of the study was to assess the long-term pacing lead performance in an unselected real-life cohort following primary DDD pacing system implantation.

Methods: A single-centre retrospective analysis of patients who underwent DDD pacing system implantation between October 1984 and December 2014 and were followed-up until August 2016 was conducted. The inclusion criterion was at least one follow-up visit after post-implant discharge. The performance of each atrial and ventricular lead implanted was evaluated during the follow-up period, and the incidence of, and predictive factors for, lead dislodgement and failure were analysed.

Results: The data of 3771 patients and 24,431.8 patient-years of follow-up were analysed. The mean follow-up of patients was 77.7 ± 61.8 months. During the study period, 7887 transvenous atrial and right ventricular pacing leads were implanted. Lead dislodgement occurred in 94 (1.2%) leads (92 [2.4%] patients), perforation in 11 (0.1%) leads (10 [0.3%] patients), and lead failure in 329 (4.2%) leads (275 [7.3%] patients). Atrial lead position was a predictive factor for lead dislodgement, while age at implantation, polyurethane 80A insulation, subclavian vein access, unipolar lead construction, and lead manufacturer were multivariate predictors of lead failure.

Conclusions: Leads with polyurethane 80A insulation, unipolar construction, and those implanted via subclavian vein puncture exhibited the worst long-term performance.

Article available in PDF format

View PDF Download PDF file



Polish Heart Journal (Kardiologia Polska)