Atrio-ventricular block in the presence of double-chamber ICD — how is that possible?

Andrzej Ząbek, Jacek Lelakowski, Jacek Łach, Barbara Małecka
Kardiol Pol 2013;71(9):969-973.

Abstract

We present description of ECG by Holter monitoring in a 26-year-old patient with double-chamber implantable cardioverter-defibrillator (ICD). Fragments of record show atrio-ventricular block, which may suggests ICD dysfunction. However, this is working mode of ICD having minimal ventricular pacing algorhythm.